K Number
K982957

Validate with FDA (Live)

Date Cleared
1999-01-13

(142 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination

Device Description

SENSIFLEX meet the requirement for surgeon's glove described by the American Standard for Testing and Material ASTM D3577 as Type 1, white in color and non-powdered. Sizes available is from 6 - 9.

AI/ML Overview

The provided text describes the 510(k) summary for the SENSIFLEX Powderfree Surgeon's Glove. Here's a breakdown of the requested information:

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (Standard)Device PerformanceOutcome
ASTM D3577 (Type 1 surgeon's glove requirements)Meets ASTM D3577Meets
ASTM D512Meets ASTM D512Meets
ASTM F 719-81 (Primary Skin Irritation in Rabbit)No irritationMeets
ASTM F 720 - 81 (86) (Dermal Sensitization Test)No sensitizationMeets
Iodine test for starch-free statusProduct is iodine tested for starch free status.Meets

2. Sample size used for the test set and the data provenance:

  • Sample size: Not explicitly stated. The document mentions "Tests conducted per ASTM D3577, ASTM D512," implying that the testing was done on a sample, but the specific number of gloves tested is not provided.
  • Data provenance: Not explicitly stated, though the tests were conducted by a Malaysian company (Terang Nusa Sdn Bhd) for a device manufactured in Malaysia. It is prospective testing as it was done to demonstrate substantial equivalence for a new device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the described tests are non-clinical, performance, and biocompatibility tests, not studies requiring expert interpretation of results to establish ground truth (e.g., medical image analysis). The ground truth for these tests is established by the specifications defined in the ASTM standards.

4. Adjudication method for the test set:

This information is not applicable for the type of non-clinical and biocompatibility tests described. Adjudication methods are typically used in clinical studies or studies involving human readers/interpreters.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, an MRMC comparative effectiveness study was not done. This device is a surgeon's glove, not an AI-powered diagnostic tool, so such a study would not be relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

No, a standalone algorithm performance study was not done. This device is a physical product (surgeon's glove), not an algorithm.

7. The type of ground truth used:

The ground truth used for the non-clinical tests is based on established industry standards and specifications (ASTM D3577, ASTM D512, ASTM F 719-81, ASTM F 720 - 81 (86)) and the requirement for starch-free status. For biocompatibility, the ground truth is the absence of irritation and sensitization.

8. The sample size for the training set:

Not applicable. This device is a physical product (surgeon's glove), not an AI model, so there is no "training set."

9. How the ground truth for the training set was established:

Not applicable. As there is no training set, there is no ground truth for it.

{0}------------------------------------------------

JAN 1 3 1999

Image /page/0/Picture/1 description: The image shows a black and white drawing of a compass rose. The compass rose has eight points, with each point being a star shape. There is a faint outline of a shape that looks like a building in the center of the compass rose. The image is simple and clear, with the compass rose being the main focus.

TERANG NUSA Sdn Bhd

510(k) Summary for SENSIFLEX Powderfree Surgeon's Glove

K982957

510(k) Summary

Submitter NameTerang Nusa Sdn Bhd
Submitter Address1, Jalan 8,Pengkalnn Chepa 2 Industrial Zone16100 Kota Bharu,Kelantan, Malaysia
Submitter Telephone+60 9 7735133
Submitter Fax+60 9 7737755
Contact PersonLOW, Chin Guan
Date of preparation15 Aug 98
Trade NameSENSIFLEX
Common NameSurgeon's Glove
ClassificationSurgeon's Glove
Legally marketed device to whichsubstantial equivalence is beingclaimed.The SENSIFLEX powderfree surgeon's glovedescribed in this 510(k) is substantially equivalentto the powdered SUR-G GLOV currently beingmarketed. The difference between the two is theadditional manufacturing process that removes thepowder from the glove.
Description of deviceSENSIFLEX meet the requirement for surgeon'sglove described by the American Standard forTesting and Material ASTM D3577 as Type 1,white in color and non-powdered. Sizes availableis from 6 - 9.
Intended Use of the deviceThese surgeon's gloves are to be worn by healthcare workers or similar personnel during procedures or in any work to prevent cross contamination between the user and the patient.
Summary of technologicalcharacteristics compared topredicate deviceThis notification describes the minor changes to the manufacturing process that removes the powder from the device.
Brief description of non-clinicaltestsTest conducted per ASTM D3577, ASTM D512 indicates that the product meet the requirements.Primary Skin Irritation in RabbitTest ASTM F 719-81 and Dermal Sensitization Test ASTM F 720 - 81 (86) indicates no sensitization or irritation.Final product is iodine tested for starch free status.
Brief description of clinical testsNot carried out
Conclusion drawn from clinical andnon clinical testsNon-clinical and bio-compatibility test indicate device meet all performance and bio-compatibility requirements.
Additional information deemednecessary by the FDANone

{1}------------------------------------------------

q82957

Image /page/1/Picture/1 description: The image shows a black and white drawing of a compass rose. The compass rose has eight points, with each point being a solid black color. There are some dashed lines around the compass rose, which may be part of the design. The image is simple and clear, with the compass rose being the main focus.

TERANG NUSA Sdn Bhd

510(k) Summary for SENSIFLEX Powderfree Surgeon's Glove

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human figures, possibly representing health, family, and community. The figures are depicted in a flowing, abstract manner.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 3 1999

Mr. Chin Guan Low Managing Director Terang Nusa Sdn. Bhd. 1. Jalan 8 Pengkalnn Chepa 2 Industrial Zone 16100 Kota Bharu, Kelantan, MALAYSIA

K982957 Re : Sensiflex Powder-Free Latex Surgeon's Gloves Trade Name: Requlatory Class: I Product Code: KGO November 9, 1998 Dated: November 12, 1998 Received:

Dear Mr. Chin Guan Low:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Chin Guan Low

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Timo II atows Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Figure/3 description: The image shows a logo with a compass-like symbol on the left and the text "TERANG NUSA Sdn Bhd" on the right. Below the text, it says "510(k) Submission for Surgeon's Gloves". The text is in a simple, bold font, and the overall design is clean and professional.

510(k) NumberK 982957
Device NameSENSIFLEX Powderfree Latex Surgeon's Glove

Indication For Use :

A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination

Concurrence of CDHR Office of Device Evaluation (ODE)

George 1. Weible for Chris S. Zin, Ph D

510(k) Number _

Prescription Use
(Per 21 CFR 801.109)

OR

Over the counterX
---------------------

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).