K Number
K974084

Validate with FDA (Live)

Date Cleared
1998-03-03

(125 days)

Product Code
Regulation Number
878.4580
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

THE PRODUCT WILL BE USED TO ILLUMINATE SURGICAL PROCEDURES WITH COLOR CORRECTED WHITE LIGHT AND HEAT FILTERING (IR). DESIGNED TO ELIMINATE SHADOWS AND ILLUMINATE DEEP CAVITY WOUNDS WITH ADEQUATE PENETRATION. IT IS INTENDED TO BE USED BY SURGEONS AND OTHER MEDICAL CARE PRACTITIONERS IN A SURGICAL SETTING. THERE ARE NO CONTRA-INDICATIONS FOR USE.

Device Description

The Getinge/Castle OptiView® surgical lights are designed to provide the high quality light required in an operating room setting. There are four major considerations that the lights have been designed for: 1. Shadow Elimination. 144 radial dioptric lenses, disposed equally on the lens glass, are used to broaden the light beam such that a 10 X 20 cm patch of light is produced on the surgical site from each lens. Should one or several of the lenses become obstructed by the surgeon or from equipment, a constant quantity of light is subtracted from each point of the light field. This virtually eliminates shadows and provides for optimum visibility. 2. Optimal Depth of Field. A central toric lens and conic mirror is used to produce very broad beams whose intersection are not points but are volumes. In some cases the volume can measure up to 70 cm in height and 20 cm across the center. The design also allows the volume of light to be moved in the axis of the beam, this is called variable focusing. 3. Heat Elimination and High Color Rendition. Through the use of a special type of glass, the emitted light is corrected for color (towards the model for broadband sunlight) and filtered for infrared. This eliminates the thermal effect of the beam and allows a color rendition index of Ra = 95 or greater. 4. Electromagnetic and Bacteriologic Neutrality. The product have been designed and tested to comply with the strictest electromagnetic compatibility standards. Also, the sealed optical unit, the flat, smooth surfaces and sensitive touch-controls provide for an extremely aseptic environment. The mechanical aspects of the lights are that they are constructed of specifically designed steel load-bearing drop tubes, yokes and suspension assemblies that are coated with a strong epoxy resin paint. These structures are mechanically connected at pivot points that incorporate counter-balances for ease of movement as well as specially designed wire-runs and fully pivoting electrical connectors. The electrical aspects of the lights are designed to comply with international safety standards such as: Low Voltage Directive (LVD) 72/23/CEE, Decree No. 95-1081 Electromagnetic Compatibility Directive (EMC) 89/336/CEE, Decree No. 95-587 Medical Device Directive (MDD) 93/42/CEE. Power Supply Units, NF EN 60-4391 (IEC 433-1), NF EN 60-439-Operating Theatre Lighting, NF EN 60-601-1 (IEC 601). The design will also be examined to UL 544 and certified by either Underwriters Laboratory or Intertek Testing Services (ITS) which should result in the UL or ETL mark. The lights were previously CE marked, while marketed in Europe, by the original manufacturer: Scialityque Industrie. The lights are electrically controlled via pressure sensitive touch controls that may be located on the lights or at remote locations. The controls allow the lights to be energized, the light level to be adjusted and the focus mechanisms to be operated. The light beam is achieved through the use of one or two (dependent upon model) tungsten-halogen 24 VDC bulbs that are contained in a sealed optical unit. The emitted light is focused through a toroidial lens and is reflected off of a conic, polished aluminum reflector and through 144 radial dioptric lenses which allows for a radial-focused broad-beam light pattern on the surgical surface. This eliminates shadows while concurrently extending the true depth of field of the beam.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Getinge/Castle OptiView® Surgical Light. It details the device's description, intended use, and a comparison to predicate devices to establish substantial equivalence. However, it does not include a formal study with acceptance criteria and reported device performance in the way a clinical trial or performance validation study would typically present it for an AI/ML device.

Instead, the "acceptance criteria" can be inferred from the comparison tables which demonstrate that the OptiView® surgical lights meet or exceed the specifications of the predicate devices for key performance characteristics. The "study" proving this is implicitly the comparison itself, showing that the new device's specifications are within acceptable ranges relative to already-marketed and cleared devices.

