K Number
K982238

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-08-21

(57 days)

Product Code
Regulation Number
866.3300
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Binax NOW™ Legionella Urinary Antigen Test is a rapid immunochromatographic assay for the qualitative detection of Legionella pneumophila serogroup 1 antigen (L. pneumophila serogroup 1 antigen) in human urine as an adjunct to culture for the presumptive diagnosis of Legionnaires' Disease. It is intended for in vitro diagnostic use.

Device Description

The Binax NOW™ Legionella Urinary Antigen Test is an immunochromatographic membrane assay to detect Legionella pneumophila serogroup 1 antigen in human urine. A test strip, containing gold-conjugated and immobilized anti-Legionella pneumophila serogroup 1 antibodies, and a swab well are mounted on opposite sides of a cardboard, book-shaped hinged test device. A dacron swab is dipped into the urine to be tested and then inserted into the swab well. A single reagent is added to the swab well from a dropper bottle before closing the test device. Soluble antigen captured by immobilized anti-Legionella pneumophila serogroup 1 antibody reacts to bind the visualizing gold-conjugate. There are no transferring steps, the sample is contained, and results are available within 15 minutes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Binax NOW™ Legionella Urinary Antigen Test, based on the provided text:

Acceptance Criteria and Reported Device Performance

CriteriaAcceptance Threshold (Implicit)Reported Device Performance (Binax NOW™)Predicate Device Performance (Binax Legionella Urinary Antigen EIA)
SensitivityNot explicitly stated97%95%
SpecificityNot explicitly stated98%94%
Cross-ReactivityNot explicitly stated98% specificity (against 200 non-Legionella pneumonia and UTI samples)Not explicitly stated (compared against the same 200 samples)
ReproducibilityNot explicitly stated100% (360 samples correctly interpreted)Not applicable (study performed on NOW™ test)
Quality ControlNot explicitly stated100% of 90 devices correctly interpreted (positive, negative, or invalid)Not applicable (study performed on NOW™ test)

Note: The acceptance thresholds are implicit as the submission focuses on demonstrating substantial equivalence to the predicate device. The performance of the NOW™ test is presented as being comparable or superior to the predicate device.


Study Details

2. Sample size used for the test set and the data provenance:
* Sample Size: 300 well-characterized frozen urine specimens were used for sensitivity and specificity determination. An additional 200 urine samples with other bacterial/fungal pneumonia or urinary tract infections were used for cross-reactivity testing. A separate 106 presumed negative urine specimens were also tested for specificity. For reproducibility, a panel of 360 coded specimens (negative, low, moderate, and high positive swabs) was used.
* Data Provenance: The specimens were "well characterized frozen urine specimens." The location of the main sensitivity and specificity testing was The University of Indiana. No country of origin is explicitly stated, but given the US submitter and testing location, it's highly likely to be US-based. The samples appear to be retrospective as they are "well characterized frozen urine specimens."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
* The document does not specify the number of experts or their qualifications used to establish the ground truth. It states that "Specimens were considered true positives when the presence of infection was confirmed by culture, RIA, DFA or IFA." This implies reliance on established laboratory methods rather than direct expert consensus on each case.

4. Adjudication method for the test set:
* The document does not explicitly state an adjudication method (like 2+1 or 3+1). Ground truth was established by laboratory confirmation (culture, RIA, DFA, or IFA) for true positives. For reproducibility, it was a blind study where individuals from diverse educational backgrounds performed testing across different days. The correctness of interpretation refers to agreement with the known status of the coded specimens.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done, nor is there any mention of AI assistance. This device is an immunochromatographic assay, not an AI-powered diagnostic tool. The performance is for the device itself.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
* Yes, this study represents a standalone performance evaluation of the Binax NOW™ Legionella Urinary Antigen Test. The test produces visually interpreted results, but the performance metrics (sensitivity, specificity) are based on the device's output compared to a "clinical diagnosis" (ground truth), not on a human interpreting the device's output and then being compared to another human or an AI. The reproducibility study directly assesses the device's consistency of output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
* The ground truth for true positives was established by laboratory confirmation (culture, RIA, DFA or IFA) of Legionella pneumophila infection, coupled with a "Clinical Diagnosis." For other aspects like cross-reactivity and general specificity, it relied on "presumed negative" status or confirmation of other bacterial/fungal pneumonia or urinary tract infections.

