K Number
K982761

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-10-22

(77 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Joint Motion Devices are an accessory to MRI diagnostic devices, including the Hitachi AIRIS™, Siemens Magnatom™, Toshiba OPART™ Picker Outlook™, Fonar, and GE Medical Systems Open MRI systems. In addition, the devices may be used in any MRI system where the physical limits of the bore size is not a constraint.

The devices are built from plastics and nonferromagnetic materials, and hold patients securely during MRI imaging sequences. The devices also constrain joint motion to one of several user-designated planes of rotation, dependant upon the joint to be imaged, and emulate the natural motion of the joint.

The Joint Motion Devices increase the clinical utility of MRI systems to provide high quality, repeatable imaging and constraint for the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies to enable fixed position imaging, and kinematic (incremental stepped) motion of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, and dynamic (continuous) motion of the knee, cervical spine, ankle, wrist , lumbar spine, hip and shoulder joint anatomies.

The devices will enable the physician to evaluate the anatomy of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, in a static mode as well as to evaluate the dynamic interaction of the different tissues (ligaments, cartilage, bone, muscle, fat). Such functional interaction may be useful in diagnosis.

Device Description

The Joint Motion Devices provide constraint for the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies to enable fixed position imaging, and kinematic (incremental stepped) motion of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, and dynamic (continuous) motion of the knee, cervical spine, ankle, wrist , lumbar spine, hip and shoulder joint anatomies. Form, fit and function, as well as a determination that the devices do not adversely affect the magnetic field for imaging has been tested.

The Joint Motion Devices add clinical utility to all known open MRI diagnostic devices, including the Hitachi AIRISTM, Siemens Magnatom™, Toshiba OPART™, and Picker, Fonar, Phillips and GE Medical Systems Open MRI systems. In addition, several of the devices have use in higher tesla magnets where bore size is not a constraint.

The devices are manually operated by the user. No active components (i.e. motors) are used and there are no ferromagnetic materials that could affect the scan field.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called "Joint Motion Devices." It does not contain information about acceptance criteria or a study proving that the device meets those criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device, which is a different regulatory pathway than proving performance against specific acceptance criteria.

The key points from the document are:

  • Device: Joint Motion Devices (for various joints like knee, cervical spine, ankle, wrist, lumbar spine, hip, and shoulder)
  • Intended Use: To provide constraint for specific anatomies during MRI imaging for fixed-position, kinematic (incremental stepped), and dynamic (continuous) motion studies. This allows physicians to evaluate anatomy in static mode and dynamic interaction of tissues.
  • Predicate Device: K953918, manufactured by CHAMCO, Inc. (the same company submitting this 510(k)). The submission states the devices are identical to the predicate device.
  • Regulatory Pathway: 510(k) premarket notification, which emphasizes substantial equivalence to a legally marketed predicate device rather than de novo demonstration of safety and effectiveness against explicit performance criteria.

Therefore, I cannot provide the requested information because the provided text describes a 510(k) submission based on substantial equivalence, not a study demonstrating adherence to specific acceptance criteria.

Here's why the specific questions cannot be answered from the provided text:

  1. Table of acceptance criteria and reported device performance: This document does not establish specific acceptance criteria for performance; it relies on the predicate device's established safety and effectiveness.
  2. Sample size, data provenance: No new study data is presented, as the claim is one of identity to a predicate device.
  3. Number of experts, qualifications: Not applicable, as there's no independent ground truth establishment for a new study.
  4. Adjudication method: Not applicable.
  5. MRMC comparative effectiveness study: Not conducted or reported.
  6. Standalone performance study: Not conducted or reported in this document. The device is an accessory to MRI diagnostic devices.
  7. Type of ground truth: Not applicable, as no new study is detailed.
  8. Sample size for training set: Not applicable, as no AI/machine learning model is being described or validated.
  9. How ground truth for training set was established: Not applicable.

{0}------------------------------------------------

OCT 22 1998

. . . . . . . .

Image /page/0/Picture/1 description: The image shows the word "CHAMCO" in a bold, serif font, with a five-pointed star above the letters "HA". To the right of the word "CHAMCO" is the abbreviation "INC.", which is smaller than the main word. The star is black and solid, and the letters are also black. The overall design is simple and corporate.

American Made Medical Products

510(k) Summary

长982761

1. Submitter Information

CHAMCO, Inc. 1.1 Submitter: 798 Clearlake Road Cocoa, Fl 32922 Owner/operator number: 9034336 David P Salvadorini 1.2 Contact: PH: 407-774-4488 FX: 407-774-0599 email: dsalvadorini@compuserve.com

  • August 4, 1998 1.3 Date:

2.0 Device Name

2.1Classification Panel:Radiology
2.2Classification Number:892.100 Magnetic Resonance Diagnostic Device
2.3Product Number:90LNH
2.4Product Nomenclature:System, Nuclear Magnetic Resonance Imaging, Accessory
2.5Device Classification:II
2.6Performance Standards:None established under section 514 of the Food, Drug and Cosmetic Act
2.7Trade Name:Joint Motion Devices
2.7.2 Knee/Cervical Joint Motion Device2.7.3 Lumbar Joint Motion Device2.7.4 Wrist Joint Motion Device2.7.5 Ankle Joint Motion Device2.7.6 Shoulder Joint Motion DeviceP/N 1026P/N 1027P/N 1028P/N 1029P/N 1030

798 Clearlake Road • Cocoa, Florida 32922 • Phone 407.6.39.3314 • Fax 407.6.39.877.4

E -

{1}------------------------------------------------

2.7.7 Hip Joint Motion DeviceP/N 1031
2.7.8 Joint Motion Device Package*P/N 1032
2.7.9 Storage CabinetP/N 1033

*Includes all devices plus the storage cabinet

3.0 Predicate Device

3.2

  • 3.1 K953918
    Hitachi Medical Systems America, Inc. Manufactured by CHAMCO. Inc.

