K Number
K973971

Validate with FDA (Live)

Date Cleared
1999-09-13

(696 days)

Product Code
Regulation Number
868.5160
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use of the KION Anesthesia System is indicated for adult or infant populations in an environment where patient care is provided by Healthcare Professionals, trained in the administration of anesthesia, when the professional determines that a device is required to assist the breathing of the patient. This device can be used to administer anesthesia while controlling the entire ventilation for patients without any ability to breathe, as well as for supporting patients with reduced ability to breathe.

Device Description

KION is a modular anesthesia system. It is available as a set of hardware and software options, which can be added to take the system from being a low-cost, general purpose anesthesia machine, to a complete state-of-the-art anesthesia workstation. This gives the opportunity to specify a system and then upgrade and modify the system as needs change in the future. The KION anesthesia system is designed to enable the operator to work with the basic components of an anesthesia system in the most suitable way for each patient case. As such, the system is mounted on wheels to allow it to be easily moved into an optimum position, where many of the components can also be individually rotated. The KION system is intended for use on adult and pediatric patients. Physically, the KION consists of three main sections: The Trolley which contains the main pneumatic inlets, power supply and power inlets, battery, communication ports, and the gas delivery system. The Column consists of a revolving Vaporizer Holder, Servo Bellows Absorber and Patient Cassette. The Integrated User Interface consists of the Control Panel and the Monitor. The KION offers the following standard ventilation capabilities: Volume controlled ventilation, Manual ventilation.

AI/ML Overview

The provided text is a 510(k) summary for the KION Anesthesia System, which is a submission to the FDA requesting clearance to market a medical device. This document focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than an in-depth study proving novel acceptance criteria for an AI/ML device. Therefore, much of the requested information about device performance, study design, and ground truth establishment is not typically found in such a document for this type of device.

However, I can extract the relevant information based on the typical content of a 510(k) submission for a non-AI/ML medical device like an anesthesia system.

Here's an analysis of the provided text with respect to your questions:


1. A table of acceptance criteria and the reported device performance

For a device like an Anesthesia System, acceptance criteria are typically related to fundamental performance characteristics (e.g., accuracy of gas delivery, ventilation parameters, alarm functionality) and compliance with international standards. The document primarily establishes equivalence to predicate devices and adherence to relevant standards.

The table below summarizes the key performance parameters mentioned in the "Table of Device Similarities and differences to predicate device" (Section 6) and implicitly considers these as performance benchmarks. However, explicit "acceptance criteria" with specific thresholds are not detailed in this summary, as is common for 510(k) summaries of non-AI devices when claiming equivalence. The "reported device performance" is essentially that the KION system meets or improves upon the performance of the predicate devices.

