K Number
K981084

Validate with FDA (Live)

Device Name
URISCREEN
Date Cleared
1998-10-07

(196 days)

Product Code
Regulation Number
866.2660
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Uriscreen is a non-quantitative rapid screen for the detection of Urinary Tract Infection. Uriscreen detects both bacteria and/or white blood cells, common indicators of Urinary Tract Infection.

Uriscreen is a non-quantitative rapid screen, which detects catalase. Urinary catalase may be associated with White Blood Cells, bacteria and other cells, some of which may be indicators of urinary tract infection. Only Adults should use the test.

Device Description

Uriscreen - Uriscreen is a rapid screen for the detection of Urinary Tract Infection. Uriscreen detects Catalase activity. Most organisms that commonly cause Urinary Tract Infection are Catalase positive. Additionally, Catalase activity is an indicator of white blood cells in urine. White blood cells are a common indicator of Urinary Tract Infection.

Uriscreen tubes contain a reagent powder, which enables Catalase detection. Urine is added to the tube to the line indicated, a few drops of hydrogen peroxide solution are added. Production of foam indicates Catalase activity. Catalase activity is indicative of bacteria and/or white blood cells in the urine. Most bacteria that commonly cause urinary tract infection are Catalase positive. The presence of bacteria and/or white blood cells in the urine are clearly indicative of Urinary Tract Infection.

AI/ML Overview

The manufacturer did not define acceptance criteria. Here's a summary of the performance data from the studies:

1. Acceptance Criteria and Reported Device Performance:

Study/ComparisonPerformance MetricReported Performance
Phase I: Uriscreen vs. Multistix Nitrite/LESensitivity100.00%
(Lay consumer)Specificity87.34%
Accuracy89.13%
Phase I: Uriscreen vs. Uri-Test Nitrite in UrineSensitivity80.00%
(Lay consumer)Specificity88.31%
Accuracy86.96%
Phase I: Uriscreen vs. Reference Culture MethodSensitivity100.00%
(Lay consumer)Specificity85.19%
Accuracy86.96%
Phase II: Consumer vs. Medical TechnologistSensitivity100.00%
Specificity100.00%
Accuracy100.00%
Phase III: Uriscreen vs. Culture (>100,000 CFU/ml)Sensitivity95.35%
(Clinical Labs)Specificity72.92%
Accuracy79.86%
Phase III: Uriscreen vs. Culture (10,000-100,000 CFU/ml)Sensitivity82%
(Clinical Labs)Specificity87%
Accuracy84%

2. Sample Size and Data Provenance (Test Set):

  • Phase I (Lay Consumer comparisons):
    • Multistix Nitrite/LE vs. Uriscreen: Total of 92 samples (13 Uriscreen+/Multistix+, 10 Uriscreen+/Multistix-, 0 Uriscreen-/Multistix+, 69 Uriscreen-/Multistix-).
    • Uri-Test Nitrite in Urine vs. Uriscreen: Total of 92 samples (12 Uriscreen+/Nitrite+, 9 Uriscreen+/Nitrite-, 3 Uriscreen-/Nitrite+, 68 Uriscreen-/Nitrite-).
    • Culture vs. Uriscreen: Total of 92 samples (11 Uriscreen+/Culture+, 12 Uriscreen+/Culture-, 0 Uriscreen-/Culture+, 69 Uriscreen-/Culture-).
    • Provenance: Not explicitly stated, but implies collected as part of a clinical trial by the sponsor, likely prospective, with consumers performing the test on their own urine.
  • Phase II (Consumer vs. Medical Technologist):
    • Sample Size: 65 blinded samples (water only and water + yeast at 10 CFU/ml).
    • Provenance: Samples tested by participants at a social club and MDC Associates. Prospective, with "blinded samples."
  • Phase III (Clinical Laboratories):
    • Uriscreen vs. Culture (>100,000 CFU/ml): 139 samples (41 Uriscreen+/Culture+, 26 Uriscreen+/Culture-, 2 Uriscreen-/Culture+, 70 Uriscreen-/Culture-).
    • Uriscreen vs. Culture (≥10,000 to >100,000 CFU/ml): 139 samples (58 Uriscreen+/Culture-Pos, 9 Uriscreen+/Culture-Neg, 13 Uriscreen-/Culture-Pos, 59 Uriscreen-/Culture-Neg).
    • Provenance: Conducted in three CLIA Registered Clinical Laboratories, including patients presenting with symptoms of Urinary Tract Infection. Likely prospective.

3. Number and Qualifications of Experts for Ground Truth (Test Set):

  • Phase I: Not explicitly stated. For "reference culture method," it implies standard laboratory procedures, but the number and qualifications of experts interpreting these are not detailed.
  • Phase II: "A trained Medical Technologist" confirmed all consumer results. Number of technologists not specified.
  • Phase III: "Culture" results were used as ground truth. This implies microbiology laboratory professionals performed and interpreted these cultures. The number and specific qualifications of these experts are not explicitly stated, but standard clinical laboratory practices would apply.

