K Number
K974510

Validate with FDA (Live)

Date Cleared
1998-04-14

(134 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, and Pulse Oximetry. The ENVOY may be used to monitor a wide range of patient conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

Device Description

ENVOY is hospital based, multiparameter patient monitor for monitoring physiological patient vital signs. The ENVOY patient monitor system consists of the following components: Main Processing Unit, Display Unit, Module Rack, Vital Signs Plug-in Modules. ENVOY vital signs modules acquire vital signs data from the patient, and display their waveforms and alarms indications on the ENVOY display unit. Vital signs and waveform information are displayed simultaneously on the ENVOY Display Unit. Up to 8 traces can be displayed at any one time. The vital signs modules interface with readily available physiologic transducers through electrically isolated patient input connections. After amplification, the signals are digitized, analyzed and displayed. All processing and alarm determination for ECG, Respiration and Invasive Blood Pressure is made using proprietary algorithms and software based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions. SpO2 and Non-Invasive Blood Pressure Modules use sofware/hardware from vendors whose products already appearing on the USA market. Information from each vital sign is presented in a separate portion of the display. Each vital sign is labeled for identification and numeric value. Displayed Vital sign information can include: Primary Vital Sign Name, Waveform, Vital Sign Numeric Value, Alarm Status Message. Operation of the ENVOY is accomplished by interaction with front panel controls on the main processor unit. A quick-knob control allows direct interaction with displayed menus for direct parameter selection and setup. Where manual entry of alphanumeric information is required, a menu keyboard menu is display. ENVOY is a reusable, software driven, patient monitor, intended for use as part of a physiological monitoring system in a hospital environment. As such it is not a life supporting, or life sustaining device; nor is it implantable and therefore sterility is not a consideration.

AI/ML Overview

This submission K974510 for the ENVOY Patient Monitor describes various validations against industry standards and the manufacturer's design specifications for different physiological parameters. However, it does not provide a table of acceptance criteria with corresponding reported device performance values in a structured format that can be easily extracted for all parameters. It also lacks detailed information on studies concerning AI assistance, multi-reader multi-case studies, or specific AI algorithms.

Here's an analysis based on the provided text, highlighting what is present and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly provide a single, comprehensive table outlining acceptance criteria and reported device performance for all vital signs. Instead, it describes general compliance with standards and internal design specifications.

