K Number
K984046

Validate with FDA (Live)

Date Cleared
1999-01-21

(70 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BIOMEDICS 55 (ocufilcon D) UV Blocking soft (hydrophilic) contact lens is indicated for daily wear in not-aphakic, normal eyes which manifest myopia, hyperopia, or astigmatism in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Spherical lenses may mask astigmatism in patients with up to 2.00 diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the eyecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.

Device Description

Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA and other monomeric ingredients including a proprietary ultraviolet absorber, crosslinked with EGDMA and other components which yield the appearance of lenses which are designed to fit over the corneal surface of the eye. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness), hyperopia (farsightedness) and astigmatism (multiple foci). Each lens provides corrective power which is to correspond to the refractive power of the eye to which it is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6mm. Secondary and tertiary curves as well as beveled edge configurations are built into the lens for the purpose of aiding in lens centration and comfort.

AI/ML Overview

Acceptance Criteria and Study for OCULAR SCIENCES BIOMEDICS® 55 (ocufilcon D) UV Blocking Daily Wear soft (Hydrophilic) Contact Lens

This document describes the acceptance criteria and supporting evidence for the Ocular Sciences BIOMEDICS® 55 (ocufilcon D) UV Blocking Daily Wear soft (Hydrophilic) Contact Lens, based on the provided 510(k) summary (K984046).

1. Table of Acceptance Criteria and Reported Device Performance

The provided document primarily focuses on demonstrating substantial equivalence to previously cleared devices and outlining the optical and physical properties of the contact lens. The acceptance criteria are implicitly met by demonstrating these properties are equivalent to the predicate devices and conform to regulatory guidelines for daily wear contact lenses, particularly regarding UV blocking.

Acceptance Criteria CategorySpecific Criteria/Performance GoalReported Device PerformanceStudy/Evidence Provided
Material EquivalenceSubstantially equivalent to predicate device material (ocufilcon D)The BIOMEDICS 55 (ocufilcon D) UV Blocking Daily Wear Contact Lens with In-Monomer Tint is equivalent to the daily wear lens of the same material as cleared in K942214, P890023/S26, P890023/S4, and K972303.Comparison to predicate devices (K942214, P890023/S26, P890023/S4, K972303).
Optical PropertiesEquivalence of physical, optical, and chemical properties to clear and tinted versions of BIOMEDICS 55 (ocufilcon D) ocufilcon D cast molded contact lenses.The subject contact lens is equivalent and has the same characteristics and properties as the clear BIOMEDICS® 55 (ocufilcon D) cast molded contact lens, cleared under K942214, and P890023/S26.Comparison to predicate devices (K942214, P890023/S26).
Indication for UseDaily wear for myopic, hyperopic, or astigmatic conditions, with specific power ranges.Indicated for daily wear in normal eyes (myopic, hyperopic, or astigmatic) in powers from -20.00 to +20.00 Diopters and astigmatic corrections from -0.25 to -10.00 Diopters.Stated indications for use in the 510(k) summary.
UV BlockingUV transmittance profile comparable to or better than predicate devices and relevant human ocular structures.Demonstrated typical transmittance profiles for Ocufilcon D Hydrophilic Contact Lens with UV, including comparison to human cornea and crystalline lens.Graphs showing UV transmittance curves (pages 3, 4, 5).
Lens Group ClassificationClassified within FDA's Lens Group IV (high water ionic group).This lens is in the Lens Group IV high water ionic group as established by the FDA.Statement in the 510(k) summary, referencing FDA guidance.
Manufacturing ProcessApproval for cast molding process.The cast molding process was approved under P890023/S4.Reference to predicate device (P890023/S4).
Disinfection RegimenCompatible with chemical disinfection.The lens may be disinfected only with a chemical disinfection regimen.Stated in the Indications for Use.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not explicitly describe a separate "test set" in the context of a clinical trial or algorithm evaluation with a specific sample size. The primary evidence presented is a comparison to predicate devices and specifications, along with transmittance curves.

  • Sample Size: Not applicable in the context of a traditional test set for performance evaluation. The data provenance for the transmittance curves is not specified (e.g., how many lenses were tested to generate the "typical" profiles).
  • Data Provenance: The document does not specify the country of origin for any specific data, nor whether it is retrospective or prospective. The human cornea and lens data used for comparison in the transmittance graphs are from "24-year-old" and "25-year-old" individuals, respectively, without further detail on their origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable or provided in the context of this 510(k) summary. The summary focuses on manufacturing processes, material equivalence, and physical/optical properties, not on diagnostic accuracy tasks that would require expert-established ground truth.