Here's an analysis based on the provided text, addressing the requested points:


Acceptance Criteria and Device Performance for Getinge/Castle OptiView® Surgical Light

The "acceptance criteria" for this device are implicitly defined by its substantial equivalence to predicate devices, particularly in key performance specifications relevant to surgical lighting. The study proving conformance is a direct comparison of these specifications.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission for a non-AI surgical light, formal acceptance criteria as typically seen for AI/ML devices (e.g., sensitivity, specificity thresholds) are not explicitly stated. Instead, the acceptance criteria are satisfied by demonstrating that the OptiView® series lights perform comparably or better than the predicate devices across relevant technical specifications.

The tables below present selected specifications for the OptiView® 500 series and 700 series alongside their predicate devices. The "Acceptance Criteria" column represents the range or value observed in the predicate devices within which the new device should fall to demonstrate substantial equivalence. The "Reported Device Performance" column shows how the OptiView® series lights meet these implicit criteria.

OptiView® 500 series vs. Predicate Devices (ALM PRC 7001 series, Berchtold C-570 series)

SpecificationAcceptance Criteria (Predicate Range/Value)Reported Device Performance (OptiView® 500 series)
Illumination Level, max ft-candles (lux)9,300 (100,000) - 13,530 (145,000) ft-candles (lux)12,300 (130,000) ft-candles (lux)
Color Temperature, °K3,500 - 4,500 °K4,300 °K
Field Size, Diameter, cm (in)18-28 (7-11) cm10-15 (3.9-5.9) cm
Depth of Field, cm (in)70 (27.6) - 80 (31) cm40 (15.7) cm
Bulb Life, hr1000 hr500 (min) hr
Heat Filtering, %99%99%
ControlsContinuous/Adjustable Dimmer, Adjustable Focus, Adjustable Field SizeVariable intensity Dimmer, Adjustable Focus, Adjustable Field Size

Note on Field Size and Depth of Field for OptiView® 500 series: While the reported values are sometimes outside the predicate range, the narrative emphasizes benefits like "optimal depth of field" (up to 70cm in some cases) and variable focusing, suggesting these differences are part of an equivalent or improved design aimed at specific functional benefits such as shadow elimination and deep cavity penetration.

OptiView® 700 series vs. Predicate Devices (ALM PRC 9001 series, AMSCO SQ240)

SpecificationAcceptance Criteria (Predicate Range/Value)Reported Device Performance (OptiView® 700 series)
Illumination Level, max ft-candles (lux @ 1 m)10,000 (107,640) - 12,000 (129,170) ft-candles (lux)11,200 (120,000) ft-candles (lux)
Color Temperature, °K3,500 - 4,400 °K4,300 °K
Field Size, Diameter, cm (in)16.5 (6.5) - 21.6 (8.5) cm17-24 (6.7-9.4) cm
Depth of Field, cm (in)66 (26) - 70 (27.6) cm50 (19.7) cm
Bulb Life, hr1000 hr500 (min) hr
Heat Filtering, %98%99%
ControlsContinuous/Variable intensity Dimmer, Fixed/Adjustable Focus, Adjustable Field SizeVariable intensity Dimmer, Adjustable Focus, Adjustable Field Size

Note for OptiView® 700 series: similar to the 500 series, specific values like bulb life (500 min hrs vs 1000 hrs for predicates) might differ, but the overall functional equivalence and safety are argued through the 510(k) process, emphasizing design features like "true depth of field of the beam."

2. Sample Size Used for the Test Set and Data Provenance

This is not a study involving a "test set" in the context of AI/ML. The "test" consists of comparing the specifications of the new device to existing predicate devices. The data provenance is from the technical specifications provided by Getinge/Castle and the manufacturers of the predicate devices. The data is based on engineering specifications and likely internal testing, not patient data.

  • Sample Size for Test Set: Not applicable in the context of an AI/ML device test set. The "test" is a comparison of product specifications.
  • Data Provenance: Technical specifications of manufactured surgical lights (OptiView series and predicate devices). Implicitly, this data is from the respective manufacturers, likely based on engineering designs and performance testing. The country of origin for Getinge/Castle is the USA (North Charleston, SC). For predicate devices, ALM, AMSCO, and Berchtold are mentioned, which are international manufacturers. The data is retrospective in the sense that it reflects established specifications of existing devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. "Ground truth" in this context is established by the documented technical specifications of the surgical lights themselves, which are derived from engineering and manufacturing standards, not expert medical opinion on a dataset.