ParameterAcceptance Criteria (Implied / Predicate Device Performance)Reported Device Performance (KION Anesthesia System)
Applications and Intended UseInfants to AdultsSame (Infants to Adults)
Fresh gas dosage RangePredicate: 500 ml/min - 40 l/min100 ml/min - 18lpm (Enhanced functionality for low flow applications)
Fresh gas dosage Accuracy+/- 10%Same (+/- 10%)
Oxygen Percent21% / 25% - 100% (Air / N₂O)21% / 28% - 100% (Air / N₂O) (Enhanced anti-hypoxic measures)
Oxygen/Nitrous Oxide RatioAnti-hypoxia device presentSame (Anti-hypoxia device present)
Tidal VolumePredicate: 2 - 1500 ml20 - 1500 ml (Altered functionality, no neonatal capability on Kion, but still within adult/pediatric range and specified)
CMV RatePredicate: 0.5 - 1206 - 99 (Enhanced functionality due to specific clinical anesthesia requirements)
I:E RatioPredicate: 1:4 - 4:1 in seven steps1:3 - 4:1 in five steps (Enhanced functionality due to specific clinical anesthesia requirements)
Pause TimePredicate: 0, 5, 10, 20, 30% of cycle10% of cycle in Volume Control Mode (except for I:E - 4:1 = 0%) (Enhanced functionality due to specific clinical anesthesia requirements)
Monitored ParametersInspired Tidal Volume, Expired Tidal Volume, Rate, Peak airway Pressure, Mean Airway Pressure, Pause Pressure, PEEP/CPAP, Airway pressure / Analogue Display (from predicate devices)Same (All listed monitored parameters are present in KION)
Ventilatory AlarmsHigh Airway Pressure, High Continuous Airway Pressure, Disconnect, Low Minute Volume, Low O₂ Concentration (from predicate device)Same, plus new alarms: Low Air Supply Wall and Tank, Low O₂ Supply Wall and Tank, Low N₂O Supply Wall and Tank, High Air Supply Wall and Tank, High O₂ Supply Wall and Tank, High N₂O Supply Wall and Tank, Low Battery Capacity (Enhanced functionality)
Standards ComplianceIEC 601-1, IEC 601-1-1, IEC 601-1-2, ISO/DIS 8835-1.2, prEN 794-1, ISO 7767KION complies to these standards. This is a critical acceptance criterion for medical devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states: "See the test results under section In-house Testion T, in this submission." This indicates that in-house testing was performed. However, the 510(k) summary does not provide details on the sample size used for this testing, nor the data provenance (e.g., country of origin, retrospective/prospective nature). For a device like an anesthesia system, "test set" would typically refer to the various simulations, bench tests, and potentially animal studies, as well as the manufacturing quality control tests conducted during development.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the 510(k) summary. For this type of device, ground truth (or verification/validation reference) would be established through engineering specifications, validated test equipment, and recognized scientific and medical principles for measurement accuracy and safety, rather than through expert consensus in the way it might be for AI/ML imaging devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the 510(k) summary. Adjudication methods are typically associated with human review processes, often in clinical studies or for establishing ground truth for AI/ML models. For an anesthesia system, verification and validation would follow engineering testing protocols.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is specific to diagnostic imaging devices, particularly those involving human interpretation (readers) and AI assistance. This device (KION Anesthesia System) is an anesthesia delivery and monitoring system and does not involve "readers" or "AI assistance" in the diagnostic interpretation sense. Therefore, this type of study was not conducted or applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This refers to the standalone performance of an algorithm, typically for AI/ML devices. The KION Anesthesia System is a physical medical device with hardware and integrated software for control and monitoring, not a standalone software algorithm in the context of AI/ML.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For an anesthesia system, the "ground truth" for performance would be engineering specifications, validated measurement standards and equipment, and established physiological parameters. For example, the accuracy of gas flow would be measured against calibrated flow meters, and alarm thresholds would be set based on clinical safety guidelines and physiological limits. The summary does not explicitly detail these "ground truth" methods but implies them through adherence to standards and in-house testing.

8. The sample size for the training set

This question is not applicable as the KION Anesthesia System is not an AI/ML device that requires a "training set" in the context of machine learning.

9. How the ground truth for the training set was established

This question is not applicable as the KION Anesthesia System is not an AI/ML device that requires a "training set" or ground truth for such a set.

{0}------------------------------------------------

K973971

510(k) SUMMARY

as required per 807.92(c)

    1. Submitters Name, Address:
      Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (508) 750-7500 Fax: (508) 777-3398

Official Correspondent: David Simard, Director Quality Assurance & Regulatory Affairs

Contact person for this submission: Wulf R. Trepte Quality Assurance Date submission was prepared: 16/10/97

Trade Name, Common Name and Classification Name: 2.

  • Trade Name: A
    KION Anesthesia System

Common Name, Classification Name, Class and Regulation Number: B.