4. Adjudication Method for the Test Set

Not applicable. There is no description of a test set requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done or reported in this 510(k) summary. The document is for a medical device (contact lens) and does not involve human readers evaluating cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a contact lens, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the claims in this submission is established through:

  • Predicate Device Equivalence: The primary method for establishing that the new device meets acceptance criteria is by demonstrating its substantial equivalence in materials, manufacturing processes, and intended use to previously cleared devices. The "truth" here is built upon the regulatory acceptance of those predicate devices.
  • Physical and Chemical Properties Measurement: Laboratory measurements of properties like hydrogel composition, water content, oxygen permeability (implied by material group), and UV transmittance.
  • Regulatory Grouping: Classification within FDA's established Lens Group IV.

No clinical outcomes data or pathology results are presented as ground truth for this 510(k) in the usual sense of a diagnostic device.

8. The Sample Size for the Training Set

Not applicable. This device is a contact lens, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This device is a contact lens, not an AI/ML algorithm that requires a training set.

{0}------------------------------------------------

JAN 2 1 1999

Image /page/0/Picture/1 description: The image shows a logo with the words "OCULAR SCIENCES" on the top line. Below that is a symbol that looks like a target, followed by the words "AMERICAN HYDRON" on the next two lines. The text and symbol are all in black and white.

K984046

510(k) Summary

SUBMITTER:

Submitted on behalf of:

Company Name: Address: Phone: Fax:

Ocular Sciences, Inc. 475 Eccles Av. South San Francisco, CA 95014 (415) 583-1400 (415) 583-1108

CONTACT PERSON:

Richard E. Lippman, O.D., F.A.A.O. Official Representative and Correspondent C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Suite 625 Rockville, MD 20852

DATE SUMMARY PREPARED: October 1, 1998

TRADE NAME:

BIOMEDICS® 55 (ocufilcon D) UV Blocking Daily Wear soft (Hydrophilic) Contact Lens

COMMON NAME: contact lens

SUBSTANTIALLY EQUIVALENT TO:

The BIOMEDICS 55 (ocufilcon D) UV Blocking Daily Wear Contact Lens with In-Monomer Tint is equivalent to the daily wear lens of the same material as cleared in K942214, P890023/S26, P890023/S4,and K972303 in-monomer visibility tint cast molded ocufilcon D for daily wear currently marketed by Ocular Sciences, Inc.

The BIOMEDICS 55 (ocufilcon D) UV Blocking Daily Wear contact lens is substantially equivalent to the indication for use of the BIOMEDICS 55 (ocufilcon D) In Monomer Tint ocufilcon D cast molded contact lens marketed for use in the U.S. by Ocular Sciences, Inc. cleared under K972303. Additionally, the subject contact lens is equivalent and has the same characteristics and properties as the clear BIOMEDICS® 55 (ocufilcon D) cast molded contact lens, cleared under K942214, and P890023/S26. The cast molding process was approved under P890023/S4. This lens is in the Lens Group IV high water ionic group as established by the FDA and located in the Guidance

000032

Corporate Offices:
475 Eccles Avenue
South San Francisco.
California 94080
800-628-5367
Ph: 650-583-1400

Canada:
110 Cochrane Drive
Markham, Ontario
L3R 9S1
Canada
Ph: 905-940-1998

England:
Reliant Close
Chandlers Ford
Eastleigh
Hampshire S053 4ND
England

Hungary:
Hársmajor u.8
1118 Budapest
Hungary
Ph: 36-1-319-6241
Fax: 36-1-319-3034

RICO:
P.O. Box 2801
Highway 538 km. 0.7
Santa Isabel Industrial Park
Santa Isabel,
Puerto Rico 00757

{1}------------------------------------------------

Document for Daily Wear Contact Lenses, Revised Edition May 1994. The physical, optical, and chemical properties of the BIOMEDICS® 55 (ocufilcon D) UV visibility tint cast molded Ocular Sciences, Inc.contact lens are substantially equivalent to both the clear and tinted versions of the BIOMEDICS 55 (ocufilcon D) ocufilcon D cast molded contact lenses.

DESCRIPTION of the DEVICE:

Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA and other monomeric ingredients including a proprietary ultraviolet absorber, crosslinked with EGDMA and other components which yield the appearance of lenses which are designed to fit over the corneal surface of the eye. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness), hyperopia (farsightedness) and astigmatism (multiple foci). Each lens provides corrective power which is to correspond to the refractive power of the eye to which it is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6mm. Secondary and tertiary curves as well as beveled edge configurations are built into the lens for the purpose of aiding in lens centration and comfort.

INDICATIONS FOR USE:

The BIOMEDICS® 55 (ocufilcon D) soft (hydrophilic) contact lens is indicated for daily wear in normal eyes whose eyes are either myopic, hyperopic, or astigmatic in powers from -20.00 to +20.00 Diopters and astigmatic corrections from -0.25 to -10.00 Diopters. Spherical lenses may mask astigmatism of up to 2.00 Diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the evecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.

