K Number
K981343

Validate with FDA (Live)

Device Name
ERCO-RIBBON
Manufacturer
Date Cleared
1998-11-24

(225 days)

Product Code
Regulation Number
876.5020
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Erco-Ribbon ™ is used to maintain penile rigidity in men with erectile dysfunction. It restricts blood outflow from the penis after the patient has obtained an erection (either with the aid of a vacuum pump, or naturally).

Men with firm but short lasting erection may use constricting device separately.

If a man is impotent to a degree that he is unable to achieve a natural erection, a vacuum constriction system may help him to overcome this dysfunction.

Erco-Ribbon ™ can be used for patients with following conditions:

  • a) separately:
    premature ejaculation,

firm but short lasting erection.

b) as a part of a vacuum constriction system:

Diabetes

Venous leakage

Prostatectomy

Hypertension

Psychogenic impotence

Impotence due to radiation therapy

Spinal cord injury

Device Description

A constriction ribbon is made of elastic and soft rubber-like material with generally rectangular cross section. A belt with release loop is used for keeping the constrictor in a prearranged tightened position and for unlocking the ribbon.

The material conforms FDA regulation for devices contacting human skin.

For unlocking of the constrictor, the belt's release loop is pulled forward. The belt slips off, allowing easy unwrapping of the constricting ribbon.

A mounting tube comprises a segment of acrylic tube.

AI/ML Overview

The provided text is a 510(k) summary for the ERCO-RIBBON™ constriction device. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, it does not contain details about specific acceptance criteria, performance studies, sample sizes, ground truth establishment, or expert involvement.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I answer questions about the study that proves the device meets such criteria based solely on the provided input.

Here's a breakdown of what CANNOT be extracted from the provided text:

  1. A table of acceptance criteria and the reported device performance: This information is completely absent. The document states a claim of "superior safety and performance" compared to predicate devices, but no metrics or specific criteria are provided.
  2. Sample sized used for the test set and the data provenance: There is no mention of any test set, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective).
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: There is no mention of ground truth or experts for performance evaluation.
  4. Adjudication method for the test set: No test set or adjudication method is described.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: No such study is mentioned.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: As this is a physical medical device, not an algorithm, this question is not applicable in the context of the provided document. Even if rephrased for device performance, no standalone performance study details are given.
  7. The type of ground truth used: No ground truth is discussed in relation to a performance study.
  8. The sample size for the training set: There is no mention of a "training set" as this is not an AI/ML device and no performance study details are given.
  9. How the ground truth for the training set was established: Not applicable, as there is no training set mentioned.

What the document does provide in relation to general performance and comparison:

  • Claim of Superiority: "Technological differences of Erco-Ribbon™ provide superior safety and performance." This is a general claim, not supported by quantitative data in this summary.
  • Comparison to Predicate Devices: The document extensively compares the Erco-Ribbon™ to predicate devices based on design and perceived user experience (e.g., ease of application, comfort, pressure control, removal). It highlights potential issues with predicate devices (excessive pressure, discomfort, difficult placement, painful removal) and suggests that the Erco-Ribbon's design (multi-turn ribbon, user-adjustable tension) addresses these. These are qualitative comparisons based on design features, not quantitative performance metrics from a study.
  • Material Conformance: "The material conforms FDA regulation for devices contacting human skin." This is a regulatory conformance statement, not a performance metric from a study.

In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence based on intended use and design characteristics, arguing for improved user experience and safety. It lacks the specific performance data and study details requested in your prompt.

{0}------------------------------------------------

NOV 2 4 1998

ERCONS, Inc. 66 Overlook Terraca, s-te 2E New York, NY 10040 Tel. 1-212-927 3275 Fax 1-212-827 7387

K981343 144

ersum7a

510(k) SUMMARY

for

ERCO-RIBBON™

Constriction Device for External Penile Rigidity System

Submitter: Dr. Yakov Altshuler, Vice President

Date: Apr. 7, 1998

Trade name - Erco-Ribbon TM

Common name - constriction device for external penile rigidity system

Equivalence is claimed to:

ErecAidOsbon Medical Systems, Ltd.K841257
ReviveRevive System Corp.K844445
ConfidencePerformance, Inc.K891125
VEDMission Pharmacal Co.K901223
VETVET-CO, Inc.K9022403.
Pos-T-VacPOS-T-VACK960828
Code of Federal Regulation (CFR) Number:Unclassified
Product Code:78 LKY

Ercons, Inc.

PATENT PENDING

Erco-Ribbon™

CONFIDENTIAL Apr. 1998

{1}------------------------------------------------

K 981343

Intended Use

Erco-Ribbon™ is used to maintain penile rigidity in men with erectile dysfunction. It restricts blood outflow from the penis after the patient has obtained an erection (either with the aid of a vacuum pump or naturally).

Men with firm but short lasting erection may use constricting device separately. If a man is impotent to a degree that he is unable to achieve a natural erection, a vacuum constriction system may help him to overcome this dysfunction.

Erco-Ribbon TM can be used for patients with following conditions:

a) separately:

premature ejaculation,

firm but short lasting erection

b) as a part of a vacuum constriction system: Diabetes, Venous leakage, Prostatectomy. Hypertension, Psychogenic impotence, Impotence due to radiation therapy, Spinal cord injury.

