K Number
K974235

Validate with FDA (Live)

Device Name
AMPLAID AM50
Manufacturer
Date Cleared
1998-02-10

(90 days)

Product Code
Regulation Number
874.1090
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Amplaid AM50 is a Programmable Admittance Meter which can:

  • Evaluate middle ear functions such as otitis media, glue ear, eardrum scar tissues, . perform myringotomy status, ossicular chain discontinuity ear canal volume, otosclerosis, stapes fixation.
  • Perform Acoustic reflex test. ●
  • Determine acoustic reflex threshold .
  • . Perform reflex decay test.

It is intended for use by a qualified/trained audiologist on both adult and pediatric subjects for measurement and diagnosis of various types of hearing losses.

Device Description

The Amplaid AM50 ADMITTANCE METER performs plan and compensated tympanometry; Programmed and manual stimuli for ipsilateral and contralateral acoustic reflex; Automatic reflex threshold; and Decay measurements.

AI/ML Overview

The provided documentation describes the Amplaid AM50 Clinical Automatic Programmable Admittance Meter and its substantial equivalence to a predicate device, the Amplaid 770 Admittance Meter, rather than outlining specific acceptance criteria or an in-depth study proving its performance against such criteria.

The submission is a 510(k) Summary of Safety and Effectiveness, which typically focuses on demonstrating that a new device is "substantially equivalent" to a legally marketed predicate device. This process often relies on comparisons of intended use, technological characteristics, and performance data to show that the new device is as safe and effective. It explicitly states, "The results of bench and user testing indicates that the new device is as safe and effective as the predicate device." and "After analyzing both bench and user testing data, it is the conclusion of Amplifon S.p.A. that the Amplaid AM50 is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate Device."

Given this, I cannot provide a table of acceptance criteria and reported device performance in the typical format of a clinical trial or performance study, because the document does not present such criteria or detailed results. Instead, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence to the predicate device, which itself would have undergone performance validation.

However, I can extract the information that is present and frame it in response to your request as best as possible given the limitations of the provided text.


1. Table of Acceptance Criteria and the Reported Device Performance

As specific, quantitative acceptance criteria are not explicitly defined in the provided 510(k) summary, the "acceptance" is understood to be the demonstration of substantial equivalence to the predicate device, the Amplaid 770. The "performance" is implicitly deemed equivalent to the predicate.

Acceptance Criteria (Implicit from Substantial Equivalence)Reported Device Performance (as stated in the 510(k) Summary)
Intended Use: Device must perform clinical auditory impedance testing for evaluation of middle ear functions, acoustic reflex tests, thresholds, and decay.Amplaid AM50 Performance: "The Amplaid AM50 is a Programmable Admittance Meter which can: - Evaluate middle ear functions such as otitis media, glue ear, eardrum scar tissues, perform myringotomy status, ossicular chain discontinuity ear canal volume, otosclerosis, stapes fixation. - Perform Acoustic reflex test. - Determine acoustic reflex threshold. - Perform reflex decay test." (Stated as "Same" as predicate in the comparison chart, implying equivalent performance for these functions).
Technical Characteristics: Device must perform plain and compensated tympanometry; Programmed and manual stimuli for ipsilateral and contralateral acoustic reflex; Automatic reflex threshold; and Decay measurements.Amplaid AM50 Performance: "The Amplaid AM50 ADMITTANCE METER performs plan and compensated tympanometry; Programmed and manual stimuli for ipsilateral and contralateral acoustic reflex; Automatic reflex threshold; and Decay measurements." (Stated as "Same" as predicate in the comparison chart, implying equivalent performance for these functions).
Safety and Effectiveness: Device must be as safe and effective as the predicate device (Amplaid 770).Amplaid AM50 Performance: "The results of bench and user testing indicates that the new device is as safe and effective as the predicate device. The same circuit technology is used in both devices." "After analyzing both bench and user testing data, it is the conclusion of Amplifon S.p.A. that the Amplaid AM50 is as safe and effective as the predicate device..."
Standards Compliance: Compliance with relevant audiometric and electrical safety standards (e.g., ANSI 1969, ISO 1975, IEC 61027, UL-544, IEC 601).Amplaid AM50 Performance: "Audiometric: (Same) [as predicate, which is] ANSI 1969, ISO 1975 for contralateral, 2 cm³ cavity for ipsilateral, IEC 61027" "Electrical safety: (Same) [as predicate, which is] UL-544, IEC 601"
Technological Differences: Differences should not raise new questions of safety or effectiveness.Amplaid AM50 Performance: The only changes noted are "packaging and control software," where "all keyboard, display, and control functions have been moved to the attached (via RS232) personal computer." "few technological differences, and has no new indications for use."

The "study" that proves the device meets (or rather, is substantially equivalent to) these criteria is referred to as "bench and user testing." No specific details on the methodology, sample sizes, or results of these tests are provided in this summary.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document only mentions "bench and user testing."
  • Data Provenance: Not specified. The manufacturer is Amplifon S.p.A. in Milan, ITALY. It is plausible the "bench and user testing" was conducted in Italy or a region accessible to the manufacturer, but this is not confirmed. The data appears to be retrospective in the sense that it supports a regulatory submission rather than a prospective clinical trial for public dissemination.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not provided. The "user testing" would likely involve qualified audiologists as the device is intended for use by a "qualified/trained audiologist," but the number and specific qualifications are not detailed.

