The following are specific indications for the Osteonics® Spinal System:
For non-pedicular fixation of the T4-S2 spine:
- Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis) .
- . Vertebral fracture or dislocation
- . Spinal stenosis
- Spondylolisthesis ●
- Degenerative disc disease (defined as back pain of discogenic origin with degeneration of . the disc confirmed by patient history and radiographic studies).
- . Previously failed fusion
- . Spinal tumor
For pedicular use:
- Additionally, when used as a pedicle screw system, the system is intended for patients: (a) . having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
The Osteonics® Spinal System is comprised of single-use, non-sterile devices manufactured from ASTM F-136-96 Titanium Alloy (Ti6Al-4V ELI). The Osteonics® Spinal System bone screws are top loading screws that are threaded distally, have a forked proximal design, and are available in both standard and extended ("extra-long" or "long arm") proximal length configurations. The cylindrical bone screws will be available in standard and extra-long versions in lengths and diameters identical to the predicate top loading tapered screws.
This document is a 510(k) Premarket Notification for the Osteonics® Spinal System Cylindrical Bone Screws. It's a regulatory submission to demonstrate substantial equivalence to a legally marketed predicate device, rather than a study proving performance against acceptance criteria for a novel device. Therefore, a direct answer to your request in the format provided is not fully applicable as the document does not contain a study with acceptance criteria and device performance in the way you've outlined for a new device's validation.
However, I can extract the relevant information from the document to address aspects of your request within the context of a 510(k) submission for substantial equivalence.
Here's an analysis based on the provided text, outlining why some requested information is not present in this type of document and what can be inferred:
Key Takeaway: The provided document is a 510(k) submission, whose primary goal is to demonstrate "substantial equivalence" of a modified device (cylindrical bone screws) to an existing, legally marketed predicate device (tapered bone screws within the same Osteonics® Spinal System), not to establish novel performance against clinical acceptance criteria through a new study.
1. Table of Acceptance Criteria and Reported Device Performance
- Not Applicable in this format: A 510(k) submission for substantial equivalence typically does not present new acceptance criteria and reported device performance in the way a clinical trial or performance study for a novel device would. The core "acceptance criteria" here are that the modified device demonstrates substantial equivalence to the predicate device in terms of intended use, materials, design, and safety/effectiveness.
- Inferred Performance: The document states: "The function of the modified spinal system bone screws remains unchanged and is identical to that of the predicate Danek TSRH bone screws." This is the core "performance" claim for a 510(k) – that the new device performs equivalently to the predicate. The "study" here would be the comparative analysis outlined in the submission.
2. Sample Size Used for the Test Set and Data Provenance
- Not Applicable (Clinical Study): This document does not describe a clinical "test set" in the sense of a patient cohort for a performance study.
- Inferred "Test": The "test" in this context refers to the comparison of the cylindrical bone screws with the tapered predicate screws. This comparison is based on design parameters, material specifications, and intended uses. The "sample size" would effectively be "one" (the modified cylindrical bone screw design) compared against "one" (the predicate tapered bone screw design).
- Data Provenance: Not applicable for a clinical study. The data provenance would relate to the design specifications, material testing (if new), and intended use statements.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not Applicable: There is no mention of experts establishing a "ground truth" for a clinical test set because a clinical performance study (as you might see for a diagnostic AI device) was not conducted or reported here.
- Inferred Experts: The "experts" involved are likely the engineering and regulatory teams at Osteonics® who designed the device and prepared the submission, and later, the FDA reviewers who assessed the submission for substantial equivalence. Their qualifications are inherent in their roles, but not explicitly stated as "experts for ground truth."
4. Adjudication Method for the Test Set
- Not Applicable: No clinical test set or adjudication method is described. The "adjudication" in this regulatory context is the FDA's review and decision on substantial equivalence.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Conducted/Applicable: This is not a study type relevant to the submission of a medical implant (bone screws) for substantial equivalence. MRMC studies are typically for diagnostic imaging devices where human readers interpret images with and without AI assistance. There is no AI component here.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Not Conducted/Applicable: This concept is not relevant to bone screws. There is no algorithm or AI component in this device.
7. Type of Ground Truth Used
- Not Applicable (Clinical): There isn't a "ground truth" derived from patient data, pathology, or outcomes in the traditional sense of a clinical performance study.
- Inferred "Ground Truth": The "ground truth" for this submission is the established safety and effectiveness of the predicate device (the tapered bone screws within the Osteonics® Spinal System) and the regulatory framework that dictates how substantial equivalence is demonstrated. The ground truth relies on the assertion that the design changes (cylindrical vs. tapered) do not alter the fundamental function, safety, or effectiveness when compared to the predicate, as evidenced by engineering principles, material science, and the lack of new risks.
8. Sample Size for the Training Set
- Not Applicable: There is no "training set" as this is not an AI/ML device or a device requiring a clinical training phase as part of its development.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable: As there is no training set, this question is not relevant.
Summary of Information from the Text Regarding Substantial Equivalence:
The "study" in this context is the comparison presented in the 510(k) to demonstrate substantial equivalence.
