K Number
K983176

Validate with FDA (Live)

Date Cleared
1998-09-24

(14 days)

Product Code
Regulation Number
866.5680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the in vitro quantitative determination of myoglobin in human serum and plasma. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Boehringer Mannheim immunoassay analyzers.

A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart and renal disease.

Device Description

The Elecsys ® Myoglobin test principle is based on the sandwich principal. Total duration of assay: 9 minutes, 37 ℃.
·1st incubation (4.5 minutes): By incubating the sample (15 uL) with a biotinylated monoclonal myoglobin-specific antibody (75 ul) and a monoclonal myoglobin -specific antibody labeled with a ruthenium-complex** (75 uL), a sandwich immunocomplex is formed, the amount of which is dependent upon the analyte concentration in the sample.
·2nd incubation (4.5 minutes): After addition of streptavidin-coated microparticles (35 uL) the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
· The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
·Results are determined via a calibration curve that is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.
**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)2+22)

AI/ML Overview

The device, Elecsys® Myoglobin STAT Assay, is an electrochemiluminescence immunoassay (ECLIA) intended for the in vitro quantitative determination of myoglobin in human serum and plasma, for use on Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers.

1. A table of acceptance criteria and the reported device performance

The provided document does not explicitly state "acceptance criteria" for the Elecsys® Myoglobin STAT Assay. Instead, it presents performance characteristics in comparison to a predicate device, the Tina-quant® Myoglobin Assay. The implicit acceptance criteria would be that the Elecsys® Myoglobin STAT Assay performs comparably or better than the predicate device, especially in aspects like precision, linearity, and method comparison.

Here's a table summarizing the reported device performance (Elecsys® Myoglobin STAT Assay) and the predicate device's performance:

FeatureElecsys® Myoglobin STAT (Reported Performance)Tina-quant® Myoglobin (Predicate Device Performance)
Precision
Level HS1 (43.0 ng/mL)Intra-assay: SD 0.89, %CV 2.1Intra-assay: SD 0.9, %CV 1.2
Total: SD 1.11, %CV 2.6Total: SD 1.6, %CV 2.3
Level HS4 (1147 ng/mL)Intra-assay: SD 39.5, %CV 3.4Not directly comparable (HS2 536.7 ng/mL)
Total: SD 46.3, %CV 4.0Not directly comparable (HS2 528.2 ng/mL)
Level HS5 (3056 ng/mL)Intra-assay: SD 161, %CV 5.3Not directly comparable (Control 53.1 ng/mL)
Total: SD 204, %CV 6.7Not directly comparable (Control 51.8 ng/mL)
Level PCC1 (82.5 ng/mL)Intra-assay: SD 1.03, %CV 1.3Not reported
Total: SD 1.31, %CV 1.6Not reported
Level PCC2 (672 ng/mL)Intra-assay: SD 12.5, %CV 1.9Not reported
Total: SD 15.6, %CV 2.3Not reported
Lower Detection Limit15 ng/mL3.0 ng/mL
Linearity15 - 3,000 ng/mL3.0 - 560 ng/mL
Method ComparisonVs Tina-quant MyoglobinN/A (this is the predicate for comparison)
Passing/Babloky = 1.01x - 0.13, r = 1.0, SEE = 4.54, N = 398
Least Squaresy = 1.0x + 1.28, r = 1.0, SEE = 7.58, N = 398
Interfering SubstancesNo interference at:No interference at:
Bilirubin 65 mg/dLBilirubin 60 mg/dL
Hemoglobin 1.4 g/dLHemoglobin 0.5 g/dL
Lipemia 2200 mg/dLLipemia 1500 mg/dL
Biotin 50 ng/mLNot reported

