K Number
K982580

Validate with FDA (Live)

Date Cleared
1998-10-22

(90 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bird Monitor Interface (BMI) is a stand-alone unit providing a means of communicating information from Bird ventilators to patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin and SpaceLab's PDMS are provided. as well as a Bird proprietary interface for use with other systems. The purpose of BMI is to deliver ventilator settings, monitored values, alarm information and waveform data to the patient monitoring system.

Device Description

The Bird Monitoring Interface (BMI) is a stand-alone unit providing a communications interface between the Bird ventilators, 8400STi, VIP and TBird, and patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin (via Hewlett-Packard's Vuelink proprietary interface) and SpaceLab's PDMS (via Spacelab's Universal Flexport proprietarv interface) are provided as well as a Bird proprietary interface, the General Purpose Computer Protocol (GPCP), for use with other systems. The purpose of the BMI is to deliver ventilator settings, monitored values, alarm information and waveform data to the patient monitoring system. The BMI has no user-operated controls or displays. This device is activated when it is attached to a 9 to 15 Volt power supply and fully functional when attached to the ventilator and patient monitoring system. Indicators are limited to a power indicator. The BMI is installed between the ventilator and the Patient Monitoring System. Connection between the ventilator and the BMI is Fiber Optic. Connection between the BMI and the Patient Monitoring System is RS-232 compatible. The BMI also contains a fiber optic feed-through for connection to Bird accessories, particularly the Bird Graphics Monitor and The Partner Volume Monitor. Serial data is received from the ventilator over the fiber optic interface. The fiber optic signal is converted to an electrical signal by a fiber optic receiver. The electrical signal then routes to a fiber optic driver for connection to an accessory device as described above. The electrical signal also routes to the serial input of the microprocessor where the serial data is converted to parallel data. The parallel data is then stored in random access memory (RAM) for future use. Program code for the microprocessor is contained in an electrically programmable read only memory (EPROM) and is accessed as necessary for program execution. Requests from the monitoring system are received as serial RS-232 signals. These signals are converted to logic levels by RS-232 receivers. These logic levels then are processed bv a universal asynchronous receiver transmitter (UART). The UART converts the serial data to parallel data and then notifies the microprocessor by using the interrupt signal. The microprocessor reads the data from the UART, forming a request packet from the monitor. Following receipt of a complete request packet from the monitor, the processor generates the appropriate response packet from the ventilator data stored in RAM. This data is written to the UART, which converts the data to serial form. The serial data is then converted to RS-232 signals and transmitted to the monitor.

AI/ML Overview

The Bird Monitor Interface (BMI) is a communication device, not a diagnostic or AI-powered system that would typically undergo studies involving expert adjudication, MRMC, or training sets in the way more complex medical imaging or AI-driven diagnostic devices would. Its "performance" is based on its ability to accurately transmit data as designed.

Here's the information based on the provided text, recognizing the nature of this device:

Acceptance Criteria and Device Performance

The acceptance criteria for the Bird Monitor Interface (BMI) are derived from its design specifications and its ability to communicate specific data points between Bird ventilators and patient monitoring systems, as well as its compliance with various electrical, environmental, and safety standards. The study performed was focused on verifying that all these specified functions operated as designed.

