K Number
K990197

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-02-04

(15 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Exactech Hip Systems are also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.

Device Description

The Exactech MCS - HA Porous Coated Total Hip System is made of similar materials and is of a similar design to prostheses that were on the market before May 28, 1976. Additionally, this Porous Coated Total Hip System is of similar design to other components on the market that have been determined to be equivalent to devices on the market prior to May 28, 1976.

AI/ML Overview

The Exactech MCS - HA Total Hip System's acceptance criteria and the study proving its performance are based on demonstrating substantial equivalence to predicate devices, primarily through material integrity and sterilization testing.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Test DescriptionReported Device PerformanceComparator/Predicate Data (if available)
Material IntegrityInterfacial shear strength (uncoated sintered bead surface)In excess of 27 N/sq.mmData meets or exceeds predicate data (K921113, K921114)
Material IntegrityHA coating tensile strength (porous coated Ti alloy)Approximately 55 MPa-
Material IntegrityHA shear strength on solid substrates (Ti-6A1-4V and Co-Cr-Mo)36.6-44.8 MPa (5300-6500 psi)-
Material IntegrityHA shear strength on porous coated substrates (Ti-6A1-4V and Co-Cr-Mo)34.5-40.7 MPa (5000-5900 psi)-
Material IntegrityHA shear fatigue strength (on Ti-6A1-4V at 10 million cycles)13.8 MPa (2000 psi)-
SterilizationSterility Assurance Level (SAL)10⁻⁴-

Note: The primary acceptance criterion here is substantial equivalence to legally marketed predicate devices, meaning the new device performs as safely and effectively as a device already on the market. The specific values reported for material integrity are considered acceptable because they either meet or exceed predicate device data or are within acceptable ranges for such materials in this application.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Material Integrity (Interfacial Shear Strength): 10 coating samples were tested.
    • Material Integrity (HA Coating Tensile/Shear/Fatigue Strength): Specific sample sizes for each of these tests are not explicitly stated in the provided text. However, the data is presented as definitive values (e.g., "approximately 55 MPa"), suggesting testing was conducted to establish these properties.
  • Data Provenance: The data appears to be retrospective laboratory testing on manufactured components. There's no mention of clinical data or human subject studies for this 510(k) submission regarding the acceptance criteria. The tests were performed on device components, not on patient data. No country of origin for the data is specified, but the submission is to the US FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable in the context of this 510(k) submission. The "ground truth" for material integrity and sterilization is established through standardized engineering and laboratory testing protocols, not through expert consensus on interpretation of images or clinical outcomes. The performance is measured against objective physical and chemical properties and established sterilization standards.

4. Adjudication Method for the Test Set

This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image interpretation studies where expert review is needed to resolve discrepancies. For bench testing of material properties, the results are objectively measured and reported.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This type of study (comparing human readers with and without AI assistance) is relevant for diagnostic imaging devices where AI assists in interpretation. This 510(k) pertains to a total hip system, which is an implantable medical device, and its evaluation focuses on material properties and bio-compatibility, not diagnostic accuracy.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No standalone performance study of an algorithm was done. This 510(k) does not involve an AI algorithm or software device. The product is a physical hip implant system.

7. The Type of Ground Truth Used

The "ground truth" for this submission is based on:

  • Physical and Mechanical Material Properties: Direct measurements of characteristics like interfacial shear strength, tensile strength, shear strength, and fatigue strength of the device's materials and coatings.
  • Sterilization Standards: Demonstration of achieving a specified Sterility Assurance Level (SAL) according to recognized guidelines (e.g., AAMI Guideline for gamma radiation sterilization).
  • Predicate Device Performance: The underlying "ground truth" for substantial equivalence is the established safety and effectiveness of the legally marketed predicate devices. The new device's performance must meet or exceed these benchmarks.

8. Sample Size for the Training Set

This section is not applicable. There is no "training set" as this 510(k) does not involve a machine learning or AI model that requires training data. The device is a physical product, and its design and manufacturing processes are validated through engineering tests.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reason as above; there is no training set.

{0}------------------------------------------------

K990197

SUMMARY OF THE SAFETY AND EFFECTIVENESS INFORMATION FOR THE PREMARKET NOTIFICATION FOR THE MODULUS COMPATIBLE STABILITY (MCS) - HA TOTAL HIP SYSTEM

Exactech, Inc. Established Registration Number 1038671

The Exactech MCS - HA Porous Coated Total Hip System is made of similar materials and is of a similar design to prostheses that were on the market before May 28, 1976. Additionally, this Porous Coated Total Hip System is of similar design to other components on the market that have been determined to be equivalent to devices on the market prior to May 28, 1976. These predicates include but are not limited to:

    1. Exactech MCS Total Hip System
    1. Charnley
    1. Zimmer Harris-Galante
    1. Depuy AML
    1. Osteonics Secur Fit HA
    1. Smith & Nephew Synergy HA
    1. OTI

Material Integrity:

Overall integrity has been previously determined in the initial 510(k) submissions K921113 and K921114. HA coating performance characteristics are further defined in the FDA Master File MAF339 (BioCoat, Inc.) Data obtained meets or exceeds predicate device data. A summary is presented below.