{2}------------------------------------------------

PARAMETERS AVAILABLE:

Base Curves:6.50 mm to 10.8 mm
Diameters:12.5 mm to 18.0 mm
Powers:-20.00 to +20.00 Diopters sphere
Cylinder Power:-00.25 to -10.00 Diopters
Center Thickness:0.025 mm to 0.27 mm depending on power
Tint:Blue Visibility tint

ULTRAVIOLET TRANSMITTANCE CURVES

The following graph represents the transmittance of ultraviolet light in the UVA and UVB range as well as light transmittance through the visual spectrum, as compared to the human cornea and crystalline lens:

Image /page/2/Picture/4 description: The image shows a scattering of small, irregularly shaped black dots against a white background. The dots vary in size, with some appearing as tiny specks and others as slightly larger, more defined shapes. The distribution of the dots is uneven, with some areas of the image having a higher concentration of dots than others, creating a sense of randomness and dispersion.

Image /page/2/Picture/5 description: The image shows the number 000034 in a bold, sans-serif font. The numbers are evenly spaced and aligned horizontally. The image is simple and only contains the number.

{3}------------------------------------------------

Typical Transmittance Profile of Ocufilcon D Hydrophilic Contact Lens with UV versus a Human Cornea and a Human Lens

Image /page/3/Figure/1 description: The image is a graph comparing the percentage of light transmission through different types of lenses at different wavelengths. The x-axis represents the wavelength in nanometers, ranging from 200 to 750. The y-axis represents the percentage of transmission, ranging from 0 to 100. The graph shows the transmission of light through an Ocufilcon D contact lens with UV protection, a human cornea from a 24-year-old person, and a human crystalline lens from a 25-year-old person.

O

COOO(

{4}------------------------------------------------

Typical Transmittance Profile of -2.50 D Ocufilcon D Hydrophilic Contact Lens with UVC versus a Human Cornea and a Human Lens

Image /page/4/Figure/1 description: The image is a graph that shows the percentage of transmission of light through different materials as a function of wavelength. The x-axis represents the wavelength of light in nanometers, ranging from 200 to 750 nm. The y-axis represents the percentage of transmission, ranging from 0 to 100%. There are three lines on the graph, representing the transmission of light through Ocufilcon D contact lens, the human cornea of a 24-year-old person, and the human crystalline lens of a 25-year-old person.

COOOO

{5}------------------------------------------------

Typical Transmittance Profile of Ocufilcon D Hydrophilic Contact Lens with UVAM versus a Human Cornea and a Human Lens

Image /page/5/Figure/1 description: The image is a graph comparing the percentage of light transmission through different parts of the human eye and a contact lens. The x-axis represents the wavelength of light in nanometers, ranging from 200 to 750. The y-axis represents the percentage of light transmission, ranging from 0 to 100. The graph shows the transmission of light through the Ocufilcon D contact lens, the human cornea of a 24-year-old, and the human crystalline lens of a 25-year-old.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 1 1999

Ocular Sciences, American Hydron, Inc. c/o Richard E. Lippman, O.D., F.A.A.O. C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Rockville, Maryland 20852

Re: K984046 Trade Name: Biomedics ® 55 (ocufilcon D) UV Blocking Soft Contact Lens for Daily Wear Regulatory Class: II Product Code: 86 LPL Dated: November 6, 1998 Received: November 13, 1998

Dear Dr. Lippman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{7}------------------------------------------------

Page 2 - Richard E. Lippman, O.D., F.A.A.O.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of - Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications Statement

510(k) Number (if known) K984046

Device Name; Biomedics 55 (ocufilcon D) UV Blocking Daily Wear Contact Lens

Indications for Use:

Additional Claims:

The BIOMEDICS 55 (ocufilcon D) UV Blocking soft (hydrophilic) contact lens is indicated for daily wear in not-aphakic, normal eyes which manifest myopia, hyperopia, or astigmatism in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Spherical lenses may mask astigmatism in patients with up to 2.00 diopters that does not interfere with visual acuity. The lens may be dispensed in a daily disposable or frequent replacement scheduled program. The lenses may be prescribed for daily wear for disposal after removal in the Disposable Lens Program or for cleaning and disinfection in the Frequent Replacement Program, as recommended by the eyecare practitioner. The lens may be disinfected only with a chemical disinfection regimen.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Presription UseX
Over-The-Counter Use

(Division Sign-Off)

Division of Ophthalmic Devices

(Oprtional Format 1-2-96)

510(k) Number_K984046

00001

N/A