1. Table of Acceptance Criteria and Reported Device Performance

SpecificationRequirementReported Device Performance
BMI interface to T-Bird, 8400STi, and VIP ventilatorsRecieve specific Settings, Monitored Values, and Alarm Status from each ventilator modelAll listed Settings, Monitored Values, and Alarm Status were verified to be received from each ventilator model. (Indicated by ☑)
BMI interface to Hewlett-Packard VueLink and SpaceLabs Universal FlexportTransmit specific data (Total Breath Rate, Peak Inspiratory Pressure, Mean Airway Pressure, Exhaled Tidal Volume, Exhaled Minute Volume, Inspiratory Time, PEEP/CPAP, Inspiratory Tidal Volume) to respective monitorsAll listed data were verified to be transmitted to both Hewlett-Packard and SpaceLabs monitors. (Indicated by ☑)
Proprietary InterfaceMake all listed T-Bird, 8400, and VIP ventilator data (Settings, Monitored Values, Alarm Status) available through the proprietary interfaceAll listed T-Bird, 8400, and VIP data were verified to be available through the proprietary interface. (Indicated by ☑)
Power On Self TestRuns at power application, verifies processor operation, EPROM operation, and RAM operationAll power-on self-test functions were verified. (Indicated by ☑)
Electrical RequirementsCompliance with IEC601-1:1988 (Electrical Safety, Dielectric Withstand, Leakage Current)Compliance with IEC601-1:1988 verified.
AC Power Grounding and PolarityOutput cable, connector cannot be reversed; reverse polarity will not damage BMIVerified that reverse polarity does not damage the device.
Electromagnetic Compatibility (EMC)Emissions: IEC 601-1-2, CISPR11 (Radiated and Conducted EMI), MIL-5TD462D, RE101 (Magnetic Fields)Verified compliance with specified EMC emissions standards.
Electromagnetic Compatibility (EMC)Immunity: IEC 601-1-2, IEC801-2, FDA (Electrostatic Discharge), FDA Reviewer's Guidance (Radiated Electromagnetic Fields, Fluctuations, Transients, Surges, Dropout, Slow Sags and Surges, Fast Surges), IEC801-4 (Fast Transient Bursts), MIL-5TD461D (Conducted EMC), MIL-5TD462D, RS101 (Magnetic Fields), FDA Reviewer's Guidance (Quasi-Static Electric Fields)Verified compliance with specified EMC immunity standards.
Mechanical and Environmental RequirementsShock (IEC 68-2-27), Sinusoidal Vibration (IEC 68-2-6), Random Vibration (IEC 68-2-34), Fluid Spill Resistance (IEC 601-1, Clause 44.6)Verified compliance with specified mechanical and environmental standards.
Operating Temperature+10 to +40°C, 30 to 75% RHDevice operated within specified temperature and humidity range.
Storage Temperature-20 to +60°C, up to 95% RHDevice met storage temperature and humidity requirements.
Surface TemperatureLess than 41 °C while in an ambient of 35°CDevice maintained surface temperature less than 41°C.

Study Details:

The provided document describes a performance testing study conducted to verify the functionality and compliance of the Bird Monitor Interface.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not specify a numerical "sample size" in terms of number of devices tested or data points collected for the functional verification. Instead, it states that "All functions were verified to operate as designed and intended." This implies comprehensive testing against each listed function rather than statistical sampling of performance.
  • Data Provenance: The testing was conducted "in the laboratory." Environmental, EMI/RFI, and Electrical Safety Standards testing were performed by "certified test facilities." This indicates a controlled, simulated environment, not clinical data from patients or specific geographical origins. The testing is retrospective in nature, verifying the device's adherence to its design specifications.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts/Qualifications: Not applicable for this type of device. The ground truth is the device's design specifications and accepted engineering and safety standards (e.g., IEC, CISPR, MIL-STD, FDA Reviewer's Guidance). The "experts" would be the engineers and technicians involved in the design and testing process, and the personnel at the certified test facilities who conduct the standard compliance tests. No clinical "expert" consensus was required as the device's function is data transmission, not clinical interpretation.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

  • Adjudication Method: Not applicable. Functional verification against engineering specifications and industry standards does not involve adjudication by multiple human observers as would be the case for diagnostic accuracy studies. The "adjudication" is essentially a pass/fail determination based on whether the device performs according to its predefined requirements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No. The Bird Monitor Interface is not an AI-powered diagnostic device or an assistive technology for human readers/clinicians, therefore, an MRMC comparative effectiveness study was not performed.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Standalone Performance: Yes, in essence. The "performance testing" described is for the device operating independently to transmit data as specified. While it interfaces with other systems (ventilators and patient monitors), its functionality is standalone in terms of correctly handling and transmitting the data according to its programmed logic. The testing verified the algorithm's (microprocessor's) ability to process and transmit data as per design.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

  • Type of Ground Truth: The ground truth for this device's performance is its design specifications and compliance with recognized national and international engineering, electrical safety, environmental, and electromagnetic compatibility standards (e.g., IEC601-1, CISPR11, MIL-5TD462D, IEC801-2, IEC801-4, IEC 68 series). The "truth" is whether the data transmitted matches the data received from the ventilator, and whether the device operates reliably and safely according to these standards.

8. The Sample Size for the Training Set:

  • Training Set Sample Size: Not applicable. The Bird Monitor Interface is a hardware device with embedded software (firmware) that performs predefined functions. It does not utilize machine learning or AI that requires a "training set" to learn patterns or make predictions. The software is programmed deterministically.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set: Not applicable, as there is no training set for this device. The "ground truth" for the device's functionality is established during its design and engineering phases, based on functional requirements derived from the intended use and compliance standards.

