K Number
K984381

Validate with FDA (Live)

Date Cleared
1999-02-19

(73 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Buretrol® Solution Sets with the proposed burette chamber will have the same intended use as sets with the current design of this component. The intended use of these solution administration sets is the administration of fluids from a container into the patient's vascular system through a vascular access device.

Device Description

Baxter markets a line of Buretrol® solution sets which are intended for the administration of infusion fluids from a container to the patient's vascular system. The sets contain a burette chamber which can be used to mix supplementary medication in a measured amount of diluent from the main container. The sets can be adjusted for either metered volume solution administration (intermittent) or continuous solution administration. During intermittent infusion, the main container is shut off above the burette chamber allowing metered (controlled) delivery of the measured volume in the chamber. Some of the Buretrol® sets contain a valve feature which automatically blocks flow after solution has emptied from the burette chamber. The set can be easily converted from the intermittent to the continuous administration mode by closing the air control lever at the top of the burette chamber and allowing continuous infusion from the main container. The subject of this submission is a change in the material composition of the burette chamber housing. The material of the housing will change from cellulose to polyester.

AI/ML Overview

This document describes a 510(k) premarket notification for modified Buretrol® Solution Sets. The modification involves a change in the material composition of the burette chamber housing from cellulose to polyester to reduce product cost. The submission aims to demonstrate that the modified device is substantially equivalent to the currently marketed Buretrol® Solution Sets.

Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria or specific numerical performance metrics. However, it broadly states that:

Acceptance Criteria CategoryReported Device Performance
Biological and Chemical Reactivity"The materials were found to be acceptable for their intended use," based on ISO Standard 10993-1 and USP Physicochemical tests.
Functional Performance"The data indicate that the proposed burette chamber meets or exceeds all functional requirements and support its suitability for use in Buretrol® Solution Sets."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It simply states that "Data regarding the functional performance of the proposed burette chamber have been generated."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The studies mentioned (biological/chemical and functional performance) are likely laboratory-based tests and do not involve human expert interpretation for "ground truth" in the typical sense of medical image analysis or diagnostic studies.

4. Adjudication Method for the Test Set

This information is not applicable and is not provided in the document. The studies are not clinical trials requiring adjudication of expert interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic or screening devices where human readers provide interpretations, and the AI's impact on their performance is evaluated. The modified Buretrol® Solution Set is an infusion administration device, not a diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a sense, "standalone" testing was performed. The biological, chemical, and functional performance tests evaluate the device itself (the modified burette chamber) without human interaction being a variable in the performance assessment. However, this is not an "algorithm-only" standalone performance as typically described in AI/ML contexts, since the device itself is a physical product, not a software algorithm.

7. The Type of Ground Truth Used

The ground truth used for this type of device modification would be based on:

  • Established standards: ISO Standard 10993-1 for biocompatibility and USP Physicochemical tests for material properties.
  • Pre-defined functional specifications: The "functional requirements" of the existing Buretrol® Solution Sets against which the modified device's performance was compared. These specifications likely include parameters like flow rate accuracy, drip chamber integrity, connection compatibility, etc.
  • Laboratory measurements and observations: Direct measurements and observations from the conducted tests.

8. The Sample Size for the Training Set

This information is not applicable and is not provided in the document. The studies described are not machine learning studies that require a training set. They are engineering and materials science tests.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for the type of device and study described.

{0}------------------------------------------------

510(k) SUMMARY

510(k) Premarket Notification Modified Buretrol® Solution Sets

984381

Modified Buretrol® Solution Sets

Submitted by:

Judy Kannenberg Baxter Healthcare Corporation I.V. Systems Division Rte. 120 and Wilson Road Round Lake, IL 60073

Date Prepared: December 7. 1998

Proposed Device: Modified Buretrol® Solution Sets

Predicate Devices: Buretrol® Solution Sets

Proposed Device Description:

Baxter markets a line of Buretrol® solution sets which are intended for the administration of infusion fluids from a container to the patient's vascular system. The sets contain a burette chamber which can be used to mix supplementary medication in a measured amount of diluent from the main container. The sets can be adjusted for either metered volume solution administration (intermittent) or continuous solution administration. During intermittent infusion, the main container is shut off above the burette chamber allowing metered (controlled) delivery of the measured volume in the chamber. Some of the Buretrol® sets contain a valve feature which automatically blocks flow after solution has emptied from the burette chamber. The set can be easily converted from the intermittent to the continuous administration mode by closing the air control lever at the top of the burette chamber and allowing continuous infusion from the main container.

The subject of this submission is a change in the material composition of the burette chamber housing. The material of the housing will change from cellulose to polyester. This change is being made to reduce product cost.

DEC 0 7 1998

{1}------------------------------------------------

Statement of Intended Use:

The Buretrol® Solution Sets with the proposed burette chamber will have the same intended use as sets with the current design of this component. The intended use of these solution administration sets is the administration of fluids from a container into the patient's vascular system through a vascular access device.

Summary of Technological Characteristics of New Device to Predicate Devices

The Buretrol® Solution Sets with the proposed burette chamber are the same as currently marketed sets, except for the material change in the burette chamber housing. The material of the housing will change from cellulose to polyester. All other aspects of the burette chamber will remain unchanged

Discussion of Nonclinical Tests and Referenced Studies Reported in Published Literature

The biological and chemical reactivity of the new polyester materials have been assessed using biological methods specified in ISO Standard 10993-1 and USP Physicochemical tests. The materials were found to be acceptable for their intended use.

Data regarding the functional performance of the proposed burette chamber have been generated. The data indicate that the proposed burette chamber meets or exceeds all functional requirements and support its suitability for use in Buretrol® Solution Sets.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a caduceus, a symbol often associated with medicine and healthcare.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 9 1999

Ms. Judy Kannenberg Manager, Regulatory Affairs Baxter Healthcare Corporation I.V. Systems Division Route 120 and Wilson Road Round Lake, Illinois 60073

Re : K984381 Modified Buretrol® Solution Sets Trade Name: Requlatory Class: II Product Code: FPA Dated: December 7, 1998 Received: December 8, 1998

Dear Ms. Kannenberg:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Ms. Kannenberg

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Premarket Notification Modified Buretrol® Sets

K984381

510(k) Number: Not Available

Modified Buretrol® Solution Sets Device Name:

Indication for Use:

The Buretrol® Solution Sets with the proposed burette chamber will have the same intended The Duction "Solution becs with and this component. The intended use of these solution administration sets is the administration of fluids from a container into the patient's vascular system through a vascular access device.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

Optional Format 1-2-96)

(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number K984381

s:\510k\buretcov

DEC 0 7 199

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.