Summary:

The evaluation of the interfacial shear strength of the uncoated sintered bead surface of the Exactech Porous System indicates excellent bonding of the beaded surface to the T1-6A1-4V substrate. Interfacial shear strengths in excess of 27 N/sq.mm can be expected with little variability based on the standard deviations observed for ten coating samples tested.

The integrity of the HA coating has been characterized. The manufacturer acknowledges the importance of coating quality to the success of the device and has collaborated with BioCoat, Inc., a well know coating specialty company, for the application and characterization of its coatings. (Additional information regarding the HA coating has been provided to the FDA as MAF339 by BioCoat)

The HA coating tensile strength on porous coated Ti alloy has been shown to be approximately 55 MPa. The shear strength of the HA on Ti-6Al-4V and Co-Cr-Mo solid

{1}------------------------------------------------

substrates has been shown to be between 5300 and 6500 psi. (36.6-44.8 MPa) while on porous coated Ti-6A1-4V and Co-Cr-Mo substrates the shear strength of the HA is between 5000 and 5900 psi. (34.5-40.7 MPa). Shear fatigue testing of the HA on Ti-6A1-4V shows a fatigue strength of 2000 psi. (13.8 MPa.) at 10 million cycles.

Sterilization

The Exactech HA - MCS Total Hip System will be sterilized by gamma irradiation. The Sterility Assurance Level (SAL) is 104. Exactech utilizes Method 3, Protocol B from the "AAMI Guideline for gamma radiation sterilization" for the sterility dose setting and validation procedure.

Utilization and Implantation

Selection of the Exactech HA - MCS Hip System depends on the judgment of the surgeon in relationship to the requirements of the patient. The surgeon should become thoroughly familiar with the technique of implantation by appropriate reading of the literature, and training in the operative skills and techniques required for total hip Arthoplasty surgery.

Indications

The Exactech HA - MCS Total Hip System is indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, osteonecrosis, congenital hip dysplasia, rhematoid arthritis, ankvlosing spondylitis and /or posttraumatic degenerative problems. It is also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.

Contraindications

Use of the Exactech HA - MCS Total Hip System is contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure to the system.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is a stylized symbol consisting of three curved lines that resemble a person embracing another person.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

4 1999 FEB

Gary J. Miller, Ph.D. Vice President of Research and Development Exactech® 2320 NW 66th Court Gainesville, Florida 32653

Re : K990197 Ha Coated MCS Total Hip System Requlatory Class: II Product Code: MEH Dated: January 15, 1999 Received: January 20, 1999

Dear Dr. Miller:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on consideration of the specific design of stem and coating composition detailed in this application. You may, therefore, market the device, subject to the general controls provisions of the Act and the following limitation:

You may not label or in any way promote these devices for "enhanced clinical or radiographic performance, enhanced fixation and/or long-term stable fixation." The data presented support equivalence with no additional claims over a conventional porous coated hip prosthesis.

Additional limitations for more specific claims of safety and effectiveness may be forthcoming. Should additional limitations be applied you will be contacted in writing to inform you of the additional labeling limitations.

The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

{3}------------------------------------------------

Page 2 - Gary J. Miller, Ph.D.

You may market your device under the above limitations as class II devices. These devices would be considered not substantially equivalent to a legally marketed predicate device if labeled with other intended uses and/or claims of safety or effectiveness. Any other intended uses or claims may cause the device to be classified into Class III under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing.

Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes 895. compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be

{4}------------------------------------------------

Page 3 - Gary J. Miller, Ph.D.

obtained from the Division of Small Manufacturers Assistance obcathed from the bivibror. or 638-2041 or (301) 443-6597 or at at Its Coll-free namber (600)- 050-2811-02-28 11:52 PM dsmamain.html".

Sincerely yours,
Neef R.P. Oyler

Celia M. Witten, Ph.D., M . Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

"INDICATIONS FOR USE"

510(k) Number (if known) _ K990197 7 ___

Device Name:

Indications for Use:

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Exactech Hip Systems are also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X

OR

Over the Counter Use

MRO

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K990197

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.