8. The sample size for the training set:
* The document does not mention a training set. This is a point-of-care immunoassay, not a machine learning model, so the concept of a training set as used in AI/ML is not applicable here.

9. How the ground truth for the training set was established:
* Not applicable, as there is no training set mentioned for this type of device.

{0}------------------------------------------------

AUG 21 1998

Binax, Inc. Binax NOW™ Legionella Urinary Antigen Test 510(k) Notification

Vital Information ... NOWTM

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS Binax NOW™ Legionella Urinary Antigen Test

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter:

Binax, Inc. 217 Read Street Portland, Maine 04103

Pamela S. Angell Attention: (207) 772-3988 Office (207) 871-5751 FAX

Binax NOW™ Legionella Urinary Antigen Test Trade Name:

Common Name: Legionella ICT

Haemophilus spp. serological reagents (per Classification Name: 21 CFR 8660.3300)

Binax Legionella Urinary Antigen EIA, 510(k) Predicate Device: number K934965/S1

The Binax NOW™ Legionella Urinary Antigen Device Description: Test is an immunochromatographic membrane detect Legionella pneumophila assay to serogroup 1 antigen in human urine. A test gold-conjugated and strip, containing anti-Legionella pneumophila immobilized serogroup 1 antibodies, and a swab well are mounted on opposite sides of a cardboard, book-shaped hinged test device. A dacron swab is dipped into the urine to be tested and then inserted into the swab well. A single reagent is added to the swab well from a dropper bottle before closing the Soluble antigen captured by test device. anti-Legionella pneumophila immobilized bind the serogroup 1 antibody reacts to visualizing gold-conjugate. There are no transferring steps, the sample is contained, and results are available within 15 minutes.

{1}------------------------------------------------

Binax, Inc. Binax NOW™ Legionella Urinary Antigen Test 510(k) Notification

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (Continued)

NOWTMLegionella Urinary Antigen Intended Use: Binax The Test is intended for in vitro diagnostic use qualitatively detect the presence of to Legionella pneumophila serogroup 1 soluble antigen in human urine. The indication statements are the same as those made for the predicate device.

Technological Characteristics : Both the Binax NOW™ Legionella Urinary and the Binax Legionella Urinary Antigen Antigen EIA tests use a solid phase coated with polyclonal rabbit antibody to detect Legionella pneumophila serogroup 1 antigen samples. However, in human urine the predicate device is an enzyme immunoassay a horseradish peroxidase employing (EIA) conjugate and a microtiter well solid phase, Binax NOW™ Legionella Urinary while the Test is an immunochromatographic Antigen assay utilizing a colloidal gold conjugate and a membrane solid phase. Furthermore, the predicate device is a multi-step assay requiring calculation of a ratio for result interpretation while the NOW™ membrane test simple and rapid test interpreted is a visually by the presence or absence of two pink-to-purple colored lines.

Performance Summary: The Binax NOW™ Legionella Urinary Antigen Test for Legionella pneumophila serogroup 1 antigen in urine is substantially equivalent predicate device, Binax Legionella to the Antigen EIA (K934965/S1) . The Urinary performance of the Binax NOW™ Legionella Test was verified by Antigen Urinary specificity, and sensitivity, reproducibility studies using well characterized frozen urine specimens and Refer to attached volunteer urine donors. CHARACTERISTICS. PERFORMANCE

Signed J.
Date 2/6/98
J. Georges Nitis, Ph.D., MBA
Director, Regulatory Affairs

{2}------------------------------------------------

Binax. Inc. Binax NOW™ Legionella Urinary Antigen Test 510(k) Notification

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (Continued)

PERFORMANCE CHARACTERISTICS

BINAX NOW™ LEGIONELLA URINARY ANTIGEN TEST

Sensitivity and Specificity:

Because of the low rate of positivity of Legionella pneumonia, test sensitivity and specificity were determined by testing 300 well characterized frozen urine specimens in both the Binax NOW™ Legionella (NOW™) and in the Binax Legionella Urinary Antigen EIA The University of Indiana conducted all testing. (EIA) Tests. EIA and NOW™ test sensitivity and specificity were calculated based on performance versus Clinical Diagnosis. Specimens were considered true positives when the presence of infection was confirmed by culture, RIA, DFA or IFA.

