K Number
K984559

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-02-26

(65 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.

Device Description

Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use), meet all of the requirements of ASTM D 3577, Type 2.
Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) meet all the current specifications for ASTM D 3577 Rubber Surgical Gloves.
Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Elite™ Powder Free Polyurethane Synthetic Surgical Gloves:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionsMeets ASTM D 3577Meets ASTM D 3577
Physical PropertiesMeets ASTM D 3577, Type 2Meets ASTM D 3577, Type 2
Freedom from holesMeets ASTM D 3577Meets ASTM D 3577
Freedom from holesMeets ASTM D 5151Meets ASTM D 5151
Powder-FreeMeets described test in Attachment VI of K973461Not more than 2 mg residue by mass.
Biocompatibility: Primary Skin Irritation in RabbitsPassPasses
Biocompatibility: Guinea Pig SensitizationPassPasses

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample size used for the test set during the performance evaluations. It generally states that the "performance test data of the non clinical tests are the same as mentioned immediately above," referring to the ASTM standards.

The data provenance is not explicitly stated in terms of country of origin, but the testing refers to ASTM (American Society for Testing and Materials) standards, suggesting testing was conducted against these established specifications, likely in the US or by labs adhering to these standards. The testing appears to be retrospective in the sense that the results are presented as already achieved, rather than for a future study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The "ground truth" for these types of device performance tests (e.g., dimensions, physical properties, freedom from holes, biocompatibility) is typically established by objective measurements and standardized procedures, not by human expert consensus in the way clinical diagnostic products might be. The ASTM standards themselves define the methods and criteria.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided. Adjudication methods like "2+1" or "3+1" are relevant for subjective interpretations, often in clinical studies where multiple human readers assess cases. For objective device performance tests against ASTM standards, there isn't typically a panel of adjudicators. The tests are designed to yield objective results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or assistive devices where human readers (e.g., radiologists) are involved in interpreting results. The Elite™ gloves are a medical device without an AI component for interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone study was not done. This question is also relevant for AI-powered diagnostic or assistive devices. The Elite™ gloves are a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used for these performance tests is based on objective measurements defined by the specific ASTM standards referenced. For example:

  • Dimensions: Measured against specified tolerances in ASTM D 3577.
  • Physical Properties: Determined through standardized tests (e.g., tensile strength, elongation) described in ASTM D 3577.
  • Freedom from Holes: Assessed using methods outlined in ASTM D 3577 and ASTM D 5151 (e.g., water leak test).
  • Powder-Free: Quantified by residue mass as per the described test in Attachment VI of K973461.
  • Biocompatibility: Evaluated through standardized animal tests (rabbits for skin irritation, guinea pigs for sensitization), where "Passes" indicates compliance with established toxicity criteria.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. The concept of a "training set" refers to data used to train machine learning models. The Elite™ gloves are a manufactured product, not an AI system.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above.

{0}------------------------------------------------

2/26/99

Ansell Perry

K984559

Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) Ansell Perry 1875 Harsh Avenue SE Massillon, Ohio 44646 330-833-2811 Telephone: 330-833-6213 Fax:

Checklist Section 21.0

  • 510 (k) Summary [1]
  • [2] Ansell Perry Inc. 1875 Harsh Avenue SE Massillon, Ohio 44646

330-833-2811 Telephone: 330-833-6213 Fax:

James R. Chatterton Contact: 330-833-2811 Telephone: Fax: 330-833-6213

December 23, 1998

  • Trade Name: Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) [3] Surgical Gloves, Green Polyurethane Common Name: Classification Name: Surgeon's Glove
  • Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use), meet all of [4] the requirements of ASTM D 3577, Type 2.
  • Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) meet all the [ર] current specifications for ASTM D 3577 Rubber Surgical Gloves.
  • Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) are sterile િશ disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.
  • [7] Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) are summarized with the following technological characteristics compared to ASTM or equivalent standards.
CharacteristicsStandard
DimensionsMeets ASTM D 3577
Physical PropertiesMeets ASTM D 3577, Type 2

Ansell Healthcare Products Inc. • 1875 Harsh Ave. S.E. • Massillon, Ohio 44646 • U.S.A. http://www.amsell.com Telephone: 330-833-2811 • Toll Free: 800-321-9752 (USA only) • Fax: 330-833-5991

{1}------------------------------------------------

K984559

Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) Ansell Perry 1875 Harsh Avenue SE Massillon, Ohio 44646 330-833-2811 Telephone: 330-833-6213 Fax:

Freedom from holes

.)

Meets ASTM D 3577 Meets ASTM D 5151

Powder-Free Meets described test in Attachment VI of K973461 Biocompatability Primary Skin Irritation in Rabbits Guinea Pig Sensitization

Meets ASTM D 6124 Not more than 2 mg residue by mass.

Passes Passes

  • The performance test data of the non clinical tests are the same as mentioned immediately above. [8]
  • (9) Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process.
  • [10] It is concluded that Elite™ Powder Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) are as safe, as effective, and perform as well as the glove performance standards referenced in Section 7 above and therefore meet:

ASTM listed standards. FDA hole requirements, and labeling claims for the product.

  • [11] This summary will include any other information reasonably deemed necessary by The FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 26 1999

Mr. James R. Chatterton Vice President Regulatory Affairs/Technical Ansell Perry, Incorporated 1875 Harsh Avenue S.E. Massillon, Ohio 44646-7199

K984559 Re : Elite Powder-Free Polyurethane Synthetic Trade Name: Surgical Gloves (Chemotherapy Use) Requlatory Class: I Product Code: KGO Dated: December 23, 1998 December 23, 1998 Received:

Dear Mr. Chatterton:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general The general controls controls provisions of the Act. provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Chatterton

through 542 of the Act for devices under the Electronic enrough Size of Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA deborizon substantial equivalence of your device to a legally Finding of babboandable results in a classification for your marketed production of mits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{4}------------------------------------------------

3.0 Indications for Use Statement:

510(K) Number (if known): __

INDICATIONS FOR USE

Ansell Perry Applicant: 1698 4559

Elite™ Powder-Free Polyurethane Synthetic Surgical Gloves (Chemotherapy Use) Device Name: _ Indications For Use:

A device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Clim S. Lin
(Division Sign Off)

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number

Prescription Use Per 21 CFR 801.109 OR

Over-The-Counter

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).