4. Adjudication Method for the Test Set

Not applicable. There is no "adjudication method" as this is not a diagnostic device with ambiguous outputs requiring clinical interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. This is a medical device (surgical light), not a diagnostic algorithm, and therefore an MRMC study is not relevant or performed in this submission.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This is a physical surgical lighting device, not an AI algorithm. "Standalone" performance testing would refer to the device's technical specifications and functionality (e.g., light intensity, color temperature, heat dissipation) as reported in the tables, independent of human interaction during those measurements. The information provided demonstrates this by listing the device's inherent performance characteristics.

7. The type of ground truth used

The "ground truth" implicitly used for comparison is the published performance specifications of legally marketed predicate surgical lights, which represent accepted performance benchmarks for such devices in a clinical setting.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that uses a "training set."

9. How the ground truth for the training set was established

Not applicable.


Summary of the "Study" (510(k) Premarket Notification):

The "study" in this submission is the comprehensive comparison of the Getinge/Castle OptiView® Surgical Light (including its 200, 500, and 700 series) with predicate devices (ALM Prismatic models, AMSCO SQ240, Berchtold Chromophare models). This comparison is based on detailed technical specifications across numerous parameters, including:

  • Physical dimensions (diameter, weight)
  • Electrical specifications (supply voltage, power, bulb type, volts, life)
  • Optical performance (color temperature, field size, depth, focal length, illumination level)
  • Control features (dimmer, focus, field size, rotation)
  • Material composition (heat filtering, filter material, reflector material)
  • Mounting types and ceiling height requirements

The premise of a 510(k) submission is to demonstrate "substantial equivalence" to a legally marketed predicate device. This is achieved by showing that the new device has the same intended use, and either has the same technological characteristics as the predicate, or has different technological characteristics but does not raise new questions of safety and effectiveness and is as safe and effective as the predicate device.

The provided tables and descriptive text fulfill this requirement by showing that the OptiView® series lights meet or exceed the performance metrics of the predicate devices for critical elements like illumination, color rendition, heat filtering, and shadow elimination, indicating that they are at least as safe and effective. The narrative explicitly states: "We are not introducing new technology. The OptiView® lights and the predicate devices use very similar technology to achieve the same basic goal..." and "We are not introducing any new indications for use." These statements, backed by the comparative data, serve as the "proof" that the device meets the implicit acceptance criteria of substantial equivalence.

{0}------------------------------------------------

FINGE Clastle

Getinge/Castle, Inc. 7371 Spartan Blvd., East North Charleston, SC 29418

ಳು

ﺎﺕ ﺍﻟﺘﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ

Tei 803.552.8652 Fax 803.552.8652

(1)Submitters NameGetinge/Castle, Inc.
Submitters Address7371 Spartan Blvd., EastNorth Charleston, SC 29418
Submitters Telephone No.(803) 552-8652
Contact NameTrevor Williamson
Summary Preparation DateOctober 3, 1997
(2)Device NameGetinge/Castle OptiView® Surgical Light
Common or Usual NameSurgical Light
Classification NameSurgical Lamp
(3)Predicate Device(s)ALM Prismatic®, model 7001 and 9001 seriesAMSCO SQ240Berchtold Chromophare®, model C-570, C-570/570 and C-570/570/570
(4)Device DescriptionThe Getinge/Castle OptiView® surgical lights are designed toprovide the high quality light required in an operating roomsetting.There are four major considerations that the lights have beendesigned for:1. Shadow Elimination. 144 radial dioptric lenses, disposedequally on the lens glass, are used to broaden the lightbeam such that a 10 X 20 cm patch of light is produced onthe surgical site from each lens. Should one or several ofthe lenses become obstructed by the surgeon or fromequipment, a constant quantity of light is subtracted fromeach point of the light field. This virtually eliminatesshadows and provides for optimum visibility.2. Optimal Depth of Field. A central toric lens and conicmirror is used to produce very broad beams whoseintersection are not points but are volumes. In some casesthe volume can measure up to 70 cm in height and 20 cmacross the center. The design also allows the volume oflight to be moved in the axis of the beam, this is calledvariable focusing.true3. Heat Elimination and High Color Rendition. Through theuse of a special type of glass, the emitted light is correctedfor color (towards the model for broadband sunlight) andfiltered for infrared. This eliminates the thermal effect ofthe beam and allows a color rendition index of Ra = 95 orgreater.

