K Number
K981899

Validate with FDA (Live)

Date Cleared
1998-07-14

(43 days)

Product Code
Regulation Number
862.1475
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CARESIDE™ HDL-Cholesterol cartridge is intended for in vitro diagnostic use in conjunction with the Exigent Diagnostics CARESIDE™ Analyzer to quantitatively measure HDL-cholesterol concentration in pre-treated serum or plasma. This product is indicated for use in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism.

Device Description

CARESIDE™ HDL-Cholesterol cartridges are used with the Exigent Diagnostics CARESIDE™ Analyzer to measure HDL-cholesterol concentration in pre-treated serum or plasma specimens. The CARESIDE™ HDL-Cholesterol cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of pre-treated serum or plasma to a dry film to initiate the measurement of HDL-cholesterol concentration. The film cartridge (patent pending) contains all reagents necessary to measure HDL-cholesterol concentration.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the CARESIDE™ HDL-Cholesterol device, based on the provided text:

Acceptance Criteria and Device Performance

CriteriaAcceptance Criteria (Predicate)Reported Device Performance (CARESIDE™ HDL-Cholesterol)
Detection Limit1 mg/dL10 mg/dL
Reportable Range1 to 110 mg/dL10 to 110 mg/dL
AccuracyNot providedMean recovery 109%
PrecisionTotal CV, 41 mg/dL, 3.2%Total CV, 33 mg/dL, 6.5%
Method Comparison (Correlation to Predicate)N/A (Predicate vs. Unknown)CARESIDE™ = 0.97 (Vitros HDLC DT) - 1.5 mg/dL, r = 0.97
LinearityNot providedLinearity by mixing and by dilution yielded slope and correlation coefficient within acceptable limits.
InterferenceAscorbic Acid interferes (non-interfering concentration undefined)No significant interference observed at tested concentration of interferent: Ascorbic Acid, 1 mg/dL; Bilirubin, 10 mg/dL; Hemoglobin, 250 mg/dL; Protein 5 - 9 g/dL; Triglycerides 1500 mg/dL

Note: The acceptance criteria are inferred from the reported performance of the predicate device (Vitros HDLC DT Slides) where specific criteria for the new device are not explicitly stated, or from the new device's demonstrated performance exceeding or meeting expectations related to the predicate.

Study Information

  1. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the sample size used for the test set.
    • The document does not explicitly state the country of origin of the data.
    • The document does not explicitly state if the data was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The study focuses on comparing the device's measurements to a predicate device, not on expert-adjudicated ground truth.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not applicable or provided. The study is a quantitative measurement comparison, not an analysis requiring adjudication of interpretations.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a MRMC comparative effectiveness study was not done. This is a study comparing an in vitro diagnostic device's quantitative measurements to a predicate device, not an AI-assisted human reader study.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, a standalone study was done. The performance data presented (accuracy, precision, method comparison, linearity, interference) describes the algorithm's performance in measuring HDL-cholesterol concentration in pre-treated serum or plasma specimens, without human-in-the-loop performance influencing the measurement itself. The "CARESIDE™ HDL-Cholesterol" device, in conjunction with the CARESIDE™ Analyzer, performs the measurement automatically.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The primary ground truth for the comparative performance study was the measurements obtained from the predicate device, Vitros HDLC DT Slides. Additionally, some tests (like linearity) might have been assessed against theoretical expectations or known concentrations of reference materials. The "Reference Method" for the CARESIDE™ is stated as "Abell Kendall," implying this might be a recognized standard against which the predicate or even the new device could be ultimately traceable, though the direct comparison in the table is against the Vitros HDLC DT.
  7. The sample size for the training set:

    • The document does not explicitly mention a "training set" or its sample size. This type of premarket notification for an IVD kit typically presents validation data, not a machine learning model's training details.
  8. How the ground truth for the training set was established:

    • As no training set is mentioned (see point 8), the method for establishing its ground truth is also not applicable or provided.

{0}------------------------------------------------

81899

Exigent Diagnostics, Inc. Page 11

CARESIDE™ HDL-Cholesterol Premarket Notification May 29, 1998

510(K) SUMMARY: CARESIDE™ HDL-CHOLESTEROL IV. SAFETY AND EFFECTIVENESS

I. Applicant Information

  • A. Applicant Name
  • Applicant/Manufacturer Address B.
  • C. Telephone Number
  • D. Contact Person
  • E. FAX Number
  • e-Mail Address F.
  • Date 510(k) Summary prepared G.