{0}------------------------------------------------

510(K) SUMMARY

K982580

Bird Monitor Interface

Bird Products Corporation

Neil Battiste Regulatory Affairs Manager Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262-6267

760.778.7341 (voice) 760.778.7274 (fax)

July 21, 1998

{1}------------------------------------------------

General Information

Bird Monitor Interface Device Trade Name:

Device Common/Classification Name:

The Bird Monitor Interface is an accessory to Bird Products Corporation ventilator products classified under 73 CBK "Continuous Ventilators"

Predicate Device:Bird RS-232 Interface Module
FDA 510(k) No:K880721
Option 20-26 of Puritan Bennett 7200 Volume Ventilator
FDA 510(k) No:K841757

Intended Use:

The Bird Monitor Interface (BMI) is a stand-alone unit providing a means of communicating information from Bird ventilators to patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin and SpaceLab's PDMS are provided. as well as a Bird proprietary interface for use with other systems. The purpose of BMI is to deliver ventilator settings, monitored values, alarm information and waveform data to the patient monitoring system.

{2}------------------------------------------------

Device Description

The Bird Monitoring Interface (BMI) is a stand-alone unit providing a communications interface between the Bird ventilators, 8400STi, VIP and TBird, and patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin (via Hewlett-Packard's Vuelink proprietary interface) and SpaceLab's PDMS (via Spacelab's Universal Flexport proprietarv interface) are provided as well as a Bird proprietary interface, the General Purpose Computer

Protocol (GPCP), for use with other systems. The purpose of the BMI is to deliver ventilator settings, monitored values, alarm information and waveform data to the patient monitoring system.

The BMI has no user-operated controls or displays. This device is activated when it is attached to a 9 to 15 Volt power supply and fully functional when attached to the ventilator and patient monitoring system. Indicators are limited to a power indicator.

Image /page/2/Picture/5 description: The image shows a close-up of a device with several labels and indicators. The device has labels such as "DATA IN", "DATA OUT", "POWER INDICATOR", "DYAD", "MONITOR INTERFACE", "TO MONITOR", and "INPUT 9/16VDC". The device also has a few ports and indicators on the top and bottom.

The BMI is installed between the ventilator and the Patient Monitoring System. Figure 3-1 illustrates a tvpical BMI installation. Connection between the ventilator and the

BMI is Fiber Optic. Connection between the BMI and the Patient Monitoring System is RS-232 compatible. The BMI also contains a fiber optic feed-through for connection to Bird accessories, particularly the Bird Graphics Monitor and The Partner Volume Monitor.

Image /page/2/Figure/8 description: The image shows a diagram of a bird monitoring system. The system consists of a graphics monitor, a bird ventilator, a bird monitor interface, and a monitoring system. The graphics monitor and bird ventilator are connected to the bird monitor interface via fiber optic cables, and the bird monitor interface is connected to the monitoring system via an RS-232 cable.

Figure 3-1 Bird Monitor Interface Installation

{3}------------------------------------------------

Serial data is received from the ventilator over the fiber optic interface. The fiber optic signal is converted to an electrical signal by a fiber optic receiver. The electrical signal then routes to a fiber optic driver for connection to an accessory device as described above.

The electrical signal also routes to the serial input of the microprocessor where the serial data is converted to parallel data. The parallel data is then stored in random access memory (RAM) for future use. Program code for the microprocessor is contained in an electrically programmable read only memory (EPROM) and is accessed as necessary for program execution.

Requests from the monitoring system are received as serial RS-232 signals. These signals are converted to logic levels by RS-232 receivers. These logic levels then are processed bv a universal asynchronous receiver transmitter (UART). The UART converts the serial data to parallel data and then notifies the microprocessor by using the interrupt signal. The microprocessor reads the data from the UART, forming a request packet from the monitor. Following receipt of a complete request packet from the monitor, the processor generates the appropriate response packet from the ventilator data stored in RAM. This data is written to the UART, which converts the data to serial form. The serial data is then converted to RS-232 signals and transmitted to the monitor.

Image /page/3/Figure/4 description: The image shows a block diagram of a ventilator interface system. The system consists of several main components, including a ventilator interface, a processor, memory, a monitor interface, and a voltage regulator. The ventilator interface includes an optical receiver and transmitter for fiber optic communication. The processor contains a microprocessor and a latch, while the memory consists of 8K RAM and 64K EPROM. The monitor interface includes a UART and an RS-232 driver/receiver for communication with a patient monitoring system.

Image /page/3/Figure/5 description: The image shows the title of a diagram. The title reads "Bird Monitor Interface Block Diagram". The image also includes the text "Figure 3-2".