NOW™ test sensitivity is 97% compared to 95% for the predicate device. Now™ test specificity is 98% compared to 94% for the predicate device.

Cross-Reactivity :

for Legionella pneumophila The 200 urines presumed negative serogroup 1 were positive for other bacterial or fungal pneumonia or urinary tract infections, allowing a comparison of the NOW™ and predicate device cross-reactivities. One hundred ninety six (196) of the 200 patient specimens produced negative results in the NOW™ test for a specificity of 98%.

further demonstrate Binax test specificity, 106 presumed To negative urine specimens were tested in the Binax NOW™ Legionella Urinary Antigen Test. All specimens tested negative.

Reproducibility:

A blind study of the NOW™ test was conducted at 3 separate sites panel of coded specimens containing negative, low using a positive, moderate positive and high positive swabs. Individuals diverse educational backgrounds performed testing on ਤੇ of different days. One hundred percent (100%) of the 360 samples tested were correctly interpreted.

Quality Control:

The ability of the Binax NOW™ Legionella Urinary Antigen Test control to indicate test failure was evaluated when 3 operators each ran 20 kit controls in a panel of 20 devices, 9 of which had been rendered inoperative. The number of defective devices and the defect itself were not apparent to the operator. All ୧୦ devices were correctly interpreted as positive, negative, or invalid.

{3}------------------------------------------------

Binax, Inc. Binax, Inc.
Binax NOW™ Legionella Urinary Antigen Test 510(k) Notification

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (Continued)

PERFORMANCE CHARACTERISTICS

BINAX NOW™ LEGIONELLA URINARY ANTIGEN TEST

Preliminary Stability:

Preliminary stability studies of the Binax NOW™ Legionella Urinary Antigen Test are ongoing.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 21 1998

Pamela S. Angell Program Manager Binax, Inc. 217 Read Street Portland, Maine 04103

K982238 Re:

Trade Name: Binax NOW™ Legionella Urinary Antigen Test Regulatory Class: II Product Code: MJH Dated: August 12, 1998 Received: August 13, 1998

Dear Ms. Angell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours.

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Binax, Inc. Binax NOW™ Legionella Urinary Antigen Test 510(k) Notification

APPENDIX B

INDICATIONS FOR USE FORM

510 (k) Number (if known) :

K 98 2238

Device Name:

Binax NOW™ Legionella Urinary Antigen Test

Indications For Use:

The Binax NOW™ Legionella Urinary Antigen Test is a rapid immunochromatographic assay for the qualitative detection of Legionella pneumophila serogroup 1 antigen (L. pneumophila serogroup 1 antigen) in human urine as an adjunct to culture for the presumptive diagnosis of Legionnaires' Disease. It is intended for in vitro diagnostic use.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

(Division Sign Off)
Division of Clinical Laboratory Devices
510(k) Number K982738

Prescription Use (Per 21 CFR 801.109

Over-The-Counter Use OR

(Optional Format 1-2-96)

§ 866.3300

Haemophilus spp. serological reagents.(a)
Identification. Haemophilus spp. serological reagents are devices that consist of antigens and antisera, including antisera conjugated with a fluorescent dye, that are used in serological tests to identifyHaemophilus spp. directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusHaemophilus and provides epidemiological information on diseases cause by these microorganisms. Diseases most often caused byHaemophilus spp. include pneumonia, pharyngitis, sinusitis, vaginitis, chancroid venereal disease, and a contagious form of conjunctivitis (inflammation of eyelid membranes).(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.