K Number
K974581

Validate with FDA (Live)

Date Cleared
1998-10-28

(324 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents whonitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and Additionally, this more and to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.

Device Description

The Millennia 3500 Series Monitor with Anesthetic Agents Monitor is a patient monitor which draws a continuous flow of sample gas from the patient's anesthesia breathing circuit. The sample is then drawn through a measurement sample cell where the agent(s) and other patient gas concentrations are measured. Measured values are displayed as numbers and waveforms on the main display of the monitor. The monitor is operated by controls on the front panel of the monitor. The major functional subsystems of the device are the power supply, the pneumatic system, the measurement sample cell, the system electronics and product software. Power is supplied by an AC power cord, or an internal set of batteries. AC power is internally conditioned by the power supply subassembly, which powers the system and charges the internal batteries. The pneumatic system draws sample gas into the device through a moisture collection system. The sample then enters the sample cell, which is a cavity within the optical bench assembly. The sample is then exhausted out the rear of the unit. The sample cell cavity allows infrared light to be passed through the sample so a detector on the other side of the sample can measure the amount of infrared light absorption. The infrared light source consists of a resistive element which radiates a relatively wide spectrum of infrared light. The amount of transmitted infrared light is measured by an optical detector, and several optical filters are used to determine the identity and concentration of the anesthetic agents, nitrous oxide and carbon dioxide. The oxygen measurements are obtained using a polarographic oxygen cell which also receives sample gas from the sample cell. The electronic system consists of an analog circuit board, and a digital circuit board. These boards are added to the existing standard Millennia 3500 Series Patient Monitoring System. The analog board takes the electrical signals from the sample cell assembly and conditions them so that measurement calculations can be performed. The digital circuit board digitizes the conditioned analog signals and performs measurement calculations. Proprietary software algorithms are used to calculate measured concentrations of anesthetic agents and patient gases in the sample gas.

AI/ML Overview

The provided text describes the Millennia 3500 Series Monitor with Anesthetic Agent Monitoring. However, it does not contain specific acceptance criteria or detailed study results that would allow for a comprehensive table of acceptance criteria and reported device performance, nor does it provide the detailed information requested regarding sample sizes, expert qualifications, or comparative effectiveness studies.

The document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to predicate devices rather than providing full study protocols and results. It states that "The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitor conforms with national and available international product safety standards for electrical, electromagnetic compatibility, and anesthetic agent monitoring." This is a general statement about compliance but not a detailed presentation of performance data against specific criteria.

Therefore, many of the requested fields cannot be filled based on the provided text.

Here's an attempt to answer based on the available information and what can be inferred:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred)Reported Device Performance (Inferred/Stated)
Anesthetic Agent Monitoring
Accuracy for Halothane Concentration"Same performance specifications as predicate devices" (Datex AS/3, Datascope Multinex, Criticare Poet II)
Accuracy for Enflurane Concentration"Same performance specifications as predicate devices"
Accuracy for Isoflurane Concentration"Same performance specifications as predicate devices"
Accuracy for Sevoflurane Concentration"Same performance specifications as predicate devices"
Accuracy for Desflurane Concentration"Same performance specifications as predicate devices"
Other Gas Monitoring
Accuracy for End-tidal CO2 Concentration"Same performance specifications as predicate devices"
Accuracy for Nitrous Oxide Concentration"Same performance specifications as predicate devices"
Accuracy for Oxygen Concentration"Same performance specifications as predicate devices"
General Performance
Conformance to national product safety standards for electrical, electromagnetic compatibility, and anesthetic agent monitoringStated as "conforms with national and available international product safety standards"

Note: The actual numerical acceptance criteria for accuracy, drift, response time, etc., are not provided in this document. The stated "same performance specifications as predicate devices" is the primary performance claim.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified. Performance testing is mentioned generally, but no details on the number of cases or samples are provided.
  • Data Provenance: Not specified. The document does not indicate the country of origin of the data or if the data was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • Not specified. The document does not describe the establishment of a "ground truth" using human experts for performance testing. Performance is likely assessed against calibrated gas mixtures or reference devices, not human expert interpretation.

4. Adjudication Method for the Test Set

  • Not applicable/Not specified. This typically applies to studies involving human interpretation or clinical outcomes, which are not detailed here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. An MRMC study was not done, as this device measures gas concentrations and does not involve human readers interpreting images or data in the same way an AI diagnostic tool would.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The device is a monitor that provides numerical readouts of gas concentrations. Its performance is inherent to the algorithm and sensors (standalone) without a human-in-the-loop directly influencing the measurement calculation itself. The "Proprietary software algorithms are used to calculate measured concentrations."

7. The Type of Ground Truth Used

  • Inferred: Calibrated gas standards or reference measurement devices. For a gas analyzer, the ground truth for "concentration" would be established by preparing gas mixtures with known, precise concentrations and then measuring how accurately the device reports these known values. This is not explicitly stated but is the standard method for such devices.

