K Number
K983702

Validate with FDA (Live)

Date Cleared
1999-02-11

(113 days)

Product Code
Regulation Number
862.3170
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Abuscreen ONLINE Benzodiazepines is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on the Hitachi 917 analyzer at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of benzodiazepine use or abuse.

Device Description

The proposed Abuscreen ONLINE Benzodiazepines test kit is specifically intended for use on the Hitachi 917 Analyzer and future similar analyzer models. It was adapted from the currently marketed Abuscreen ONLINE Benzodiazepines test kit. The labeling and packaging have been changed for use on the Hitachi 917 Analyzer as well as an addition of a surfactant to the diluent. This modified test kit is not a replacement to the currently marketed kit. The Hitachi 917 Analyzer System is a fully automatic, computer-controlled system for clinical chemistry. It was conceived for both quantitative and qualitative in vitro determination using a large variety of tests for analysis, e.g. in serum and urine. Integrated in the system is an ion-selective unit for determination of electrolytes. The throughput per hour is 800 tests for clinical chemistry (1200 with electrolytes). The system consists of the analyzer which performs all functions required for fully automatic sample and test processing. Beginning with the automatic recording of patient samples - provided that they are supplied in barcode-labeled vessels - up to the photometric measurement and results transmission to the computer unit.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Abuscreen ONLINE® Benzodiazepines device, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The provided document details various performance characteristics used to demonstrate substantial equivalence to the predicate device. The acceptance criteria are implicitly set by matching or surpassing the predicate's performance and demonstrating acceptable levels of precision and accuracy at different cut-off concentrations.

Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Predicate K914509)Reported Device Performance (Proposed Device)
Qualitative Precision (100 ng/mL Cutoff)
% Negative at 80 ng/mL>95%>95% negative at 75 ng/mL
% Positive at 120 ng/mL>95%>95% positive at 125 ng/mL
Quantitative Precision (100 ng/mL Cutoff) - Within Run CV%
50 ng/mL6.9%2.5%
75 ng/mL3.6%1.9%
100 ng/mL3.1%1.0%
125 ng/mL2.9%1.3%
150 ng/mL1.7%1.0%
Quantitative Precision (100 ng/mL Cutoff) - Day-to-Day CV%
50 ng/mL10.1%3.2%
75 ng/mL5.0%2.1%
100 ng/mL4.1%1.7%
125 ng/mL3.1%1.6%
150 ng/mL2.7%1.5%
Qualitative Precision (200 ng/mL Cutoff)Not explicitly stated for predicate in summary>95% negative at 150 ng/mL>95% positive at 250 ng/mL
Quantitative Precision (200 ng/mL Cutoff) - Within Run CV%Not applicable for predicate100 ng/mL: 0.7%, 150 ng/mL: 1.3%, 200 ng/mL: 1.6%, 250 ng/mL: 1.8%, 300 ng/mL: 1.8%
Quantitative Precision (200 ng/mL Cutoff) - Day-to-Day CV%Not applicable for predicate100 ng/mL: 2.7%, 150 ng/mL: 1.5%, 200 ng/mL: 1.5%, 250 ng/mL: 2.6%, 300 ng/mL: 3.1%
Qualitative Precision (300 ng/mL Cutoff)Not explicitly stated for predicate in summary>95% negative at 225 ng/mL>95% positive at 375 ng/mL
Quantitative Precision (300 ng/mL Cutoff) - Within Run CV%Not applicable for predicate150 ng/mL: 1.7%, 225 ng/mL: 2.6%, 300 ng/mL: 1.9%, 375 ng/mL: 1.3%, 450 ng/mL: 1.2%
Quantitative Precision (300 ng/mL Cutoff) - Day-to-Day CV%Not applicable for predicate150 ng/mL: 2.2%, 225 ng/mL: 2.7%, 300 ng/mL: 3.1%, 375 ng/mL: 2.9%, 450 ng/mL: 2.4%
Accuracy (100 ng/mL Cutoff)N= 48 Confirmed Pos.47 Pos. 1 Neg.N= 50 Confirmed Pos.: 50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff: 10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff: 0 Pos. 10 Neg.
Accuracy (200 ng/mL Cutoff)Not applicable for predicateN= 50 Confirmed Pos.: 50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff: 10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff: 0 Pos. 10 Neg.
Accuracy (300 ng/mL Cutoff)Not applicable for predicateN= 50 Confirmed Pos.: 50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff: 10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff: 0 Pos. 10 Neg.
Limit of Detection61 ng/mL (clinical sensitivity)100 Cutoff - 13 ng/mL200 Cutoff - 16 ng/mL300 Cutoff - 28 ng/mL