Parameter/AspectAcceptance Criteria (from recognized standards or design specs)Reported Device Performance (as stated in the document)
ECG/Heart Rate ResponseAAMI/ANSI Cardiac Monitors, Heart Rate Meters, and Alarms Standards."ECG/Heart Rate response was evaluated against the AAMI/ANSI Cardiac Monitors, Heart Rate Meters, and Alarms Standards." Implicitly met, as the clinical trial for NIBP is separate.
Other Vital SignsMennen Medical's Design Specifications."Testing of other Vital signs was performed against Mennen Medical's Design specifications. Measurements were within published specifications." (Specific values not provided).
Non-Invasive Blood Pressure (NIBP)IEC601-2-30 Requirements for automatic cycling indirect blood pressure monitoring. AAMI SP-10 Electronic or Automated Sphygmomanometers Standard."The performance of the ENVOY was compared against another automated non-invasive blood pressure device (Mennen Medical HORIZON XL)." "The study followed the AAMI Electronic or Automate Sphygmomanometers Standard SP-10 (Refer to Part 8 for NIBP clinical trials and results)." (Specific results are referenced to "Part 8" but not provided in the summary).
Pulse Oximetry(Not explicitly stated in this summary regarding specific standard limits)"Clinical validation studies of the Pulse Oximetry Module are present in Part 8 of the submittal." (Specific results are referenced to "Part 8" but not provided in the summary).
Patient Safety (Electrical)EN60601-1 and AAMI/ANSI Safe Current Limits for Electromedical Apparatus standard."Patient safety was confirmed by testing EN60601-1 and AAMI/ANSI Safe Current Limits for Electromedical Apparatus standard." (Implicitly met).
Environmental (Flammability, Mechanical, Temp, Humidity)IEC 601-1 requirements, IEC-68 series (2-1, 2-2, 2-3, 2-56, 2-13, 2-6, 2-36, 2-27), MIL-STD-810E."ENVOY complies with IEC 601-1 requirements for flammability, mechanical abuse, temperature and humidity. Environmental testing was conducted pursuant to IEC-68-2-1 (Cold), IEC-68-2-2 (Dry Heat), IEC-68-2-3 and IEC-68-2-56 (Damp heat steady state), IEC-68-2-13 (Low air pressure), IEC-68-2-6 (Vibration - sinusoidal), IEC-68-2-36 (Random vibration wide band), IEC 68-2-27 (Shock) and MIL-STD-810E." (Implicitly met).
EMCIEC 601-1-2. IEC 801-1/2/3/4 and 5. EN 55011 and CISPR 16."EMC testing, pursuant to IEC 601-1-2. IEC 801-1/2/3/4 and 5. EN 55011 and CISPR 16 was conducted and compliance verified." (Implicitly met).
Software ValidationMennen Medical's test plan protocol (verified functionality, user input/output, error recovery, hidden functionality)."Software validation verified the functionality of the Mennen Medical ENVOY from a 'black box' approach." "The results met expectations and the software was approved for release..." (Specific quantitative metrics not provided).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in this summary. The text mentions "Simulated inputs were used to test the vital signs monitored by the ENVOY" and "Simulated patient waveforms were generated using a DNI Navada Inc. 217A Patient simulator and Mennen Medical simulators." For NIBP and Pulse Oximetry, it refers to "clinical trials" in "Part 8" but doesn't detail the sample size or provenance in this summary.
  • Data Provenance:
    • Simulated Data: For general vital signs (ECG, Respiration, Invasive BP), "proprietary algorithms and software based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions."
    • Clinical Data: For NIBP and Pulse Oximetry, "clinical validation studies" are mentioned, with results in "Part 8." The country of origin and whether data was retrospective or prospective is not specified in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document does not mention the use of experts to establish ground truth for the test set explicitly. The ground truth for simulated inputs would be the known parameters set on the simulators. For the clinical studies (NIBP, Pulse Oximetry), the "ground truth" would typically come from reference devices or expert-interpreted data, but details are not provided in this summary.

4. Adjudication Method for the Test Set

  • The document does not describe any adjudication method for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study is not indicated or described in this document. The device is a patient monitor, not an AI-assisted diagnostic tool for human readers.
  • Effect size of human readers improving with AI vs. without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device for human interpretation improvements in the context of MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

  • Yes, the core of the validation described for ECG/Heart Rate and "other Vital signs" (excluding SpO2 and NIBP modules that use vendor software/hardware) implicitly involves standalone algorithm performance. The device's "proprietary algorithms and software" for ECG, Respiration, and Invasive Blood Pressure were tested using simulated inputs. The performance for these parameters against AAMI/ANSI standards (for ECG/HR) and Mennen Medical's Design specifications (for others) indicates a standalone evaluation.
  • It's important to note that "standalone" here refers to the device's algorithms processing data and displaying results without human intervention in the processing itself, which is standard for patient monitors. However, the clinical use always involves a human-in-the-loop (healthcare provider) to interpret the displayed information.

7. The Type of Ground Truth Used

  • For ECG/Heart Rate, Respiration, Invasive BP: The ground truth came from "Simulated patient waveforms generated using a DNI Navada Inc. 217A Patient simulator and Mennen Medical simulators," and "well known and accepted data bases." This is a form of simulated data/known input ground truth.
  • For Non-Invasive Blood Pressure and Pulse Oximetry: These parameters underwent "clinical validation studies" and their results are in "Part 8," implying reference device comparisons (e.g., against an established, accurate NIBP device or pulse oximeter) or potentially directly measured physiological data. The summary doesn't specify if it was pathology, outcomes data, or expert consensus for these clinical trials. Given the nature of vital sign monitoring, it's most likely comparison against a gold-standard measurement.

8. The Sample Size for the Training Set

  • The document does not specify a sample size for a "training set." The proprietary algorithms for ECG, Respiration, and Invasive Blood Pressure were "based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases." This suggests the algorithms were likely developed and refined (trained/tuned) using these historical "data bases," but the size of these databases is not provided. Modern AI/ML terminology like "training set" might not have been standard practice in medical device submissions in 1997.