Common NameClassificationn NumberClassRegulation Number
Gas-machine, Anesthesia73BSZII21 CFR 868.5160
Gas-machine, Analgesia73ELIII21 CFR 868.5160
Cardiac monitor74DRTII21 CFR 870.2300
Pulse rate monitor74BWSII21 CFR 870.2300
Pulse oximeter74DQAII21 CFR 870.2700
Breathing Frequency Monitor73BZQII21 CFR 868.2375
Clinical Electronic Thermometer80BWXII21 CFR 880.2910
Indwelling Blood Pressure Monitor74CAAII21 CFR 870.1110
Heart Rate Monitor, Neonatal74FLOII21 CFR 870.2300
Ventilatory Effort Monitor (Apnea Detector)73FLSII21 CFR 868.2375
Monitor Blood Pressure, Neonatal, Invasive74FLPII21 CFR 870.1110
Arrhythmia detector & Alarm74DSIIII21 CFR 870.1025
Medical Cathode-Ray Tube Display74DXJII21 CFR 870.2450
ST Segment Monitor with Alarm74MLDIII21 CFR 870.1025
Non-indwelling Blood Pressure Monitor74DXNII21 CFR 870.1130
End-tidal Carbon-Dioxide Monitor73CCKII21 CFR 868.1400

COMPANY CONFIDENTIAL

Sicmens-Elema AB Electromedical Systems Division

Röntgenvägen 2 S-171 95 Solna SWEDEN

Tel. (46) 8 730 7228 Fax. (46) 8 98 63 05

{1}------------------------------------------------

3. Predicate Device Identification:

Legally marketed devices to which equivalence is being claimed510(k) #
Siemens Servo Ventilator 900CK841529
Servo Ventilator 900C/D Electronic gas supply unitK871354
Servo Ventilator 900C/D CAV 1.0 Master Program Flow & Prs, CO2 Lung MonitorK861707
Servo Ventilator 900C/D P.C. InterfaceK852771
Servo Ventilator 900C/D CAV 4.0 Software Servo GraphicsK923444/A
Servo Ventilator 900C/D Man. Vent. Acc.K854036
Servo Ventilator 900C/D/E CAV 3.0 SoftwareK890932
Siemens Servo Anesthesia Circle 985K893786
Siemens Vaporizers 950/951/952K841157
Ohmeda Tec5 VaporizerK942091
Siemens Servo Ventilator 300K960010
Siemens SC9000/SC9015 Bedside Monitoring SystemK946306

4. Device Description:

KION is a modular anesthesia system. It is available as a set of hardware and software options, which can be added to take the system from being a low-cost, general purpose anesthesia machine, to a complete state-of-the-art anesthesia workstation. This gives the opportunity to specify a system and then upgrade and modify the system as needs change in the future. The KION anesthesia system is designed to enable the operator to work with the basic components of an anesthesia system in the most suitable way for each patient case. As such, the system is mounted on wheels to allow it to be easily moved into an optimum position, where many of the components can also be individually rotated. The KION system is intended for use on adult and pediatric patients

Physically, the KION consists of three main sections:

  • The Trolley which contains the main pneumatic inlets, power supply and power 1. inlets, battery, communication ports, and the gas delivery system.
  • The Column consists of a revolving Vaporizer Holder, Servo Bellows Absorber 2. and Patient Cassette.
  • The Integrated User Interface consists of the Control Panel and the Monitor. 3.

The KION offers the following standard ventilation capabilities:

  • . Volume controlled ventilation
  • . Manual ventilation

5. Intended Use:

The intended use for the KION is as a modular anesthesia system for use on Adult and Infant patient populations. The unit is designed to be used in the operating room or any other hospital location where administration of anesthesia is necessary.