4. Adjudication Method (Test Set):

  • The document does not describe an adjudication method for conflicting results between experts.
  • In Phase II, a single "trained Medical Technologist" confirmed consumer results, implying a direct comparison rather than an adjudication process for conflicting expert opinions.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No MRMC comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance. The Uriscreen is a rapid diagnostic test, not an AI-assisted diagnostic tool for human readers. Phase II did compare consumer interpretation to a medical technologist, but this was to validate consumer use, not human reader improvement with AI.

6. Standalone (Algorithm Only) Performance:

  • The Uriscreen itself is a standalone test that produces a result (foam for positive, no foam for negative). The performance data presented (e.g., Sensitivity, Specificity, Accuracy) are representative of the device's standalone performance compared to various benchmarks. There is no separate "algorithm only" component as it's a reagent-based test.

7. Type of Ground Truth Used:

  • Phase I:
    • Predicate devices (Bayer Multistix Leukocyte esterase/Nitrite and Uri-Test Nitrite in Urine).
    • Reference culture method.
  • Phase II: Blinded samples (water/yeast) with results confirmed by a "trained Medical Technologist."
  • Phase III: Reference culture method, with a positive culture defined as >100,000 CFU/ml or ≥10,000 to >100,000 CFU/ml based on the specific analysis. This is a definitive laboratory standard for bacterial infection.

8. Sample Size for the Training Set:

  • This device is a rapid diagnostic test using visual interpretation of a chemical reaction. It is not an algorithm or AI model that typically requires a "training set" in the computational sense. Therefore, there is no training set mentioned or applicable for this type of device.

9. How Ground Truth for the Training Set Was Established:

  • As stated above, there is no "training set" for this type of device. The development of the Uriscreen likely involved R&D and optimization of the chemical formulation and reaction parameters based on scientific principles of catalase detection, rather than machine learning training on a dataset with established ground truth.

{0}------------------------------------------------

OCT 7 1998

510(k) Summary

981084

Date of Summary: March 21, 1998

Product Name: Uriscreen

Sponsor's Name: Diatech Diagnostics, Inc. 90 Windom Street Boston, MA 02134

CORRESPONDENT: MDC ASSOCIATES

Fran White Regulatory Consultant 15 Oak Street Beverly Farms, MA 01915

SUBSTANTIALLY EQUIVALENT DEVICES:

Product: MULTISTIX- Reagent Strip for Urinalysis Manufactured by: Bayer Diagnostics

Product: Uri-Test - Nitrite in Urine Test Manufactured by: Technical Chemical and Product Inc.

PRODUCT DESCRIPTION:

Uriscreen - Uriscreen is a rapid screen for the detection of Urinary Tract Infection. Uriscreen detects Catalase activity. Most organisms that commonly cause Urinary Tract Infection are Catalase positive. Additionally, Catalase activity is an indicator of white blood cells in urine. White blood cells are a common indicator of Urinary Tract Infection.

INTENDED USE:

Uriscreen is a non-quantitative rapid screen for the detection of Urinary Tract Infection. Uriscreen detects both bacteria and/or white blood cells, common indicators of Urinary Tract Infection.

{1}------------------------------------------------

SUMMARY OF TECHNOLOGY:

Uriscreen tubes contain a reagent powder, which enables Catalase detection. Urine is added to the tube to the line indicated, a few drops of hydrogen peroxide solution are added. Production of foam indicates Catalase activity. Catalase activity is indicative of bacteria and/or white blood cells in the urine. Most bacteria that commonly cause urinary tract infection are Catalase positive. The presence of bacteria and/or white blood cells in the urine are clearly indicative of Urinary Tract Infection.

PERFORMANCE DATA:

A clinical trial was done to compare the performance of the Uriscreen to Bayer Multistix Leukocyte esterase/Nitrite and the Uri-Test Nitrite in Urine. These data clearly demonstrate that the performance of the Uriscreen test is substantially equivalent to the predicate devices.

Phase I

Multistix Nitrite/LE vs. Uriscreen

Multistix +Multistix -
Uriscreen +1310
Uriscreen -069
Sensitivity =100.00%
Specificity =87.34%
Accuracy =89.13%

Uri-Test Nitrite in Urine vs. Uriscreen

Nitrite +Nitrite -
Uriscreen +129
Uriscreen -368
Sensitivity = 80.00%
Specificity = 88.31%
Accuracy = 86.96%

{2}------------------------------------------------

Uriscreen was also compared to reference culture method.

Culture vs. Uriscreen
Culture +Culture -
Uriscreen +1112
Uriscreen -069
Sensitivity =100.00%
Specificity =85.19%
Accuracy =86.96%

Uriscreen was performed by consumers on their own urine. Data confirm that Uriscreen is easy to use and can be effectively used by consumers with no training or instructions.

Phase II

To demonstrate the ease-of-use and the ability for the lay consumer to perform and interpret the Uriscreen test, we asked participants at a social club and MDC Associates to test blinded samples. We recruited 65 consumers to read the Uriscreen Package Insert, perform the test on a blinded sample and interpret the results. The blinded samples consisted of a negative, which contained water only and the positive sample contained water and yeast at 10 CFU/ml. All consumer results were confirmed by a trained Medical Technologist. 100% accuracy was observed.