{1}------------------------------------------------

*GETINGE

Lancer

Getinge/Castle. Inc. 7371 Spartan Blvd.. East North Charleston, SC 29418 Tel 803.552.8652 Fax 803.552.8652

  • Electromagnetic and Bacteriologic Neutrality. The product 4. Device Description (con't) (4) have been designed and tested to comply with the strictest electromagnetic compatibility standards. Also, the sealed optical unit, the flat, smooth surfaces and sensitive touch-controls orovide for an extremely aseptic environment. The mechanical aspects of the lights are that they are constructed of specifically designed steel load-bearing drop tubes, yokes and suspension assemblies that are coated with a strong epoxy resin paint. These structures are mechanically connected at pivot points that incorporate counter-balances for ease of movement as well as specially designed wire-runs and fully pivoting electrical connectors. The electrical aspects of the lights are designed* to comply with international safety standards such as: ഗ് Low Voltage Directive (LVD) 72/23/CEE, Decree No. 95-1081 Directive (EMC) 6. Electromagnetic Compatibility 89/336/CEE, Decree No. 95-587 Device Directive (MDD) 93/42/CEE. Power 7. Medical Supply Units, NF EN 60-4391 (IEC 433-1), NF EN 60-439-Operating Theatre Lighting, NF EN 60-601-1 (IEC 601) 8. The design will also be examined to UL 544 and certified by either Underwriters Laboratory or Intertek Testing Services (ITS) which should result in the UL or ETL mark. *NOTE: The lights were previously CE marked, while marketed in Europe, by the original manufacturer: Scialityque Industrie. The lights are electrically controlled via pressure sensitive touch controls that may be located on the lights or at remote locations. The controls allow the lights to be energized, the light level to be adjusted and the focus mechanisms to be operated. The light beam is achieved through the use of one or two (dependent upon model) tungsten-halogen 24 VDC bulbs that are contained in a sealed optical unit. The emitted light is focused through a toroidial lens and is reflected off of a conic. polished aluminum reflector and through 144 radial dioptric lenses which allows for a radial-focused broad-beam light pattern on the surgical surface. This eliminates shadows while concurrently extending the true depth of field of the beam.

{2}------------------------------------------------

*GETINGE asdc

Getinge/Castle, Inc. 7371 Spartan Blvd., East North Charleston, SC 29418

(5)Statement of Intended UseThe Getinge/Castle OptiView Surgical Lights are to be usedin the same, (general type and configuration), surgicaloperatory theatres, and out-patient surgery centers thatcurrently employ various models of the referenced predicatedevices.
(6)Summary of Similar TechnologySee attached product comparison tables. The specific modelsshown in the tables were chosen on the basis of the similarityof the lamp head designs, illumination and other performancedata.

The OptiView® 500 series lights are identified as follows:

ModelDescriptionPredicate
500Ceiling mount - one (1) 500 series lamp headALM PRC 7001, Berchtold C-570/570
525Ceiling mount - one (1) 200 series lamp head and one (1)500 series lamp headALM PRC 7001, Berchtold C-570/570
550Ceiling mount - two (2) 500 series lamp headsALM PRC 5501 DF, BerchtoldC-570/570
555Ceiling mount - three (3) 500 series lamp headsALM PRC 5551, Berchtold C-570/570
590Floor stand mount - one (1) 500 series lamp headALM PRC 7001, Berchtold C-570/570

The OptiView® 700 series lights are identified as follows:

ModelDescriptionPredicate
700Ceiling mount - one (1) 700 series lamp headALM PRC 9001, AMSCOSQ240, Berchtold C-570/570/570
750Ceiling mount - one (1) 500 series lamp head and one (1)700 series lamp headALM PRC 9501 DF, AMSCOSQ240, Berchtold C-570/570/570
755Ceiling mount - two (2) 500 series lamp heads and one(1) 700 series lamp headALM PRC 9551, AMSCOSQ240, Berchtold C-570/570/570

The basic assumptions that we are basing this pre-market notification on, and why we believe that the GetingelCastle OptiView® Surgical Light is substantially equivalent to the predicate devices, is that we believe that:

  • We are not introducing new technology. The OptiView® lights and the predicate devices use very similar . technology to achieve the same basic goal; to illuminate the patient area and concentrate light at the surgical incision with deep cavity penetration. The use of the toric lens, conic reflector and radial dioptric lens in the 500 and 700 series has been standard practice for Scialytique (the original manufacturer) for over ten (10) years in Europe.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "GETINGE" in bold, black letters next to a star-like symbol. To the right of "GETINGE" is another word that is difficult to read due to the image quality, but it appears to end in "...tic". The background is white behind "GETINGE" and the star, but it is black behind the other word. The overall impression is that this is a logo or brand name.