{4}------------------------------------------------

Comparison to Predicate Device

The Bird Monitor Interface™, (referred to as BMI), is a stand-alone unit providing a means of communicating information from Bird ventilators to patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin and SpaceLab's PDMS are provided, as well as, a

Image /page/4/Picture/3 description: The image shows a cassette tape with a white label on top of a black cassette. The label has the word "SWITCH" in large, bold letters. Below the word "SWITCH" is the word "ENTERPRISES" in smaller letters. The cassette tape has a few buttons and holes on the bottom.

Bird Monitor Interface

Bird proprietary interface for use with other systems. The purpose of the Bird Monitor Interface is to deliver ventilator settings. monitored values. alarm information, and waveform data to the patient monitoring system.

The BMI is not significantly different from the predicate device the Bird RS 232 Interface. Both devices are microprocessor controlled systems that accept the fiber optic output

from the 8400STi® and V.I.P Bird® ventilators. Both interfaces act as intermediaries

linking a Bird ventilator with a central computer. The means of communicating the data between the devices is a commonly used platform.

Image /page/4/Picture/9 description: The image shows a piece of electronic equipment, possibly a computer or server. The device has a boxy shape with a front panel that includes several slots or bays. There is a logo or label on the top right of the device. The device is sitting on a surface with a grid pattern.

RS-232 Interface Module

The development of the BMI involves four changes to the currently marketed Interface Module RS232:

    1. The BMI interfaces directly with the Hewlett-Packard Merlin and SpaceLab's PDMS systems;
    1. The BMI accepts input from the TBird ventilator family;
    1. The BMI does not contain any alarms or indicate any alarm conditions;
    1. The BMI does not interface directly to a printer

First Modification:

Direct output to the Hewlett-Packard Merlin and SpaceLabs PDMS systems. Both the Hewlett-Packard Merlin and SpaceLab's PDMS are popular patient monitoring svstems with an established hospital customer base.

Each system employs a unique, proprietary communications protocol (Hewlett-Packard VueLink and SpaceLab's Universal Flexport). The protocol, maintained and supplied by each company, describes the exact structure and hierarchy for transmitted data. Information received in the proper format can be displayed on the monitor's display screen.

The BMI functions as a data organizer for each of these monitoring systems. The Bird ventilator sends out a stream of data in a format designed by Bird. The BMI gathers this

{5}------------------------------------------------

information and structures it into the required format, one that is acceptable to the monitoring system. This does not entail substantive modification of the data. Rather, the BMI describes the data in terms that are understandable and usable by the monitoring system before sending it on.

The Data Interface-RS232 performed a similar function but formatted the data for printing or into a more universal format call the General Purpose Computer protocol. The Data Interface-RS232 could not be connected directly to a monitoring system.

Second Modification;

The BMI accepts information from the TBird family of ventilators.

The Data Interface-RS232 was designed to accept information from the 6400ST, 8400STi, and the VIP Bird. The product scope of the BMI is enhanced in comparison. The BMI can accept and transmit information from the following list of Bird ventilators: 8400STi, VIP Bird, and the TBird Series. The TBird family of ventilators use the same fiber optic output as was used in the previously listed models. All of the connections, cables, and protocols are common to the entire line.

Third Modification;

The BMI does not contain any alarms or indicate any alarm conditions.

In addition to receiving information on alarm conditions, the Data Interface-RS232 contained the means of displaying this information. The front panel of the Data Interface-RS232 housed LED indicators for an alarm condition and an alarm disable. There is an audible alarm generator located inside of the Data Interface-RS232's housing. Any of the ventilator's alarm conditions would engage the Data Interface-RS232's slave alarm system. An alarm condition would trigger a red indicator lamp and sound the audible alarm. The user could manually disable these remote alarm functions. When operating in this condition a yellow indicator lamp would be illuminated.

The BMI acts solely as a means of communicating data from the Bird ventilator to the monitoring system or hospital computer system. All alarm information and alarm conditions are relayed through the BMI to the patient monitoring system for visual and audible display by the monitoring system. The BMI does not contain any audible alarm generators or visual alarms LED indicators. The single LED indicator on the BMI indicates that a power source has been successfully attached.

Fourth Modification - The BMI does not interface directly to a printer.

The Data Interface-RS232 contained the ability to send the ventilator information to a serial printer. The user had to select the proper DIP switch position prior to activating the printer switch. A serial printer and cable were also required.

