K Number
K981839

Validate with FDA (Live)

Date Cleared
1998-08-12

(78 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chiron Diagnostics ACS:Centaur PSA2 Immunoassay is for the quantitative determination of prostate specific antigen in serum to aid in the management of cancer patients in whom changing concentrations of PSA are observed using the Chiron Diagnostics ACS:Centaur Automated Chemiluminescence Systems.

Device Description

Prostate-specific antigen (PSA) is a single-chain glycoprotein normally found in the cytoplasm of the epithelial cells lining the acini and ducts of the prostate gland. PSA is a neutral serine protease of 240 amino acids involved in the lysis of seminal coagulum.

PSA is detected in the serum of males with normal, benign hypertrophic, and malignant prostate tissue. PSA is not detected in the serum of males without prostate tissue (because of radical prostatectomy or cystoprostatectomy) or in the serum of most females. The fact that PSA is unique to prostate tissue makes it a suitable marker for monitoring men with cancer of the prostate. PSA is also useful for determining possible recurrence after therapy when used in conjunction with other diagnostic indices.

The Chiron Diaqnostics ACS:Centaur PSA2 immunoassay is a two-site immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies. The first antibody, in the Lite Reagent, is a purified polyclonal sheep anti-PSA antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-PSA antibody covalently coupled to paramagnetic particles. A direct relationship exists between the amount of PSA present in the patient sample and the amount of relative light units (RLU's) detected by the system.

AI/ML Overview

Acceptance Criteria and Device Performance for Chiron Diagnostics ACS:Centaur PSA2 Immunoassay (K981839)

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
SensitivityMinimum detectable concentration for PSA measurement0.06 ng/mL (measures up to 135 ng/mL)
Accuracy (Correlation with Predicate Device)High correlation (e.g., r > 0.95 or similar) with the predicate device (ACS:180 PSA2) to demonstrate equivalence.ACS:Centaur PSA2 = 0.97 (ACS:180 PSA2) + 0.19 ng/mL; Correlation coefficient (r) = 0.99

Note: The document does not explicitly state numerical acceptance criteria in a dedicated section. The "Performance Data" section details the device's capabilities, and the high correlation with the predicate device (r = 0.99) strongly implies that this level of agreement was the acceptance criteria for accuracy to demonstrate substantial equivalence. The minimum detectable concentration demonstrates the lower limit of reliable measurement, implying an acceptance of this level of analytical sensitivity.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 287 samples were used for the accuracy study comparing the ACS:Centaur PSA2 with the ACS:180 PSA2.
  • Data Provenance: The document does not specify the country of origin of the data. It is a retrospective analysis comparing the performance of the new device against the predicate device using existing samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This device is an immunoassay for quantitative determination of PSA levels. The "ground truth" in this context is the quantitative measurement of PSA itself. Therefore, human experts are not directly involved in establishing the ground truth for individual samples in the same way they would be for image interpretation.

The ground truth for the comparison was established by the predicate device, the ACS:180 PSA2 Immunoassay. The performance of this predicate device would have been validated previously, likely involving method comparison studies with established reference methods or clinical samples.

4. Adjudication Method for the Test Set

The concept of an adjudication method (e.g., 2+1, 3+1) is not applicable here as the test set involves quantitative measurements from an immunoassay, not subjective interpretations requiring expert consensus. The comparison is a direct numerical correlation between the new device and the predicate device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was conducted or is applicable for this type of device. This device is an automated immunoassay, meaning there are no human "readers" in the diagnostic process to compare or improve.

6. Standalone Performance Study

Yes, a standalone performance study was conducted. The "Performance Data" section describes the sensitivity and accuracy of the ACS:Centaur PSA2 Immunoassay as a standalone device, as it measures PSA concentration independently and then compares its output to the predicate device. The correlation study directly assesses the performance of the algorithm (the immunoassay technology) without human intervention in the measurement process.

7. Type of Ground Truth Used

The ground truth used for the comparison was measurements from a legally marketed predicate device (ACS:180 PSA2 Immunoassay). This is a form of comparative ground truth, where the performance of the new device is validated against an existing, accepted method. While not pathology or outcomes data directly, the predicate device's measurements are considered the established "truth" for this comparison of analytical performance. The clinical utility of PSA measurements (which pathology and outcomes data would inform) is assumed to be established by the predicate device's history and the broader medical literature.

8. Sample Size for the Training Set

The document does not specify a training set or its sample size. Immunoassays are generally developed and optimized through extensive experimentation and reagent optimization, rather than a "training set" in the machine learning sense. The performance data presented are for the validated, final device.

9. How the Ground Truth for the Training Set Was Established

As no explicit training set is mentioned in the context of the immunoassay, the concept of establishing ground truth for it as per AI/ML terminology is not applicable. The development of the immunoassay involves chemical and biological engineering principles, with validation steps along the way, not a "ground truth" derived from expert consensus on a training dataset. The established "truth" for such systems typically comes from meticulously prepared reference materials, calibrators, and robust analytical methods.

