(240 days)
IMMULITE Thyroglobulin is a two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Analyzer and designed for the quantitative measurement of thyroglobulin in serum or heparinized plasma. It is intended strictly for in vitro diagnostic use as an aid in monitoring patients who have undergone thyroidectomy.
IMMULITE Thyroglobulin is a clinical device for use with the IMMULITE Automated Immunoassay Analyzer. It is a two-site chemiluminescent enzyme-labeled immunometric assay, based on ligand-labeled monoclonal antibody and separation by anti-ligand-coated solid phase.
Here's an analysis of the provided 510(k) summary, structured to address your specific questions about acceptance criteria and the supporting study:
The provided document describes the IMMULITE® Thyroglobulin device, an in vitro diagnostic assay. It directly compares its performance to a predicate device, the ORGenTec Thyroglobulin ELISA, to demonstrate substantial equivalence, rather than setting explicit "acceptance criteria" against generalized standards.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't define explicit acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or overall agreement that the device must meet. Instead, it presents the results of a method comparison study between the IMMULITE Thyroglobulin and a predicate device (Kit A, which is ALPCO's ORGenTec). The implied acceptance is based on demonstrating comparable performance to the predicate.
| Performance Metric | Implied "Acceptance Criteria" (via predicate comparison) | Reported Device Performance (vs. Predicate Kit A) |
|---|---|---|
| Overall Agreement | Comparable to predicate | 94.0% |
| Sensitivity | Comparable to predicate | 100.0% (95% CI: 69.2% - 100%) |
| Specificity | Comparable to predicate | 93.6% (95% CI: 88.2% - 97%) |
| Linear Regression | Good correlation (r value close to 1) | IMMULITE = 2.16 x Kit A + 0.76 ng/mL; r = 0.95 |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): 150 samples.
- Data Provenance: The document does not explicitly state the country of origin. The data is retrospective, as it involves samples from patients with existing conditions (Hashimoto's disease, Graves disease, post-thyroidectomy patients) and euthyroid individuals, which were then tested with both devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This type of in vitro diagnostic device (an immunoassay) does not typically rely on "experts" to establish a ground truth in the same way an imaging AI device might. Instead, the "ground truth" for the test set is effectively derived from the predicate device's measurements (Kit A) and the clinical diagnoses of the patients from whom the samples were obtained.
- Number of Experts: Not applicable in the context of human interpretation of data for ground truth.
- Qualifications of Experts: Not applicable. The comparison is between two quantitative assays. The clinical status of the patients (e.g., Hashimoto's disease, post-thyroidectomy) provides the context for evaluating the assay's performance, which would have been established by medical professionals, but this isn't detailed as "experts establishing ground truth for the test set" in the document's context.
4. Adjudication Method for the Test Set
Not applicable. This is a direct comparison between two quantitative laboratory assays. There is no human interpretation or adjudication of individual test results in the way there would be for, say, diagnostic image review. The comparison is based on the numerical outputs of each assay.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices that assist human interpretation (e.g., AI for radiology). The IMMULITE Thyroglobulin is a standalone diagnostic assay that provides a quantitative measurement, not an AI assistance tool for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance study was done for the IMMULITE Thyroglobulin. The entire method comparison study evaluates the performance of the IMMULITE device on its own, comparing its readings directly against another standalone device (the predicate). There is no human "in the loop" for the assay's performance itself; it's an automated immunoassay.
7. The Type of Ground Truth Used
The "ground truth" in this context is established by:
- Predicate Device Results: The ORGenTec Thyroglobulin ELISA (Kit A) serves as the primary reference for comparison, essentially acting as the standard against which the new device's measurements are judged.
- Clinical Diagnoses/Patient Status: The samples were collected from patients with known clinical conditions (Hashimoto's disease, Graves disease, post-thyroidectomy) and euthyroid individuals, providing a clinical context for evaluating the agreement, sensitivity, and specificity of the assays. This implies that the clinical status, confirmed by other diagnostic means (though not explicitly detailed in the summary), contributes to the "ground truth" for categorizing samples.
8. The Sample Size for the Training Set
Not applicable. The IMMULITE Thyroglobulin is an immunoassay, not an AI/machine learning algorithm that requires a "training set" to develop its model. Its performance is based on its chemical and biological reactions, and its calibration is typically established through internal laboratory procedures using known standards, rather than a data-driven training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no "training set" in the context of an immunoassay. The device is calibrated using internal standards and controls as per standard laboratory assay development practices, not through a ground-truthed training dataset for an AI model.
