(154 days)
The product is an image processing workstation software package designed to run on standard PC hardware. The hardware is all "off-the-shelf" standard computer components and may be purchased independently by the end user or supplied by NIT. The NIT InSight™ software receives image files from medical scanning devices, such as CT or MRI and performs real time viewing, image manipulation, 3D and 4D visualization, communication, and archiving.
The InSight™ Diagnostic Imaging Workstation receives image files from medical scanning devices, such as CT or MRI and performs real-time viewing, image manipulation, three and four dimensional visualization, communication, and archiving. All of the functions are supported on standard personal computer platform for ease of cost and maintenance. The use of Microsoft Windows NT™ operating system makes the InSight™ Diagnostic Imaging Workstation easy to use and capable of being integrated with your other computer needs.
This document primarily focuses on the 510(k) submission for the Neo Imagery Technologies Inc., InSight™ Diagnostic Imaging Workstation, asserting its substantial equivalence to previously cleared devices. It does not contain information about acceptance criteria or a dedicated study proving the device meets specific performance criteria.
The device is being cleared based on its equivalence to existing PACS systems, which implies it operates under similar performance expectations but doesn't detail new, specific performance metrics or studies to validate them for this particular submission.
Therefore, requested information, such as specific acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, types of ground truth, and training set information (items 1, 2, 3, 4, 5, 6, 7, 8, 9), is not available within the provided text.
The document states:
- Device Description and Intended Use: The InSight™ Diagnostic Imaging Workstation receives image files from medical scanning devices, performs real-time viewing, image manipulation, three and four-dimensional visualization, communication, and archiving.
- Substantial Equivalence: The submission argues that the InSight™ device is substantially equivalent to K965179 (Neo Imagery Technologies Inc., InSight™ Diagnostic Imaging Workstation) and K961023 (AccuImage, AIDP).
- Software Certification: Neo Imagery Technologies Inc. certifies that the InSight™ software is designed, developed, tested, and validated according to written procedures.
- Hazard Analysis: The potential hazards are deemed "no different than those of other PACS components" and are primarily related to computer system component failure, not expected to "materially contribute to patient death or injury." The "Level of Concern" is "Minor."
In essence, the document confirms the device's regulatory pathway (510(k)) and its intended use, but it doesn't offer the detailed study information regarding acceptance criteria for device performance that you requested.
{0}------------------------------------------------
1. Company Identification
Neo Imagery Technologies, Inc. 2315 S. Birch Log Way Hacienda Heights, CA 91745 Tel. (626) 333-3633 Fax (626) 961-0080
2. Official Correspondent
Gary J. Allsebrook Regulatory Affairs
3. Date of Submission
July 16, 1998
4. Device Name
| Classification Name: | Computed Imaging Device |
|---|---|
| Common/Usual Name: | Picture Archiving and Communications System(PACS) |
| Proprietary Name: | Neo Imagery Technologies Inc., InSight™Diagnostic Imaging Workstation |
5. Substantial Equivalence
InSight™, Diagnostic Imaging Workstation, K965179; Acculmage, AIDP (Acculmage Image Display Processor), K961023
6. Device Description and Intended Use
The InSight™ Diagnostic Imaging Workstation receives image files from medical scanning devices, such as CT or MRI and performs real-time viewing, image manipulation, three and four dimensional visualization, communication, and archiving. All of the functions are supported on standard personal computer platform for ease of cost and maintenance. The use of Microsoft Windows NT™ operating system makes the InSight™ Diagnostic Imaging Workstation easy to use and capable of being integrated with your other computer needs.
{1}------------------------------------------------
7. Software
Neo Imagery Technologies Inc. certifies that the InSight™ Diagnostic Imaging Workstation software is designed, developed, tested and validated according to written procedures. These procedures identify individuals within the organization responsible for developing and approving product specifications, coding and testing, validation testing and field maintenance.
8. Hazard Analysis
Hazard analysis on this product has been performed throughout the definition, design, coding and testing phases of product development and implementation. This process has emphasized:
- . identification of potential hazards, their causes, and their effects;
- development of methodologies to control the occurrence of hazards and to . constrain their effects; and
- Determine any effect on patient safety and system effectiveness. .
