K Number
K981498

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-08-21

(116 days)

Product Code
Regulation Number
866.3780
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diamedix Is-Toxoplasma IgG Test Kit is an Enzyme Immunoassay (EIA) for the qualitative and quantitative determination of IgG antibodies in human serum to aid in the assessment of the patient's immunological response to infection with Toxoplasma gondii and in the determination of the immune status of individuals, including females of child-bearing age. The evaluation of acute and convalescent sera can aid in the diagnosis of primary or reactivated infection with Toxoplasma gondii. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor. This product is not FDA cleared for use in screening blood and plasma donors.

Device Description

The Is-Toxoplasma IgG Test System is an enzyme-linked immunosorbent assay (ELISA) for the qualitative and quantitative detection of IgG to Toxoplasma gondii in human serum

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Is-Toxoplasma IgG Test System:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative, pre-defined manner for the comparison studies. Instead, it presents the results of these comparisons and implicitly suggests that these results were deemed acceptable for market clearance. For the purpose of this table, I will infer the "acceptance criteria" based on common expectations for such devices and the reported "predicate device performance" as the target. The "reported device performance" will be the "Is-Toxoplasma IgG" results.

MetricAssumed/Inferred Acceptance Criterion (Based on Predicate Performance)Is-Toxoplasma IgG Test System Reported Performance (Range across sites)
Qualitative Detection
Relative SensitivityHigh (e.g., >85-90%)87.5% - 93.3% (manual), 92.2% - 92.5% (Diamedix, manual & MAGO+)
Relative SpecificityHigh (e.g., >95%)98.7% - 100.0% (manual), 99.5% (Diamedix, manual & MAGO+)
Overall AgreementHigh (e.g., >90%)92.1% - 99.4% (manual), 97.5% - 97.9% (Diamedix, manual & MAGO+)
Precision (Interassay CV%)Low (e.g., typically <15-20% for quantitative assays)Range across various serum samples and sites: <1% - 82.69%
Linearity (R-squared)High (e.g., >0.95 or >0.98)0.9969 (Manual), 0.9973 (MAGO Plus)
Correlation to WHO Standard (R-squared)High (e.g., >0.95 or >0.98)0.9932
Cross-ReactivityNo significant cross-reactivity with common interfering analytes0 positive results out of 50 samples positive for various viruses/ANA

2. Sample Sizes and Data Provenance:

  • Test Set for Comparison Studies:
    • Site #1 (Miami, FL): 200 samples (37% fresh, 73% frozen). Samples from the S. Florida area. Purposefully selected for equal positive/negative results. Retrospective/Prospective status not explicitly stated for individual samples, but likely a mix.
    • Site #2 (Salt Lake City, Utah): 179 samples (all fresh). Samples from the West region. Submitted for ToRCH screening. Prospective.
    • Site #3 (Diamedix Corp., Miami, FL): 242 samples (all frozen) for manual testing, 241 for MAGO Plus testing. Samples from S. Florida blood donors. Retrospective.
    • Total for comparison studies: 200 + 179 + 242 (manual) / 241 (MAGO Plus) = 621-622 samples total across all sites/methods.
  • Test Set for Cross-Reactivity: 50 sera, negative for T. gondii IgG, but positive for one or more of VZV, HSV, CMV, Rubella, EBV, anti-DNA, ANA. Provenance not specified, but implied to be human serum samples.
  • Test Set for Precision: 6 serum samples, 50 IU/ml kit Standard, Low Positive Control, and Negative Control. Tested in triplicate across 3 days at 3 sites (Site #1, Site #2, Site #3).
  • Test Set for Linearity: Several strongly positive serum samples (number not specified).
  • Test Set for WHO Standard Correlation: Several dilutions of the WHO 3rd International Standard for Anti-Toxoplasma Serum (code TOXM).
  • Test Set for Quantitative Data/Serum Pairs: Multiple two-fold dilutions of several strongly positive sera (number not specified).
  • Test Set for Expected Values (Prevalence): 100 healthy South Florida donors (52 female, 48 male). 37 female donors of child-bearing age (18-45 years) from this group. 45 sera from pregnant females (15 from each trimester) (separate from above 100, but unclear if they overlap with the "outside and in-house clinical studies" mentioned later). 216 samples from females of childbearing age identified in "outside and in-house clinical studies" (includes the 45 pregnant females).