DESCRIPTION

A constriction ribbon is made of elastic and soft rubber-like material with generally rectangular cross section. A belt with release loop is used for keeping the constrictor in a prearranged tightened position and for unlocking the ribbon.

The material conforms FDA regulation for devices contacting human skin.

For unlocking of the constrictor, the belt's release loop is pulled forward. The belt slips off, allowing easy unwrapping of the constricting ribbon.

A mounting tube comprises a segment of acrylic tube.

{2}------------------------------------------------

PHONE NO. : 9277387

K981343 રત ખ

CONFIDENTIAL Apr. 1998 Ercons, Inc. PATENT PENDING Erco-Ribbon™

For arrangement of Erco-Ribbon ™ penile constriction device, the ribbon is wrapped with the tension by multyple turns over the segment of the proximal end of the mounting tube. (or a vacuum chamber, when Erco-Ribbon ™ is used as a part of a vacuum constriction system).

SUBSTANTIAL EQUIVALENCY COMPARISON

Comparison of design and performance of marketable constriction rings and Erco-Ribbon is made with consideration of the impact of device's design on its safety andeffectiveness. Considered marketable Predicate devices and Erco-Ribbon™ are constriction devices with the same intended use, but have different technological characteristics.

Predicate devices:

To provide sufficient inward pressure on erected penis, one or more of constriction rings have to be placed at the edge of an open end of the vacuum chamber. Rings are molded of natural or synthetic rubber with different durometer numbers, usually in the range of 30 to 60 and are manufactured in large, medium and small sizes. To provide acceptable inward pressure, selection for individual users is based on size, durometer and a number of constriction rings used together. Despite high cost due to a number of expensive molds, there is no way to provide smooth control of the pressure which happens to be excessive and causes discomfort, numbness, bruises.

Placement of rings presents certain technical difficulties. Recognizing this problem, special cone-shaped applicators are added to the system. Applicators facilitate the problem, but complicate the system and increase its cost.

Discomfort and pain may be caused by twisting of doubled rings during their transfer onto the penis and because of intertwining with pubic hair.

Removal of constriction ring from erected penis could be painful too, especially when two or more rings are used together. In this case after removal of the first ring the penis is still engorged, as the remaining ring prevents blood outflow.

{3}------------------------------------------------

K 951343

Erco-Ribbon ™ :

There are countless possibilities of providing proper pressure of the constricting device during fabrication: by selection of constricting ribbon's material, cross section, length.

With the constrictor of given properties and dimensions, the user may prearrange desirable pressure by changing the tension and number of turns during The present design radically reduces major technical difficulties of wrapping. constriction ring associated with its placement on the vacuum chamber. A multytum constriction ribbon is much easier to apply than a solid constriction ring as the force necessary to extend multyturn constrictor is much lower than for solid ring.

SUMMARY:

Erco-Ribbon ™ has the same intended use as Predicate devices. Technological differences of Erco-Ribbon™ provide superior safety andperformance.

Erco-Ribbon™ is substantially equivalent to constriction devices that have been marketed for many years.

Ercons, Inc. PATENT PENDING Erco-Ribbon™ Apr. 1998

র্য

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a partial view of a logo and the beginning of the word "DEPARTMENT". The logo is a stylized graphic with three curved shapes stacked on top of each other. The word "DEPARTMENT" is in a bold, sans-serif font and is cut off, suggesting there is more text to the right of the frame. The image appears to be a close-up of a document or sign.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 4 1998

Yakov Altshuler, Ph.D. Vice President Ercons, Inc. 66 Overlook Terrace, Suite 2E New York, NY 10040

Re: K981343 Erco - Ribbon™ - OTC Dated: October 9, 1998 Received: October 13, 1998 Regulatory Class: Unclassified Unclassified/Procode: LKY 78

Dear Dr. Altshuler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, narket the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findine of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely your

Lillian Yin, Ph.D.

Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for use:

Erco-Ribbon ™ is used to maintain penile rigidity in men with erectile dysfunction. It restricts blood outflow from the penis after the patient has obtained an erection (either with the aid of a vacuum pump, or naturally).

Men with firm but short lasting erection may use constricting device separately.

If a man is impotent to a degree that he is unable to achieve a natural erection, a vacuum constriction system may help him to overcome this dysfunction.

Erco-Ribbon ™ can be used for patients with following conditions:

  • a) separately:
    premature ejaculation,

firm but short lasting erection.

b) as a part of a vacuum constriction system:

Diabetes

Venous leakage

Prostatectomy

Hypertension

Psychogenic impotence

Impotence due to radiation therapy

Spinal cord injury

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 8)

Over-The-Counter Use

QR
hiilian

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices

510(k) Number. K981343

§ 876.5020 External penile rigidity devices.

(a)
Identification. External penile rigidity devices are devices intended to create or maintain sufficient penile rigidity for sexual intercourse. External penile rigidity devices include vacuum pumps, constriction rings, and penile splints which are mechanical, powered, or pneumatic devices.(b)
Classification. Class II (special controls). The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: External Penile Rigidity Devices.” See § 876.1(e) for the availability of this guidance document.