The BMI does not contain the software or hardware to interface with a printer. The information being transmitted from the BMI can only go to the specified monitoring systems or a computer.

{6}------------------------------------------------

Summary of Performance Testing

Performance testing was conducted in the laboratory to confirm input and output of data to and from the Bird Monitor Interface. Testing to Environmental, EMI/RFI and Electrical Safety Standards were performed by certified test facilities. The following table specifies all Bird Monitor Interface functions. All functions were verified to operate as designed and intended, as denoted by the checkmark.

SpecificationRequirement
BMI interface to the T-Bird, 8400STi and VIP ventilatorsThe following data is received from the T-Bird ventilator:
Settings
Mode☑ Pressure Support ☑
Tidal Volume☑ %02 ☑
Peak Flow☑ Pressure Control ☑
Breath Rate☑ Inspiratory Pause ☑
Sensitivity☑ Inspiratory Time ☑
PEEP☑ Waveform type ☑
Vol. Assured Press. Support☑ Low Peak Pressure Alarm ☑
High Pressure Limit Alarm☑ Low Minute Volume Alarm ☑
Monitored Values
Flow☑ Mean Airway Pressure ☑
Proximal Pressure☑ Exhaled Minute Volume ☑
Exhaled Tidal Volume☑ PEEP ✔
Breath Rate☑ Inspiratory Time ☑
Peak Inspiratory Pressure
Alarm Status
High Pressure Limit☑ Apnea ☑
Low Minute Volume☑ Low Peak Pressure ✔
The following data is received from the 8400 ventilator:
Settings
Mode☑ Pressure Support ☑
Tidal Volume☑ Waveform type ☑
Peak Flow☑ High Pressure Limit Alarm ☑
Breath Rate☑ Low Peak Pressure Alarm ☑
Sensitivity☑ Low Minute Volume Alarm ☑
PEEP☑ Low PEEP/CPAP Alarm ☑
Apnea Interval☑ High Breath Rate Alarm ☑
Back Up Breath Rate
Monitored Values
Flow☑ Peak Inspiratory Pressure ☑
Proximal Pressure☑ Mean Airway Pressure ☑
Exhaled Tidal Volume☑ Exhaled Minute Volume ☑
Breath Rate☑ PEEP ☑
I:E Ratio☑ Inspiratory Time ☑
Alarm Status
High Pressure Limit☑ Low Peak Pressure ☑
Low Minute Volume☑ Low PEEP/CPAP ☑
Apnea
The following data is received from the VIP ventilator:
Settings
Tidal Volume☑ Pressure Support ☑
Peak Flow☑ Termination Sensitivity ☑
Breath Rate☑ High Pressure Limit ☑
Sensitivity☑ Low Peak Pressure Alarm ☑
PEEP☑ Low PEEP/CPAP Alarm ☑
Monitored Values
Flow☑ Peak Inspiratory Pressure ☑
Proximal Pressure☑ Total Breath Rate ☑
Exhaled Tidal Volume☑ Exhaled Minute Volume ☑
I:E Ratio☑ Mean Airway Pressure ☑
Inspiratory Time
Alarm Status
High Pressure Limit☑ Low Peak Pressure ☑
Apnea☑ Low PEEP/CPAP ☑

