K Number
K981066

Validate with FDA (Live)

Date Cleared
1998-08-31

(160 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use of this device is indicated where a physician or healthcare professional requires information about the electrical events within the myocardium using waveforms recorded from the body's surface during depolarization and repolarization of the atrium and ventricles. The FX-4010 is indicated where a clinician seeks advisory interpretive measurement, and diagnostic statements that would require confirmation by a physician. Also, can provide measurement data, R to R timing measurements with trend information and interpretation of arrhythmia to the clinician.

The basic EKG and recording features of this device are intended to be used on any patient; neonate, pediatric, or adult; where the placement of EKG electrodes does not interfere with or complicate the treatment of the patient. The interpretive feature of this device is intended to be used on any patient, at least one (1) year old, where the placement of EKG electrodes does not interfere with or complicate the treatment of the patient. This device is not intended for home use.

Device Description

The Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph is a portable, multi channel, interpretive, automatic or manual electrocardiograph. This electrocardiograph is designed to produce a thermally printed recording of the electrical signals produced by the heart. The size of the unit is 37.8(W) x 33.8(D) x 10.1(H) cm and weighs approximately 8 kg.

AI/ML Overview

The provided text is a 510(k) summary for the Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph. It describes the device, its intended use, and generally outlines the testing performed, but it does not contain detailed acceptance criteria or a specific study demonstrating performance against those criteria.

Instead, the submission primarily focuses on demonstrating substantial equivalence to a predicate device (FCP-2155) through general compliance with relevant standards and risk analysis. The text states that "Laboratory testing was conducted to validate and verify the Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph met all design specifications and was substantially equivalent to the FUKUDA DENSHI model FCP-2155," but it doesn't provide the specifics of these "design specifications" as measurable acceptance criteria or the numerical results of those tests.

Therefore, I cannot populate the table or answer most of the questions based on the provided text alone.

Here's a breakdown of what can and cannot be answered:

1. A table of acceptance criteria and the reported device performance

CriteriaAcceptance CriteriaReported Performance
General- Met all design specifications. - Substantially equivalent to FUKUDA DENSHI model FCP-2155."met all design specifications and was substantially equivalent" (No specific performance data provided to quantify "met" or "substantially equivalent" beyond general claims.)
Safety- Compliance with ANSI/AAMI ES1-1993 for safe current limits. - Compliance with UL 2601, CSA 22.2, and AAMI standards for electrical safety to prevent excessive electrical leakage current."demonstrate compliance with ANSI/AAMI ES1-1993" "device is designed to meet UL 2601, CSA 22.2 and AAMI standards for electrical safety" (No specific test results or values of leakage current provided.)
Efficacy (Diagnostic Accuracy)- Inferred: Prevent misdiagnosis due to inadequate design of signal processing, measurement circuitry, or software."The design of the FX-4010 has taken into account all the above." (No direct performance metrics like sensitivity, specificity, or agreement rates for interpretive features are provided.)
Performance (Diagnostic Features)- Compliance with ANSI/AAMI EC11-1991 for diagnostic electrocardiographic devices."demonstrate compliance with [...] ANSI/AAMI EC11-1991" (No specific test results related to ECG signal acquisition, measurement accuracy, or interpretive statement accuracy are provided.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided. The document refers to "Laboratory testing" but does not specify sample sizes or data provenance for any performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not provided. The document does not describe any specific ground truth establishment processes involving experts for performance testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not provided. The device is an "interpretive electrocardiograph," implying it provides automated interpretations, but there is no mention of a comparative effectiveness study involving human readers with and without its assistance. This type of study is more common for AI-driven diagnostic aids. This ECG device is a much older technology.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Implied, but no specific data presented. The device "can provide ECG interpretation" and "measurement and interpretive analysis." The "Laboratory testing" would have assessed the device's functional integrity as a standalone unit, but no specific performance metrics for the interpretive algorithm are given.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not provided. For a device claiming interpretive capabilities, ground truth would typically be established by expert cardiologists. However, the document does not elaborate on how ground truth was determined for any internal testing related to its interpretive features.

8. The sample size for the training set

  • Not applicable / Not provided. For older ECG devices with built-in interpretive algorithms, the algorithms are typically rule-based or derived from extensive analysis of ECG patterns, rather than "trained" in the modern machine learning sense. The document does not mention any training data or its size.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. See point 8.

{0}------------------------------------------------

K981066

AUG 3 1 1998

Section 2 510(k) Summary Fukuda Denshi model FX-4010

Multi Channel Electrocardiograph

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR part 807.92.