{0}------------------------------------------------
3/12/99
: ・
510(k) Summary Safety and Effectiveness
This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Name: | Diagnostic Products Corporation |
|---|---|
| Address: | 5700 West 96th StreetLos Angeles, California 90045-5597 |
| Telephone Number: | (310) 645-8200 |
| Facsimile Number: | (310) 645-9999 |
| Contact Person: | Edward M. Levine, Ph.D.Director of Clinical Affairs |
| Date of Preparation: | February 12, 1999 |
| Device Name:Trade: | IMMULITE Thyroglobulin |
| Catalog Number: | LKTYZ (50 tests), LKTY1 (100 tests), LKTY5 (500 tests) |
| Common: | Reagent system for the determination of thyroglobulin inserum or heparinized plasma. |
| Classification: | Class II device, 82-JZO (21CFR 866.5870) |
| Manufacturer: | EURO/DPC Ltd. (Manufacturing under a Quality System –ISO9001/EN29001/BS 5750 |
| Sole U.S. Importer: | Diagnostic Products Corporation5700 West 96th StreetLos Angeles, California 90045-5597 |
| Establishment RegistrationNumber | EURO/DPC: Not applicableDPC: 2017183 |
| Substantially EquivalentPredicate Device: | ORGenTec Thyroglobulin ELISA (K972190)Manufactured by ORGenTec, and distributed in the USAby ALPCO, Windham, NH |
| Description of Device: | IMMULITE Thyroglobulin is a clinical device for use withthe IMMULITE Automated Immunoassay Analyzer |
{1}------------------------------------------------
Intended Use of the Device:
IMMULITE Thyroglobulin is a two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Analyzer and designed for the quantitative measurement of thyroglobulin in serum or heparinized plasma. It is intended strictly for in vitro diagnostic use as an aid in monitoring patients who have undergone thyroidectomy.
Performance Equivalence:
Diagnostic Products Corporation (DPC) asserts that the IMMULITE Thyroglobulin produces substantially equivalent results to other commercially marketed thyroglobulin assays, such as the ORGenTec Thyroglobulin ELISA. Each product is intended strictly for in vitro diagnostic use to aid in the clinical diagnosis of thyroid diseases.
Summary and Explanation of the Test:
Thyroglobulin (TG) is a heterogeneous iodoglycoprotein which has a molecular mass of approximately 660,000 daltons. Thyroglobulin is normally synthesized in the follicular cells of the thyroid gland, under the influence of thyrotropin, and represents the precursor to thyroxine and the other iodothyronines.
The expected upper limit of normal for circulating thyroglobulin is approximately 40 to 60 ng/mL, with a median of 5 to 10 ng/mL. Somewhat higher values are encountered in newborns and during the third trimester of pregnancy. Thyroglobulin levels also tend to be elevated in regions of endemic goiter.
The major clinical applications for measurement of this prohormone derive from the fact that functioning thyroid tissue, whether normal or neoplastic, appears to be the only source of circulating thyroglobulin. Accordingly, thyroglobulin determinations have been widely used to complement radioiodine scanning and other techniques (such as ultrasound or immunohistochemical staining) as an aid in identifying the presence or absence of functioning thyroid tissue, or an increase in such tissue relative to an individually established baseline. The differential diagnosis of congenital hypothyroidism and the management of nonmedullary differentiated thyroid carcinoma constitute two well-established contexts of use for this application of serum thyroglobulin measurements. ..
Congenital Hypothyroidism
Thyroglobulin determinations have been used, sometimes in conjunction with ultrasound and radioiodine scanning, to help clarify the type of thyroid defect in previously diagnosed congenital hypothyroidism. Very low or undetectable thyroglobulin levels are expected in infants born without thyroid tissue (thyroid agenesis), whereas higher, but widely varying levels are generally encountered in infants with hypoplastic thyroid glands, ectopic thyroid tissue, dyshormonogenic goiter, congenital TBG deficiency or transient hypothyroidism.
{2}------------------------------------------------
Summary and Explanation of the Test (continued):
Other Applications
Thyroglobulin measurements may also be of value in helping to distinguish subacute thyroiditis from thyrotoxicosis caused by covert administration of thyroid hormones. In the latter event, low levels of thyroglobulin are expected due to thyroid hormone suppression of thyrotropin.
Technological Comparison to Predicate:
IMMULITE Thyroglobulin is a chemiluminescent enzyme-labeled immunometric assay, based on ligand-labeled monoclonal antibody and separation by anti-ligand-coated solid phase.
