K Number
K982156

Validate with FDA (Live)

Date Cleared
1998-07-29

(41 days)

Product Code
Regulation Number
888.3800
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pasco MIC and MIC/ID panels are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement of category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

This 510(k) notification is for the addition of Cefdinir to Pasco panels at concentrations of 0.03 to 8 mcg/ml for use in determining the susceptibility of methicillin susceptible strains of S. aureus. Clinical testing of Streptococci spp. including S. pneumoniae with Cefdinir has not been performed with the Pasco system.

Device Description

Varying concentrations of antimicrobial agents (usually in twofold dilutions) are dispensed into the Pasco panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours and then observed for visible growth or color changes as described in the package insert. The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Changes in pH and production of specific metabolites from growth in biochemical substrates are interpreted as described in the package insert for conventional tubed media.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and study details for the Pasco MIC and MIC/ID Panels with the addition of Cefdinir:

Acceptance Criteria and Device Performance for Pasco MIC and MIC/ID Panels (Cefdinir)

The device in question is the Pasco MIC and MIC/ID Panels, with the specific modification being the addition of Cefdinir for susceptibility testing. The study aimed to demonstrate substantial equivalence for the Cefdinir component.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state pre-defined "acceptance criteria" as clear numerical targets like "Essential Agreement must be >= 95%." Instead, it reports performance that is then qualified as "acceptable." The FDA's "Review Criteria For Assessment Of Antimicrobial Susceptibility Devices" (May 1991) is referenced as the guideline. Based on the reported results, the implicit acceptance criteria relate to achieving "acceptable Essential Agreement (EA)" and "acceptable Category Agreement (CA)" with minimal major and very major errors. Reproducibility is also assessed.

Performance MetricGram-Negative Strains (N=334)Gram-Positive Strains (N=294)Reproducibility
Essential Agreement (EA)97.6% (initial)97.7% (initial)N/A
98.8% (retesting)99.4% (retesting)
Very Major (VM) Errors7 (initial), 4 (after retest)1 (initial & retest)N/A
Major (M) ErrorsN/A3 (initial), 1 (after retest)N/A
Category Agreement (CA)97.3%95.6%N/A
Minor Errors5 (random)9 (random)N/A
ReproducibilityN/AN/A98.8% within +/- 1 dilution

2. Sample Sizes Used for the Test Set and Data Provenance:

  • Gram-Negative Strains: 334 strains
  • Gram-Positive Strains: 294 strains
  • Reproducibility Testing: 9 organisms at each of the two test sites.
  • Data Provenance: The study used "CDC challenge strains and clinical isolates." This indicates a mix of reference strains for quality control/standardization and real-world clinical samples, suggesting a robust evaluation. The origin (country) of the clinical isolates is not specified, but the testing was performed at two sites (also not explicitly named, but one appears to be Pasco itself, and the other is referred to as "CMI"). The study appears to be prospective in nature for the comparative testing, as panels were "prepared in-house at Pasco using routine manufacturing procedures" and then tested.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not specify the number or qualifications of experts used to establish the ground truth. It refers to comparing the Pasco test panel to a "reference panel" and uses "CDC challenge strains." This strongly implies that the ground truth was established by a reference method (e.g., broth microdilution or agar dilution as per CLSI/NCCLS standards for antimicrobial susceptibility testing) and potentially confirmed by expert microbiologists, but the details are absent.

4. Adjudication Method for the Test Set:

The document mentions "retesting" for discrepant results. For example:

  • "Seven Very Major (VM) errors were observed... one E. aerogenes which was resolved upon retest, three P. rettgeri with one of the VM errors resolved upon retest..."
  • "...two E. faecium strains, with one M error resolved upon retest and one E. faecalis strain which resolved upon retest."

