IS BORRELIA BURGDORFERI IGG/IGM ELISA TEST SYSTEM

K983605 · Columbia Bioscience, Inc. · LSR · Dec 16, 1998 · Microbiology

Device Facts

Record IDK983605
Device NameIS BORRELIA BURGDORFERI IGG/IGM ELISA TEST SYSTEM
ApplicantColumbia Bioscience, Inc.
Product CodeLSR · Microbiology
Decision DateDec 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3830
Device ClassClass 2

Intended Use

For the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease. The test can be performed either manually or in conjunction with the MAGO TM PLUS Automated EIA Processor.

Device Story

The Borrelia burgdorferi IgG/IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) used to detect antibodies to B. burgdorferi in human serum. The device utilizes microtiter wells coated with partially purified B. burgdorferi antigen. Patient serum is added; if specific antibodies are present, they bind to the antigen. After incubation and washing, an enzyme-labeled anti-human immunoglobulin conjugate is added, followed by a substrate solution. The resulting color intensity is measured photometrically, providing an indirect detection of the specific antibody. The test is performed in clinical laboratory settings, either manually or using the MAGO PLUS Automated EIA Processor. Healthcare providers use the results as a presumptive diagnostic tool; positive or equivocal findings must be confirmed via standardized Western blot. This two-step testing approach supports clinical diagnosis of Lyme disease by identifying exposure to the pathogen.

Clinical Evidence

Performance evaluated using CDC-characterized serum panels (n=47) and clinical lab panels (n=72). Prospective study (n=173) compared the subject device to another commercial EIA, with positive/equivocal results confirmed by Western Blot. Subject device showed 78.6% Western Blot positivity among EIA positive/equivocal samples compared to 38.9% for the comparator. Precision studies (intra- and inter-assay) conducted across three sites showed CVs generally ranging from 4% to 20%. Correlation between manual and MAGO PLUS automated methods yielded a Pearson correlation coefficient of 0.991.

Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) using microtiter wells coated with partially purified Borrelia burgdorferi antigen. Employs enzyme-labeled anti-human immunoglobulin conjugate and substrate for photometric detection. Compatible with manual or automated (MAGO PLUS) processing. No specific material standards or software algorithm classes (e.g., ML/DL) are specified; operation is based on standard photometric absorbance measurement.

Indications for Use

Indicated for qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum for patients exhibiting signs and symptoms consistent with Lyme disease. Equivocal or positive results require supplemental Western blot testing.