9. How the Ground Truth for the Training Set Was Established

  • The document states that the proprietary algorithms were "based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions."
  • This implies the ground truth for these "data bases" was established through:
    • Reference device comparison: Likely, the older Mennen Medical devices and the data they collected were compared against established, validated reference measurements for physiological parameters and artifact detection.
    • Expert knowledge/manual annotation: For "waveform artifact," it's highly probable that experts (e.g., cardiologists, intensivists) annotated or identified these artifacts in the "well known and accepted data bases."
    • The exact methodology for establishing ground truth for these historical "data bases" is not detailed in this summary.

{0}------------------------------------------------

K974510

ENNE

MEDICAL LTD.

APR 1 4 1998

(

MENNEN MEDICAL LTD. Kiryat Weizmann Science Park P.O.B. 102 Rehovot 76100 Israel Tel: 972-8-938-3030 Fax: 972-8-940-6519

  • 30 October 1997 Date:
    Food and Drug Administration To: Center for Devices and Radiological Health Document Control Center (HFZ-401) 1390 Piccard Drive Rockville MD 20850

Document Control Clerk Attn.:

Kenneth Raichman From: Director of Regulatory Affairs

510(k) Summary Topic: ENVOY Patient Monitor Safety and Effectiveness

Product Name

ENVOY Proprietary: Physiological Patient Monitor Common:

Establishment Registration Number and Address

9611022 Registration Number Mennen Medical LTD. Ken Raichman, Director of Regulatory Affairs Kiryat Weizmann Science Park Rehovot 76100 Israel Tel: 972-8-938-3030

Classification (Section 513)

Classification: Panel:

Class III Anesthesiology and Respiratory/Cardiology

Performance Standards

None promulgated

Voluntary Standards

AAMI/ANSI ES1 Safe Current Limits for Electromedical Devices AAMI/ANSI EC13, Cardiac Monitors, Heart Rate Meters and Alarms AAMI/ANSI SP-10, Electronic or Automated Sphygmomanometers IEC 601-1 Medical Electrical Equipment IEC 601-2-27 Safety of electrocardiographic monitoring IEC601-2-30 Requirements for automatic cycling indirect blood pressure monitoring IEC601-2-34 Requirements for Invasive Blood Pressure monitoring equipment

Predicate Devices

MENNEN MEDICAL HORIZON XL (K911616). HEWLETT-PACKARD M1175/26A COMPONENT MONITORING SYSTME (K941811).

Date Prepared: 30 October 1997

{1}------------------------------------------------

510(k) Summary, ENVOY Safety & Effectiveness

Device Description

ENVOY is hospital based, multiparameter patient monitor for monitoring physiological patient vital signs.

The ENVOY patient monitor system consists of the following components:

Main Processing Unit Display Unit Module Rack Vital Signs Plug-in Modules

ENVOY vital signs modules acquire vital signs data from the patient, and display their waveforms and alarms indications on the ENVOY display unit. Vital signs and waveform information are displayed simultaneously on the ENVOY Display Unit. Up to 8 traces can be displayed at any one time.

The vital signs modules interface with readily available physiologic transducers through electrically isolated patient input connections. After amplification, the signals are digitized, analyzed and displayed. All processing and alarm determination for ECG, Respiration and Invasive Blood Pressure is made using proprietary algorithms and software based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions. SpO2 and Non-Invasive Blood Pressure Modules use sofware/hardware from vendors whose products already appearing on the USA market.

Information from each vital sign is presented in a separate portion of the display. Each vital sign is labeled for identification and numeric value. Displayed Vital sign information can include: Primary Vital Sign Name, Waveform, Vital Sign Numeric Value, Alarm Status Message.

Operation of the ENVOY is accomplished by interaction with front panel controls on the main processor unit. A quick-knob control allows direct interaction with displayed menus for direct parameter selection and setup. Where manual entry of alphanumeric information is required, a menu keyboard menu is display.

ENVOY is a reusable, software driven, patient monitor, intended for use as part of a physiological monitoring system in a hospital environment. As such it is not a life supporting, or life sustaining device; nor is it implantable and therefore sterility is not a consideration.

