Intended Use

The Apolipoprotein A1 assay is used for the quantitation of apolipoprotein A1 in human serum or plasma. Measurement of A lipoproteins, the major proteins found in HDL (high density lipoprotein), is used in the diagnosis and treatment of premature coronary artery disease, hyper-a-lipoproteinemia, and hypo-α-lipoproteinemia.

Device Description

Apolipoprotein A1 is an in vitro diagnostic assay for the quantitative determination of apolipoprotein A1 in human serum or plasma. Antibodies to apolipoprotein A1 combine with apolipoprotein A1 in the sample to form insoluble immune complexes. The immune complexes cause an increase in light scattering (turbidity). The resulting increase in sample turbidity, measured at 804 nm, is directly proportional to the concentration of apolipoprotein A1 in the sample.

AI/ML Overview

This document describes the validation of an in vitro diagnostic assay for Apolipoprotein A1.

  1. Table of Acceptance Criteria and Reported Device Performance:
Performance CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
Method CorrelationAcceptable correlation with predicate deviceCorrelation coefficient = 0.9919, slope = 0.976, Y-intercept = 6.236 mg/dL (vs. K-ASSAY® Apo A1 on Hitachi® 717 Analyzer)
Precision (Total %CV)Not explicitly stated, but expected to be low.Level 1 control: 2.1% (within-run, between-run, between-day) Level 2 control: 2.1% (within-run, between-run, between-day)
Assay RangeComparable to predicate deviceUp to 464 mg/dL
Sensitivity (LoQ)Comparable to predicate device0.525 mg/dL
  1. Sample sizes used for the test set and data provenance:

    • Test Set Sample Size: Not explicitly stated for method correlation. Precision studies were conducted using two levels of control material, but the number of unique samples or runs is not specified.
    • Data Provenance: Not explicitly stated, but typically these studies are conducted by the manufacturer at their facilities. No information on country of origin or retrospective/prospective nature is provided.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is an in vitro diagnostic device for quantitative measurement, and "ground truth" is established through comparison with a legally marketed predicate device using quantitative performance metrics rather than expert consensus on interpretive tasks.
  3. Adjudication method for the test set:

    • Not applicable. The "ground truth" for the test set is derived from the objective measurements obtained from the predicate device (for method correlation) and the device itself (for precision and sensitivity). There is no human adjudication process involved as would be for diagnostic image interpretation.
  4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical images, and the AI system acts as an assist. This submission is for an in vitro diagnostic assay with quantitative results.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance characteristics described (method correlation, precision, assay range, sensitivity) represent the standalone performance of the Apolipoprotein A1 assay on the AEROSET™ System. There is no human-in-the-loop component for the measurement and reporting of Apolipoprotein A1 levels by this automated assay.
  6. The type of ground truth used:

    • For method correlation, the "ground truth" was the quantitative results obtained from the predicate device (K-ASSAY® Apo A1 on the Hitachi® 717 Analyzer). The new device's measurements were compared against these established measurements.
    • For precision and sensitivity, the "ground truth" refers to the inherent performance characteristics of the assay itself, demonstrating its consistency and lower detection limit.
  7. The sample size for the training set:

    • Not applicable. This is a submission for an in vitro diagnostic chemistry assay, not a machine learning or AI algorithm that requires a separate training set. The assay's performance is based on its chemical and immunological principles, reagents, and instrument calibration, not on data-driven training in the sense of AI.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set in the context of a machine learning algorithm for this type of device. The assay's analytical performance is developed and optimized through standard laboratory practices, reagent formulation, and instrument design.

{0}------------------------------------------------

KG8 3289

JAN 2 9 1399

ーー

ra

Section II

510(k) Summary

Apolipoprotein A1 510(k)
September 17. 1998
ApoA_Vi.lwp

Section II

Image /page/0/Picture/6 description: The image shows a sequence of binary digits. The sequence consists of six zeros, followed by a one, and then another zero. The digits are bold and slightly distorted, giving them a textured appearance.

{1}------------------------------------------------

510(k) Summary

Submitter's Name/Address Contact Person Abbott Laboratories Linda Morris Senior Regulatory Specialist MS 1-8 1920 Hurd Drive Irving, Texas 75038 Regulatory Affairs (972) 518-6711 Fax (972) 753-3367

Date of Preparation of this Summary:Month, 1998
Device Trade or Proprietary Name:ApoA
Device Common/Usual Name or Classification Name:Apolipoprotein A1
Classification Number/Class:Class II

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _ (4832 89

Test Description:

Apolipoprotein A1 is an in vitro diagnostic assay for the quantitative determination of apolipoprotein A1 in human serum or plasma. Antibodies to apolipoprotein A1 combine with apolipoprotein A1 in the sample to form insoluble immune complexes. The immune complexes cause an increase in light scattering (turbidity). The resulting increase in sample turbidity, measured at 804 nm, is directly proportional to the concentration of apolipoprotein A1 in the sample.