Regulatory Classification

Identification

Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K98365 # 510k Summary of Safety and Effectiveness s summary of 510(k) safety and effectiveness information is being submitted in accordance with the . Juirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: ### Applicant Information: | Date Prepared: | Oct 11, 1998 | |----------------|--------------------------------| | Name: | Columbia Bioscience, Inc. | | Address: | 8775 M Centre Park Drive, #559 | | | Columbia, MD 21045 | | Contact Person: | Norman Jenkins | |-----------------|----------------| | PhoneNumber. | 410-995-0450 | | Fax Number. | 410-995-0448 | #### Device Information: | Trade Name: | Borrelia burgdorferi IgG/IgM ELISA Kit | |----------------------|----------------------------------------| | Common Name. | Borrelia burgdorferi EIA Test | | Classification Name; | Borrelia Serological Reagent | livalent Device: Jus Lyme Elisa Device Description: The & Borrelia burgdorferi IgG/IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG/IgM to Borrelia burgdorferi antigen in human serum. Intended Use: For the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia bargdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease. The test can be performed either manually or in conjunction with the MAGO TM PLUS Automated EIA Processor. Principle of Procedure: The & Borrelia burgdorferi IgG/IgM ELISA Kit is an enzymelinked immunosorbent assay to detect IgG/IgM to Borrelia burgdorferi in human serum. Partially purified Borrelia burgdorferi antigen is attached to a solid phase (microtiter well). Diluted test sera an added to each well. If antibodies which recognize the, Borrelia burgdorferi antigen are present in the patient sample they will bind to the antigen in the well. After incubation, the wells are washed to remove unbody. An enzyme labeled anti-human immunoglobulin (conjugate) the to each test well. If antibody is present the enzyme-linked antibody will bind to it. After incubation, wells are washed to remove unbound conjugate. A substrate solution is then added to each well. If enzyme {1}------------------------------------------------ is present from prior step, the reaction is stopped and the, color intensity is measured photometrically producing an indirect detection of the specific antibody present in the patient sample. ## rformance Characteristics ### 1. Clinical Sensitivity and Specificity : The lollowing information is from a panel of characterized sera obtained from the CDC (Centers for Disease Control and Prevention) r no toned by Diamedix Corp. using the Is-anti B.burgdorferi IgG/IgM Test Kit. The panel consists of 5 normal sera with . clinical diagnosis of Lyme disease and obtained at different times from onset of disease. The results are means to convey further information on the performance of the assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC. Table I illustrates the performance of the assay with this serum panel. | Elapsed Time<br>From Onset | Positive | Equiv. | Negative | Total | % Agreement | |----------------------------|----------|--------|----------|-------|-------------| | > 1 Yr | 8 | 0 | 0 | 8 | 100.0% | | 3 - 12 Months | 9 | 3 | 8 | 20 | 60.0% | | 1 - 2 Months | 4 | 1 | 4 | 9 | 55.6% | | <1 Month | 3 | 0 | 2 | 5 | 60.0% | | Negatives | 0 | 0 | 5 | 5 | 100.0% | | Total | 24 | 4 | 19 | 47 | 70.2% | Table 1 : Results of the CDC Serum Panel Stratified by Time After Onset Note that equivocal samples were considered positive for the above calculations due to the fact that all equivocal samples would be : ested by immunoblotting in a 2-step system. : following information is from a panel of characterized sera obtained from a clinical lab in Wisconsin and assayed by Diamedix o using the Is-ani B.burgdorferi IgG/IgM Test Kit. The panel consists of 72 sera with a clinical diagnosis of Lyme disease and otained at different times from onset of disease. Table 2 illustrates the performance of the assay with this serum panel. Table 2 . Results of the Characterized Lyme Sera Stratified by Time After Onset | Elapsed Time<br>From Onset | + | E | - | Total | % Agreement | |----------------------------|----|----|----|-------|-------------| | > 1 Yr | 3 | 0 | 0 | 3 | 100% | | 3 - 12 Months | 11 | 2 | 0 | 13 | 100% | | 1 - 2 Months | 8 | 2 | 3 | 13 | 76.9% | | <1 Month | 24 | 8 | 11 | 43 | 74.4% | | Total | 46 | 12 | 14 | 72 | 80.6% | Note that equivocal samples were considered positive for the above calculations due to