ENVOY complies with IEC 601-1 Medical Electrical Equipment, IEC 601-2-27 Safety of electrocardiographic monitoring, IEC 601-2-30 Requirements for automatic cycling indirect blood pressure monitoring, AAMI/ANSI SP-10 Electronic or Automated Sphygmomanometers, IEC 601-2-34 Requirements for invasive blood pressure monitoring, AAMI/ANSI ESI Safe Current Limits for Electromedical Apparatus, and AAMI/ANSI EC13, Cardiac Monitors, Heart Rate Meters and Alarms. ENVOY is not a kit, does not contain any drug or biological products and is not for prescription use.

{2}------------------------------------------------

Indications for Use:

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, and Pulse Oximetry. The ENVOY may be used to monitor a wide range of patient conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

{3}------------------------------------------------

510(k) Summary, ENVOY Safety & Effectiveness

Substantial Equivalency Information:

The following tables summarize data on the Mennen Medical ENVOY, the Mennen Medical The forlowing actions of animals ), and the Hewlett-Packard M1175/26A Component Monitoring System (K941811), both are substantially equivalent devices, available in the U.S. market.

DisplayedParametersMennen MedicalENVOYPatient MonitorMennen MedicalHORIZON XLPatient MonitorH-PM1175/26AComponentMonitoring System
ECGYesYesYes
Heart RateYesYesYes
Invasive BPYesYesYes
Noninvasive BPYesYesYes
Pulse OximetryYesYesYes
RespirationYesYesYes
TemperatureYesYesYes
Alarm Indications:Mennen MedicalENVOYPatient MonitorMennen MedicalHORIZON XLPatient MonitorH-PM1175/26AComponentMonitoring System
ECGVisual & SoundVisual & SoundVisual & Sound
Heart RateVisual & SoundVisual & SoundVisual & Sound
Invasive BPVisual & SoundVisual & SoundVisual & Sound
Noninvasive BPVisual & SoundVisual & SoundVisual & Sound
Pulse OximetryVisual & SoundVisual & SoundVisual & Sound
RespirationVisual & SoundVisual & SoundVisual & Sound
TemperatureVisual & SoundVisual & SoundVisual & Sound
Graded according toSeverity: audio, visualYESYESYES
Technical Alarms(INOPS)YESYESYES
Resetting/SuspendingAlarms - silence tone,automatic reactivateafter set intervalYESYES?YES

{4}------------------------------------------------

510(k) Summary, ENVOY Safety & Effectiveness

MeasurementPrincipleMennen MedicalENVOYPatient MonitorMennen MedicalHORIZON XLPatient MonitorH-PM1175/26AComponentMonitoring System
Heart RateFrom ECG, PulseOxymetry, BloodPressureFrom ECGFrom ECG
Invasive BPPressure TransducerPressure TransducerPressure Transducer
Noninvasive BPArm CuffArm CuffArm Cuff
Pulse OximetryInfra-red SensorInfra-red SensorInfra-red Sensor
RespirationFrom ECGFrom ECGFrom ECG
TemperatureTemp probeTemp probeTemp probe
Auxiliary FunctionsMennen MedicalENVOYPatient MonitorMennen MedicalHORIZON XLPatient MonitorH-PM1175/26AComponentMonitoring System
Change ECG LeadSelectionYESYESYES
Display of ArrhythmiaInformationYESYESYES
Change BP Range/ScaleYESYESYES
Data Review: TrendsYESYESYES
Data Review: TabularYESYESYES
User definedConfiguration SetupYESYES
User defined DefaultSettingsYESYES

Alternative Practices and Procedures:

The information displayed by ENVOY is acquired by similar techniques used by other commercially available equipment within the same product category. Like the referenced predicate products, ENVOY provides vital signs information for a patient. This enables a single operator to monitor several parameters for significant clinical events. Whenever an alarm is detected the parameter and alarm type are identified. Viewing the patient's current physiological waveforms on ENVOY allows the user to qualify the severity of the alarm condition.

ENVOY integrates the measured vital signs information for a patients into a single display, and provides a uniform method for interaction and comparison.

{5}------------------------------------------------

510(k) Summary, ENVOY Safety & Effectiveness

Marketing History

ENVOY has no marketing history.

Summary of Validation

ENVOY is a physiologic patient monitor. Device validation studies were performed to verify that conditions detected by the "bedside device" were completely and accurately reported and displayed on the ENVOY.