The potential hazards associated with this software product are no different than those of other PACS components . These are primarily related to failure of computer system components, and may be variously obviated by decisions taken by the customers of this product. None of these failures are expected to materially contribute to patient death or injury.
It is our conclusion that there is no hardware or software component, operating in a properly configured environment, whose failure or latent design defect would be expected to result in death or injury of a patient. Thus the "Level of Concern" is "Minor".
9. Safety Concerns
The hardware is "off-the-shelf" and complies with applicable electrical safety standards for standard PC hardware and peripherals.
{2}------------------------------------------------
10. Substantial Equivalence
The following product provides functions, which are substantially equivalent to this product:
| Manufacturer: | Neo ImageryTechnologies | Neo ImageryTechnologies | Acculmage |
|---|---|---|---|
| Product Name: | InSight™ | InSight™ | AIDP |
| 510(k) Number: | K965179 | K961023 | |
| Computer Platform: | Pentium/WindowsNT | Pentium/WindowsNT | Pentium/Win95 |
| Image Format In: | Imatron ProprietaryACR NEMA 2.0DICOM 3.0 | Imatron ProprietaryACR NEMA 2.0DICOM 3.0 | Imatron ProprietaryACR NEMA 2.0DICOM 3.0 |
| Image Format Out: | DICOM 3.0TIFF, BMP | DICOM 3.0TIFF, BMP | BMP |
| Image Archive: | Magneto-OpticalDrive, 2.6 GB | Magneto-OpticalDrive, 2.6 GB | IDE Disk Drive, 1+G |
| Image Display: | Color/gray scale, CRTor Laptop PC LCD,up to 1024x1024, 12bits | Color/gray scale, CRTor Laptop PC LCD,up to 1024x1024, 12bits | Color/gray scale, CRTor Laptop PC LCD,up to 512x512, 12bits |
| Image Processing: | Window-Level,Zoom, Variablesmooth filter, Cinedisplay | Window-Level,Zoom, Variablesmooth filter, Cinedisplay | Window level,Zoom, variable Filter:edge, sharpen,smoothCine Display |
| Image Edit: | ManualSegmentation bydrawing a contour.Segmentation by CTnumber threshold.(MIP, MinIP,radiographicprojection),Surface rendering,Quick volumerendering,Volume rendering | ManualSegmentation bydrawing a contour.Segmentation by CTnumber threshold(MIP), Surfacerendering | ManualSegmentation bydrawing a contour.Segmentation by CTnumber( MIP,MinIP), Surfacerendered, Depthencoded surface,Quick Mip |
| Volume Rendering: | Maximum, orminimum IntensityProjection (MIPMinIP) Radiographicprojection, Surfacerendering, Quickvolume rendering,Volume Rendering | Maximum IntensityProjection(MIP),surface rendering | Max or Min IntensityProjection (MIP,MinIP), SurfaceRendered Depthencoded surface,Quick Mip |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract image of an eagle with three heads.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 1998
Tina Young CEO Neo Imagery Technologies, Inc. 2315 South Birch Log Way Hacienda Heights, CA 91745
Re: K982535
InSight™ Diagnostic Imaging Workstation Dated: November 3, 1998 Received: November 4, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Ms. Young:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notfication"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K982535
Neo Imagery Technologies, Inc., InSight™ Diagnostic Device Name: Imaging Workstation
Indications For Use:
The product is an image processing workstation software package designed to run on standard PC hardware. The hardware is all "off-the-shelf" standard computer components and may be purchased independently by the end user or supplied by NIT. The NIT InSight™ software receives image files from medical scanning devices, such as CT or MRI and performs real time viewing, image manipulation, 3D and 4D visualization, communication, and archiving.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Arvil H. Calhounge of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
1
510(k) Number .
Prescription Use_ (Per 21 CFR 901.109)
Over-the-Counter Use
(Optional Format 1-2-96)
N/A