II. Device Information

  • A. Device Name (Trade)
  • Device Name (Classification) B.
  • C. Device Classification

Exigent Diagnostics, Inc.

6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 asarchk @ worldnet.att.net Mav 29, 1998

CARESIDE™ HDL-Cholesterol

HDL-Cholesterol test system Clinical chemistry panel HDL-Cholesterol test system Regulation Number: 21 CFR 862.1475 Regulatory Class I Classification Number: 75LBS None applicable

  • D. Special controls and performance standards

III. Substantial Equivalence Claim

  • A. General equivalency claim
    The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.

HDL-Cholesterol in vitro diagnostic products, in both dry film and other formats, are already on the U.S. market, including HDL-cholesterol products which utilize enzymatic (cholesterol oxidase) generation of hydrogen peroxide from free and liberated cholesterol which reacts with chromogens in a peroxidase catalyzed reaction to form a dye.

B. Specific equivalency claim

This CARESIDE™ HDL-Cholesterol test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of HDL-cholesterol on the Vitros DT 60 II.

Name of Predicate Device:Johnson and Johnson's (formerly Eastman Kodak,Inc.) Vitros HDLC Slides for Johnson and Johnson'sVitros DT 60 (formerly Eastman Kodak's DT 60 II).
Predicate Device 510K number:K912844/A
Product Code:75LBS

{1}------------------------------------------------

IV. Device Description

CARESIDE™ HDL-Cholesterol cartridges are used with the Exigent Diagnostics CARESIDE™ Analyzer to measure HDL-cholesterol concentration in pre-treated serum or plasma specimens. The CARESIDE™ HDL-Cholesterol cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of pre-treated serum or plasma to a dry film to initiate the measurement of HDL-cholesterol concentration. The film cartridge (patent pending) contains all reagents necessary to measure HDL-cholesterol concentration.

Explanation of Device Function A.

Each Exigent Diagnostics CARESIDE™ HDL-Cholesterol cartridge consists of a HDL-cholesterol-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the pre-treated specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the Exigent Diagnostics CARESIDE™ Analyzer.

Once loaded, the CARESIDE™ analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37°C, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. Ten microliters of sample remains in the metering passage. Any excess sample flows into an overflow well.

The ten microliters of pre-treated sample is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the sample evenly on the film, and the sample moves through a reflection layer into the detection layer. Cholesterol is dissociated from protein by the action of lipoprotein lipase (LPL) and a surfactant. Cholesterol esterase (CHE) hydrolyzes the cholesterol esters (CHOL-ester) to liberated cholesterol. Endogenous cholesterol and the liberated cholesterol react with oxygen and water to produce hydrogen peroxide in a cholesterol oxidase (CO) catalyzed reaction. Peroxidase (POD) then catalyzes the reaction of the hydrogen peroxide with a diarylimidazole leuco dye to produce a blue chromogen. The color intensity of the resulting blue dye, as measured by the amount of reflected light at 615 nanometers directly relates to the HDL-cholesterol concentration of the specimen.

Test Reaction Sequence:

LPL> CHOL + CHOL-Ester+ Protein HDL-Cholesterol -CHOL-Ester - CHE > CHOL (free) -> H2O2 + Cholesten-3one CHOL + O2 + H2O -POD Fe+2 Blue dye Leuco dye + H2O2 -

As the cartridges spin, a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) at a fixed time. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate HDL-cholesterol concentration.

{2}------------------------------------------------

B. Test Summary

HDL-Cholesterol in blood comprises all of the cholesterol found in various lipoproteins. Cholesterol is the major component of the low-density lipoprotein fraction, and a minor component of the very low-density lipoprotein and high-density lipoprotein fractions. Cholesterol measurements are used in classifying patients according to risk of coronary heart disease, in the diagnosis and treatment of various primary or secondary hyperlipidemias, and to monitor changes in cholesterol levels resulting from treatment. Elevated LDLC has consistently been associated with incidence of atherosclerosis. There is also a strong correlation between considerably elevated cholesterol levels and an increased tendency for atherosclerosis. However, HDL- cholesterol concentration and cardiovascular disease risk are inversely related. Measurement of total and HDLcholesterol in serum is useful in evaluating cardiovascular disease risk.

