K Number
K983857

Validate with FDA (Live)

Device Name
AUTOVENT 1500
Date Cleared
1998-11-04

(2 days)

Product Code
Regulation Number
868.5925
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pulmonary resuscitation during respiratory and/or cardiac arrest. Short term ventilatory support for both inter and intra-hospital transport of non-breathing patients.

Device Description

The Autovent 1500 is an automatic time cycled resuscitator designed to provide respiratory support to patients in either respiratory arrest or respiratory distress. The device incorporates the features of Automatic Ventilation, Manual Ventilation, and Demand Breathing into a control module with separate hose and patient valve. This device is identical in terms of features to the Flynn Genesis product - 510K # K913039/B except in the range of automatic settings of the device which have been extended and chosen to provide the output requirements of the latest AHA recommendations.

AI/ML Overview

The document describes the Autovent 1500, an automatic time-cycled resuscitator, and its clearance based on substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the international standard for these devices, ISO 8382-1988, which the predicate device's performance data addressed. The Autovent 1500's performance was then compared against the predicate device.

Acceptance Criteria (Based on ISO 8382-1988 as demonstrated by Predicate)Reported Device Performance (Autovent 1500)
Performance within specifications as defined by ISO 8382-1988Performed within specification when compared with the predicate device.
Automatic VentilationSupported
Manual VentilationSupported
Demand BreathingSupported
Pneumatically powered, requiring 50-90 psi gas supplySupported
Time cycled ventilationSupported
Range of automatic settingsExtended and chosen to provide output requirements of latest AHA recommendations (compared to predicate)
Patient airway over pressure alarm (pre-set to 60 cmH2O) with audible signal and ventSupported

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a "test set" in the context of human data. The assessment of performance data relies on non-clinical test data.

  • Sample Size: Not applicable. The testing was non-clinical.
  • Data Provenance: The standard is "international standard for these devices - ISO 8382-1988." The testing was undertaken by the manufacturer, O-Two Systems International Inc. The document does not specify a country of origin for the test data itself, beyond the device manufacturer's location in Canada. The data is retrospective in the sense that it evaluates a completed device design against established standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. The performance assessment was based on non-clinical testing against an international standard, not on expert-established ground truth from a human dataset.

4. Adjudication Method for the Test Set

Not applicable, as the evaluation was based on non-clinical testing against a standard and comparison to a predicate device, not on expert adjudication of a test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on device performance against a standard and equivalence to a predicate, not on human reader performance with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

Yes, a standalone study was done in the sense that the device's technical specifications and performance were evaluated independently as a piece of equipment ("algorithm only" in the context of device function, not AI). The "Assessment of Performance Data" section states that the predicate device submission incorporated significant non-clinical test data to support its claim of substantial equivalence, and this new device was verified against that performance. This testing would be considered standalone evaluation of the device's functional performance characteristics.

7. The Type of Ground Truth Used

The ground truth used was technical specifications and performance characteristics defined by the international standard ISO 8382-1988, as demonstrated by the legally marketed predicate device (Flynn® Genesis® II Adult/Child - 510K # K913039/B).

8. The Sample Size for the Training Set

Not applicable. This device is not an AI/ML device that requires a training set in the typical sense. Its performance is based on its mechanical and pneumatic design.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

K983857

NOV 4 1998

SAFETY AND EFFECTIVNESS STATEMENT.

TRADE NAME: Autovent 1500

COMMON NAME: AUTOMATIC RESUSCITATOR

REGULATORY CLASS: II

TO BE SUPPLIED 2983857 510K # :

LEGALLY MARKETED DEVICE TO WHICH EQUIVALENCE IS CLAIMED:

GENESIS® II Adult/Child -

DESCRIPTION OF THE DEVICE:

The Autovent 1500 is an automatic time cycled resuscitator designed to provide respiratory support to patients in either respiratory arrest or respiratory distress.

The device incorporates the features of Automatic Ventilation, Manual Ventilation, and Demand Breathing into a control module with separate hose and patient valve.

This device is identical in terms of features to the Flynn® Genesis® product - 510K # K913039/B except in the range of automatic settings of the device which have been extended and chosen to provide the output requirements of the latest AHA recommendations.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

The device is a pneumatically powered, requiring no batteries or other power sources apart from a compressed gas supply of either air or oxygen (as specified by the customer) and a regulated output from the gas source between 50 and 90 psi.

The device is a time cycled ventilator with a range of 12, preset automatic settings in which both the tidal volume and frequency of ventilation are adjusted simultaneously by the rotation of a vertically mounted rotary control.

{1}------------------------------------------------

A manual button is provided to enable the operator to control the patient's breathing parameters. Also provided is a separate switch to select manual or automatic cycling which, in the predicate device, is integrated into the frequency/flow control.

Should the patient commence spontaneous breathing while being ventilated, their inspiratory effort will cause the device to provide breathing on demand while simultaneously causing the automatic cycling to cease. Should the patient cease to breathe spontaneously the automatic cycling will recommence without intervention by the operator being necessary.

The patient's airwav is protected by an over pressure alarm, pre-set to 60 cmHzO. which gives an audible signal of over pressure and vents the remaining tidal volume delivery for that breath to atmosphere.

ASSESSMENT OF PERFORMANCE DATA

The predicate device submission incorporated a significant amount of non-clinical test data to support the claim of substantial equivalence. This test data reviewed the performance of the device against international standard for these devices - ISO 8382-1988.

CONCLUSIONS BASED ON PERFORMANCE DATA

The testing undertaken verified that the Autovent 1500, when compared with the performance of the device to which substantial equivalence is claimed, performed within specification.

CONTACT PERSON:KEVIN BOWDEN,DIRECTOR OF PRODUCT TECHNOLOGYO-TWO SYSTEMS INTERNATIONAL INC.7575 KIMBEL STREETMISSISSAUGAONTARIOCANADAL5S 1C8TEL: 905 677 9410FAX:905 67 2035
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

E-MAIL: kevinb@otwo.com

Doc. #:SER/ATV1500/OCT1998

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services USA. The seal features the department's logo, which is a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and are connected by a flowing ribbon-like shape at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the logo.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 4 1998

Mr. Kevin Bowden O-Two Systems International Inc. 7575 Kimbel Street Mississauga, Ontario Canada L5S 1C8

Re: K983857 Autovent 1500 Regulatory Class: II (two) Product Code: 73 BTL October 30, 1998 Dated: November 2, 1998 Received:

Dear Mr. Bowden:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any bligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Kevin Bowden

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,
Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

.ﺗﺎﺳﻪ

Enclosure

{4}------------------------------------------------

Page 1 of of 1

510K Number (if known):

Device Name: AUTOVENT 1500 RESUSCITATOR

Indications for Use:

  • Pulmonary resuscitation during respiratory and/or cardiac arrest. []
  • Short term ventilatory support for both inter and intra-hospital transport of [] non-breathing patients.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OROver-The-Counter Use
(Division Sign-Off)Marle Kramer (Optional Format 1-2-96)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) NumberK983857

§ 868.5925 Powered emergency ventilator.

(a)
Identification. A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).