Substantial Equivalence:

The Apolipoprotein A1 assay is substantially equivalent to the K-ASSAY® Apo A1 (K964292) on the Hitachi® 717 Analyzer.

Apolipoprotein A1 510(k) September 17, 1998 ApoA_VI.Iwp

Section II Page 1

Image /page/1/Picture/11 description: The image shows a sequence of numbers. The numbers are 0000011. The numbers are printed in a bold, slightly distressed font.

{2}------------------------------------------------

Both assays yield similar Performance Characteristics.

Similarities:

  • . Both assays are in vitro immunoassay methods.
  • Both assays can be used for the quantitative determination of apolipoprotein A1. .
  • . Both assays vield similar clinical results.
  • . Both assays are based on the formation of immune complexes.

Differences:

  • There is a difference between the assay range. .

Intended Use:

The Apolipoprotein A1 assay is used for the quantitation of apolipoprotein A1 in human serum or plasma.

Performance Characteristics:

Comparative performance studies were conducted using the AEROSET™ System. The Apolipoprotein A1 assay method comparison yielded acceptable correlation with the K-ASSAY Apo A1 on the Hitachi 717 Analyzer. The correlation coefficient = 0.9919, slope = 0.976, and Y-intercept = 6.236 mg/dL. Precision studies were conducted using the Apolipoprotein A1 assay. Within-run, between-run, and between-day studies were performed using two levels of control material. The total %CV for Level 1/Panel 401 is 2.1% and Level 2/Panel 402 is 2.1%. The Apolipoprotein A1 assay range is up to 464 mg/dL. The limit of quantitation (sensitivity) for the Apolipoprotein A1 assay is 0.525 mg/dL. These data demonstrate that the performance of the Apolipoprotein A1 assay is substantially equivalent to the performance of the K-ASSAY Apo A1 on the Hitachi 717 Analyzer.

Apolipoprotein A1 510(k) September 17, 1998 ADOA VI.Iwn

Section II Page 2

Image /page/2/Picture/14 description: The image shows the number 000012 in a bold, serif font. The numbers are black and appear to be printed on a white background. The number is slightly distorted, with some of the numbers appearing slightly blurred.

{3}------------------------------------------------

Conclusion:

The Apolipoprotein A1 assay is substantially equivalent to the K-ASSAY Apo A1 on the Hitachi 717 Analyzer as demonstrated by results obtained in the studies.

:"

,

Apolipoprotein A1 510(k) September 17. 1998 ApoA_V1.lwp

Section II Page 3

Image /page/3/Picture/4 description: The image shows a sequence of digits. The digits are 0000013. The numbers are printed in a bold, sans-serif font.

{4}------------------------------------------------

Public Health Service

Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three wing-like shapes, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle symbol.

JAN 29 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Linda Morris Senior Regulatory Specialist Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038

Re: K983289 Trade Name: AEROSET Apolipoprotein A1 Regulatory Class: II Product Code: DER Dated: December 3, 1998 Received: December 4, 1998

Dear Ms. Morris:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

, "

510(k) Number (if known): ---

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The Apolipoprotein A1 assay is used for the quantitation of apolipoprotein A1 in human serum or plasma. Measurement of A lipoproteins, the major proteins found in HDL (high density lipoprotein), is used in the diagnosis and treatment of premature coronary artery disease, hyper-a-lipoproteinemia, and hypo-α-lipoproteinemia.

(Division Sign-Off) ( Division of Clinical Laboratory Levices
$10(k) Number_ _ K a Clinical Laboratory Levices
$10(k) Number_ _ K a Clinical Laboratory Levices
$10(k) Number_ _ K a Clinical Labo 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use Prescription Use V OR (Per 21 CFR 801.109)

(Optional Format 1-2-96)

0

§ 866.5580

Alpha -1-lipoprotein immunological test system.(a)
Identification. Analpha -1-lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques thealpha- 1-lipoprotein (high-density lipoprotein) in serum and plasma. Measurement ofalpha- 1-lipoprotein may aid in the diagnosis of Tangier disease (a hereditary disorder of fat metabolism).(b)
Classification. Class II (performance standards).