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set:
    • Precision: For each precision level (HS1, HS4, HS5 for Elecsys; HS1, HS2, Control for Tina-quant; PCC1, PCC2 for Elecsys), the sample size was N = 60 for the Elecsys device and N = 21 for the Tina-quant device. The details of the "Modified NCCLS" method likely specify the number of replicates and runs that sum up to this 'N' value.
    • Method Comparison: N = 398 samples were used for the method comparison study between the Elecsys Myoglobin STAT Assay and the Tina-quant Myoglobin Assay.
  • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the context of a 510(k) summary for a US FDA submission, it is typically expected that studies are conducted in a manner consistent with US regulatory standards, but this is not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This product is an in vitro diagnostic (IVD) assay for quantitative measurement of a biomarker (myoglobin). The "ground truth" for such assays is typically established by reference methods, independent laboratory testing, or the established values of commercially available controls and calibrators, rather than by human expert review in the sense of image interpretation. The document does not mention the use of human experts to establish ground truth. Instead, the ground truth for method comparison is the measurement obtained from the predicate device (Tina-quant Myoglobin Assay) which is itself an IVD.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As described above, the ground truth for this type of IVD is based on quantitative measurements and comparison to a predicate device, not on interpretive human expert review that would require an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a fully automated in vitro diagnostic assay, not an AI-assisted diagnostic tool that involves human readers or interpretation of medical images. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the performance data presented (precision, linearity, lower detection limit, interfering substances, and method comparison) represents the standalone performance of the Elecsys® Myoglobin STAT Assay algorithm (the entire assay system including reagents and analyzer). The device is fully automated, and the results are quantitatively generated without human interpretation or intervention in the measurement process itself.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for the method comparison study was the measurements obtained from the predicate device, the Tina-quant® Myoglobin Assay. For other performance characteristics like precision, linearity, and detection limit, the ground truth is based on the inherent analytical capabilities of the system when processing known control materials or serially diluted samples.

8. The sample size for the training set

The document does not specify a separate "training set" or its sample size. For IVD devices like this, the development process typically involves internal studies to optimize assay parameters and establish analytical performance. This often uses various samples (e.g., patient samples, spiked samples, control materials) during the development phase. However, the provided 510(k) summary focuses on the validation studies, which are akin to the "test set" in AI/ML terminology.

9. How the ground truth for the training set was established

Since a "training set" is not explicitly mentioned or quantified in the document in the context of AI/ML, the method for establishing its ground truth is not described. For traditional IVD development, the "ground truth" during development (analogous to training) would generally involve reference methods, highly characterized samples, and internal laboratory standards to ensure the assay's accuracy and reliability.

{0}------------------------------------------------

KG83176

ੋਕੇ ਮ SEP 2 4 1998

and and the state state and and and

510(k) Summary

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1. Submitter name, address, contactBoehringer Mannheim Corporation9115 Hague RoadIndianapolis, IN 46250(317) 845 - 3723
Contact Person: Priscilla A. Hamill
Date Prepared: September 08, 1998
2. Device nameProprietary name: Elecsys® Myoglobin STAT Assay
Common name: Electrochemiluminescence assay for the determination of Myoglobin.
Classification name: Myoglobin immunological test system
3. Predicate deviceThe Boehringer Mannheim Elecsys Myoglobin is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Boehringer Mannheim Tina-quant Myoglobin Assay.

continued on next page

{1}------------------------------------------------

The Elecsys ® Myoglobin test principle is based on the sandwich principal. 4.Device Description Total duration of assay: 9 minutes, 37 ℃. ·1st incubation (4.5 minutes): By incubating the sample (15 uL) with a biotinylated monoclonal myoglobin-specific antibody (75 ul) and a monoclonal myoglobin -specific antibody labeled with a ruthenium-complex** (75 uL), a sandwich immunocomplex is formed, the amount of which is dependent upon the analyte concentration in the sample. ·2nd incubation (4.5 minutes): After addition of streptavidin-coated microparticles (35 uL) the complex becomes bound to the solid phase via interaction of biotin and streptavidin. · The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame). ·Results are determined via a calibration curve that is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code. **Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)2+22)