Study Details

  1. Sample size used for the test set and the data provenance:

    • Precision Studies: The specific sample sizes for "Within Run" and "Day-to-Day" precision studies are provided in terms of replicates for different concentrations (e.g., 50 ng/mL, 75 ng/mL, etc.), but the total number of individual samples is not explicitly given. Each CV% calculation likely represents a set of replicates.
    • Accuracy Studies:
      • 100 ng/mL Cutoff: N=50 Confirmed Positive samples, N=10 samples diluted within 25% above cutoff, N=10 samples diluted within 25% below cutoff.
      • 200 ng/mL Cutoff: N=50 Confirmed Positive samples, N=10 samples diluted within 25% above cutoff, N=10 samples diluted within 25% below cutoff.
      • 300 ng/mL Cutoff: N=50 Confirmed Positive samples, N=10 samples diluted within 25% above cutoff, N=10 samples diluted within 25% below cutoff.
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It describes the data as "clinical and nonclinical studies performed," which typically implies prospective data collection for performance evaluation of a new device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    The document does not provide information about the number or qualifications of experts used to establish the ground truth. For drug screening tests, "confirmed positive" usually refers to confirmation by a highly sensitive and specific method like GC/MS (Gas Chromatography-Mass Spectrometry), which serves as the analytical ground truth.

  3. Adjudication method for the test set:
    The document does not describe an adjudication method for the test set. For in vitro diagnostic tests, especially for drug screening, the "ground truth" is typically the result from a definitive confirmatory method (e.g., GC/MS), not subjective interpretation by human readers requiring adjudication.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers. It is an automated assay and does not involve human readers interpreting results in the way an imaging diagnostic device might. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    Yes, this entire submission is a standalone performance evaluation of the automated assay (the "algorithm only") on the Hitachi 917 Analyzer. The device itself is an in vitro diagnostic test, which by nature operates as a standalone system to detect analytes in a sample.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    For the accuracy studies, the ground truth is referred to as "Confirmed Pos." (Confirmed Positive), which in the context of drug screening tests typically means confirmation by a highly specific and sensitive analytical method such as Gas Chromatography-Mass Spectrometry (GC/MS). This is an analytical ground truth.

  7. The sample size for the training set:
    The document does not explicitly state a "training set" sample size. For in vitro diagnostic devices like this, the development process involves reagent formulation and optimization that leverages proprietary internal data and methods, but it's not typically described in terms of a "training set" in the way machine learning algorithms are. The provided data focuses on the validation of the finalized assay.

  8. How the ground truth for the training set was established:
    As no explicit "training set" is mentioned in the machine learning sense, the method for establishing its ground truth is not described. The analytical methods used to establish the reference values for calibrators and controls used during the development and validation of the assay would serve a similar purpose to ground truth in a broader sense.

{0}------------------------------------------------

K 983702

Roche

510(k) Summary

Abuscreen ONLINE® Benzodiazepines

In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.

The assigned 510(k) number is: K983702

I. Identification of 510(k) Sponsor:

Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

510(k) Submission dated October 20, 1998

Contact: Rita Smith Senior Regulatory Affairs Associate Phone: (908) 253-7545 (908) 253-7547 Fax:

Roche Diagnostic Systems, Inc.

Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

{1}------------------------------------------------

II. Device Name:

The device name, including both the trade/proprietary name and the classification name are provided in the table below.

Product NameClassificationNameProductCodeCFRNumber andRegulatory Class
Abuscreen ONLINE forEnzyme Immunoassay,JXM862.3170
BenzodiazepinesBenzodiazepineClass II

Identification of the legally marketed device to which the 510(k) sponsor claims III. equivalence:

The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.

14able4
-------------------
Product NamePredicate ProductName510(k) Number and DatePredicate Cleared
Abuscreen ONLINE forBenzodiazepinesAbuscreen ONLINE forBenzodiazepinesK914509 11/1/91

{2}------------------------------------------------

Description of the Device/Statement of Intended Use: IV.

Abuscreen ONLINE Benzodiazepines is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of benzodiazepine use or abuse.