Getinge/Castle. Inc. 7371 Spartan Blvd., East North Charleston, SC 29418

Tel 803.552.8652 Fax 803.552.8652

  • We are not introducing new or radical manufacturing methods. The Getinge/Castle OptiView® surgical lights . will be manufactured using production methods readly available to the general medical device industry. These practices are already in use for existing domestically marketed products and will be used for the manufacture of the OptiView® 500 and 700 series lights.
  • We are not introducing any new indications for use. As stated earlier, the GetingelCastle OptiView® surgical . lights are to be used for the same purposes and in the same manner as the predicate devices. The lights will be used as the primary and secondary methods of illuminating areas of the human body such that deep cavities can be penetrated and shadows eliminated during surgical procedures. Our intended customer base includes various health care professionals working in medium to large operatory settings.

We consider the OptiView® lights to be a single product line and we intend to market them in that manner. We have included them together for this notification because we believe that, like the ALM Chromophare® product line, which incorporates several different lamp head configurations under the same FDA clearance number, the Optiview lights are contemporary and complimentary to each other. These lights will be used in different configurations as stand alone units but also will be combined together (i.e. the 500 series lamp paired with a 700 series lamp on the same suspension package) for total operatory coverage.

{4}------------------------------------------------

Getinge/Castle Getinge/Castle

Additional Information in Support of K974084

Part 2:

Predicate devices for the Optiview Series 200, with the requested specifications:

SpecificationOptiview 200 SeriesModelALM IlluminatorSpotlightBerchtold ChromophareC-300
Diameter, cm (in)
Lens230 (9.0)Not specified365 (10.4)
Lighthead316 (12.4)330 (13)300 (11.8)
Weight, kg (lb.)2.5 (5.5)Not specifiedNot specified
Mounting TypeWall, Ceiling, FloorStandWall, Ceiling, FloorStandWall, Ceiling, FloorStand
Supply Voltage, VAC120/230120110/220
Power, W5515055
Bulb, TypeTungsten HalogenQuartz HalogenHalogen
Volts24Not specified24
Life, hr500Not specifiedApprox. 1000
Number of Light Heads111
Number of Bulbs111
Color Temp. °K4300, NominalNot specified4500
Field Size, cm (in)
Diameter100-150 (3.9-5.9)Not specified170 (6.7)
Depth400 (15.7)Not specifiedNot specified
Focal Length, cm (in)350 (13.8)Not specifiedNot specified
Illumination Level, ft-candles (Lux)5,670 (61,000) at4300°K3717 (40,000)3253 (35,000)
Controls
DimmerVariable IntensityNot specifiedNot specified
Volts24Not specified24
FocusAdjustableNot specifiedAdjustable
Field SizeAdjustableNot specifiedAdjustable
Rotation, °360Not specified360
Vertical AdjustmentRange, cm (in)1792 (70.5)Not specified1190 (46.9)
Heat Filtering, %99Not specifiedNot specified
Filter MaterialGlassNot specifiedYes
Reflector MaterialPolished AluminumNot specifiedNot specified
SatellitesOptionalOptionalNot specified
Sterilizable HandleYesNot specifiedNot specified
Min. Ceiling Height,cm (in)2591 (102)Not specifiedNot specified
Spec.ModelOptiView®,500 seriesALM, PRC 7001seriesBerchtold, C-570series
Diameter, cm (in)44.3 (17.5)71.1 (28)50 (20)
Lens54.2 (21.3)71.1 (28)57 (22.5)
Lighthead12 (26.4) lamphead only73.9 (163) lamphead only63 (138) single,95 (209) dual
Weight, kg (lb.)Ceiling, wall,floor standCeilingCeiling, wall, floorstand
Mounting Type120/230120120/230/240
Supply Voltage,VAC140240150
Power, WTungstenHalogenQuartz HalogenXenon Halogen
Bulb, Type242424
Volts500 (min)10001000
Life, hr1, 2 or 311,2 or 3
Number of Lightheads1 or 221 main, 1 reserve
Number of Bulbs4,3003,5004,500
Color Temp, °K
Field Size, cm (in)10-15 (3.9-5.9)20.3 (8)18-28 (7-11)
Diameter40 (15.7)70 (27.6)80 (31)
Depth90-140 (35-55)Not specified70-140 (27.5-55)
Focal Length, cm(in)12,300 (130,000)9,300 (100,000)9,300 (100,000)13,530 (145,000)
Illumination Level,max ft-candles (lux)
ControlsVariable intensityContinuousContinuous
Dimmer24120120
VoltsAdjustableFixedAdjustable
FocusAdjustableFixedAdjustable
Field Size360360360
Rotation, °122.6 (48.3)83.8 (33)115 (45)
Vert. AdjustmentRange, cm (in)99Not specified99
Heat Filtering, %GlassGlassGlass/film
Filter MaterialAluminumN/A (refractedthrough prisms)Aluminum
Reflector MaterialOptionalOptionalOptional
SatellitesYesYesYes
Sterilizable Handle245 (100)264.2 (104)260 (102)
Min. Ceiling Height,cm (in)
ModelSpec.OptiView®,700 seriesALM, PRC 9001seriesAMSCO, SQ240
Diameter, cm (in)
Lens58.5 (23)91.4 (36)55.9 (22)
Lighthead75.6 (29.8)91.4 (36)60.9 (24)
Weight, kg (lb.)28 (61.6)21.8 (48) light unitonly67 (148) single,113 (250) dual
Mounting TypeCeilingCeilingSingle or dual arm,track, optional AV arm
Supply Voltage,VAC120/230120120
Power, W140120220
Bulb, TypeTungsten HalogenQuartz HalogenQuartz Halogen
Volts242422
Life, hr500 (min)10001000
Number of Lightheads1,2 or 311,2 or 3
Number of Bulbs1 or 231 main, 1 reserve perlighthead
Color Temp, °K4,3003,5004,200; 4,400
Field Size, cm (in)
Diameter17-24 (6.7-9.4)20.3 (8)16.5 (6.5); 17.8 - 21.6(7-8.5)
Depth50 (19.7)70 (27.6)66 (26)
Focal Length, cm(in)50 (19.7)Not specified106.7 (42)
Illumination Level,max ft-candles(lux @ 1 m)11,200 (120,000)11,160 (120,000)10,000 (107,640)12,000 (129,170
Controls
DimmerVariable intensityContinuousVariable intensity
Volts2412024
FocusAdjustableFixedFixed
Field SizeAdjustableFixedAdjustable
Rotation, °360360360
Vert. AdjustmentRange, cm (in)122.6 (50)134.6 (53)74.9 - 200.1 (29.5 -79)
Heat Filtering, %99Not specified98
Filter MaterialGlassGlassCold/hot mirror; coldmirror
Reflector MaterialAluminumN/A (refracted throughprisms)Plastic
SatellitesOptionalOptionalOptional
Sterilizable HandleYesYesYes
Min. Ceiling Height,cm (in)245 (100)281.9 (111)257.8 (101.5) single269.2 (106) dual