3. Number of Experts and Qualifications for Ground Truth:

  • The document does not explicitly state that experts were used to establish the ground truth for the entire test set in the primary comparison studies.
  • Instead, comparison was done against "their currently used testing method" or "another marketed EIA method." The performance of these comparator assays served as the initial "relative ground truth."
  • For discordant samples, "further resolution" was performed by testing them in a "referee EIA method." The qualifications of the individuals running or interpreting this referee method are not specified, nor is the number of experts.

4. Adjudication Method for the Test Set:

  • For the comparison studies, there wasn't a formal adjudication method of multiple independent readers or experts establishing a primary ground truth against which both devices were judged.
  • Instead, for discordant results between the Is-Toxoplasma IgG Test Kit and the comparator EIA, a "referee EIA method" was used for resolution (essentially a third-party tie-breaker). For example:
    • Site #1: 12 Is-Toxoplasma IgG negative, Other EIA positive -> referee negative. 1 Is-Toxoplasma IgG positive, Other EIA negative -> referee negative.
    • Site #2: 1 Is-Toxoplasma IgG negative, Other EIA positive -> referee negative.
    • Site #3 (manual): 5 Is-Toxoplasma IgG negative, Other EIA positive -> referee negative. 1 Is-Toxoplasma IgG positive, Other EIA negative -> referee positive.
    • Site #3 (MAGO Plus): 4 Is-Toxoplasma IgG negative, Other EIA positive -> referee negative. 1 Is-Toxoplasma IgG positive, Other EIA negative -> referee positive.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No. This document describes the performance of a diagnostic assay (ELISA) which typically does not involve human "readers" interpreting images or complex data in the same way an MRMC study would apply to an AI for image analysis. The device itself is designed to provide a result (qualitative or quantitative IU/ml), not to assist human interpretation of complex medical cases. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable here.

6. Standalone (Algorithm Only) Performance:

  • Yes, in essence. The entire evaluation focuses on the performance of the "Is-Toxoplasma IgG Test System" itself, both in manual operation and when integrated with an automated EIA processor (MAGO Plus). This is a standalone device providing diagnostic results. Human involvement is in plate loading, reagent addition (for manual), and result interpretation (based on established cut-offs), but the "algorithm" (the ELISA's chemical reactions and spectrophotometric reading) is what determines the result. The comparison studies directly assess this standalone performance against predicate devices.

7. Type of Ground Truth Used:

  • For the primary comparison studies, the "ground truth" was established by comparison to "another marketed EIA method" (predicate devices).
  • For discordant results, a "referee EIA method" was used.
  • The document explicitly states: "NOTE: Please be advised that 'relative' refers to the comparison of the assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be comparison assay's accuracy to predict disease." This clarifies that the ground truth was based on agreement with other laboratory tests, not clinical outcomes or pathology data directly.
  • For the WHO Correlation study, the ground truth was the WHO International Standard for Anti-Toxoplasma Serum.

8. Sample Size for the Training Set:

  • The document does not specify a separate "training set" in the context of machine learning. This is a traditional in vitro diagnostic device (ELISA).
  • The manufacturer would have performed internal development and validation studies, which involve implicitly "training" the assay (e.g., optimizing reagents, establishing cut-offs) based on a broad range of samples, but these are not explicitly termed "training sets" with a specific sample size in the context of this 510(k) summary. The clinical studies described are for validation, not training.

9. How the Ground Truth for the Training Set Was Established:

  • As there's no explicitly defined "training set" in the machine learning sense, this question isn't directly applicable.
  • For the development of the assay, the manufacturer would have used various characterized samples (positive, negative, equivocal) to establish assay parameters and cut-offs. The characterization of these samples would have likely relied on a combination of known clinical status and/or results from established reference assays, similar to how the ground truth for the validation studies was approached (i.e., using predicate devices and a referee method).