The proposed Abuscreen ONLINE Benzodiazepines test kit is specifically intended for use on the Hitachi 917 Analyzer and future similar analyzer models. It was adapted from the currently marketed Abuscreen ONLINE Benzodiazepines test kit. The labeling and packaging have been changed for use on the Hitachi 917 Analyzer as well as an addition of a surfactant to the diluent. This modified test kit is not a replacement to the currently marketed kit.

The Hitachi 917 Analyzer System is a fully automatic, computer-controlled system for clinical chemistry. It was conceived for both quantitative and qualitative in vitro determination using a large variety of tests for analysis, e.g. in serum and urine. Integrated in the system is an ion-selective unit for determination of electrolytes. The throughput per hour is 800 tests for clinical chemistry (1200 with electrolytes). The system consists of the analyzer which performs all functions required for fully automatic sample and test processing. Beginning with the automatic recording of patient samples - provided that they are supplied in barcode-labeled vessels - up to the photometric measurement and results transmission to the computer unit. Additional detailed information about the Hitachi 917 Analyzer is contained in volume II of the premarket notification (K953239) cleared on September 25, 1995.

V. Summary of the technological characteristics of the new device in comparison to those of the predicate.

Tables 3 outlines the technological characteristics (methodologies) of the Abuscreen ONLINE Benzodiazepines test kit in comparison to that of the legally marketed predicate product.

{3}------------------------------------------------

Brief discussion of the clinical and nonclinical tests relied on for a determination of VI. substantial equivalence:

Tables 3 demonstrates the results of clinical and nonclinical studies performed using the Abuscreen ONLINE Benzodiazepines test kit. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this chart. This information concludes that the performance of this device is essentially equivalent to the legally marketed predicate device.

{4}------------------------------------------------

ଜନ୍ମ

Abuscreen ONLINE Benzodiazepines for Hitachi 917 Table 3

Proposed:Previously Cleared: (K914509)
Abuscreen ONLINEAbuscreen ONLINE
Benzodiazepines for Hitachi 917Benzodiazepines (1000 Test Kit)
MethodologyKinetic interaction of microparticlesin a solution as measured bychanges in light transmissionKinetic interaction of microparticlesin a solution as measured bychanges in light transmission
Sample typeurineurine
Intended Usequalitative and semiquantitativedetection of benzodiazepinesqualitative detection ofbenzodiazepines
CalibratorAbuscreen ONLINE CalibrationPack or Abuscreen ONLINE BENZ200 Cal Pack or Abuscreen ONLINEBENZ 300 Cal PackAbuscreen ONLINE CalibrationPack or Abuscreen ONLINECalibrator Level 3
Cutoff(s)100, 200, 300 ng/mL100 ng/mL
Reagent (activeingredients)1. Ab reagent: benzodiazepinespolyclonal (sheep) antibody in buffer2. Microparticle reagent:Conjugated benzodiazepinederivative microparticles in buffer3. Diluent: Buffer1. Ab reagent: benzodiazepinespolyclonal (sheep) antibody in buffer2. Microparticle reagent:Conjugated benzodiazepinederivative microparticles in buffer3. Diluent: Buffer
Performance Characteristics:
Precision Qualitative ( 100 ng/mL Cutoff ):
>95% negative at 75 ng/mL>95% negative at 80 ng/mL
>95% positive at 125 ng/mL>95% positive at 120 ng/mL
Within RunMean (OD)CV%
50 ng/mL85112.4
75 ng/mL67282.6
100 ng/mL55042.6
125 ng/mL44011.6
150 ng/mL36411.4
Day-to-DayMean (OD)CV%
50 ng/mL85532.1
75 ng/mL67322.3
100 ng/mL55102.3
125 ng/mL43521.6
150 ng/mL35712.4
Proposed:Previously Cleared: (K914509)
Abuscreen ONLINEAbuscreen ONLINE
Benzodiazepines for Hitachi 917Benzodiazepines (1000 Test Kit)
Precision Qualitative ( 200 ng/mL Cutoff ):
>95% negative at 150 ng/mL
>95% positive at 250 ng/mL
Within RunMean (OD)CV%
100 ng/mL104300.7
150 ng/mL86211.3
200 ng/mL68461.6
250 ng/mL53111.8
300 ng/mL41701.8
Day-to-DayMean (OD)CV%
100 ng/mL102362.7
150 ng/mL84821.5
200 ng/mL67361.5
250 ng/mL51712.6
300 ng/mL40153.1
Precision Qualitative ( 300 ng/mL Cutoff ):
>95% negative at 225 ng/mL
>95% positive at 375 ng/mL
Within RunMean (OD)CV%
150 ng/mL95261.7
225 ng/mL69982.6
300 ng/mL54071.9
375 ng/mL41731.3
450 ng/mL33051.2
Day-to-DayMean (OD)CV%
150 ng/mL93682.2
225 ng/mL68802.7
300 ng/mL52183.1
375 ng/mL40032.9
450 ng/mL32002.4
Precision Quantitative (100 ng/mL Cutoff ):
Within RunMean (ng/mL)CV%Mean (ng/mL)CV%
50 ng/mL512.5506.9
75 ng/mL801.9813.6
100 ng/mL1011.0943.1
125 ng/mL1281.31102.9
150 ng/mL1491.01511.7
Proposed:Previously Cleared: (K914509)
Abuscreen ONLINEAbuscreen ONLINE
Benzodiazepines for Hitachi 917Benzodiazepines (1000 Test Kit)
Precision Quantitative (100 ng/mL Cutoff ):
Day-to-DayMean (ng/mL)CV%Mean (ng/mL)CV%
50 ng/mL503.25310.1
75 ng/mL802.1845.0
100 ng/mL1011.7974.1
125 ng/mL1291.61123.1
150 ng/mL1501.51542.7
Precision Quantitative ( 200 ng/mL Cutoff ):
Within RunMean (ng/mL)CV%
100 ng/mL964.4
150 ng/mL1412.7
200 ng/mL2001.9
250 ng/mL2511.5
300 ng/mL3151.2
Day-to-DayMean (ng/mL)CV%
100 ng/mL1005.3
150 ng/mL1453.3
200 ng/mL2052.1
250 ng/mL2582.2
300 ng/mL3202.0
Precision Quantitative ( 300 ng/mL Cutoff ):
Within RunMean (ng/mL)CV%
150 ng/mL1522.1
225 ng/mL2241.2
300 ng/mL3011.4
375 ng/mL3881.1
450 ng/mL4650.7
Day-to-DayMean (ng/mL)CV%
150 ng/mL1542.7
225 ng/mL2281.8
300 ng/mL3062.0
375 ng/mL3941.7
450 ng/mL4704.3