Continued on next page

{2}------------------------------------------------

ﻣﻴﺪ ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤ

5.Intended useImmunoassay for the in vitro quantitative determination of myoglobin inhuman serum and plasma.The electrochemiluminescence immunoassay "ECLIA" is intended for use onthe Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers.
6.Comparison topredicatedeviceThe Boehringer Mannheim Elecsys® Myoglobin is substantially equivalent toother products in commercial distribution intended for similar use. Mostnotably it is substantially equivalent to the currently marketed Tina-quant®Myoglobin Assay (K972513).The following table compares the Elecsys® Myoglobin with the predicatedevice, Tina-quant® Myoglobin Assay. Specific data on the performance ofthe test have been incorporated into the draft labeling in attachment 5.Labeling for the predicate device in provided in attachment 6.Similarities:•Intended Use: Immunoassay for the in vitro quantitative determinationof Myoglobin•Sample type: Serum and plasma

Continued on next page

{3}------------------------------------------------

Comparison to predicate
device cont.

Differences:

FeatureElecsys® Myoglobin STATTina-quant® Myoglobin
Reaction testprincipleElectrochemiluminescenceImmunoturbidity
InstrumentrequiredElecsys analyzerHitachi analyzer

Performance Characteristics:

FeatureElecsys® Myoglobin STATTina-quant® Myoglobin
PrecisionLevelModified NCCLS (ng/mL)HS1 HS4 HS5Modified NCCLS (ng/mL)HS1 HS2 Control
N60 60 6021 21 21
Intra-assay43.0 1147 305673.3 536.7 53.1
SD0.89 39.5 1610.9 1.6 1.1
%CV2.1 3.4 5.31.2 0.3 2.1
Total43.0 1147 305670.8 528.2 51.8
SD1.11 46.3 2041.6 7.7 1.9
%CV2.6 4.0 6.72.3 1.5 3.7
PrecisionLevelModified NCCLS (ng/mL)PCC1 PCC2
N60 60
Intra-assay82.5 672
SD1.03 12.5
%CV1.3 1.9
Total82.5 672
SD1.31 15.6
%CV1.6 2.3
LowerDetectionLimit15 ng/mL3 ng/mL

Continued on next page

{4}------------------------------------------------

Performance Characteristics:

Comparison to predicate
device, (cont.)

FeatureElecsys® Myoglobin STATTina-quant® Myoglobin
LowerDetectionLimit15 ng/mL3.0 ng/mL
Linearity15 - 3,000 ng/mL3.0 - 560 ng/mL
MethodComparisonVs Tina-quant MyoglobinPassing/Bablok$y =1.01x - 0.13$$r = 1.0$$SEE = 4.54$$N=398$Least Squares$y =1.0x + 1.28$$r = 1.0$$SEE = 7.58$$N=398$
InterferingsubstancesNo interference at:Bilirubin 65 mg/dLHemoglobin 1.4 g/dLLipemia 2200 mg/dLBiotin 50 ng/mLNo interference at:60 mg/dL0.5 g/dL1500 mg/dL

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle.

SEP 2 4 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Priscilla A. Hamill Regulatory Affairs Consultant Boehringer Mannheim Corporation 9115 Haque Road P.O. Box 50457 Indianapolis, Indiana 46250-0457

Re: K983176 Elecsys® Myoglobin STAT Regulatory Class: II Product Code: DDR September 8, 1998 Dated: Received: September 10, 1998

Dear Ms. Hamill:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

14983176 510(k) Number (if known): N/A

Device Name: Elecsys® Myoglobin STAT Assay

Indications For Use:

For the in vitro quantitative determination of myoglobin in human serum and plasma. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Boehringer Mannheim immunoassay analyzers.

A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart and renal disease.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K983176

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter Use

(Optional Format 1-2-96)

§ 866.5680 Myoglobin immunological test system.

(a)
Identification. A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.(b)
Classification. Class II (performance standards).