{7}------------------------------------------------

SpecificationRequirement
BMI interface to the Hewlett-Packard VueLink and SpaceLabs Universal Flexport
Hewlett-Packard VueLink
The following datais transmitted to theHewlett-Packard monitor:Total Breath RateExhaled Tidal Volume ☑
Peak Inspiratory PressureExhaled Minute Volume ☑
Mean Airway PressureInspiratory Time ☑
Inspiratory Tidal VolumePEEP/CPAP ☑
SpaceLabs Universal Flexport
The following datais transmitted to theSpaceLabs monitor:Total Breath RateExhaled Minute Volume ☑
Inspiratory TimeMean Airway Pressure ☑
Peak Inspiratory PressureExhaled Tidal Volume ☑
PEEP/CPAPInspiratory Tidal Volume ☑
Proprietary Interface
All TBird data listed below is available through the proprietary interface.
SettingsModePressure Support ☑
Tidal Volume%02 ☑
Peak FlowPressure Control ☑
Breath RateInspiratory Pause ☑
SensitivityInspiratory Time ☑
PEEPWaveform type ☑
Vol. Assured Press. SupportLow Peak Pressure Alarm ☑
High Pressure Limit AlarmLow Minute Volume Alarm ☑
Monitored ValuesFlowPeak Inspiratory Pressure ☑
Proximal PressureMean Airway Pressure ☑
Exhaled Tidal VolumeExhaled Minute Volume ☑
Breath RatePEEP ☑
Inspiratory Time
Alarm StatusHigh Pressure LimitApnea ☑
Low Minute VolumeLow Peak Pressure ☑
All 8400 data listed in below is available through the proprietary interface.
SettingsModePressure Support ☑
Tidal VolumeWaveform type ☑
Peak FlowHigh Pressure Limit Alarm ☑
Breath RateLow Peak Pressure Alarm ☑
SensitivityLow Minute Volume Alarm ☑
PEEPLow PEEP/CPAP Alarm ☑
Apnea IntervalHigh Breath Rate Alarm ☑
Back Up Breath Rate
Monitored ValuesFlowPeak Inspiratory Pressure ☑
Proximal PressureTotal Breath Rate ☑
Exhaled Tidal VolumeExhaled Minute Volume ☑
I:E RatioPEEP ☑
Alarm StatusHigh Pressure LimitLow Peak Pressure ☑
Low Minute VolumeLow PEEP/CPAP ☑
ApneaHigh Breath Rate ☑
All VIP data listed below is available through the proprietary interface.
SettingsModePEEP ☑
Tidal VolumePressure Support ☑
Peak FlowHigh Pressure Limit ☑
Breath RateLow Peak Pressure Alarm ☑
SensitivityLow PEEP/CPAP Alarm ☑
Monitored ValuesFlowPeak Inspiratory Pressure ☑
Proximal PressureTotal Breath Rate ☑
Exhaled Tidal VolumeExhaled Minute Volume ☑
I:E Ratio
Alarm StatusHigh Pressure LimitLow Peak Pressure ☑
ApneaLow PEEP/CPAP ☑

{8}------------------------------------------------

SpecificationRequirement
Power On Self TestRuns at power applicationverifies processor operationverifies EPROM operationverifies RAM operation
Electrical Requirements
Electrical SafetyIEC601-1:1988
Dielectric WithstandIEC 601-1:1988, clause 20
AC Power Grounding and Polarityoutput cable, connector cannot be reversedreverse polarity will not damage the BMI
Leakage CurrentIn accordance with IEC601-1, 1988
Electromagnetic Compatibility
EmissionsIEC 601-1-2
Radiated and Conducted EMIIn accordance with CISPR11
Magnetic FieldsIn accordance with MIL-5TD462D, RE101
ImmunityIEC 601-1-2
Electrostatic DischargeIn accordance with IEC801-2, FDA
Radiated Electromagnetic FieldsFDA Reviewer's Guidance
Fluctuations, Transients, Surges
Steady State Voltage120VAC +10%,-20%
DropoutFDA Reviewer's Guidance
Slow Sags and SurgesFDA Reviewer's Guidance
Fast Transient BurstsIEC801-4 for table top equipment
Fast SurgesFDA Reviewer's Guidance
Conducted EMCIn accordance with MIL-5TD461 D
Magnetic FieldsIn accordance with RS101 of MIL-5TD462D
Quasi-Static Electric FieldsFDA Reviewer's Guidance
Mechanical and Environmental Requirements
ShockIn accordance with IEC 68-2-27
Sinusoidal VibrationIn accordance with IEC 68-2-6
Random VibrationIn accordance with IEC 68-2-34
Fluid Spill ResistanceIn accordance with IEC 601-1, Clause 44.6
Operating Temperature+10 to +40°C, 30 to 75% RH
Storage Temperature-20 to +60°C, up to 95% RH
Surface TemperatureLess than 41 °C while in an ambient of 35°C

Hardware and software performance testing verified that the Bird Monitor Interface meets all of it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other. The faces are abstract and appear to be connected by a flowing line.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 22 1998

Mr. Neil Battiste Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262-8099

K982580 Re: Bird Monitor Interface Regulatory Class: II (two) Product Code: 73 CBK Dated: July 22 1998 Received: July 24, 1998

Dear Mr. Battiste:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{10}------------------------------------------------

Page 2 -- Mr. Neil Battiste

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Statement of Indications for Use

The Bird Monitor Interface (BMI) is a stand-alone unit providing a means of communicating information from Bird ventilators to patient monitoring systems. Specific interfaces to the Hewlett-Packard Merlin and SpaceLabs PDMS are provided, as well as a Bird proprietary interface for use with other systems. The purpose of BMI is to deliver ventilator settings, monitored values, alarm information and waveform data to the patient monitoring system.

Kramer

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).