K Number
K983197

Validate with FDA (Live)

Date Cleared
1998-10-26

(45 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The components of the Surgical Dynamics Rodding System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the confirmed by history and radiographic studies, disc spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachment are the lumbar, thoracic and cervical spine and the sacrum and illium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e. scoliosis, kyphosis, lordosis), turnors, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the SDRS is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; {c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, in the noncervical spine of skeletally mature patients, the SDRS is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Device Description

The components of the Surgical Dynamics Rodding System are indicated for spinal fixation. The material used is implant grade material that conforms to ASTM F136 and ISO 5835/3 standards for wrought Titanium alloy (Ti-6Al-4V).

AI/ML Overview

I am sorry, but based on the provided text, there is no information about acceptance criteria or a study that proves a device meets those criteria. The document is a 510(k) premarket notification for the "Surgical Dynamics Rodding System," which details the device's intended use, indications for use, and a letter from the FDA confirming its substantial equivalence to predicate devices. It does not contain any performance data, clinical study results, or specific acceptance criteria for the device's functionality or safety.

{0}------------------------------------------------

OCT 26 1998Surgical Dynamics Rodding SystemK9831971of2
IX.Summary of Safety and Effectiveness
SUBMITTER:United States Surgical Corporation150 Glover AvenueNorwalk, CT 06856
CONTACT PERSON:Sharon L. Murphy
DATE PREPARED:October 19, 1998
CLASSIFICATION NAME:Spinal Intervertebral Body Fixation OrthosisSpinal Interlaminal Fixation OrthosisSpondylolisthesis Spinal Fixation Device System
COMMON NAME:Titanium Spinal Rod System
PROPRIETARY NAME:Surgical Dynamics Rodding System
PREDICATE DEVICES:USSC's Surgical Dynamics Rodding System - K970635,K974213, K974734, K980862 (originally submitted as theAurora Titanium Spinal Rod System)
INTENDED USE:The components of the Surgical Dynamics Rodding System areindicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the confirmed by history and radiographic studies, disc spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachment are the lumbar, thoracic and cervical spine and the sacrum and illium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e. scoliosis, kyphosis, lordosis), turnors, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

{1}------------------------------------------------

Surgical Dynamics Rodding System

P. 06

K983197

2 of 2

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the SDRS is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; {c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, in the noncervical spine of skeletally mature patients, the SDRS is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

These device components are for single use. Cementless or cement fixation is not applicable to these components.

MATERIALS: The material used is implant grade material that conforms to ASTM F136 and ISO 5835/3 standards for wrought Titanium alloy (Ti-6Al-4V).]

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The eagle's head is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 6 1998

Ms. Sharon L. Murphy Regulatory Affairs Program Manager United States Surqical Corporation 150 Glover Avenue Norwalk, Connecticut 06856

Re: K983197 Trade Name: Surgical Dynamics Rodding System Regulatory Class: II Product Codes: MNI, KWP, MHN, and KWQ Dated: September 9, 1998 September 11, 1998 Received:

Dear Ms. Murphy:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಕ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Sharon L. Murphy

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}------------------------------------------------

Surgical Dynamics Rodding System

IV. Indications for Use

510(k) Number (if known): _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Surgical Dynamics Rodding System Device Name:

Indications For Use:

The components of the Surgical Dynamics Rodding System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacralliliac screws the levels of attachment are the lumbar, thoracic and cervical spine and the sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e. scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the SDRS is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the SDRS is intended to provide immobilization and stabilization of spinal seqments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: - degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use:XOR Over-The-Counter
Use:
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK9831

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.