Consumer vs. Medical Technologist

Uriscreen + (Med Tech)Uriscreen -(Med Tech)RowTotal
Uriscreen +(consumer)32032
Uriscreen -(consumer)03333
Total323365
Sensitivity =100.00%
Specificity =100.00%
Accuracy =100.00%

{3}------------------------------------------------

Phase III

)

To demonstrate the performance of the Uriscreen test, an Accuracy and Outcome Analysis study was done in three CLIA Registered Clinical Laboratories. Only patients presenting with symptom of Urinary Tract Infection were included in the study.

Uriscreen test results were compared to culture, and based on our protocol criteria a positive culture is considered >100,000 CFU/ml.

Culture +Culture -Row Total
Uriscreen +412667
Uriscreen -27072
Total4396139

Uriscreen vs. Culture (>100,000)

Sensitivity =95.35%
Specificity =72.92%
Accuracy =79.86%
_eukocyte esterase (LE) vs. Culture
-------------------------------------------
(>100,000)
Culture +Culture -RowTotal
LE +384381
LE -45458
Total4297139
Sensitivity =90.48%
Specificity =55.67%
Accuracy =66.19%

s

{4}------------------------------------------------

Nitrite vs. Culture(>100,000)
Culture +Culture -RowTotal
Nitrite +16218
Nitrite -2794121
Total4396139

Accuracy =

Based on our protocol criteria, a positive culture would have significant growth of >100,000. Uriscreen demonstrates 95% Sensitivity when compared to positive culture. Leukocyte esterase only demonstrates 90% Sensitivity when compared to positive culture and Nitrite is only 37% Sensitive.

79.14%

Uriscreen vs. Culture (≥10,000 to
>100,000)
Culture - PositiveCulture - NegRow Total
Uriscreen - Pos58967
Uriscreen - Neg135972
Total7168139
Sensitivity =82%
Specificity =87%
Accuracy =84%

Uriscreen demonstrates a specificity of 87% and a sensitivity of 82% at 103 - 103 CFU/ml.

{5}------------------------------------------------

There are 9 specimens with negative culture results (No Growth) and positive Uriscreen results. The Uriscreen test will produce a positive result if other cells are present in the urine. Of the 9 positive Uriscreen/No Growth cultures, 6 were reported to have White blood cells or blood in the urine.

Uriscreen is a better indicator of UTI than either Leukocyte esterase (Sensitivity 90%) or Nitrite (Sensitivity 37%) based on positive culture results (>100,000). When you look at culture growth from >10,000 to >100,000 Uriscreen sensitivity is 82%, but the specificity is 87%.

STATEMENT OF SAFETY AND EFFICACY:

When compared to a Bayer Multistix and the Uriscreen was performed by untrained consumers, the product performance was as follows:

Sensitivity - 100% Specificity - 87 Accuracy - 89%

When the Uriscreen test performed by the lay consumer and was compared to Uri-Test Nitrite in Urine, the performance of the Uriscreen test was as follows:

Sensitivity - 80% Specificity - 88% Accuracy - 87%

When the Uriscreen test performed by the lay consumer was compared to culture, performance was as follows:

Sensitivity - 100% Specificity - 85% Accuracy - 87%

When Uriscreen was tested in three clinical laboratories, comparing Uriscreen test results to reference culture results, performance was as follows:

Sensitivity - 95% Specificity - 73% Accuracy - 80%

These data clearly demonstrate the safety and efficacy of Uriscreen.

Uriscreen was performed by consumers on their own urine. Data confirm that Uriscreen is easy to use and can be effectively used by consumers with no training or instructions.

Diatech Diagnostics, Inc. confirm that any/all data provided in this submission may be released upon request.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 2098 Gaither Road, Rockville MD 20850. The text is black and the background is white. The text is left-aligned.

OCT 7 1998

Diatech Diagnostics, Inc. c/o Fran White Regulatory Consultant MDC Associates 15 Oak Street Beverly Farms, MA 01915

Re: K981084

Trade Name: Uriscreen - A Rapid Screen for Catalase Regulatory Class: I Product Code: JXA Dated: September 1, 1998 Received: September 2, 1998

Dear Ms. White:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health -------

Enclosure

{8}------------------------------------------------

510(k) Submission Uriscreen™ Diatech Diagnostics, Inc. K981084 Addendum 2 8-31-98

Page 4 of 4

Statement Indication of Use

510(k) Number: K981084 Uriscreen - a rapid screen for catalase. Device Name:

Indication for Use:

Uriscreen is a non-quantitative rapid screen, which detects catalase. Urinary catalase may be associated with White Blood Cells, bacteria and other cells, some of which may be indicators of urinary tract infection. Only Adults should use the test.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K981084

J Prescription Use (Per 21 CFR 801.109)

OR

Image /page/8/Picture/11 description: The image shows the text "Over-The-Counter Use" in bold font, followed by the text "(Optional Format 1-2-96)" in a smaller font size. To the left of the text is a symbol that looks like a square with an X through it. The text and symbol appear to be part of a label or instruction.

CONFIDENTIAL

§ 866.2660 Microorganism differentiation and identification device.

(a)
Identification. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.