8. The Sample Size for the Training Set

  • Not applicable/Not specified. The document does not describe the development or training of AI/ML models in the modern sense that would require a "training set" of data labels. The device uses "Proprietary software algorithms" for calculations, but these are likely deterministic algorithms based on physical principles (infrared absorption) rather than statistical or neural network models trained on large datasets.

9. How the Ground Truth for the Training Set was Established

  • Not applicable/Not specified. As above, the concept of a training set ground truth, as used for AI/ML, does not appear relevant to the described device.

{0}------------------------------------------------

OCT 28 1998

510(k) Summary of Safety and Effectiveness

Submitter Information:

Invivo Research Inc. 12601 Research Parkway Orlando, FL 32826 407-275-3220 Contact: Mr. Francis Casey

Product Name:

Proprietary:Millennia 3500 Series Monitor with Anesthetic Agent Monitoring
Common:Anesthetic Gas Analyzer
Classification:Anesthetic Gas Analyzer, enflurane - (73 CFR/21CFR 868.1500 Class II)
Anesthetic Gas Analyzer, halothane - (73 CFR/21CFR 868.1620 Class II)
Oxygen Gas Analyzer - (73 CFR/21CFR 868.1720 Class II)

Predicate Devices:

The predicate device(s) are the Datex AS/3 Anesthesia Agent Monitor (510(k) numbers K832113, K903104, K923276, K932098), the Datascope Multinex Anesthetic Agent Monitor (510(k) numbers K883254, K903104), and the Criticare Poet II Anesthesia Agent Monitor (510(k) numbers K892646, K942737). These predicate devices have the same performance specifications as the Millennia 3500 Series Monitor with Anesthetic Agents Monitor.

Device Description:

The Millennia 3500 Series Monitor with Anesthetic Agents Monitor is a patient monitor which draws a continuous flow of sample gas from the patient's anesthesia breathing circuit. The sample is then drawn through a measurement sample cell where the agent(s) and other patient gas concentrations are measured. Measured values are displayed as numbers and waveforms on the main display of the monitor. The monitor is operated by controls on the front panel of the monitor. The major functional subsystems of the device are the power supply, the pneumatic system, the measurement sample cell, the system electronics and product software. Power is supplied by an AC power cord, or an internal set of batteries. AC power is internally conditioned by the power supply subassembly, which powers the system and charges the internal batteries. The pneumatic system draws sample gas into the device through a moisture collection system. The sample then enters the sample cell, which is a cavity within the optical bench assembly. The sample is then exhausted out the rear of the unit. The sample cell cavity allows infrared light to be passed through the sample so a detector on the other side of the sample can measure the amount of infrared light absorption. The infrared light source consists of a resistive

{1}------------------------------------------------

element which radiates a relatively wide spectrum of infrared light. The amount of transmitted infrared light is measured by an optical detector, and several optical filters are used to determine the identity and concentration of the anesthetic agents, nitrous oxide and carbon dioxide. The oxygen measurements are obtained using a polarographic oxygen cell which also receives sample gas from the sample cell. The electronic system consists of an analog circuit board, and a digital circuit board. These boards are added to the existing standard Millennia 3500 Series Patient Monitoring System. The analog board takes the electrical signals from the sample cell assembly and conditions them so that measurement calculations can be performed. The digital circuit board digitizes the conditioned analog signals and performs measurement calculations. Proprietary software algorithms are used to calculate measured concentrations of anesthetic agents and patient gases in the sample gas.

Intended Uses:

The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents monitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and oxygen gas concentrations. The need to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.

Technological Comparison to Predicate Device(s):

The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitor uses a polychromatic non-dispersive infrared light (NDIR) system to identify and measure the anesthetic gases and carbon dioxide. This technology is identical to the other cited predicate devices listed above.

Summary of Performance Testing:

The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitor conforms with national and available international product safety standards for electrical, electromagnetic compatibility, and anesthetic agent monitoring.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 28 1998

Mr. Francis X. Casey Invivo Research, Inc. 12601 Research Parkway Orlando, FL 32826

Re : K974581 Millennia 3500 Series Patient Monitor with Anesthetic Agent Regulatory Class: II (two) Product Code: 74 DRT Dated: August 3, 1998 Received: August 4, 1998

Dear Mr. Casey:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Francis X. Casey

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labelinq requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Intended Uses

The Invivo Research Inc. Millennia 3500 Series Monitor with Anesthetic Agent monitoring is intended for general hospital or clinical use by medical professionals whenever it is required to monitor concentrations of anesthetic gases. The agents whonitored include halothane, enflurane, isoflurane, sevoflurane and desflurane. Additionally, this monitor also monitors end-tidal carbon dioxide, nitrous oxide and Additionally, this more and to monitor these anesthetic agents is most commonly encountered in the operating room during administration of anesthesia. This device is available for sale only upon the order of a physician or other related licensed medical professional.

Kramer

(Division Sign-Division of Care and Neurologic 510(k) Num

prescriptions use ✓

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).