2 COMPANY CONFIDENTIAL

Siemens-Elema AB Electromedical Systems Division

Röntgenvägen 2 S-171 95 Solna SWEDEN

Tel. (46) 8 730 7228 Fax. (46) 8 98 63 05

{2}------------------------------------------------

6. Table of Device Similarities and differences to predicate device

ManufacturerSiemens-Elema AB,Electromedical SystemsDivisionSiemens-Elema AB,Electromedical SystemsDivision-
510(k) numberPredicate devicesSiemens Servo ventilator 900C - K841529 Siemens Servo Ventilator 300 - K960010 Servo Anesthesia Circle 985 - K893786. Siemens Vaporisers 950, 951, 952 - K841157. Ohmeda Tec 5 vaporiser - K942091. Siemens Servo Screen - K960168Modified DeviceKlon Anesthesia SystemK number to be decided-
Parameters GeneralPredicate devices: AllModified DeviceKlon Anesthesia SystemExplanation of the differences comparedto the Predicate devices
Applications and Intended UseInfants to AdultsSame-
Gas Delivery SystemPneumatic Powered ServoControlled "Bag in Bottle"system.Same-
Battery Back upNoYesBattery backup for the 900C is only availablewith an external system.
Parameters Gas DosagePredicate devicesSiemens Servo ventilator 900C- K841529Servo Anesthesia Circle 985 -K893786.Modified DeviceKion Anesthesia SystemExplanation of the differences comparedto the Predicate devices
Fresh gas dosageRangeRange:500 ml/min - 40 l/minRange: 100 ml/min - 18lpmEnhanced functionality on Kion for low flowapplications
Accuracy+/- 10%Same-
Oxygen Percent21% / 25% - 100%(Air / N2O)21% / 28% - 100%(Air / N2O)Enhanced anti-hypoxic measures
Gas Mixture SettingO₂/Air or O₂/N₂OSame-
Anti-hypoxia device(Limits O₂/N₂O fresh gas ratio)YesSame-
Ventilator IntegralYesSame-
Anesthetic Vapour DeliveryModulesPredicate devicesSiemens Vaporizers950/951/952 and OhmedaTec5Modified DeviceKion Anesthesia SystemVaporizersExplanation of the differences comparedto the Predicate devices
Operating PrincipleHigh pressure precisioninjectorSame-

Magazine for three

vaporizers mounting

3

Single vaporizer mounting

COMPANY CONFIDENTIAL

Siemens-Elema AB Electromedical Systems Division

Vaporizer mounting

Röntgenvägen 2 S-171 95 Solna SWEDEN

Tel. (46) 8 730 7228 Fax. (46) 8 98 63 05

Enhanced functionality.

{3}------------------------------------------------

Breathing SystemsPredicate devicesSV 900 and ServoAnesthetic Circle 985Modified DeviceKion Anesthesia SystemExplanation of the differences compared to thePredicate devices
Circle SystemYesSame
Auxiliary Fresh Gas OutletNoYesEnhanced functionality to support external breathingsystems (e.g. Bain, T-piece)
Adjustable Pressure Regulating(APL) valveNoAdjustable to 90 cm H2OImproved functionality
Ventilation ModesPredicate devicesSV 900CModified DeviceKion Anesthesia SystemExplanation of the differences compared to thePredicate devices
ManualYesSame
Volume Controlled VentilationYesSame
Pressure Controlled VentilationYesNoPlanned for future enhancement in Kion
Pressure SupportYesNoPlanned for future enhancement in Kion
Ventilatory Control SettingsPredicate devicesSV 300Modified DeviceKion Anesthesia SystemExplanation of the differences compared to thePredicate devices
Preset Minute Volume0.5 - 45LPMSame
Tidal Volume2 - 1500 ml20 - 1500 mlAltered functionality, no neonatal capability on Kion.
CMV Rate0.5 - 1206 - 99Enhanced functionality due to specific clinicalanesthesia requirements
I:E Ratio1:4 - 4:1 in seven steps.1:3 - 4:1 in five steps.Enhanced functionality due to specific clinicalanesthesia requirements
Pause Time0, 5, 10, 20, 30% of cycle10% of cycle in VolumeControl Mode ( except for I:E- 4:1 = 0%)Enhanced functionality due to specific clinicalanesthesia requirements
Inspiratory Gas FlowMaximum 180 lpmSame
Monitored Ventilatory &Respiratory parametersPredicate DevicesSV 900 and Servo Screen390Modified DeviceKion Anesthesia SystemExplanation of the differences compared to thePredicate devices
Inspired Tidal VolumeYesSame
Expired Tidal VolumeYesSame
RateYesSame
Peak airway PressureYesSame
Mean Airway PressureYesSame
Pause PressureYesSame-
PEEP/CPAPYesSame-
Airway pressure / AnalogueDisplayYesSame-
Ventilatory AlarmsPredicate DeviceSV900Modified DeviceKion Anesthesia SystemExplanation of the differences compared to thePredicate devices
This device has the followingSameEnhanced functionality. In addition Kion also have thefollowing alarms
This device has the followingSameEnhanced functionality. In addition Kion also have the
alarms: High AirwayPressure, High Continuousfollowing alarms:
Airway Pressure, Disconnect,Low Air Supply Wall and Tank, Low 02 Supply Wall
Low Minute Volume, Low 02and Tank, Low 02 Supply Wall and Tank, Low N20
ConcentrationSupply Wall and Tank, High Air Supply Wall and
Tank, High 02 Supply Wall and Tank, High N20
Supply Wall and Tank, Low Battery Capacity