{0}------------------------------------------------

AUG 2 / 1998

K981498

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

Applicant Information:

Date Prepared:April 24, 1998
Name:Diamedix Corporation
Address:2140 N. Miami AvenueMiami, FL 33127
Contact Person:Dr. Lynne Stirling
Phone Number:305-324-2354
Fax Number:305-324-2585

Device Information:

Trade Name:Is-Toxoplasma IgG Test System
Common Name:Toxoplasma gondii EIA Test
Classification Name:Enzyme linked immunosorbent assay, Toxoplasma gondii(866.3780)

Equivalent Device:

Incstar Toxoplasma IgG ELISA Kit

Device Description: The Is-Toxoplasma IgG Test System is an enzyme-linked immunosorbent assay (ELISA) for the qualitative and quantitative detection of IgG to Toxoplasma gondii in human serum

Intended Use: The assay is intended for use in detecting IgG antibodies to T. gondii antigen in human serum. The results of the assay can be used as an aid in the assessment of the patient's immunological response to infection with T. gondii and in the determination of immune status of individuals, including females of child-bearing age. The evaluation of paired sera can aid in the diagnosis of primary or reactivated infection.

Principle of the Procedure: The Is-Toxoplasma IgG Test System is an enzyme-linked immunosorbent assay to detect IgG to Toxoplasma gondii in human serum. Partially purfied T. gondii antigens are attached to a solid phase microtiter well. Diluted test sera are added to each well. If antibodies which recognize the T. gondii antigens are present in the patient sample they will bind to the antigens on the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human immunoglobulin (conjugate) is added to each test well. If antibody is present the enzyme-linked antibody will bind to it. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is then added to each well. If enzyme is present from the prior step, the reaction is stopped and the color intensity is measured photometrically producing an indirect measure of the specific antibody present in the patient sample.

000246

{1}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Performance Characteristics

A. Comparison Testing

A total of six hundred and twenty one sera were tested for the presence of toxoplasma IgG antibodies using the Diamedix Is-Toxoplasma IgG Test Kit and three other marketed tests at two independent sites (site #1, Miami, FL and site #2, Salt Lake City, Utah) as well as at Diamedix Corp., Miami, FL (site #3). At site #3 testing was performed both manually and using the MAGO Plus Automated EIA Processor.

Site #1 tested 200 samples (37% fresh and 73% frozen). Samples were obtained from the S. Florida area. Site #1 purposefully selected their sample population in order to provide an equal number of positive and negative results. Table 1 compares the results obtained for the Is-Toxoplasma IgG Test Kit and their currently used testing method.

Site #2 tested 179 samples (all fresh) submitted for ToRCH screening. Samples were obtained from the West region. Table 2 compares the results obtained for the Is-Toxoplasma IgG Test Kit and their currently used testing method.

TABLE 1

Is-Toxoplasma IgG - Site #1

Is-Toxoplasma IgG - Site #2

Equivocal 0

0

0

95% CI* 68.0 - 99.8

97.8 - 100.0

96.9 - 100.0

PositiveNegativeEquivocalPositiveNegativeEquivocal
OtherEIAsPositive98 [23]14 [4]3Positive14 [6]1
Negative177 [29]0Negative0164 [65]
Equivocal061Equivocal00
Relative Sensitivity 98/112 = 87.5%Relative Specificity 77/78 = 98.7%Overall Agreement** 175/190 = 92.1%Relative Sensitivity 14/15 = 93.3%Relative Specificity 164/164 = 100.0%Overall Agreement** 178/179 = 99.4%
95% CI*81.4 - 93.693.1 - 100.087.3 - 95.5

For Site #1, further resolution of the discordant samples was performed by testing such samples in a referee EIA method. Twelve of the samples negative by the Is-Toxoplasma IgG Test Kit and positive by the other EIA were negative by the referee method; the remaining two sera were equivocal. The sample that was positive in the Is-Toxoplasma IgG Test Kit and negative in the other EIA was negative when tested by the referee method.

For Site #2, further resolution of the discordant sample was performed in a similar manner. The sample that was negative in the Is-Toxoplasma IgG Test Kit and positive by the other EIA was negative in the referee EIA method.

Site #3 (Diamedix Corp.) tested 242 samples (all frozen) by the manual method and 241 of these samples (one being QNS) by the MAGO Plus method. Samples were obtained from S. Florida blood donors. Tables 3 and 4 compare the results obtained for the Is-Toxoplasma IgG Test Kit and another marketed EIA method.

