K Number
K983031

Validate with FDA (Live)

Date Cleared
1998-12-18

(109 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITROS AFP Reagent Pack - For the in vitro quantitative measurement of alpha-fetoprotein (AFP) in human serum to aid in the management of patients with non-seminomatous testicular cancer. VITROS AFP Calibrators - For in vitro use in the calibration of the VITROS Immunodiagnostic System for the quantitative measurement of AFP in human serum.

Device Description

The VITROS Immunodiagnostic System uses luminescence as the signal in the quantitative and semi-quantitative determination of selected analytes in human body fluids, commonly serum. Coated microwells are used as the solid phase separation system. The system is comprised of three main elements: The VITROS Immunodiagnostic Products (in this case VITROS Immunodiagnostic Products AFP Reagent Pack, VITROS Immunodiagnostic Products AFP Calibrators, which are combined by the VITROS Immunodiagnostic System to perform the VITROS AFP assay). The VITROS Immunodiagnostic System - instrumentation, which provides automated use of the immunoassay kits. Common reagents used by the VITROS System in each assay.

AI/ML Overview

Here's an analysis of the provided text regarding the VITROS AFP assay's acceptance criteria and the study proving its performance:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in the typical "pass/fail" format. Instead, it demonstrates "substantial equivalence" to a predicate device. The performance is primarily shown through a correlation study and analytical performance characteristics.

Acceptance Criterion (Implicit)Reported Device Performance (VITROS AFP assay)
Substantial Equivalence to Predicate Device (Abbott AxSYM AFP)Correlation: VITROS AFP assay = 0.889 x [AxSYM AFP assay] + 2.60 (IU/mL) with a correlation coefficient of 0.991. Clinical Utility (Serial Monitoring): "The serial monitoring study demonstrated the clinical utility of the VITROS AFP assay as an aid in the management of patients with non-seminomatous testicular cancer." Distribution of Results: "In clinical studies of apparently healthy individuals, patients with cancer and patients with a variety of non-malignant diseases, the VITROS AFP assay exhibited distribution results that parallel expected distributions for these patient types."
Analytical Sensitivity (Specificity, Precision, Dilution)"tests were performed to obtain analytical sensitivity, specificity, precision, dilution and expected values. Refer to the VITROS AFP assay package insert for VITROS AFP assay results." (Specific values are not provided in this summary.)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document mentions "samples from serially monitored patients diagnosed and treated for testicular cancer" and "patient specimens covering the normal, therapeutic and diagnostic range." It also refers to "clinical studies of apparently healthy individuals, patients with cancer and patients with a variety of non-malignant diseases."
    • Specific sample sizes are NOT provided in this summary.
  • Data Provenance: The data appears to be retrospective as it compares the new device results with those from an already approved predicate device using existing "patient specimens." The country of origin is not specified, but the submission is to the US FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This is a lab assay, not a diagnostic imaging device that typically uses expert readers for ground truth.

  • Not applicable in the context of this in vitro diagnostic (IVD) assay for AFP levels. The "ground truth" for an IVD assay typically refers to the accuracy of its measurement of the analyte. In this case, the predicate device (Abbott AxSYM System AFP) served as the reference standard for the comparison.

4. Adjudication Method for the Test Set:

  • Not applicable for this type of IVD assay. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) where there can be disagreement among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC study was not done. This type of study is relevant for AI-assisted diagnostic tools where human readers are involved. The VITROS AFP assay is an automated in vitro diagnostic test.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, in essence, the study is a standalone performance assessment. The VITROS AFP assay is an automated system designed to quantitatively measure AFP levels. Its performance is evaluated directly against a predicate device without human interpretation being the primary variable. The results are compared directly to the predicate device's output.

7. The Type of Ground Truth Used:

  • The primary "ground truth" or reference standard used for comparison was the predicate device, the Abbott AxSYM System AFP assay. The performance of the VITROS AFP assay was evaluated based on its correlation and agreement with readings obtained from this established, FDA-approved device. Additionally, the clinical utility was assessed through "serial monitoring" of actual cancer patients, implying clinical outcomes were considered in relation to AFP levels.

8. The Sample Size for the Training Set:

  • Not applicable/Not explicitly stated. This document describes a traditional medical device (an immunoassay kit and system), not a machine learning or AI-driven device that requires a distinct "training set" in the common AI sense. The development of such assays involves extensive R&D and calibration, but not typically a separate "training set" as defined for AI models.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable/Not explicitly stated. As this is not an AI/ML device, the concept of a training set ground truth doesn't directly apply in the same way. The assay's analytical characteristics (calibration, sensitivity, etc.) are established through laboratory procedures and reference materials.

{0}------------------------------------------------

VITROS AFP assay

ADMINISTRATIVE 1

510(k) Summary 1.1

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is: 1983031

1. Submitter name, address, contact

Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, New York 14626-5101 (716) 453-3607

Contact Person: Charles Morganson

Date 510(k) prepared: August 27, 1998

2. Device Name

Trade or Proprietary Name: VITROS Immunodiagnostic Products AFP assay Common Name: Alpha-Fetoprotein assay Classification Name: Alpha-Fetoprotein test kit for testicular cancer

3. Predicate Device

The VITROS Immunodiagnostic Products AFP assay is substantially equivalent to the Abbott AxSYM System AFP (P820060S8).