4. Adjudication method for the test set

  • Not specified.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This device is not an AI-powered diagnostic tool, but rather an acoustic admittance meter. Therefore, an MRMC comparative effectiveness study involving "AI vs without AI assistance" is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is a medical instrument and requires human operation ("intended for use by a qualified/trained audiologist"). Therefore, a standalone (algorithm only) performance study is not applicable and was not performed. The performance evaluated was of the device in a user-testing context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The concept of "ground truth" as typically applied to diagnostic AI models is not directly applicable here. For an admittance meter, "ground truth" would likely refer to the accuracy and reliability of its physical measurements against known standards or against the established performance of the predicate device. This would involve calibrated equipment and possibly clinical comparisons, but the specifics are not detailed.

8. The sample size for the training set

  • This device does not utilize a "training set" in the context of machine learning or AI. Its functionality is based on established physical principles and circuit technology, not on learned patterns from a dataset.

9. How the ground truth for the training set was established

  • As there is no training set for this type of device, this question is not applicable.

{0}------------------------------------------------

K974235

EXHIBIT 2 Amplifon S.p.A. Via Ripamonti, 133 20141 Milan, ITALY Tel ++39-2-57472482 Fax ++39-2-57409427 Contact: Giovanni Rollier, President

FEB | 0 1998

November 7, 1997

510(k) Summary of Safety and Effectiveness

    1. Identification of the Device: Proprietary-Trade Name: Amplaid AM50 Classification Name: Auditory Impedance Tester 77ETY Common/Usual Name: Clinical Admittance Meter
    1. Equivalent legally marketed devices This product is similar in design and function to the Amplaid 709 Admittance Meter (K802097) and the 770 Admittance Meter (K903066)
    1. Indications for Use (intended use) The Amplaid AM50 is a Programmable Admittance Meter which can:
    • Evaluate middle ear functions such as otitis media, glue ear, eardrum scar tissues, . perform myringotomy status, ossicular chain discontinuity ear canal volume, otosclerosis, stapes fixation.
    • Perform Acoustic reflex test. ●
    • Determine acoustic reflex threshold .
    • . Perform reflex decay test.

It is intended for use by a qualified/trained audiologist on both adult and pediatric subjects for measurement and diagnosis of various types of hearing losses.

    1. Description of the Device: The Amplaid AM50 ADMITTANCE METER performs plan and compensated tympanometry; Programmed and manual stimuli for ipsilateral and contralateral acoustic reflex; Automatic reflex threshold; and Decay measurements.
    1. Safety and Effectiveness, comparison to predicate device. The results of bench and user testing indicates that the new device is as safe and effective as the predicate device. The same circuit technology is used in both devices. The only change is packaging and control software, whereby all keyboard, display, and control functions have been moved to the attached (via RS232) personal computer.

{1}------------------------------------------------

CharacteristicPredicate device:New device:
Amplaid 770Amplaid AM50
Intended Use:Clinical auditory impedancetesting applications(Same)
Technical characteristicsPer K903066(Same)
Physical characteristics:
Computer interfaceRS232 Transmit onlyRS232 Bi-directional
DisplayBuilt-in liquid crystalVia attached computer
Control interfaceBuilt-in keyboardVia attached computer ordedicated keyboard.
Size/weight17.3" W x 19.3" D x 7" H,11.4" W x 12"D x 4" H
28 lbs.4.4 lbs.
Energy Source:115/230 Vac, ± 10%, 50-60 Hz(Same)
Hardcopy Output:Built-in Thermal printerVia attached computer
Standards and Safetycharacteristics:
Audiometric:ANSI 1969, ISO 1975 forcontralateral, 2 $cm³$ cavity foripsilateral, IEC 61027(Same)
Electrical safety:UL-544, IEC 601(Same)

6. Substantial Equivalence Chart

7. Conclusion

After analyzing both bench and user testing data, it is the conclusion of Amplifon S.p.A. that the Amplaid AM50 is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate Device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a seal for the Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH, HUMAN" around the top left of the circle. The right side of the seal contains a symbol that looks like three stylized human profiles stacked on top of each other.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Amplifon S.p.A. c/o Kamm & Associates Daniel Kamm. P.E., Regulatory Engineer P. O. Box 7007 Deerfield, Illinois 60015

K974235 Re: Amplaid AM50 Clinical Automatic Programmable Admittance Meter Dated: November 7, 1997 Received: November 12, 1997 Regulatory class: II Procode: 77 ETY. 21 CFR 874.1090

FEB | 0 |998

Dear Mr. Kamm:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendnents, or to devices that have been reclassified in accordance with the provisions of the FederallPood, y Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the West general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing off a practice, labeling, and prohibitions against misbranding and adulteration. 11:48 11

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Revolutions Sittle 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Pracice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR-Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply of with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation i you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions . or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of t substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yilgo diagnostic devices), please contact the Office of Compliance at (301) 594-613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled "Misbranding by reference to premaiter notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely vours.

h.Nliau Yhi

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

i) Indications for Use

510(k) Number 5 974 2 3 5

Device Name: Amplaid AM50 Clinical Automatic Programmable Admittance Meter

Indications for Use: The Amplaid AM50 is a Programmable Admittance Meter which can

  • Evaluate middle ear functions such as otitis media, glue ear, eardrum scar tissues, 1. perform myringotomy status, ossicular chain discontinuity ear canal volume, otosclerosis, stapes fixation.
  • Perform Acoustic reflex test. 2.
    1. Determine acoustic reflex threshold
    1. Perform reflex decay test.

It is intended for use by a qualified/trained audiologist on both adult and pediatric subjects for measurement of acoustic impedance.

Concurrence of CDRH, Office of Device Evaluation (ODE)

David h. Seppanen

OR

(Division Sign-Off) ivision of Reproductive, Abdominal, ENT and Radiological Devi 510(k) Number

Prescription Use V

Over the Counter Use (Per 21 CFR 801.109)

§ 874.1090 Auditory impedance tester.

(a)
Identification. An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.(b)
Classification. Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.