TABLE 3Is-Toxoplasma IgG - Site #3 : ManualTABLE 4Is-Toxoplasma IgG - Site #3 : MAGO Plus
PositiveNegativeEquivocalPositiveNegativeEquivocal
OtherEIAPositive47 [16]52Positive4941
Negative1186 [73]0Negative11850
Equivocal010Equivocal010
Relative Sensitivity47/52 = 92.2%95% CI*81.5-97.9Relative Sensitivity 49/53 = 92.5%95% CI*81.8-97.9
Relative Specificity186/187 = 99.5%97.1-100.0Relative Specificity 185/186 = 99.5%97.0-100.0
Overall Agreement**233/239 = 97.5%94.6-99.1Overall Agreement** 234/239 = 97.9 %95.2-99.3

[ ] denotes samples from females of child-bearing age.

  • 95% Confidence Intervals (CI) calculated by the Exact Method.

** Equivocal results were excluded from calculations

{2}------------------------------------------------

For Site #3 (manual testing), further resolution of the discordant sera revealed that the 5 sera negative in the Is-Toxoplasma IgG Test Kit but positive in the other EIA were also negative by a referee EIA method. The serum that was positive in the Is-Toxoplasma IgG Test Kit and negative in the other EIA was positive by the referee method. For MAGO Plus testing, the 4 sera that were negative in the Is-Toxoplasma IgG Test Kit but positive in the other EIA were also negative by a referee EIA method. The serum that was positive in the Is-Toxoplasma IgG Test Kit and negative in the other EIA was positive by the referee method.

NOTE : Please be advised that 'relative' refers to the comparison of the assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be comparison assay's accuracy to predict disease.

B. Correlation of Manual and MAGO Plus Results

The Is-Toxoplasma IgG Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus Procedures, the results of 211 serum samples tested in the comparison studies were plotted. Thirty highly reactive samples exceeded the reportable range and were excluded from this comparison. A scattergram and regression line of the results obtained with 95% confidence intervals is shown in Figure 1.

Image /page/2/Figure/4 description: The image is a scatter plot titled "FIGURE 1: Manual vs MAGO Plus Correlation". The x-axis is labeled "MANUAL IU/ML" and ranges from 0 to 250. The y-axis is labeled "MAGO PLUS IU/ML" and ranges from -50 to 300. The plot shows a positive correlation between the two variables, with the equation of the regression line being Y = 1.4991 + 1.0769 X and r = 0.9631.

C. Linearity

Several strongly positive serum samples were serially dilutions were assayed, in duplicate, in the Is-Toxoplasma IgG Test Kit both manually and using the MAGO Plus Automated EIA Processor. Representative linear regression graphs and scattergrams of the mean results with 95% confidence intervals are presented in Figures 2 and 3 for one patient sample. The results demonstrate a high degree of linearity throughout the reportable range of the assay when samples are tested either manually or by MAGO Plus.

Image /page/2/Figure/7 description: The image contains two line graphs, the first titled "FIGURE 2: Manual Linearity" and the second titled "FIGURE 3: MAGO Plus Linearity". The first graph has a slope of 993.54, a y-intercept of -7.89, and an R-squared value of 0.9969. The second graph has a slope of 865.69, a y-intercept of -6.71, and an R-squared value of 0.9973.

000248

{3}------------------------------------------------

D. Correlation to WHO Standard

The Is-Toxoplasma IgG Test Kit has been calibrated against the WHO 3rd International Standard for Anti-Toxoplasma Serum (code TOXM). To demonstrate the accuracy of the quantitative procedure, several dilutions of the WHO Standard were prepared and assayed manually in triplicate in two different runs on two different days versus the Is-Toxoplasma IgG Test Kit standard curve. The linear regression graph and scattergram of the mean results with 95% Confidence Intervals is shown in Figure 4.

FIGURE 4 : Dilutions of WHO Standard

Image /page/3/Figure/2 description: The image is a graph titled 'assayed against Is-Toxoplasma IgG Standards'. The x-axis is labeled 'WHO IU/ML' and ranges from 0 to 250. The y-axis is labeled 'Is-Toxoplasma IU/ml' and ranges from 0 to 350. The graph shows a linear relationship with the equation Y = 4.3333 + 1.1946 X and an R-squared value of 0.9932.

E. Quantitative Data

Serum pairs were obtained by preparing multiple two-fold dilutions of several strongly positive sera. Ratios for dilutions representing a four-fold difference in antibody level were evaluated as a serum pair both manually and using the MAGO Plus. Overall, it was estimated that a 3.9 to 6.2 fold (mean 5.1 fold ) increase in Is-Toxoplasma IgG IU/ml values corresponded to a four-fold titer increase in Toxoplasma IgG antibody levels.

