K Number
K982563

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-10-06

(75 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed.

Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.

Device Description

Ultrapro is a tool for a medical imaging practice. It can acquire medical images from any standard medical video source. Digital data in the DICOM format or other popular PC formats such as JPEG, TIFF, etc. can also be acquired.

Images may be acquired in uncompressed or compressed format (lossless or lossy). The lossy compression ratio is selected by viewing the original image and the lossy compressed image side by side and increasing the ratio while retaining clinically significant characteristics. Images that have been lossy compressed have the text "LOSSY xx:1" where xx is the compression ratio placed in the visible, non-diagnostic area of the image.

Related patient information can also be entered or scanned in with a TWAIN compliant scanner. Once acquired, Ultrapro serves as a digital storage for these images. Images are stored on the computer. Images may be communicated to other Ultrapro workstations or to third-party DICOM 3.0 compliant PACS components. The radiologist may review the images with Ultrapro and create reports. The display of images (contrast, brightness, zoom, etc.) may be adjusted at any time. Annotations may be drawn on images. Distance measurements may be made on images. These image parameters may be saved along with the image while still preserving the original image. This information can be communicated to other workstations through a network or modem. Finally, all of the information may be archived onto optical media for permanent storage.

AI/ML Overview

The provided text is a 510(k) Summary for the "Ultrapro 98" Picture Archiving and Communication System (PACS) from RamSoft Inc. It focuses on establishing substantial equivalence to a predicate device and does not contain information about acceptance criteria or specific studies proving device performance against those criteria. The document explicitly states:

"The determination of substantial equivalence is not based on an assessment of performance tests."

Therefore, I cannot extract the requested information regarding acceptance criteria and performance studies from this document. The sections you asked about are not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo on the left and a date on the right. The logo is circular and contains the text "RamSoft". The date is "OCT 6 1998". The text is in a bold, sans-serif font.

K982 56.3

510(k) Summary for Ultrapro 98 1

1.1 General Information

1.1.1 Name and Address of Manufacturer

RamSoft Inc. 37 Bankview Circle Toronto. ON M9W 6S6, CANADA Tel: (416) 674-1347 Fax: (416) 674-7147 Email: ram@ramsoft.on.ca Contact Person: Naganatha S Ramanathan

1.1.2 Establishment Registration Number

9006410

1.1.3 Common and Proprietary names of Device

Trade or Proprietary Name: Ultrapro™ Common Name: Picture Archiving and Communication System

1.1.4 Class

Systems, digital image communications - Class | 892.2020 LMD

12 Substantially Equivalent Devices

ALI UltraPACS (K925965/A) with ALI comPACS data compression module (K963610) Manufactured by: ALI Technologies, Inc

CEMAX-ICON Archive Manager 2.0 and Autorad modules (K955092) Manufactured by: CEMAX-ICON Inc.

1.3 Device Description

Ultrapro is a tool for a medical imaging practice. It can acquire medical images from any standard medical video source. Digital data in the DICOM format or other popular PC formats such as JPEG, TIFF, etc. can also be acquired.

Images may be acquired in uncompressed or compressed format (lossless or lossy). The lossy compression ratio is selected by viewing the original image and the lossy compressed image side by side and increasing the ratio while retaining clinically significant characteristics. Images that have been lossy compressed have the text "LOSSY xx:1" where xx is the compression ratio placed in the visible, non-diagnostic area of the image.

Related patient information can also be entered or scanned in with a TWAIN compliant scanner. Once acquired, Ultrapro serves as a digital storage for these images. Images are stored on the computer. Images may be communicated to other Ultrapro workstations or to third-party DICOM 3.0 compliant PACS components. The radiologist may review the images with Ultrapro and create reports. The display of images (contrast, brightness, zoom, etc.) may be adjusted at any time. Annotations may be drawn on images. Distance measurements may be made on images. These image parameters may be saved along with the image while still preserving the original image. This information can be communicated to other workstations through a network or modem. Finally, all of the information may be archived onto optical media for permanent storage.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for a company called "RalliSoft". The logo features the company name in a stylized font inside of an oval shape. The oval is positioned in the center of a globe graphic. The globe is depicted with curved lines to represent the continents and oceans.

1.4 Intended Use Statement

Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use with FDA approved imaging equipment that provides a video output or a digital output.

ALI UltraPACS with ALI comPACSRamSoft Ultrapro
Image AcquisitionVideo frame grabber or DICOMinputVideo frame grabber, FDAapproved film scanner or DICOMinput
Compatible Video SignalsNTSC (up to 30 frames/s,640x480, 256 grayscale, 16.7million colors)PAL (up to 25 frames/s, 768x576.256 grayscale, 16.7 million colors)NTSC (up to 30 frames/s,640x480, 256 grayscale, 16.7million colors)PAL (up to 25 frames/s, 768x576,256 grayscale, 16.7 million colors)
Cine Capability (motion capture)YesYes
Image CompressionLossless RLE and Lossy JPEGLossless RLE and Lossy JPEG
Lossy Compression RatioRange4:1 to 30:14:1 to 60:1
Uses JPEG lossy compressionin viewing, transmitting andarchiving still imagesYesYes
Uses Motion JPEG lossycompression for viewing,transmitting and archiving cineimagesYesYes
DICOM 3.0 compliantYesYes
Imaging ModalitiesMulti-modalityMulti-modality
Image Display CapabilityYesYes
Remote and Local Area NetworkCommunicationYesYes
Image Annotation CapabilityYesYes
Archives Image on OpticalMediaYesYes

1.5 Companson of Technological Characteristics

1.6 Conclusion

In conclusion, Ultrapro is substantially equivalentitititi to the ALI product. The determination of substantial equivalence is not based on an assessment of performance tests.

55555 : 35453 :

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with its wings spread, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 6 1998

Naganatha S. Ramanathan, Ph.D. President Ramsoft, Inc. 37 Bankview Circle Toronto, Ontario Canada M9W 6S6

Re:

K982563 "Ultrapro 98, Digital Image Communications System Dated: July 22, 1998 Received: July 23, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ

Dear Dr. Ramanathan:

We have reviewed your Section 5100% notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments on to devices that have been reclassified in accordance with the Feceral Food Drues and Cosmetic Act (Act). You may, therefore, narket the device, subject to the general controls of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical-Devices: Generaliation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition; FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odch/dsmamain.html"

Sincerely yours,

Lillian Yin, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and -Radiological Health

Enclosure

{3}------------------------------------------------

Page of

t

510(k) Number (if known):K982563
Device Name:ULTRAPRO 98

Indicatic is For Use:

Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed.

Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.

(PL) ASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Gimel li. Syson
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number

OR

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter Use

(Optional Format 1-2-96)

N/A