K Number
K984320

Validate with FDA (Live)

Date Cleared
1999-01-28

(56 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bioabsorbable Interference screw is used for fixation of bone-tendonbone or soft-tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

Indications:

  1. The Bioabsorbable Interference Screw is used for fixation of bone-tendon-bone or softtissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.
  2. The screw is indicated for single use only.
Device Description

The Bioabsorbable Interference Screw is a screw that has a range of diameters from 7 to 9 mm and a length that ranges from 20 mm to 30 mm. The material of construction for both screws is polylactic acid.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (Bioabsorbable Interference Screw) and its substantial equivalence to a predicate device. However, it does not contain the detailed information needed to fill out all the requested categories regarding acceptance criteria and a study proving device performance in the way a modern AI/ML device submission would.

This document predates modern AI/ML device regulations and focuses on demonstrating substantial equivalence through comparable design, materials, and intended use, rather than a performance study with specific quantitative acceptance criteria as would be expected for AI/ML.

Here's an attempt to extract what is available and note what is missing based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied)Reported Device Performance
Mechanical Performance: Equivalent to predicate device."Both screws had results which are recorded in the test results located in attachment one." (Details not provided in this extract)
Material: Polylactic Acid (PLA).Device is made of PLA.
Dimensions: Diameter 7-9mm, Length 20-30mm.Device has diameters 7-9mm and lengths 20-30mm.
Intended Use: Fixation of grafts to bone during ACL reconstruction.Device is for fixation of grafts to bone during ACL/PCL reconstruction, expanding to soft-tissue.
Sterilization: Ethylene Oxide.Device is sterilized by Ethylene Oxide.
Labeling: Sterile, Single Use Only.Device is labeled Sterile, Single Use Only.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified in the provided text. The text only states "Both screws had results which are recorded in the test results located in attachment one."
  • Data Provenance: Not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable to this type of device submission. This information is typically relevant for interpretative devices, especially those using AI/ML, where expert consensus often establishes ground truth. This document pertains to a mechanical fixation device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable to this type of device submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a MRMC comparative effectiveness study was not done. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, a standalone performance study was not done in the context of an AI/ML algorithm. This is a mechanical device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • The "ground truth" for this type of device would be based on engineering performance metrics (e.g., pull-out strength, degradation rates, biocompatibility) tested against established biomechanical standards and compared to the predicate device. The specific metrics are not detailed in the provided summary but are implied to be in "attachment one".

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device, so there is no "training set."

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

JAN 28 1999

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is : K984320

This summary was prepared on December 1, 1998

A. Submitter

Smith & Nephew, Inc., Endoscopy Division 130 Forbes Boulevard Mansfield, MA 02048

Company Contact B.

Demetrios Tsakonas Clinical/Regulatory Specialist

C. Device Name

Trade Name:Bioabsorbable Interference Screw
Common Name:Screw, Fixation, Bone, Orthopedics
Classification Name:Screw: 87HWC

D. Predicate/Legally Marketed Devices

Arthrex Bio-Interference Screw (510K # K 945303) Arthrex, Inc. 3050 North Horseshoe Drive Naples, Florida

{1}------------------------------------------------

E. Device Description

The Bioabsorbable Interference Screw is a screw that has a range of diameters from 7 to 9 mm and a length that ranges from 20 mm to 30 mm. The Arthrex PLLA Bio-Screw is provided in a range of diameters from 7 to 9 mm and one length, 23 mm. The material of construction for both screws is polylactic acid.

F. Performance

The Bioabsorbable Interference screw has been tested against the Arthrex PLLA Bio-Screw. Both screws had results which are recorded in the test results located in attachment one.

G. Intended Use

Both screws have similar intended uses. The screws are used to fixate grafts to bone during anterior cruciate ligament reconstruction surgery. The Bioabsorbable is expanding that indication to soft-tissue.

H. Substantial Equivalence

The Bioabsorbable Interference Screw and the Arthrex PLLA have similar designs. Both are used in fixating grafts to bone. The material for both screws are the same and the screws have similar dimensions.

Risks to health have been addressed through the specified materials, processing controls, quality assurance, and compliance to the Medical Device Good Manufacturing Practices requlations.

A summary comparison of the characteristics of the Bioabsorbable Interference Screw and the substantially equivalent devices is presented in the table below.

Current ProductSubstantially Equivalent Product
Attribute↓Bioabsorbable Interference ScrewArthrex PLLA Bio-Screw
IndicationBone-Tendon-Bone or Soft Tissue GraftFixation for ACL/PCLReconstructionBone-Tendon-Bone or Soft Tissue GraftFixation for ACL Reconstruction
DimensionsDiameter: 7 - 9 mmDiameter: 7 - 9 mm
Length: 20 - 30 mmLength: 23 mm
MaterialPLAPLA
SterilizationEthylene OxideNA
LabelingSterile, Single Use OnlyNA
ApplicantDemetrios Jackson
Date12/2/98

Applicant

Date 12/2/98

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized depiction of a human figure embracing a globe. The emblem is positioned in the center of the circle. Encircling the emblem are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 28 1999

Mr. Demetrios Tsakonas Clinical/Requlatory Specialist Smith & Nephew Endoscopy 130 Forbes Boulevard Mansfield, Massachusetts 02048

K984320 Re : Bioabsorbable Interference Screw Trade Name: Requlatory Class: II Product Codes: HWC and MAI Dated: December 2, 1998 Received: December 3, 1998

Dear Mr. Tsakonas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Demetrios Tsakonas

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}------------------------------------------------

510 (k) Number (If Known):

Device Name: Bioabsorbable Interference Screw

Intended Use: Bioabsorbable Interference screw is used for fixation of bone-tendonbone or soft-tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

Indications:

    1. The Bioabsorbable Interference Screw is used for fixation of bone-tendon-bone or softtissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.
    1. The screw is indicated for single use only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Picedo

Prescription Use
(Per 21 CFR 801.109)

or Over-The-Counter Use ______________________________________________________________________________________________________________________________________________________

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.