K Number
K990333

Validate with FDA (Live)

Date Cleared
1999-04-28

(84 days)

Product Code
Regulation Number
882.1480
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Introducer Sheath is a device used to make a channel through the brain into the ventricular system.

Device Description

The Introducer Sheath includes two parts, a peelaway sheath and a blunt ended obturator, which fits inside the sheath. The Introducer Sheath provides an atraumatic passage into the ventricular system.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called the "Medtronic PS Medical Endoscope Introducer." This type of document focuses on demonstrating substantial equivalence to a previously legally marketed device, rather than providing detailed acceptance criteria and a specific study proving the device meets those criteria.

Therefore, many of the requested details about performance, sample sizes, ground truth establishment, and comparative effectiveness studies are not available in this document. The document's purpose is to show that the new device is as safe and effective as existing ones, based on similar design and intended use.

Here's an analysis based on the information available in the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not provide a table of acceptance criteria or specific performance metrics that the Medtronic PS Medical Endoscope Introducer was tested against in a formal study.

Instead, the document asserts technological comparison and substantial equivalence to predicate devices (Navigational F Sheath and Codman Peel Away Sheath) based on:

  • Device configuration/contents: Peel Away Sheath, Obturator
  • Sterility Method: EtO (Ethylene Oxide)
  • Sterile: Sterile single-use device
  • Intended Use: To make a channel through the brain into the ventricular system.

The "reported device performance" is implicitly that it performs like the predicate devices, enabling the creation of a channel through the brain for ventricular access. No quantitative performance data is provided.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. As a 510(k) submission based on substantial equivalence, it relies on the established safety and effectiveness of the predicate devices. There is no mention of a specific test set or data derived from testing the Medtronic PS Medical Endoscope Introducer.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided. Ground truth establishment for a test set is typically part of a dedicated performance study, which is not detailed in this 510(k) submission.

4. Adjudication Method for the Test Set

This information is not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done (or at least, not reported in this document). The 510(k) process for this device relies on demonstrating equivalence to predicate devices, not on a human-in-the-loop study with AI.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

This device is a physical medical instrument, not an algorithm or AI. Therefore, the concept of "standalone (algorithm only)" performance is not applicable to this product.

7. The Type of Ground Truth Used

The "ground truth" here is not clinical outcome data or expert consensus on performance of the new device. Instead, the "ground truth" for this 510(k) submission is implicitly the established safety and effectiveness of the predicate devices (Navigational F Sheath (K931973) and Codman Peel Away Sheath (K883607)), which allow for the "creation of a channel through the brain into the ventricular system." The new device is deemed substantially equivalent in its design, materials, and intended use.

8. The Sample Size for the Training Set

This information is not applicable and not provided. This is a physical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided.

{0}------------------------------------------------

4/28/99

X. Safety and Effectiveness Summary

K990332

This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 882.4545

Establishment Registration Number:2021898
Address of Manufacturer:Medtronic PS Medical Corporation125 Cremona DriveGoleta CA, 93117(805) 968-1546 ext. 1776Fax: (805) 968-5038
Contact Person:Jeffrey Henderson
Date:January 29, 1998
Trade or Proprietary Name:Medtronic PS Medical Endoscope Introducer
Common usual or Classification Name:Instrument, Shunt System(882.4545)

Navigational F Sheath (K931973) Codman Peel Away Sheath (K883607)

Description:

The Introducer Sheath includes two parts, a peelaway sheath and a blunt ended obturator, which fits inside the sheath. The Introducer Sheath provides an atraumatic passage into the ventricular system.

The Introducer Sheath is a device used to Intended Use: make a channel through the brain into the ventricular system

Intended Use predicate device: The Disposable Neuroview Endoscope is intended for use in direct visualization, diagnosis of disease and therapeutic applications for intracranial procedures (e.g., biopsy, tumor resection, coagulation of choroid plexus, cyst fenestration, shunt placement, etc.),

Technological comparison: Medtronic PS Medical submits that the materials of fabrication, intended uses, performance characteristics and design specifications of the Endoscope Introducer are substantially equivalent to those of the predicate devices. Based upon the summary above, Medtronic PS Medical determines substantial equivalence, safety, and efficacy of the Endoscope Introducer based upon the predicate and currently marketed devices.

{1}------------------------------------------------

FeatureMedtronic PS MedicalEndoscope IntroducerNeuro Navigational Peel AwayIntroducer SheathCodman Peel Away Sheath
Deviceconfiguration/contents• Peel Away Sheath• Obturator• Peel Away Sheath• Obturator• Peel Away Sheath• Obturator
Sterility MethodEtOnot specifiednot specified
SterileSterile single use deviceSterile device single useSterile single use device
Intended UseThe Introducer Sheath is a device used to make achannel through the brain into the ventricular system.The Disposable Neuroview Endoscope is intendedfor use in direct visualization, diagnosis of diseaseand therapeutic applications for intracranialprocedures (e.g., biopsy, tumor resection,coagulation of choroid plexus, cyst fenestration,shunt placement, etc.).not specified

1

and the comments of the comments of the country of

.

and the state of the state of the states and

and the control control control control control controllers and

とコ

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, stylized with thick, curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 8 1999

Mr. Jeffrey Henderson Vice President, Quality Medtronic PS Medical 125 Cremona Drive Goleta, California 93117

Re: K990333 Trade Name: Endoscope Introducer Regulatory Class: II Product Code: GWG Dated: January 29, 1999 Received: February 3, 1999

Dear Mr. Henderson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Jeffrey Henderson

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Colin M. Wittwer, Ph.D., MD

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Device Name: Endoscope Introducer

Indications for Use:

The Introducer Sheath is a device used to make a channel through the brain into the ventricular system.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

colla

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number L990373

Over the Counter Use: or Prescription Use: (Per 21 CFR 801.109)

(optional format 1-2-96)

Prescription Use __
(Per 21 CFR 801.109)

§ 882.1480 Neurological endoscope.

(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).