F. Cross Reactivity

Sera containing IgG antibodies to viruses potentially cross-reactive to T. gondit have been tested in the Is-Toxoplasma IgG Test Kit. Fifty sera negative for IgG antibodies to 7. gondii in the Is-Toxoplasma IgG Test Kit as well as in another marketed test but positive for one or more vinuses were evaluated. In additon, nine of these sera were positive for antinuclear antibodies (ANA) and two were positive for anti-DNA. The data in the following table suggest that no crossreactivity should be expected with the Is-Toxoplasma IgG Test Kit from these analytes.

AnalyteToxoplasma IgGVZV IgGHSV IgGCMV IgGRubella IgGEBV IgGanti-DNAANA
No.of Pos. Samples0484737494829

TABLE 5

{4}------------------------------------------------

G. Precision

Six serum samples, spanning the reportable range, as well as the 50 IU/ml kit Standard and kit Low Positive and Negative Ontrols were tested quantitatively and values calculated from IU/ml results. Sites #1 and #2 tested samples in triplicate in three separate runs on three different days. Site #3 (Diamedix Corp.) tested samples in triplicate in two separate runs on three different days both manually and using the MAGO Plus Automated EIA Processor. The results obtained are shown in Tables 6-9.

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=9)
MEAN IU/mlSDCV%MEAN IU/mlSDCV%MEAN IU/mlSDCV%MEAN IU/mlSDCV%
A7.91.3517.098.22.0525.007.92.3629.878.01.7121.38
B31.31.404.4730.71.464.7626.22.9011.0729.43.0010.20
C159.74.512.82173.65.213.00147.27.014.76160.212.457.77
D170.813.327.80177.414.388.11165.429.4917.83171.218.4510.78
E274.728.5510.39259.123.319.00257.114.665.70263.621.528.16
F83.110.2812.3789.612.9414.4481.119.9024.5484.613.5015.96
50 STD54.64.468.1754.95.6010.2053.58.3515.6154.35.5310.18
LPC128.11.561.22138.13.152.28146.418.4212.58137.712.168.83
NC13.20.403.0312.90.513.9515.71.6510.5113.91.5811.37

TABLE 6 : Site #1 - Intra-Assay and Interassay Precision

TABLE 7 : Site #2 - Intra-Assay and Interassay Precision

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=9)
MEANIU/mlSDCV%MEANIU/mlSDCV%MEANIU/mlSDCV%MEANIU/mlSDCV%
A10.11.4614.465.41.0319.071.11.47133.605.54.0774.00
B25.93.5613.7525.02.7911.1621.01.155.4824.03.2513.54
C145.18.325.73136.35.153.7870.34.796.81117.235.8330.57
D162.88.925.48132.93.702.7895.12.963.11130.329.8122.88
E250.00.000.00237.99.964.19189.28.924.71225.728.6812.71
F56.86.6111.6475.52.763.6631.03.8012.2654.519.7736.28
50 STD48.76.5713.4951.73.797.3341.80.761.8247.45.8212.28
LPC134.919.6014.53123.111.229.11123.111.229.11127.013.9210.96
NC11.81.6814.2411.51.9316.7811.51.9316.7811.61.6113.88

TABLE 8 : Site #3-Intra-Assay and Interassay Precision (Manual)

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=18)
MEANIU/mlSDCV%MEANIU/mlSDCV%MEANIU/mlSDCV%MEANIU/mlSDCV%
A2.61.5158.081.90.4423.162.40.7029.172.30.9842.61
B17.42.0811.9519.32.2811.8119.21.417.3418.72.0410.91
C115.38.597.45105.69.889.36105.622.0420.87108.814.6913.50
D141.65.543.91141.66.904.87137.710.867.89140.37.825.57
E238.720.568.61246.816.696.76240.719.488.09242.118.177.51
F55.15.219.4646.31.623.5058.08.2214.1753.37.3213.73
50 STD56.78.1214.3255.48.9916.2357.33.536.1656.56.8912.19
LPC138.810.997.92139.08.075.81148.68.295.58142.19.876.95
NC14.10.966.8112.91.219.3815.42.0913.5714.11.7612.48

TABLE 9 : Site #3- Intra-assay and Interassay Precision (MAGO Plus)