4

COMPANY CONFIDENTIAL

Siemens-Elcma AB Electromedical Systems Division

Röntgenvägen 2 S-171 95 Solna SWEDEN

Tcl. (46) 8 730 7228 Fax. (46) 8 98 63 05

{4}------------------------------------------------

    1. Assessment of non-clinical performance data for equivalence: See the testresults under section In-house Testion T, in this submission.
    1. Assessment of clinical performance data for equivalence: Clinical trials ongoing.
    1. Biocompatability: Not applicable
  1. Sterilization: Not applicable

11. Standards and Guidances:

KION complies to the following standards:

  • · IEC 601-1 (EN 60601-1) Medical Electrical Equipment, part 1: General requirements for safety.
  • · IEC 601-1-1 (EN 60601-1-1) Medical Electrical Equipment, part 1.1: Safety Requirements for medical electrical systems.
  • · IEC 601-1-2 (EN 60601-1-2) Medical Electric Equipment, part 1.2: Electromagnetic Compatibility- Requirements and Tests
  • · ISO/DIS 8835-1.2 (Forthcoming IEC 60601-2-13, 2nd Ed.) Anesthesia Workstations and Their Components.
  • · prEN 794-1 Lung ventilators-Part 1: Particular Requirements for Critical Care Ventilators.
  • · ISO 7767 Oxygen monitors for monitoring patient breathing mixtures- Safety Requirements.

COMPANY CONFIDENTIAL

Siemens-Elema AB Electromedical Systems Division

Röntgenvägen 2 S-171 95 Solna SWEDEN

Tel. (46) 8 730 7228 Fax. (46) 8 98 63 05

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 3 1999

Mr. David Simard Simens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923

Re: K973971 KION Anesthesia System III (three) Requlatory Class: Product Code: 73 BSZ and 74 DSI July 20, 1999 Dated: Received: July 21, 1999

Dear Mr. Simard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2 - Mr. David Simard

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

· Sincerely yours,

Thomas J. Callahon

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page_1_of_1__

K973971 510(k) Number (if known):__

Device Name: ________________________________________________________________________________________

Indications for Use:

Use of the KION Anesthesia System is indicated for adult or infant populations in an environment where patient care is provided by Healthcare Professionals, trained in the administration of anesthesia, when the professional determines that a device is required to assist the breathing of the patient. This device can be used to administer anesthesia while controlling the entire ventilation for patients without any ability to breathe, as well as for supporting patients with reduced ability to breathe.

MRI Compatibility Statement:

Siemens KION is not compatible for us in a MRI magnetic field.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------

signature

Prescription UseOROver-The-Counter Use
(Per 21 CFR 801.109)

(Optional Format 1-2-96)

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) NumberK973971

§ 868.5160 Gas machine for anesthesia or analgesia.

(a)
Gas machine for anesthesia —(1)Identification. A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.(2)
Classification. Class II (performance standards).(b)
Gas machine for analgesia —(1)Identification. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).(2)
Classification. Class II (performance standards).