This suggests a form of post-hoc adjudication or re-testing of discrepant results. The term "resolved" implies that after re-testing, the initial discrepancy either disappeared or was understood in the context of the reference method. There's no explicit mention of an a priori adjudication protocol like 2+1 or 3+1 expert consensus, but rather a re-evaluation of outliers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

This type of study is not applicable to the device described. The Pasco MIC and MIC/ID Panels are in vitro diagnostic devices for antimicrobial susceptibility testing, which typically involve automated or semi-automated reading of growth in wells, not human image interpretation or AI assistance in the way an MRMC study would evaluate. Therefore, no MRMC study was performed, and no effect size would be relevant in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, the performance reported (Essential Agreement, Category Agreement, and reproducibility of the device itself) represents standalone performance of the Pasco panels with Cefdinir. While human technicians perform the inoculation and read the results (or interpret automated readers), the "device performance" metrics are for the panel's ability to accurately determine susceptibility, independent of any AI insights or human "improvement." The "Summary/Conclusion of Substantial Equivalence Testing" directly addresses the panel's output against a reference standard.

7. The Type of Ground Truth Used:

The ground truth was established by comparing the Pasco test panel results to a reference panel. This reference panel would typically be a gold standard method for antimicrobial susceptibility testing, such as the broth microdilution or agar dilution methods prescribed by standards organizations like NCCLS (now CLSI). The use of "CDC challenge strains" further reinforces the use of highly characterized reference organisms.

8. The Sample Size for the Training Set:

The document does not provide information on a training set. The described study appears to be the validation or testing phase for the Cefdinir addition. For an in vitro diagnostic device like this, the "training" phase would typically involve the manufacturer's internal development and optimization of the panel's formulation and concentrations using various strains and reference methods, but these details are not part of this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established:

As no training set is described in the provided text, the method for establishing its ground truth is not available.

{0}------------------------------------------------

K982156

(2)

JUL 29 1998

510 (k) SUMMARY (page 1 of 3)

DATE:

JUNE 16, 1998

Linda K. Dillon CONTACT PERSON: Barbara C. Lamoureux

Pasco MIC and MIC/ID Panels TRADE NAME OF DEVICE:

Antimicrobial Susceptibility Test COMMON NAME:

Class II Antimicrobial Susceptibility CLASSIFICATION NAME: Test Microbiology Panel #83

SUBSTANTIAL EQUIVALENCE:

In review of previous 510(k) notifications for the Pasco MIC and MIC/ID panels (most recently K980955 May 18, 1998 RE: Trovafloxacin; K974362, February 12, 1998 RE: Cefepime; K973317, November 14, 1997 RE: Cefpodoxime; K973695, November 5, 1997 RE: November 14, 1997 km² Sulpedo.im97 RE: Sparfloxacin; K971951, August 15, 1997 RE: Levofloxacin; and K946126, January 17, 1995 Detection of Resistant pneumococci), the FDA has determined RE: the Pasco panels to be substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments.

The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.

DESCRIPTION OF THE DEVICE:

Varying concentrations of antimicrobial agents (usually in twofold dilutions) are dispensed into the Pasco panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours and then observed for visible growth or color changes as described in the package insert.

{1}------------------------------------------------

The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Changes in pH and production of specific metabolites from growth in biochemical substrates are interpreted as described in the package insert for conventional tubed media.

INTENDED USE FOR THE PASCO MIC AND MIC/ID PANELS: PASCO MIC AND MIC/ID PANELS are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement or category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

SUMMARY/CONCLUSION OF SUBSTANTIAL EQUIVALENCE TESTING: Test panels containing Cefdinir at concentrations ranging from 8 to 0.03 mcg/ml were prepared in-house at Pasco using routine manufacturing procedures. Comparative testing of the Pasco test panel to a reference panel was performed at two sites using CDC challenge strains and clinical Isolates.