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=18)
MEAN IU/mlSDCV%MEAN IU/mlSDCV%MEAN IU/mlSDCV%MEAN IU/mlSDCV%
A2.01.5477.002.02.60130.003.91.8948.462.62.1582.69
B17.92.3413.0712.38.7070.7320.36.2230.6416.86.8740.89
C150.05.993.99154.35.073.29156.620.3112.97153.612.147.90
D149.520.4313.67163.819.3511.81181.616.158.89165.022.1913.45
E>250N/AN/A>250N/AN/A>250N/AN/A>250N/AN/A
G40.91.954.7740.42.425.9941.82.656.3441.02.305.61
50 STD65.74.246.4579.210.4513.1978.49.3511.9374.410.1913.70
LPC170.911.346.64195.88.834.51208.815.987.65191.819.9410.40
NC15.82.1813.8015.14.8231.9224.83.5214.1918.55.7030.81

{5}------------------------------------------------

Expected Values

The prevalence of Toxoplasma IgG antibodies in the normal population can vary depending on a number of factors such as age, geographical location, socio-economic status, race and type of test used. It has been esttimated that in the United States 8-20% of the normal population has anti-toxoplasma IgG antibodies ( 1). For finales of child-bearing age and for pregnant females prevalence rates from 16 to 50% have been reported (3).

In the present studies sera from 100 healthy South Florida donors (52 female and 48 male) were evaluated in the Is-Toxoplasma IgG Test Kit. Of the 100 samples, 22 (22%) were found to be positive and 78 (78%) were negative. Age distribution, geographic location and prevalence is provided in Table 10. Histograms demonstrating the distribution of IU/ml values are shown in Figures 5 and 6.

Thirty-seven of the female donors were of child-bearing age (18-45 years). Of the sera from these donors, 2 (5%) were positive and 35 (95%) were negative. A total of 45 sera from pregnant females (15 from each trimester) were also tested in the Is-Toxoplasma IgG Test Kit. Nine (20%) were positive and 36 (80%) were negative for anti-toxoplasma IgG. In addition, a total of 216 samples from females of childbearing age were identified in the outside and in-house clinical studies (these included the 45 sera from pregnant females already referenced). Of these samples, 45 (21%) were positive and 171 (79%) were negative for anti-toxoplasma IgG when evaluated in the Is-Toxoplasma IgG Test Kit.

Number of donorsPrevalence
Total Number10022%
Geographiclocation :South Eastern US10022%
Age
10-191315.4%
20-292317.4%
30-394017.5%
40-491315.4%
50-59540.0%
60-69683.0%

TABLE 10

Image /page/5/Figure/6 description: The image contains two bar charts, labeled as "FIGURE 5" and "FIGURE 6", which display the frequency of Is-Toxoplasma IgG positive and negative populations, respectively. In Figure 5, the x-axis represents IU/mL values, and the y-axis represents frequency, with the >250 IU/mL category having the highest frequency at approximately 14. In Figure 6, the x-axis also represents IU/mL values, and the y-axis represents frequency, with the 0-10 IU/mL category having the highest frequency at approximately 75.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 21 1998

Lynne Stirling, Ph.D. Vice President, Regulatory Affairs Diamedix Corporation 2140 N. Miami Ave. Miami. FL 33127

Re: K981498

Trade Name: Is-Toxoplasma IgG Test System Regulatory Class: II Product Code: LGD Dated: June 23, 1998 Received: June 25, 1998

Dear Dr. Stirling:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours.

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Appendix G. Indications for Use Statement

INDICATIONS FOR USE STATEMENT

510(K) NUMBER : _ K 98 14 98

DEVICE NAME : Is-Toxoplasma IgG Test System

Indications for Use : The Diamedix Is-Toxoplasma IgG Test Kit is an Enzyme Immunoassay (EIA) for the qualitative and quantitative determination of IgG antibodies in human serum to aid in the assessment of the patient's immunological response to infection with Toxoplasma gondii and in the determination of the immune status of individuals, including females of child-bearing age. The evaluation of acute and convalescent sera can aid in the diagnosis of primary or reactivated infection with Toxoplasma gondii. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor. This product is not FDA cleared for use in screening blood and plasma donors.

Woody Dubois

510(k) Number

000253

§ 866.3780

Toxoplasma gondii serological reagents.(a)
Identification. Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toToxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyToxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoanToxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.(b)
Classification. Class II (performance standards).