K Number
K983217

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-01-20

(128 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rowe Dual-Check Valve device is intended for use with IV cannula or extension tube sets for simultaneous or alternate administration of IV drugs and fluids. The check valve is used in conjunction with a standard Luer taper syringe.

Device Description

The Rowe Dual-Check Valve is a disposable, inexpensive device that has been designed to infuse controlled volumes of fluid quickly. The Rowe Dual-Check Valve is a combination of two onc-way flow valves joined by a "T" fitting. The arrangement allows a syringe to be placed on the medial female luer-lock port. The syringe is then used to draw fluid from the source and then infuse the fluid into the body. The Rowe Dual-Check Valve uses one-way valves that are normally open. Therefore, when the system is placed in a normal intravenous system, gravity driven flow can pass through the device.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implied)Reported Device Performance
Pressure TestThe device must withstand specified pressures without failure."assure reliable design and performance"
Flow TestThe device must allow for specified fluid flow rates."assure reliable design and performance"
Torque TestThe device must withstand specified torques without failure or disengagement."assure reliable design and performance"
BiocompatibilityThe device must be biocompatible for its intended use (implied, per general medical device standards)."assure reliable design and performance"

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for each test. The data provenance is not explicitly stated but is implicitly from in-house testing conducted by COOK INCORPORATED. There is no indication of retrospective or prospective study design, but rather laboratory testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. The tests performed are engineering and biocompatibility tests, not clinical performance tests requiring expert assessment of ground truth.

4. Adjudication Method for the Test Set

Not applicable. The tests performed are objective measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a passive medical device (a valve) and does not involve AI or human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a passive medical device and does not involve an algorithm.

7. The Type of Ground Truth Used

The ground truth for these tests would be objective physical measurements and adherence to predefined engineering specifications and biocompatibility standards. For example:

  • Pressure Test: The ability of the device to hold specified pressure without leakage or rupture.
  • Flow Test: The measured flow rate of fluid through the device under given conditions.
  • Torque Test: The measured torque at which connections maintain integrity or fail.
  • Biocompatibility: Results from established in vitro and/or in vivo biocompatibility assays (e.g., cytotoxicity, irritation, sensitization).

8. The Sample Size for the Training Set

Not applicable. This device does not involve machine learning or a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable. This device does not involve machine learning or a "training set."

{0}------------------------------------------------

JAN 2 0 1999

K983217

510(k) Premarket Notification Rowe Dual-Check Valve COOK INCORPORATED

Safety and Effectiveness Information

Submitted By:April Lavender, RACVice President, Regulatory AffairsCOOK INCORPORATED925 South Curry PikeP.O. Box 489Bloomington, In 47402(812) 339-2235September 11, 1998
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Device:Trade Name:Proposed Classification Name:Rowe Dual-Check Valve
I.V. Connector
Predicate Device:Dual Check ValveMarketed by BurronDistributed by COOK INCORPORATED

Device Description

The Rowe Dual-Check Valve is a disposable, inexpensive device that has been designed to infuse controlled volumes of fluid quickly. The Rowe Dual-Check Valve is a combination of two onc-way flow valves joined by a "T" fitting. The arrangement allows a syringe to be placed on the medial female luer-lock port. The syringe is then used to draw fluid from the source and then infuse the fluid into the body.

The Rowe Dual-Check Valve uses one-way valves that are normally open. Therefore, when the system is placed in a normal intravenous system, gravity driven flow can pass through the device.

Indications for Use

The Rowe Dual-Check Valve is intended for use with IV cannula or extension tube sets for simultaneous or alternate administration of IV drugs and fluids. The check valve is used in conjunction with a standard Luer taper syringe.

Substantial Equivalence

The Rowe Dual-Check Valve is similar to another COOK INCORPORATED check valve which was found substantially equivalent under FDA DC #790062. The similar indications for use and technological characteristics of the Rowe Dual-Check Valve as compared to the predicate device supports a determination of substantial equivalency.

11.1

{1}------------------------------------------------

510(k) Premarket Notification Rowe Dual-Check Valve COOK INCORPORATED

Test Data

The Rowe Dual-Check Valve was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests include:

  • Pressure Test �
  • � Flow Test
  • � Torque Test
  • � Biocompatibility

The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as an I.V. connector.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 20 1999

Ms. April Lavender, RAC Vice President, Regulatory Affairs Cook Incorporated 925 South Curry Pike P.O. Box 489 47402 Bloomington, Indiana

Re: K983217 Rowe Dual-Check Valve Trade Name: Regulatory Class: II Product Code: FPA Dated: November 2, 1998 November 4, 1998 Received:

Dear Ms. Lavender:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Ms. Lavender

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K983217

2

510(k) Premarket Notification Rowe Dual-Check Valve COOK INCORPORATED

510(k) Number (if known): K983217

Device Name:

:

Rowe Dual-Check Valve

Indications for Use:

The Rowe Dual-Check Valve device is intended for use with IV cannula or extension The Rowe Dual-Check Varve device administration of IV drugs and fluids. The check valve is used in conjunction with a standard Luer taper syringe.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)


OR

Over-the-Counter Use ___________

(Division Sign-Off)B. Balder
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number1 K 983217

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.