V. Intended Use

  • A. Intended Use
    The CARESIDE™ HDL-Cholesterol cartridge is intended for in vitro diagnostic use in conjunction with the Exigent Diagnostics CARESIDE™ Analyzer to quantitatively measure HDL-cholesterol concentration in pre-treated serum or plasma.

  • B. Indications for Use
    This product is indicated for use in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism.

{3}------------------------------------------------

VI. Technological Characteristics

  • Similarities A.
CARESIDE™HDL-CholesterolVitros HDLC DT Slides
Intended UsePrimarily to aid in the diagnosisand treatment of patients withdisorders of lipid andlipoprotein metabolism.Same
IndicationsFor in vitro diagnostic use.For professional laboratory.For in vitro diagnostic use
MeasurementQuantitativeSame
Method PrincipleDry film based already on theU.S. market, includingHDL-cholesterol productswhich utilize cholesterolesterase sample treatmentfollowed by cholesterol oxidasegeneration of hydrogenperoxide from free andliberated cholesterol whichreacts with a leuco dye in aperoxidase catalyzed reaction toform a blue dye.Same
Specimen dilutionNot requiredSame
MaterialsLipoprotein lipase, cholesterolesterase, cholesterol oxidase,peroxidase, and diarylimidazoleleuco dyeCholesterol esterase,cholesterol oxidase, peroxidase,and diarylimidazole leuco dye
DetectorReflectance (615 nm)Reflectance (660 nm)
Test timeApprox. 4 minute warm-up (on-board) plus 6 minute test time.15 minutes slide warm-up (off-line) plus 5 minutes test time.
Reference MethodAbell KendallNot known
Sample TypePre-treated serum or plasmaSame
Specimen volume10 µl test volume(85 ± 15 µl applied volume)10 µl
CalibrationCalibration information bar-coded on each cartridge.Calibration information maychange with each lot.Run Vitros DT II calibratorswhenever a new slide lot isused or when necessary.
Quality Control2 levelsSame
Reporting Unitsmg/dL or mmol/LSame
Reaction Temp.37 °CSame

B. Differences


CARESIDE™HDL-CholesterolVitros HDLC DT Slides
Specimen pre-treatmentRequiredRequired
Reportable range10 to 110 mg/dL1 to 110 mg/dL
AccuratepipettingNot requiredRequired
Reagent pre-warmingNot requiredRequired

{4}------------------------------------------------

Comparative Performance Characteristics C.

CARESIDE™HDL-CholesterolVitros HDLC DT Slides
Detection limit10 mg/dL1 mg/dL
Reportable range10 to 110 mg/dL1 to 110 mg/dL
AccuracyMean recovery 109%Not provided
PrecisionTotal CV, 33 mg/dL, 6.5%Total CV, 41 mg/dL, 3.2%
MethodcomparisonCARESIDE™ = 0.97 (Vitros HDLC DT) - 1.5 mg/dL, r = 0.97
LinearityLinearity by mixing and bydilution yielded slope andcorrelation coefficient withinacceptable limits.Not provided
InterferenceNo significant interferenceobserved at testedconcentration of interferent:Ascorbic Acid, 1 mg/dLBilirubin, 10 mg/dLHemoglobin, 250 mg/dLProtein 5 - 9 g/dLTriglycerides 1500 mg/dLAscorbic Acid interferes(non-interfering concentrationundefined)

D. Conclusion

The nonclinical and clinical data provided demonstrate that the CARESIDE™ HDL-Cholesterol product is as safe, effective, and performs as well as or better than the legally marketed predicate device

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circle's perimeter.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 1 4 1998

Kenneth B. Asarch, Ph.D. . VP Quality Systems and Regulatory Affairs Exigent Diagnostics Inc. 6100 Bristol Parkway 90230 Culver City, California

K981899 Re : CARESIDE™ HDL-Cholesterol Regulatory Class: I Product Code: LBS Dated: May 29, 1998 Received: June 1, 1998

Dear Dr. Asarch:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸੋ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2 -

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655 .

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" Other general information on your (21 CFR 807.97). responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

VI. INDICATIONS FOR USE

510(k) Number:

Device Name:

CARESIDE™ HDL-Cholesterol

For in vitro diagnostic use with Exigent Diagnostics' CARESIDE™ Analyzer to Indications for use: measure HDL-cholesterol from pre-treated serum or plasma specimens to aid in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism.

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK981899

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 862.1475 Lipoprotein test system.

(a)
Identification. A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.