K Number
K980899

Validate with FDA (Live)

Date Cleared
1998-06-11

(93 days)

Product Code
Regulation Number
880.5580
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To pierce the skin in the practice of acupuncture by qualified practitioners as determined by the States. The device is intended for single use only.

Device Description

JIN acupuncture needles are EtO sterilized, surgical grade stainless steel with pyrogen free for single-use only. JIN needles comply with the GMP quality and safety requirements. JIN needles use tightly secured metal handle for safe, excellent handling and are easy to use.

AI/ML Overview

This document describes a 510(k) submission for JIN Acupuncture Needles, arguing for their substantial equivalence to legally marketed predicate devices. The submission does NOT contain acceptance criteria or a study proving the device meets acceptance criteria in the manner one would typically expect for an AI/software medical device.

The provided text describes a medical device (acupuncture needles), not an AI/software device. Therefore, the concepts of acceptance criteria, performance studies, expert ground truth, multi-reader multi-case studies, standalone performance, and training/test sets are not applicable in the context of AI/software for this document.

Instead, the submission focuses on demonstrating substantial equivalence to existing predicate devices based on:

  • Intended Use: Identical to predicates.
  • Labeling: Identical warnings and information to predicates.
  • Specifications: "Very identical" in size and gauge, with minor differences due to different gauge systems.
  • Materials: Similar to two predicates (surgical grade stainless steel for needle body and handle), but one predicate uses plastic for the handle.
  • Historical Safety and Effectiveness: Based on over 60 years of use in Japan by the manufacturer, TAIHO.

The FDA's response confirms substantial equivalence based on the provided information, allowing the device to be marketed.

Therefore, I cannot extract the requested information as it pertains to AI/software performance studies, because the provided document is not about an AI/software medical device.

{0}------------------------------------------------

510(k) Summary [As required by Section 807.92(a)]

Kgyoбgg

  1. The Name of Submitter:

Kenshin Trading Corporation 1815 West 213th Street Suite 180 Torrance, CA 90501 Phone: (310)212-3199 Fax: (310)212-3299 Contact person: Kunio Suzuki Date of preparation: May 15, 1998

    1. The Name of Device: JIN Acupuncture Needle Single Use Sterile Acupuncture Needles Class: II Product Code: MQX Panel Code: General Hospital
    1. Identification of the legally marketed device to which the submitter claims equivalence:
510(k) NumberTrade NameManufacturer
1) K963194ITOIto Co. (Japan)
2) K962809SEIRINSeirin Kasei (Japan)
3) K961339CARBODynasty Medical (China)
    1. Description of the Device:
      JIN acupuncture needles are EtO sterilized, surgical grade stainless steel with pyrogen free for single-use only.

JIN needles comply with the GMP quality and safety requirements.

JIN needles use tightly secured metal handle for safe, excellent handling and are easy to use.

    1. The intended use for the JIN needles is to pierce the skin in the practice of acupuncture by qualified practitioners as determined by the States.
      And the indication for use is to pierce the skin in the practice of acupuncture by qualified practitioners as determined by the States. The device is intended for single use only.
    1. The device package will have the following warnings:
    • A) Caution: Federal law restricts this device to sale by or on the order of qualified practitioners of acupuncture as determined by the States.
    • B) Discard any unused needles in bulk packages after treatment session.

{1}------------------------------------------------

    1. JIN needles, to be sold in America, will in no way differ in quality, materials, or manufacturing from the Japanese needles currently being sold in Japan by TAIHO.

Description of the Device .

  • A) Intended use for the device:
    The intended use is for the practice of acupuncture by qualified practitioners as determined by the States.

Each model of JIN needles is used the same way.

B) Packaging:

All JIN needles are packaged in bulk.

  • C) Brief description of the JIN needles:
    Model ST needles (Bulk package): Sterile disposable, surgical stainless steel Japanese acupuncture needle with pressed pipe type metal handle.

5 needles and 1 plastic insertion tube per sterile sealed package separately.

Model PT needles (Bulk package): Sterile disposable, surgical stainless steel press tack is secured within the tape.

Model ID needles (Bulk package): Sterile disposable, surgical stainless steel intradermal skin needles with flat handle.

Comparison to a Legally Marketed Device

  • A) Comparison Information
    #2-SEIRIN #3-CARBO Subject-JIN #1-ITO

    1. Intended use: JIN is identical to #1, #2, and #3: "The intended use is for the practice of acupuncture by qualified practitioners as determined by the States."
    1. Labeling: JIN is identical to #1, #2, and #3: The label states A) Single Use Sterile Acupuncture Needle, B) "Caution: Federal law restricts this device to sale by or on the order of qualified practitioners of acupuncture as determined by the States", C) Bulk package: "Discard any unused needles in bulk packages after treatment session", D) "Pyrogene Free", E) Package size -

{2}------------------------------------------------

individually or in bulk, F) Model name, size, lot no. and expiration date, G) the origin of product.

    1. Specifications: Size and gauge of JIN are very identical to #1, #2, and #3 though there might be slight differences, because Chinese and Japanese gauge systems are not exactly the same
    1. Materials: JIN is identical to #1, and #3: Both needle body and handle are made of surgical grade stainless steel but different from #2, SEIRIN uses plastic for the handle.
  • B) Analysis of the comparable safety and effectiveness:

The comparable safety and effectiveness to the legally marketed devices is demonstrated historically in Japan.

JIN needles (manufactured by TAIHO) are equal or superior in safety and effectiveness to #1, #2, and #3 in use in Japan for over a period of more than 60 years. They enjoy a large market share (over 10 million disposable needles a year) as well as a good reputation that sustains being one of the more expensive needles.

TAHIO pioneered and introduced stainless steel acupuncture needles to Japan in the late 1940's before anyone else.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized depiction of a human figure embracing a globe. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the emblem.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 11 1998

Mr. Kunia Suzuki · President Kenshin Trading Corporation 1815 West 213the Street Suite 180 Torrance, California 90501

Re : K980899 JIN™ Acupuncture Needles, Models ST, GL, Trade Name: SI, PT, and ID Regulatory Class: II Product Code: MOX Dated: May 19, 1998 May 20, 1998 Received:

Dear Mr. Suzuki:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Reqister. this response to your premarket notification submission does not affect any obligation you might have under sections 531

{4}------------------------------------------------

Page 2 - Mr. Suzuki

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA acobing of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours!

Timothy A. Ulatowski

A. Ulatowski Timo Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure ____________________________________________________________________________________________________________________________________________________________________

{5}------------------------------------------------

STATEMENT OF INDICATIONS FOR USE

510(k) Notification Class II Product Code: MQX Panel Code: General Hospital Single Use Sterile Acupuncture Needles

510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________

Single Use Sterile Acupuncture Needles Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

To pierce the skin in the practice of acupuncture by qualified practitioners as determined by the States. The device is intended for single use only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(

(Optional Format 1-2-96)

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

ર-I

Kenshin Trading Corporation - 510(k) Notification

§ 880.5580 Acupuncture needle.

(a)
Identification. An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.(b)
Classification. Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.