the fact that all equivocal samples would be tested by immunoblotting in a 2-step system. {2}------------------------------------------------ ### 2. Prospective Sample Study ^ ye hundred and seventy three prospective sera from patients of various ages and gender from an endemic area that were submitted to Inical laboratory for B. burgdorferi antibody testing were assayed using the Is-anti-B. burgdorferi IgG/IgM test kit and another nmercially available EIA kit. Positive and equivocal results from both assays were supplemented by testing with a Western Blot method. Table 3, and the summary that follows, shows the prevalence of positive and equivocal results obtained in both EIAs (firststep) and percentage of these sera positive by the Western Blot method (second-step). | | | Western Blot | | |--------------|-------|--------------|-----| | | | Pos | Neg | | Is-<br>EIA | Pos | 7 | 2 | | | Equiv | 4 | 1 | | | | | | | Other<br>EIA | Pos | 11 | 12 | | | Equiv | 3 | 10 | Table 3 : Prospective Sample Study Results The results from Table 3 are summarized as follows : | Result | Is-EIA | (95% CI) | Other EIA | (95% CI) | |-------------------------------------------------|----------------|---------------|----------------|----------------| | EIA Pos. or Equiv. | 14/173 = 8.09% | (3.95-12.20%) | 36/173 = 20.8% | (14.6-27.0%) | | Pos. or Equiv.<br>Western Blot Pos. | 11/173 = 6.36% | (2.65-10.07%) | 14/173 = 8.09% | (3.95-12.24%) | | % Western Blot Pos.<br>among EIA Pos. or Equiv. | 11/14 = 78.6% | (56.6%-100%) | 14/36 = 38.9% | (22.64-55.14%) | #### 3. Precision 1 To determine the precision of the Is-ani-B burgdorferi IgGllgM Test Kit, four positive and two negative sera were assayed ten times each in three different runs at three different sites. The intra- and interassy precision obtained at each site is shown in Tables 4, 5 and (). {3}------------------------------------------------ ### TABLE 4 : Is-anti-B. burgdorferi IgG/IgM Precision Site 1 | RUN 1 | | | RUN 2 | | RUN 3 | | INTER ASSAY | | |---------|---------------|--------|---------------|--------|---------------|--------|---------------|--------| | SERUM | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | | 1 (POS) | 1.31 | 6.03% | 1.38 | 10.55% | 1.20 | 6.80% | 1.29 | 9.80% | | 2 (POS) | 1.40 | 6.86% | 1.49 | 5.03% | 1.24 | 6.31% | 1.38 | 9.62% | | 3 (POS) | 1.34 | 8.17% | 1.42 | 6.07% | 1.19 | 4.49% | 1.32 | 9.69% | | 4 (POS) | 2.26 | 5.00% | 2.36 | 5.98% | 1.82 | 7.53% | 2.15 | 12.52% | | 5 (NEG) | 0.20 | 16.26% | 0.24 | 22.30% | 0.20 | 14.76% | 0.21 | 20.68% | | 6 (NEG) | 0.16 | 14.78% | 0.17 | 21.19% | 0.20 | 15.81% | 0.18 | 19.20% | | | | | | | | CAL | 0.97 | 12.81% | | | | | | | | PC | 1.29 | 11.06% | | | | | | | | NC | 0.29 | 11.95% | | $11 = 30$ | | |-----------|---------| | PC and NC | $n = 3$ | | CAL | $n = 9$ | ### TABLE 5 : Is-anti-B. burgdorferi IgG/IgM Precision Site 2 RUN 2 #### RUN 1 RUN 3 INTER ASSAY : . | SERUM | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | |---------|---------------|--------|---------------|--------|---------------|--------|---------------|--------| | (POS) | 1.34 | 6.19% | 1.24 | 6.18% | 1.19 | 8.13% | 1.25 | 8.26% | | 2 (POS) | 1.52 | 8.80% | 1.36 | 3.51% | 1.33 | 5.18% | 1.40 | 8.55% | | 3 (POS) | 1.50 | 8.56% | 1.28 | 7.50% | 1.26 | 9.53% | 1.35 | 11.64% | | 4 (POS) | 2.60 | 4.81% | 2.30 | 7.27% | 2.27 | 6.56% | 2.39 | 8.66% | | 5 (NEG) | 0.21 | 18.40% | 0.20 | 12.31% | 0.20 | 8.72% | 0.20 | 13.38% | | 6 (NEG) | 0.15 | 9.21% | 0.14 | 14.02% | 0.15 | 16.87% | 0.15 | 13.51% | | | | | | | | CAL | 1.00 | 3.22% | | | | | | | | PC | 1.42 | 8.54% | | | | | | | | NC | 0.33 | 6.34% | n = 30 PC and NC n = 6 (`AL n = 9 {4}------------------------------------------------ #### TABLE 6 : Is-anti-B. burgdorferi IgG/IgM Precision Site 3 | | RUN 1 | | RUN 2 | | RUN 3 | | INTER ASSAY | | |---------|---------------|--------|---------------|--------|---------------|--------|---------------|--------| | SERUM | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | | 1 (POS) | 1.27 | 5.54% | 1.31 | 3.64% | 1.35 | 4.86% | 1.31 | 5.21% | | 2 (POS) | 1.42 | 4.55% | 1.44 | 6.30% | 1.49 | 6.84% | 1.45 | 6.18% | | 3 (POS) | 1.36 | 7.01% | 1.40 | 6.80% | 1.42 | 6.69% | 1.39 | 6.79% | | 4 (POS) | 2.26 | 5.64% | 2.29 | 5.58% | 2.31 | 6.77% | 2.29 | 5.91% | | 5 (NEG) | 0.22 | 15.92% | 0.22 | 12.49% | 0.23 | 18.81% | 0.23 | 15.45% | | 6 (NEG) | 0.18 | 6.84% | 0.16 | 5.67% | 0.17 | 8.49% | 0.17 | 8.70% | | | | | | | | CAL | 1.00 | 8.04% | | | | | | | | PC | 1.39 | 7.19% | | | | | | | | NC | 0.33 | 12.37% | #### n = 30 PC and NC n = 3 CAL n = 9 ### 4. Specificity with Potentially Cross-Reactive Sera To evaluate the performance of the Is-anti-B.burgdorferi IgGl1gM Test Kit with potentially cross reactive sera, a group of sera with laboratory results that may cross-react or interfere with the assay were tested. Table 7 summarizes the results obtained. ﺳﺴﺴﺴ | Laboratory Test | Lab Results | N | # equivocal | # positive | |------------------------|------------------|----|-------------|------------| | RPR + | 1:2 - 1:32 | 20 | 7 | 1 | | ds-DNA + | 52 - 1072 IU | 15 | 0 | 1 | | RF + | 245 - 338 IU | 5 | 0 | 0 | | Lipemic + | +++ | 5 | 1 | 1 | | Bilirubin + | 2.8 - 11.2 mg/dl | 5 | 0 | 0 | | Elevated ESR | 43-78 | 4 | 0 | 0 | | Elevated CRP | 4.2-22.2 mg/dl | 5 | 0 | 0 | | EBV + | + | 7 | 1 | 1 | | CMV + | 0.72 - 2.31 OD | 6 | 2 | 1 | | Rocky MT Spotted Fever | 1:64 G | 4 | 0 | 0 | #### Table 7 : Results with Potentially Cross-Reactive Sera. {5}------------------------------------------------ ### 5. Correlation of Manual and MAGO Plus Results s ls-anti-B.burgdorferi IgG/IgM Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence ne manual and MAGO Plus procedures, the results of 296 sera tested by both methods were plotted. Figure 3 illustrates the correlation between manual and MAGO Plus results. The data indicate good correlation with a Pearson Correlation Coefficient of 0.991. Image /page/5/Figure/2 description: This image is a scatter plot comparing Mago Plus Index Values to Manual Index Values. The x-axis represents Manual Index Values, ranging from 0.00 to 8.00, while the y-axis represents Mago Plus Index Values, also ranging from 0.00 to 8.00. The data points are clustered tightly along a diagonal line, indicating a strong positive correlation between the two sets of index values. The correlation coefficient, r, is given as 0.991, further supporting the strong positive relationship. Image /page/5/Figure/3 description: The image shows the title of a figure. The title is "Figure 3 : Correlation of Mago Plus and Manual Results". The title is written in a bold, sans-serif font. #### Mago Plus Precision 6. The precision of the assay when performed on the Mago Plus Automated EIA Processor was determined by assaying 6 sera 10 times each in three different runs. Table 8 shows the intra-and interassay precision obtained using the MAGO Plus. | | RUN 1 | | RUN 2 | | RUN 3 | | INTER ASSAY | | |---------|---------------|--------|---------------|--------|---------------|--------|---------------|--------| | SERUM | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | Mean<br>Index | CV | | 1 (POS) | 1.11 | 6.65% | 1.15 | 8.45% | 1.30 | 8.11% | 1.19 | 10.32% | | 2 (POS) | 1.31 | 9.82% | 1.39 | 6.30% | 1.51 | 5.80% | 1.41 | 9.26% | | 3 (POS) | 1.24 | 11.53% | 1.28 | 4.94% | 1.47 | 5.60% | 1.33 | 10.66% | | 4 (POS) | 2.13 | 5.88% | 2.29 | 5.39% | 2.40 | 4.39% | 2.27 | 7.05% | | 5 (NEG) | 0.14 | 36.89% | 0.19 | 16.64% | 0.21 | 15.06% | 0.18 | 26.90% | | 6 (NEG) | 0.11 | 28.75% | 0.13 | 37.16% | 0.18 | 23.42% | 0.14 | 35.59% | | | | | | | | CAL | 0.98 | 9.67% | | | | | | | | PC | 1.23 | 4.68% | | | | | | | | NC | 0.3 | 0.00% | TABLE 8 : MAGO Plus Is-anti-B. burgdorferi IgG/IgM Precision {6}------------------------------------------------ Image /page/6/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # DEC 1 6 1998 Norman Jenkins President Columbia Bioscience, Inc. 8775 M Centre Park Drive #559 Columbia. MD 21045 Re: K983605 > Trade Name: Is-Borrelia burgdorferi IgG/IgM ELISA Test Regulatory Class: II Product Code: LSR Dated: October 11, 1998 Received: October 14, 1998 Dear Mr. Jenkins: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {7}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ Page 1 of 1 510(k) Number: not known ## Device Name: Borrelia burgdorferi IgG/IgM ELISA Test Indications For Use: For the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western blot Positive supplemental results are supportive evidence of exposure to B. procedure. burgdorfer i and can be used to support a clinical diagnosis of Lyme disease. The test can be performed either manually or in conjunction with the MAGO TM PLUS Automated EIA Processor. ### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Y (Per 21 CFR 801.109) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ OR Over-The Counter Use (Optional Format 1-2-96) Woody Dubois on of Clinical Laboratory Devices 510(k) Number
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%