{0}------------------------------------------------

K981839

AUG / 2 1998

Company Confidential Summary of Safety and Effectiveness

As required by 21 CFR 807.92, the following 510(k) Summary is provided:

1. Submitter Information

William J. Pignato Contact person:

Address:

Chiron Diagnostics Corporation 63 North Street Medfield, MA 02052

Phone: (508) 359-3825 (508) 359-3885 FAX: e-mail: william.pignato@chirondiag.com

Date Summary Prepared:

May 18, 1998

2. Device Information

ACS:Centaur PSA2 Immunoassay Proprietary Name: PSA Immunoassay Common Name: Reclassified to Class II, classification number Classification Name: unknown

3. Predicate Device Information

Name:ACS:180 PSA2 Immunoassay
Manufacturer:Chiron Diagnostics
510(k) Number:P920030 & P020030/S1 (note reclassified to class II)

4. Device Description

Prostate-specific antigen (PSA) is a single-chain glycoprotein normally found in the cytoplasm of the epithelial cells lining the acini and ducts of the prostate gland. PSA is a neutral serine protease of 240 amino acids involved in the lysis of seminal coagulum.

PSA is detected in the serum of males with normal, benign hypertrophic, and malignant prostate tissue. PSA is not detected in the serum of males without prostate tissue (because of radical prostatectomy or cystoprostatectomy) or in the serum of most females. The fact that PSA is unique to prostate tissue makes it a suitable marker for monitoring men with cancer of the prostate. PSA is also useful for determining possible recurrence after therapy when used in conjunction with other diagnostic indices.

{1}------------------------------------------------

Company Confidential

Measurement of serum PSA levels is not recommended as a screening procedure for the diagnosis of cancer because elevated PSA levels also are observed in patients with benign prostatic hypertrophy. However, studies suggest that the measurement of PSA in conjunction with digital rectal examination (DRE) and ultrasound provide a better method of detecting prostate cancer than DRE alone.

PSA levels increase in men with cancer of the prostate, and after radical prostatectomy PSA levels routinely fall to the undetectable range. If prostatic tissue remains after surgery or metastasis has occurred, PSA appears to be useful in detecting residual and early recurrence of tumor. Therefore, serial PSA levels can help determine the success of prostatectomy, and the need for further treatment, such as radiation, endocrine or chemotherapy, and in the monitoring of the effectiveness of therapy.

5. Statement of Intended Use

The Chiron Diagnostics ACS:Centaur PSA2 Immunoassay is for the quantitative determination of prostate specific antigen in serum to aid in the management of cancer patients in whom changing concentrations of PSA are observed using the Chiron Diagnostics ACS:Centaur Automated Chemiluminescence Systems.

Summary of Technological Characteristics 6.

The Chiron Diaqnostics ACS:Centaur PSA2 immunoassay is a two-site immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies. The first antibody, in the Lite Reagent, is a purified polyclonal sheep anti-PSA antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-PSA antibody covalently coupled to paramagnetic particles. A direct relationship exists between the amount of PSA present in the patient sample and the amount of relative light units (RLU's) detected by the system.

7. Performance Data

Sensitivity

The ACS:Centaur Immunoassay measures PSA concentration up to 135 ng/mL with a minimum detectable concentration of 0.06 ng/mL.

Accuracy

For 287 samples in the range of .07 to 111.25 ng/mL, the correlation between the ACS:Centaur PSA2 and the ACS:180 PSA2 is described by the equation:

ACS:Centaur PSA2 = 0.97 (ACS:180 PSA2) + 0.19 ng/mL

Correlation coefficient (r) = 0.99

ACS:Centaur PSA2 510(k)

22 May, 1998

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing right, with flowing lines beneath them.

AUG | 2 |998

Mr. Thomas F. Flynn Manager, Requlatory Affairs & Compliance Chiron Diagnostics Corporation 63 North Street Medfield, Massachusetts 02052

Re: к981839/51 ACS:Centaur PSA2 Immunoassay Trade Name: Regulatory Class: II Product Code: LTJ Dated: August 3, 1998 Received: August 4, 1998

Dear Mr. Flynn:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Company Confidential

Page _ of _

510(k) Number (if known): K981839

Device Name: Chiron Diagnostics ACS:Centaur PSA2 Immunoassay

Indications for Use:

The Chiron Diagnostics ACS:Centaur PSA2 Immunoassay is for the quantitative determination of prostate specific antigen in serum to aid in the management of cancer patients in whom changing concentrations of PSA are observed using the Chiron Diagnostics ACS:Centaur Automated Chemiluminescence Systems.

Peter E. Maurer

(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

ACS:Centaur PSA2 510(k)

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.