K981066 The assigned 510(k) number is:

Submitter:

FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St. Bldg C Redmond, WA 98052-4911 Tel: 425/881-7737 425/869-2018 Fax:

Contact Person:

Richard A. Queen Regulatory Affairs Specialist FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St. Bldg C Redmond, WA 98052-4911 425/881-7737 Tel: Fax: 425/869-2018

Date Prepared:

March 10, 1998

Device Name:

Proprietary Name:

Fukuda Denshi: CardioMax, model FX-4010 Multi Channel Electrocardiograph

Common Name:

Multi Channel Interpretive Electrocardiograph

Classification Name:

Electrocardiograph

Legally Marketed Device:

FCP-2155, Multi Channel Interpretative ECG (K971440)

FX4010 Reve Additional Information 40102a

{1}------------------------------------------------

Description:

The Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph is a portable, multi channel, interpretive, automatic or manual electrocardiograph. This electrocardiograph is designed to produce a thermally printed recording of the electrical signals produced by the heart. The size of the unit is 37.8(W) x 33.8(D) x 10.1(H) cm and weighs approximately 8 kg.

Intended Use:

This Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph is intended to be used for the evaluation of the cardiovascular system. The FX-4010 will acquire and record ECG waveforms. Also, the FX-4010 can provide ECG interpretation. The FX-4010 is to be used by or on the order of a physician or similarly qualified health care professional. The FX-4010 may be used in all hospital environments; ER, OR, ICU, etc .; doctors' offices; clinics; or similar settings. This device is intended to be used on any patient; neonate, pediatric, or adult; where the placement of EKG electrodes does not interfere with or complicate the treatment of the patient.

The interpretive feature is to be used on any patient, at least one (1) year old, where the placement of EKG electrodes does not interfere with of complicate the treatment of the patient. This device is not intended for home use.

Technological Characteristics

The Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph incorporates the microprocessor, thermal printer, input and output channels, measurement and interpretive analysis, and LCD technology similar to the predicate device. The FX-4010 has a port for plugging in a 2-megabyte memory card for storing data. Also, the 4010 has the ability to have accessory cartridges and modules to enables its basic ECG functions. A rechargeable battery is available on both devices.

These technological differences do not affect the safety or efficacy of the device. Any safety issues that may be raised by a software controlled medical device are addressed in the system's hazard analysis and the system validation.

{2}------------------------------------------------

Testing:

Laboratory testing was conducted to validate and verify the Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph met all design specifications and was substantially equivalent to the FUKUDA DENSHI model FCP-2155. This testing consisted of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Additional testing was performed to demonstrate compliance with ANSI/AAMI ES1-1993, "Safe current limits for electromedical apparatus," ANSI/AAMI EC11-1991, "Diagnostic Electrocardiographic Devices". Finally, a hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operation.

Although the device is neither life supporting nor life sustaining, diagnostic information derived from the use of the device may be critical to the proper management of the patient.

So, the areas of risk for this device are the same as the predicate device and other devices in this class, and are the following:

  • . Electrical shock
    Excessive electrical chassis leakage current can disturb the normal electrophysiology of the heart, and possibly leading to the onset of cardiac arrhythmias.

  • Misdiagnosis .

    • Inadequate design of the signal processing and measurement circuitry or program can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
    • Inadequate design of the device's software, used to make various measurements, can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.

The design of the FX-4010 has taken into account all the above. The device is designed to meet UL 2601, CSA 22.2 and AAMI standards for electrical safety for medical equipment to prevent the possibility of excessive electrical leakage current to the patient.

{3}------------------------------------------------

Conclusion:

The conclusions drawn from clinical and laboratory testing of Fukuda Denshi model FX-4010 Multi Channel Electrocardiograph demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the Fukuda Denshi model FCP-2155 Multi Channel Electrocardiograph (K971440).

FX4010 Reve Additional Information 40102a

:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping heads and flowing lines representing the body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 31 1998

Mr. Richard Queen Fukuda Denshi America Corporation 17725 NE 65th Street Redmond, WA 98052

K981066 Re: Fukuda Denshi CardioMax Model FX-4010, Multi Channel Electrocardiograph Requlatory Class: III (three) Product Code: 74 LOS Dated: July 8, 1998 Received: July 15, 1998

Dear Mr. Queen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{5}------------------------------------------------

Page 2 -- Mr. Richard Queen

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahon

Thomas J. Chllahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

K981066

Device Name:Fukuda Denshi: CardioMax, model FX-4010,Multi Channel Electrocardiograph
Indications For Use:Use of this device is indicated where a physician or healthcareprofessional requires information about the electrical events withinthe myocardium using waveforms recorded from the body'ssurface during depolarization and repolarization of the atrium andventricles. The FX-4010 is indicated where a clinician seeksadvisory interpretive measurement, and diagnostic statements thatwould require confirmation by a physician. Also, can providemeasurement data, R to R timing measurements with trendinformation and interpretation of arrhythmia to the clinician.
Indications For Use:The basic EKG and recording features of this device are intendedto be used on any patient; neonate, pediatric, or adult; where theplacement of EKG electrodes does not interfere with or complicatethe treatment of the patient. The interpretive feature of this deviceis intended to be used on any patient, at least one (1) year old,where the placement of EKG electrodes does not interfere with orcomplicate the treatment of the patient. This device is not intendedfor home use.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kramer

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K981066 510(k) Number_

Prescription Use (Per 21 CRF 801.109)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

N/A