Test results of the 334 gram-negative strains demonstrated acceptable Essential Agreement (EA) of 97.6% upon initial testing and 98.8% upon retesting. Seven Very Major (VM) errors were observed with the following strains: one E. aerogenes which was resolved upon retest, three P. rettgeri with one of the VM errors resolved upon retest, two P.vulgaris and one P. stuartii which was an EA error upon retest. Category Agreement (CA) was 97.3% with five random minor errors noted

Further testing was conducted in a effort to determine the reason for the Very Major discrepancies which occurred with the P. vulgaris, P. rettgeri and P. stuartii strains at both test sites. To determine if the errors resulted from inoculum concentration differences, the inoculum for both the Pasco test method and reference panels were carefully adjusted to 5 x 10 CFU/ml, however, the differences were not resolved. Since these errors occurred at both sites, primarily with the same strains, technique and equipment related problems were ruled out. Cefdinir is not used in treatment of infections with these organisms as these species are not included in the Indications and Usage statements of the PARKE-DAVIS package insert.

{2}------------------------------------------------

510 (k) SUMMARY (page 3 of 3)

Test results of the 294 gram-positive strains demonstrated acceptable Essential Agreement (EA) of 97.7% upon initial testing and 99.4% upon retesting. Three Major (M) errors were observed: two E. faecium strains, with one M error resolved upon retest and one E. faecalis strain which resolved upon retest. One Very Major (VM) error occurred with a coagulase negative oxacillin (R) staph strain which was not resolved upon retest. Category Agreement (CA) was 95.6% with nine random minor errors noted.

Further testing of the Very Major discrepancy with the coagulase negative oxacillin (R) staph was conducted as previously described above, however, the cause of the discrepancy was not determined. Since oxacillin-resistant Staph. are resistant to all currently available ß-lactam antibiotics there is no indication of clinical use of this agent for this organism based on the PARKE-DAVIS package insert and NCCLS document M100-S8. Additionally, Cefdinir is not effective clinically for Enterococcus spp. (Major errors listed above) and therefore should not be reported for cephalosporins (NCCLS document M100-S8).

QC endpoints from both the reference and Pasco panels through 25 days of testing at CMI and 26 days of testing at Pasco were within the recommended acceptable ranges listed in the product information and NCCLS with the exception of one test panel for S. aureus ATCC 29213 and one test panel for M. morganii ATCC 7028 both of which were performed at CMI.

Reproducibility testing of 9 organisms at each site provided 5 onscale endpoints. Overall reproducibility data demonstrated 98.8% within the acceptable plus or minus 1 dilution.

The results of the clinical testing, reproducibility testing and QC performance testing supports Substantial Equivalence as outlined in the FDA draft document "Review Criteria For Assessment Of Antimicrobial Susceptibility Devices" (May 1991).

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles incorporated into its design. The eagle's head is facing to the right.

The 29 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Linda K. Dillon Technical Manager Pasco Laboratories, Inc. 12750 West Forty-Second Avenue Wheat Ridge, Colorado 80033

Re: K982156

Trade Name: Pasco MIC and MIC/ID Panels/Celdinir Regulatory Class: II Product Code: JWY Dated: June 16, 1998 Received: June 18, 1998

Dear Ms. Dillon:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Device Name:

PASCO MIC and MIC/ID Panels; Inclusion of Cefdinir

Indication For Use:

Pasco MIC and MIC/ID panels are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement of category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

This 510(k) notification is for the addition of Cefdinir to Pasco panels at concentrations of 0.03 to 8 mcg/ml for use in determining the susceptibility of methicillin susceptible strains of S. aureus. Clinical testing of Streptococci spp. including S. pneumoniae with Cefdinir has not been performed with the Pasco system.

Woody Dubois

Division Sig Division of Clinical Laboratory Devices K98215 510(k) Number_

Prescription Use (Per 21 CFR 801 OR

Over-The Counter Use

(Optional Format 1-2-96)

§ 888.3800 Wrist joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A wrist joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a wrist joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have either a one-part radial component made of alloys, such as cobalt-chromium-molybdenum, with an ultra-high molecular weight polyethylene bearing surface, or a two-part radial component made of alloys and an ultra-high molecular weight polyethylene ball that is mounted on the radial component with a trunnion bearing. The metallic portion of the two-part radial component is inserted into the radius. These devices have a metacarpal component(s) made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.