{5}------------------------------------------------

Abuscreen ONLINE Benzodiazepines for Hitachi 917 Table 3 (Continued)

{6}------------------------------------------------

Abuscreen ONLINE Benzodiazepines for Hitachi 917 Table 3 (Continued)

{7}------------------------------------------------

Accuracy
100 ng/mL CutoffN= 50 Confirmed Pos.50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff concentration10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff concentration0 Pos. 1 0 Neg.N= 48 Confirmed Pos.47 Pos. 1 Neg.
200 ng/mL CutoffN= 50 Confirmed Pos.50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff concentration10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff concentration0 Pos. 1 0 Neg.
300 ng/mL CutoffN= 50 Confirmed Pos.50 Pos. 0 Neg.N= 10 diluted within 25% above cutoff concentration10 Pos. 0 Neg.N= 10 diluted within 25% below cutoff concentration0 Pos. 1 0 Neg.
Limit of Detection100 Cutoff - 13 ng/mL200 Cutoff - 16 ng/mL300 Cutoff - 28 ng/mL61 ng/mL (clinical sensitivity)

and the same of the same of the same and the same and the same and the same and the search and

{8}------------------------------------------------

FEB 1 1 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Rita Smith Senior Regulatory Affairs Associate Roche Diagnostic Systems, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

Re: K983702 Trade Name: Abuscreen ONLINE Benzodiazepine Regulatory Class: II Product Code: JXM Dated: January 18, 1999 Received: January 19, 1999

Dear Ms. Smith:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/8/Picture/9 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird design.

{9}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

510(k) Number (if known): K983702

Device Name: Abuscreen ONLINE® Benzodiazepines

Indications for Use:

Abuscreen ONLINE Benzodiazepines is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on the Hitachi 917 analyzer at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of benzodiazepine use or abuse.

Roan Cooper

(Division Sign-Off) Division of Clinical Laboratory Devices K983702 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 862.3170 Benzodiazepine test system.

(a)
Identification. A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.(b)
Classification. Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).