- Acceptance Criteria for Substantial Equivalence:
- Intended Uses: Must be identical to the predicate device.
- Material: Must be identical or equivalent (same medical-grade titanium alloy, ASTM F-136-96 Titanium Alloy (Ti6Al-4V ELI)).
- Design: Changes (cylindrical vs. tapered) must not introduce new questions of safety or effectiveness and the fundamental function must remain unchanged.
- Reported Device "Performance" (Comparison to Predicate):
- Intended Uses: "The intended uses of the subject bone screws are identical to those of the predicate bone screws."
- Material: "The Osteonics® Spinal System is manufactured from ASTM F-136-96 titanium alloy (Ti6Al-4V ELI)." (Implied identical to predicate).
- Design: "The design of the modified cylindrical bone screw differs from the predicate bone screw in that the modified bone screw distal diameter is the same as the proximal diameter. The function of the modified spinal system bone screws remains unchanged and is identical to that of the predicate Danek TSRH bone screws."
This 510(k) submission successfully argued substantial equivalence based on these points, leading to FDA clearance (K984302).
{0}------------------------------------------------
Special 510(k) Premarket Notification
1 1 8 1498
510(k) Premarket Notification Summary of Safety and Effectiveness for the
Osteonics® Spinal System Cylindrical Bone Screws
Submission Information
Name and Address of the Sponsor of the 510(k) Submission:
Contact Person:
Date of Summary Preparation:
Osteonics Corporation 59 Route 17 Allendale, NJ 07401-1677
Kate Sutton Regulatory Affairs Specialist
November 30, 1998
Device Identification
Proprietary Name:
Common Name:
Classification Name and Reference:
Osteonics® Spinal System Cylindrical Bone Screws
Spinal fixation appliance
Spinal Interlaminal Fixation Orthosis 21 CFR 8888.3050
Predicate Device Identification
The Osteonics® Spinal System, which includes bone screws, was determined to be substantially equivalent via 510(k) #K951725. The proposed cylindrical bone screws of the Osteonics® Spinal System are substantially equivalent to the tapered bone screws in the Osteonics® Spinal System.
Device Description
The Osteonics® Spinal System is comprised of single-use, non-sterile devices manufactured from ASTM F-136-96 Titanium Alloy (Ti6Al-4V ELI). The Osteonics® Spinal System bone screws are top loading screws that are threaded distally, have a forked proximal design, and are available in both standard and extended ("extra-long" or "long arm") proximal length configurations. The cylindrical bone screws will be available in standard and extra-long versions in lengths and diameters identical to the predicate top loading tapered screws.
{1}------------------------------------------------
Intended Use:
The following are specific indications for the Osteonics® Spinal System:
For non-pedicular fixation of the T4-S2 spine:
- Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis) .
- . Vertebral fracture or dislocation
- . Spinal stenosis
- Spondylolisthesis ●
- Degenerative disc disease (defined as back pain of discogenic origin with degeneration of . the disc confirmed by patient history and radiographic studies).
- . Previously failed fusion
- . Spinal tumor
For pedicular use:
- Additionally, when used as a pedicle screw system, the system is intended for patients: (a) . having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
Statement of Technological Comparison:
The substantial equivalence of the Osteonics® Spinal System cylindrical bone screws to the predicate bone screws in the Osteonics® Spinal System, in terms of intended use and design features, is based on the following:
Intended Uses:
The intended uses of the subject bone screws are identical to those of the predicate bone screws.
Material:
The Osteonics® Spinal System is manufactured from ASTM F-136-96 titanium alloy (Ti6Al-4V ELI).
Design:
The design of the modified cvlindrical bone screw differs from the predicate bone screw in that the modified bone screw distal diameter is the same as the proximal diameter. The function of the modified spinal system bone screws remains unchanged and is identical to that of the predicate Danek TSRH bone screws.
Summary
Based on the similarities presented above and the supporting analyses, the substantial equivalence of the subject cylindrical bone screws to the predicate tapered bone screws of the legally marketed Osteonics® Spinal System is demonstrated.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping silhouettes suggesting a sense of community and support. The figure is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 18 1998
Ms. Kate Sutton Regulatory Affairs Specialist Stryker Osteonics® 59 Route 17 Allendale, New Jersey 07401-1677
Re: K984302
Cylindrical Bone Screws - part of the Osteonics® Spinal System Regulatory Class: II Product Codes: MNH and KWP Dated: November 30, 1998 Received: December 2, 1998
Dear Ms. Sutton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Kate Sutton
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely vours.
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K984-302
Device Name: Osteonics® Spinal System
Indications For Use:
The uses for the legally marketed Osteonics® Spinal System are as follows:
Non-Pedicular Use; fixation of the T4-S2 spine:
- Long and short curve scoliosis . .
- Vertebral fracture or dislocation .
- Spondylolisthesis
- Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).
- Previously failed fusion .
. - Spinal tumor
Pédicular Use:
- When used as a pedicle screw system, the system is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (C
coll
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number_ K984302
Prescription Use X
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)