Validation:

ENVOY was tested against Mennen Medical's Design Specifications, in accordance with the procedures identified in Part 7 Section 2, with results as presented in Part 7 Section 3.

Simulated inputs were used to test the vital signs monitored by the ENVOY. The objective of this study was to compare the performance of the ENVOY against voluntary industrial standards where appropriate, and Mennen Medical design standards where appropriate, or where industrial standards did not exist.

Software validation was tested against Mennen Medical's test plan protocol. Simulated patient waveforms were generated using a DNI Navada Inc. 217A Patient simulator and Mennen Medical simulators.

ECG/Heart Rate response was evaluated against the AAMI/ANSI Cardiac Monitors, Heart Rate Meters, and Alarms Standards.

Testing of other Vital signs was performed against Mennen Medical's Design specifications. Measurements were within published specifications.

Patient safety was confirmed by testing EN60601-1 and AAMI/ANSI Safe Current Limits for Electromedical Apparatus standard.

ENVOY complies with IEC 601-1 requirements for flammability, mechanical abuse, temperature and humidity. Environmental testing was conducted pursuant to IEC-68-2-1 (Cold), IEC-68-2-2 (Dry Heat), IEC-68-2-3 and IEC-68-2-56 (Damp heat steady state), IEC-68-2-13 (Low air pressure), IEC-68-2-6 (Vibration - sinusoidal), IEC-68-2-36 (Random vibration wide band), IEC 68-2-27 (Shock) and MIL-STD-810E.

EMC testing, pursuant to IEC 601-1-2. IEC 801-1/2/3/4 and 5. EN 55011 and CISPR 16 was conducted and compliance verified.

Non-invasive blood pressure was evaluated for IEC601-2-30 Requirements for automatic cycling indirect blood pressure monitoring. The performance of the ENVOY was compared against another automated non-invasive blood pressure device (Mennen Medical HORIZON XL). The study followed the AAMI Electronic or Automate Sphygmomanometers Standard SP-10 (Refer to Part 8 for NIBP clinical trials and results).

Clinical validation studies of the Pulse Oximetry Module are present in Part 8 of the submittal.

{6}------------------------------------------------

Page 7 of 7.

510(k) Summary, ENVOY Safety & Effectiveness

Conclusions Drawn from Validation Studies:

The results of the validation studies indicate that ENVOY is safe, effective and poses no new risks when compared against the Mennen Medical Design Standards, and equipment already in clinical use.

Software Validation

Software validation verified the functionality of the Mennen Medical ENVOY from a "black box" approach. Validation was performed by a persons other than those involved in the design of the system. This independent audit confirmed that the system (both hardware and software) met the specified requirements.

Software Validation Plan:

The software validation plan tested the following areas:

Verified that the system performed according to specified requirements.

Verified the system by exercising user input and assuring correct output.

Checked for hidden functionality.

Verified that the system recovers from errors.

The SOFTWARE VALIDATION PLAN was reviewed and approved by Mennen Medical's Engineering and Quality Assurance Departments, insuring that the tests were both valid and thorough.

The SOFTWARE VALIDATION PLAN was executed and the results were analyzed by the Quality Assurance and Engineering Departments. The results met expectations and the software was approved for release, pending clearance of the 510(k) process.

MENNEN MEDICAL LTD.

K. Bardin

Kenneth Raichman, Director of Regulatory Affairs

Enclosures.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, with flowing lines beneath them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 4 1998

Mr. Kenneth Raichman Director of Regulatory Affairs MENNEN MEDICAL Ltd. Kiryat Weizmann Science Park P.O.B. 102 Rehovot 76100 Israel

Re: K974510 Envoy Patient Monitor Regulatory Class: III (three) Product Code: 74 DSI Dated: March 5, 1998 Received: March 12, 1998

Dear Mr. Raichman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{8}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Page 1 of 1

510(k) Number (if known)

Kora 774510

Device Name

ENVOY PATIENT MONITOR

Indications for Use:

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, and Pulse Oximetry. The ENVOY may be used to monitor a wide range of patient ' conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

MRI Compatibility Statement:

The ENVOY is not compatible for use in a MRI magnetic field.

(DO NOT WRITE BELOW THIS LINE)

Concurrence of CDRH, Office of Device Evaluation (ODE)

or

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K974510

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter Use (Optional Form 1-2-96)

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.