4. Device Description

The VITROS Immunodiagnostic System uses luminescence as the signal in the quantitative and semi-quantitative determination of selected analytes in human body fluids, commonly serum. Coated microwells are used as the solid phase separation system.

The system is comprised of three main elements:

The VITROS Immunodiagnostic Products (in this case VITROS Immunodiagnostic Products AFP Reagent Pack, VITROS Immunodiagnostic Products AFP Calibrators, which are combined by the VITROS Immunodiagnostic System to perform the VITROS AFP assay).

The VITROS Immunodiagnostic System - instrumentation, which provides automated use of the immunoassay kits. The VITROS Immunodiagnostic System was cleared for market by a separate 510(k) pre-market notification (K962919).

{1}------------------------------------------------

Common reagents used by the VITROS System in each assay. The VITROS Immunodiagnostic Products Signal Reagent and VITROS Immunodiagnostic Products Universal Wash Reagent were cleared as part of the VITROS Immunodiagnostic Products Total T3 510(k) pre-market notification (K984310).

The VITROS System and common reagents are dedicated specifically only for use with the VITROS Immunodiagnostic Products range of immunoassay products.

5. Device Intended Use

The VITROS AFP assay is intended for the in vitro quantitative measurement of albhafetoprotein (AFP) in human serum, to aid in the management of patients with non-seminomatous testicular cancer.

6. Comparison to Predicate Device

The VITROS Immunodiagnostic Products AFP assay is substantially equivalent to Abbott AxSYM System AFP (predicate device), which was approved by FDA (P820060S8) for IVD use.

The relationship between the VITROS AFP assay and the predicate device, determined by Deming's Regression, is:

VITROS AFP assay = 0.889 x [AxSYM AFP assay] + 2.60 (IU/mL), with a correlation coefficient of 0.991.

Comparisons of the VITROS AFP assay and the predicate device were performed with samples from serially monitored patients diagnosed and treated for testicular cancer.

In addition to the studies mentioned above, tests were performed to obtain analytical sensitivity, specificity, precision, dilution and expected values. Refer to the VITROS AFP assay package insert for VITROS AFP assay results.

Table 1 lists the similarities and differences of the device characteristics between the VITROS AFP assay with the predicate device. Abbott AxSYM AFP assay:

{2}------------------------------------------------

Table 1 List of the assay characteristics

DeviceCharacteristicVITROS AFP assayPredicateDevice
Calibration range0 - 500 IU/mL (1st IRP72/225)0 - 520 ng/mL0 - 350 ng/mL
Basic principleSolid phase immunoassaySolid phase immunoassay
TracerEnzyme labeledEnzyme labeled
InstrumentationVITROSImmunodiagnostic SystemAbbott AxSYM System
Sample typeSerumSerum, plasma (heparin, citrate or EDTA)
Antibody1) Sheep polyclonal anti-AFP antibody inbiotinylated antibodyreagent2) Mouse monoclonal anti-AFP antibody in conjugatereagent1) mouse monoclonalanti-AFP antibody coatedonto microparticles2) mouse monoclonalanti-AFP antibody inconjugate
Sample volume25 µL150 µL
Incubation time andtemperatureFirst incubation 16minutes at 37°C withshakingSecond incubation 16minutes at 37°C withshakingDetails not listed inpackage insert

7. Conclusions

The data presented in the premarket notification demonstrate that the VITROS AFP assay performs substantially equivalent to the predicate device, for which there is an approved PMA.

Equivalence was demonstrated using currently commercially available reagents along with patient specimens covering the normal, therapeutic and diagnostic range. In clinical studies of apparently healthy individuals, patients with cancer and patients with a variety of non-malignant diseases, the VITROS AFP assay exhibited distribution results that parallel expected distributions for these patient types.

The serial monitoring study demonstrated the clinical utility of the VITROS AFP assay as an aid in the management of patients with non-seminomatous testicular cancer.

The data presented in the premarket notification provide a reasonable assurance that the VITROS AFP assay is safe and effective for the stated intended use.

{3}------------------------------------------------

2098 Gaither Road Rockville MD 20850

Food and Drug Administration

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of an eagle or bird, with three curved lines forming its body and wings. The logo is in black and white.

DEC 18 1998

Ortho-Clinical Diagnostics, Inc. C/O Mr. Charles Morganson, Jr. Requlatory Affairs Associate Requlatory Affairs MC00882 100 Indigo Creek Drive Rochester, New York 14626

K983031 Re:

Vitros Immunodiagnostic Products AFP Reagent Pack Trade Name: Regulatory Class: Product Code: LOJ Dated: Auqust 27, 1998 Received: August 31, 1998

Dear Mr. Morqanson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure -

{5}------------------------------------------------

1 .3 Indications For Use Statement

Page 1 of 1

510(k) Number (if known):K983031
Device Name:VITROS Immunodiagnostic Products AFP Reagent PackVITROS Immunodiagnostic Products AFP Calibrators
Indications for Use:VITROS AFP Reagent Pack - For the in vitro quantitativemeasurement of alpha-fetoprotein (AFP) in human serum to aidin the management of patients with non-seminomatous testicularcancer.
VITROS AFP Calibrators - For in vitro use in the calibration ofthe VITROS Immunodiagnostic System for the quantitativemeasurement of AFP in human serum.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK983031
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use

(Optional Format 1-2-96)

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.