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for GETINGE, a company that provides equipment and systems for the healthcare and life sciences industries. To the left of the company name is a star-like symbol. To the right of the company name is the word "Castle" in a smaller font.

ﺎ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ

OptiView® 500 series vs. Predicate Devices

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for GETINGE and Listic. The GETINGE logo is a stylized asterisk followed by the word "GETINGE" in bold, sans-serif font. To the right of GETINGE is the word "Listic" in a smaller, less distinct font.

OptiView® 700 series vs. Predicate Devices

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized representation of three human figures, possibly representing health and well-being.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 3

Mr. Peter L. Koste, Jr. benior Manager, Quality Assurance and Regulatory Affairs Getinge/Castle, Incorporated 7371 Spartan Boulevard, East (29418) P.O. Box 40488 North Charleston, South Carolina 29423-0488

Re: K974084 Getinge/Castle Optiview Series Surgical Trade Name: Light Regulatory Class: II Product Code: FSY January 27, 1998 Dated: Received: February 13, 1998

Dear Mr. Koste:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{8}------------------------------------------------

Page 2 - Mr. Koste

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

fcella
coll
Celia M. Witten, Ph.D., M.D.
Director
Division of Coporal
and

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

003

510(k) Number (if known): K979084 Device Name: Officient Sungican 61514

Indications For Use:

THE PRODUCT UNIL BE USED TO ILLUMNATE SURGICAL PROCEDURES WITH COLOR CORRECTED ITE LIGHT AND HEAT FILTERING (FF.). DESIGNED TO ELIMINATE SHADOWS AND 14 DEEP CALITY COUNDS WITH ADGRATE VENEFRATE 61947 115 IS INTENDED TO BE USED 137 THE PRODUCT SURGEONS AND OTHER MEDICIN CARE PRACTICING OF THERE ARE NO IN A SURGICAL SETTING FOR USE. CONTRA-INDICATIONS

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K974084

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Formal 1-2-96)

D

§ 878.4580 Surgical lamp.

(a)
Identification. A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.(b)
Classification. Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.