K Number
K983297

Validate with FDA (Live)

Date Cleared
1998-12-18

(88 days)

Product Code
Regulation Number
878.3610
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release in indicated for use in the treatment of esophageal obstructions produced by malignant neoplasms.

Device Description

The Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release is self-expandable metal stent which is mounted on a delivery catheter. The stent is manufactured from a nickel titanium (Nitinol) material and is comprised of a center stenting section of adjacent coils and two flared ends which enhance fixation. The delivery catheter is manufactured from a thermoplastic material with a tapered end and flexible tip. The stent is wound in a reduced diameter configuration and held in place by a restraining means. The restraining means in connected to the "slow release" handle on the proximal end of the catheter. The slow release allows the stent to be released from the catheter in a fixed and controlled speed manner.

AI/ML Overview

Here's an analysis of the provided text regarding the Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release, addressing your specific questions.

Important Note: The provided document is a 510(k) submission, which primarily focuses on demonstrating "substantial equivalence" to a predicate device, rather than detailed performance studies with acceptance criteria and statistical analysis typical of a new PMA application or extensive clinical trial. As such, many of your requested items, particularly those related to quantitative performance metrics, sample sizes for specific studies, and detailed expert ground truth processes, are not present in this document. The document states, "Testing was performed to evaluate the safety and effectiveness of the slow release mechanism. No other testing was deemed necessary for the modification of this device." This implies a focus on engineering and functional testing of the new mechanism, rather than a broad clinical efficacy study.


1. A table of acceptance criteria and the reported device performance

Based on the provided text, there are no explicit quantitative acceptance criteria or reported device performance metrics in the format of a table. The submission states that "Testing was performed to evaluate the safety and effectiveness of the slow release mechanism." However, the results of this testing, specific metrics like release speed, force profiles, or deployment success rates against numerical acceptance criteria, are not detailed in this document. The conclusion is a qualitative statement of substantial equivalence.

Acceptance Criteria (Not Explicitly Stated)Reported Device Performance (Implied)
(Likely related to controlled stent release, safety, and functional equivalence to predicate)"Testing was performed to evaluate the safety and effectiveness of the slow release mechanism." "Based on the testing performed and the nature of the change, it was concluded that the modified device is substantially equivalent to the currently marketed device."

2. Sample sized used for the test set and the data provenance

The document does not specify a sample size for any "test set" in the context of a clinical or performance study where ground truth would be established. The testing mentioned appears to be focused on the "slow release mechanism" itself, likely engineering or bench testing. No information on data provenance (country of origin, retrospective/prospective) related to patient data is provided, as no clinical study with patient data appears to be described for this specific modification.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. Given the nature of the modification (a change in the release mechanism), any "ground truth" for the testing performed would likely relate to engineering specifications or functional performance, not clinical diagnoses requiring expert consensus.

4. Adjudication method for the test set

This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not conducted or described. This device is a medical stent, not an AI diagnostic tool, so this type of study is not applicable.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

No, a standalone algorithm or AI performance study was not performed. This device is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the type of ground truth used for the testing of the "slow release mechanism." Given the "safety and effectiveness" evaluation of a mechanical feature, the ground truth would likely be based on:

  • Engineering specifications: For release speed, force, integrity of the stent upon deployment, etc.
  • Bench testing results: Demonstrating consistent and controlled release compared to design intent.
  • Predicate device comparison: Functional equivalence to the previously cleared device's release characteristics.

There is no mention of ground truth based on patient outcomes, pathology, or expert clinical consensus for this specific modification.

8. The sample size for the training set

This information is not applicable and not provided. This device does not involve a "training set" in the context of machine learning.

9. How the ground truth for the training set was established

This information is not applicable and not provided.

{0}------------------------------------------------

A.

K965297

510(k) NOTIFICATION AND SUMMARY

Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release

General Provisions
Submitter's NameMedtronic InStent
Submitter's Address6271 Bury DriveEden Prairie, Minnesota 55346
Contact PersonNoel MessengerVice President, Regulatory Affairs
Classification NameEsophageal Endoprosthesis 21 CFR, Part 876.3610
Common or Usual NameEsophageal Endoprosthesis
Proprietary NameEsophaCoil Esophageal Stent System withSlow Release

B. Name of Predicate Device

EsophaCoil Esophageal Stent System and Variable Release K941487 and K955041 respectively

C. Device Description

The Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release is self-expandable metal stent which is mounted on a delivery catheter. The stent is manufactured from a nickel titanium (Nitinol) material and is comprised of a center stenting section of adjacent coils and two flared ends which enhance fixation. The delivery catheter is manufactured from a thermoplastic material with a tapered end and flexible tip. The stent is wound in a reduced diameter configuration and held in place by a restraining means. The restraining means in connected to the "slow release" handle on the proximal end of the catheter. The slow release allows the stent to be released from the catheter in a fixed and controlled speed manner.

D. Intended Use

The Medtronic InStent EsophaCoil Esophageal Stent System with Slow Release in indicated for use in the treatment of esophageal obstructions produced by malignant neoplasms.

E. Summary of Technological Characteristics

The modified EsophaCoil Stent System with Slow Release is identical to the currently marketed EsophaCoil Stent System with Variable Release with the exception of the slow release modification. The modified slow release handle allows a fixed, controlled (slower) release sequence, where the variable release allowed the stent to be released in which ever sequence was desired by the physician without the slower control mechanism.

{1}------------------------------------------------

Non-Clinical and Clinical Test Summary F.

Testing was performed to evaluate the safety and effectiveness of the slow release mechanism. No other testing was deemed necessary for the modification of this device.

G. Conclusions

Based on the testing performed and the nature of the change, it was concluded that the modified device is substantially equivalent to the currently marketed device.

Other H.

No other information was deemed necessary for the determination of substantial equivalence of the EsophaCoil Stent System with Slow Release.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 8 1998

Mr. Noel Messenger Vice President Regulatory Affairs Medtronic InStent 6271 Bury Drive Eden Prairie, Minnesota 55346

K983297 Re:

K985277
Trade Name: EsophaCoil Esophageal Stent System with Slow Release Regulatory Class: II Product Code: ESW Dated: September 4, 1998 Received: September 21, 1998

Dear Mr. Messenger:

We have reviewed your Section 510(k) notification of intent to market the device referenced we have levice your boomer be device is substantially equivalent (for the indications for above und we have dosure) to devices marketed in interstate commerce prior to use stated in the enactment date of the Medical Device Amendments, or to devices that May 20, 1770, the chasance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Costicity for (Tree). "The general controls provisions of the Act include requirements for provisions of the Fea. Its ing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major (1 remailed if for rollar)) was device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the I was ood Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory sation. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Noel Messenger

This letter will allow you to begin marketing your device as described in your 510(k) I mis letter will anow your to ough finding of substantial equivalence of your device to a premiured predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and Compilance at (301) 59 - e., please contact the Office of Compliance at (301) 594-4639. advertising of your as result from entitled, "Misbranding by reference to premarket Also, prease flow the regareral information on your responsibilities under the nother (21 OF CF CF CF Com the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}------------------------------------------------

Page

510(k) Number (if known):___

EsophaCoil Esophageal Stent Device Name: System with Slow Release

Indications For Use:

The EsophaCoil Slow Release esophageal prosthesis inc indicated for use in the treatment of esophageal obstructions produced by malignant neoplasms.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

୦୨

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

to coela to

(Division Fon-Off Division of General 510(k) Number

§ 878.3610 Esophageal prosthesis.

(a)
Identification. An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”