  • 3.3 Sept. 12, 1995

  • 3.4 Regulatory Class II

  • 3.5 21 CFR 892.1000/Procode: 90LNH

Joint Motion Devices

  • These devices are identical to the predicate device. CHAMCO and Hitachi 3.6 Medical Systems America, developed these devices together. CHAMCO has been the only manufacturer of these devices. CHAMCO intends to market these devices under its own name, and has adequate information demonstrating its legal right to distribute these devices.

4.0 Device Description

  • 4.1 Function
    The Joint Motion Devices provide constraint for the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies to enable fixed position imaging, and kinematic (incremental stepped) motion of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, and dynamic (continuous) motion of the knee, cervical spine, ankle, wrist , lumbar spine, hip and shoulder joint anatomies. Form, fit and function, as well as a determination that the devices do not adversely affect the magnetic field for imaging has been tested.

The Joint Motion Devices add clinical utility to all known open MRI diagnostic devices, including the Hitachi AIRISTM, Siemens Magnatom™, Toshiba OPART™, and Picker, Fonar, Phillips and GE Medical Systems Open MRI systems. In addition, several of the devices have use in higher tesla magnets where bore size is not a constraint.

  • 4.2 Scientific Concepts
    Identical to the predicate device

{2}------------------------------------------------

4.3 Physical and Performance Characteristics

Identical to the predicate device

5.0 Device Intended Use

Identical to the predicate devices will enable the physician to evaluate the anatomy of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, in a static mode as well as to evaluate the dynamic interaction of the different tissues (ligaments, cartilage, bone, muscle, fat). Such functional interaction may be useful in diagnosis.

6.0 Technological Characteristics

Identical to the predicate device. The devices are manually operated by the user. No active components (i.e. motors) are used and there are no ferromagnetic materials that could affect the scan field.

{3}------------------------------------------------

Image /page/3/Picture/16 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, which is a symbol of medicine and healing. The caduceus is composed of three human profiles.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 2 1998

David P. Salvadorini Director of Marketing Chamco. Inc. 798 Clearlake Road Cocoa, FL 32714

Re:

K982761 Joint Motion Devices Package for MRI Dated: August 4, 1998 Received: August 6, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH

Dear Mr. Salvadorini:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for-use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device; please contact the Office of Compliance at (301) 594-4639. Also, please mote the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html",

Sincerely yours,

William Yip

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Ver/3- 4/24/96Indications for Use Statement
Applicant:CHAMCO, Inc.
798 Clearlake Road
Cocoa, Fl 32922
Owner/operator number: 9034336
David P Salvadorini
PH: 407-774-4488
FX: 407-774-0599
email: dsalvadorini@compuserve.com
510 (k) Number:K982761
Device Name:Joint Motion Device
Knee/Cervical Joint Motion DeviceP/N 1026
Lumbar Joint Motion DeviceP/N 1027
Wrist Joint Motion DeviceP/N 1028
Ankle Joint Motion DeviceP/N 1029
Shoulder Joint Motion DeviceP/N 1030
Hip Joint Motion DeviceP/N 1031
Joint Motion Device Package*P/N 1032
Storage CabinetP/N 1033

*Includes all devices plus the storage cabinet

Indications for Use:

The Joint Motion Devices are an accessory to MRI diagnostic devices, including the Hitachi AIRIS™, Siemens Magnatom™, Toshiba OPART™ Picker Outlook™, Fonar, and GE Medical Systems Open MRI systems. In addition, the devices may be used in any MRI system where the physical limits of the bore size is not a constraint.

The devices are built from plastics and nonferromagnetic materials, and hold patients securely during MRI imaging sequences. The devices also constrain joint motion to one of several user-designated planes of rotation, dependant upon the joint to be imaged, and emulate the natural motion of the joint.

The Joint Motion Devices increase the clinical utility of MRI systems to provide high quality, repeatable imaging and constraint for the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies to enable fixed position imaging, and kinematic (incremental stepped) motion of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, and

{5}------------------------------------------------

dynamic (continuous) motion of the knee, cervical spine, ankle, wrist , lumbar spine, hip and shoulder joint anatomies.

The devices will enable the physician to evaluate the anatomy of the knee, cervical spine, ankle, wrist, lumbar spine, hip and shoulder anatomies, in a static mode as well as to evaluate the dynamic interaction of the different tissues (ligaments, cartilage, bone, muscle, fat). Such functional interaction may be useful in diagnosis.

The intended patients are primarily drawn from a pool of those subjects undergoing diagnostic evaluation by physicians who are skilled in diagnosis and treatment of the disease process(es) under consideration.

David C. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devi 510(k) Number

Concurrence of CDRH, Office of Device Evaluation, (ODE)

Prescription use

(Per 21 CFR 801.109) (Optional Format 1-2-96)

D-2

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.