The patient sample, a ligand-labeled anti-thyroglobulin monoclonal antibody and an alkaline phosphatase-labeled anti-thyroglobulin polyclonal antibody are simultaneously introduced into the Test Unit containing immobilized anti-ligand, and incubated for approximately 60 minutes at 37°C with intermittent agitation. During this time, thyroglobulin in the sample forms an antibody sandwich complex which, in turn, binds to anti-ligand on the solid phase. Unbound conjugate is removed by a centrifugal wash; substrate is then added and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex - and thus also the photon output, as measured by the luminometer - is proportional to the concentration of thyroglobulin in the sample.
The ORGenTec Thyroglobulin assay is an indirect solid phase enzyme immunometric assay based on precoated microplates in strip format (12x8 wells), ideal for smaller runs, or economical batch processing. The Thyroglobulin assay takes place in three separate phases:
Phase 1:
Microplate wells are coated with highly specific anti-TG antibodies. Standards, controls and undiluted patient samples are pipetted into the wells of the first microplate. Sample buffer is added to appropriate wells. Recovery control is added to patient sample for recovery test. Anv thyroglobulin molecules present bind to the inner well surfaces. After a 60 minute incubation, the microplate is washed with wash buffer to remove non-reactive serum components.
Phase 2:
An anti-thyroglobulin peroxidase conjugate solution is pipetted into the microplate wells which recognizes the thyroglobulin bound to the immobilized antibody. After a 60 minute incubation, excess conjugate is washed away.
、
{3}------------------------------------------------
Technological Comparison to Predicate (continued):
Phase 3:
A chromogenic substrate solution containing TMB (3,3',5,5'-tetramethyl-benzidine) is pipetted into the microplate wells. During a 15 minute incubation, the solution changes to a blue color. 1 M hydrochloric acid is added to stop color development.
The amount of color is directly proportional to the concentration of TG present in the original The optical density for each standard may be graphically plotted against the sample. concentration of TG and unknowns extrapolated from the curve. Optical density is read at 450 nm. Bichromatic measurement with a 650 nm reference is recommended.
Method Comparison:
The IMMULITE Thyroglobulin procedure was compared to Kit A (ALPCO's ORGenTec), a commercially available indirect solid phase enzyme immunometric assay for thyroglobulin, on 150 samples from patients with Hashimoto's disease, Graves disease, patients who had undergone thyroidectomy, as well as euthyroid individuals. The thyroglobulin concentrations of these specimens covered the entire calibration range of the data were tabulated in reference to the respective assays' suggested cutoffs, with the following results:
| IMMULITE | N | 150 | |||
|---|---|---|---|---|---|
| Kit A | >50 | 10 | 0 | Agreement | 94.0% |
| <=50 | 9 | 131 | Sensitivity | 100.0% | |
| >55 | <=55 | Specificity | 93.6% |
95% Confidence Limits for Relative Sensitivity and Specificity. respectively: 69.2% - 100%, 88.2% - 97%.
The same data, excluding one specimen in the above analysis due to its thyroglobulin level exceeding both assays' upper calibration limits, vielded the following result in a linear regression analysis:
$$\text{I} \textbf{M} \textbf{M} \textbf{U} \textbf{L} \textbf{T} \textbf{E} = \text{2.16} \times \textbf{K} \textbf{t} \text{ A} + \text{0.76 } \textbf{n} \textbf{g} \text{/m} \textbf{L} \qquad \qquad \textbf{r} = \textbf{0}. \textbf{95} \text{ s}$$
{4}------------------------------------------------
Conclusion:
The data presented in this summary of safety and effectiveness is the data that the Food and Drug I no add probential in alle in alle substantial equivalence for IMMULITE Thyroglobulin.
Edward L. Ferin
Edward M. Levine, Ph.D. Director of Clinical Affairs
2/12/99
Date
. .
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, composed of three curved lines that suggest wings and a head.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 1 2 1999
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, CA 90045-5597
K982466 Re: Trade Name: IMMULITE® Thyroglobulin Model LKTY Regulatory Class: II Product Code: MSW Dated: February 12, 1999 Received: February 16, 1999
Dear Dr. Levine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): Device Name: IMMULITE® Thyroglobulin
Indications For Use:
IMMULITE Thyroglobulin is a two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Analyzer and designed for the quantitative measurement of thyroglobulin in serum or heparinized plasma. It is intended strictly for in vitro diagnostic use as an aid in monitoring patients who have undergone thyroidectomy.
Peter E. Madsen
ision of Clinical Laboratory De
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
§ 866.6010 Tumor-associated antigen immunological test system.
(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.