K Number
K973613

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-04-14

(204 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Citizen CH-611B and CH-611C digital wrist sphygmomanometers (blood pressure meters) are intended to be used for the oscillometric measurement of systolic and diastolic blood pressure and pulse. They are intended to be sold over-the-counter and to health care professionals, and not primarily under the order or direction of a physician.

Device Description

The Citizen CH-611B and CH-611C digital wrist sphygmomanometers are small handheld noninvasive blood pressure measurement systems that measure systolic and diastolic blood pressure and pulse from the user's left wrist. The units are contained in a hard plastic housing that contains a user interface panel to which an adjustable wrist cuff is attached by rubber tubing. The user interface panel has a power switch, and a liquid crystal display ("LCD") for displaying the systolic and diastolic blood pressure and pulse, and a memory switch. The device measures blood pressure through the use of an automatically-inflating wrist cuff. The cuff automatically deflates during blood pressure measurement. The only difference between the CH-611B and the CH-611C is that the CH-611B has a memory function which is able to display the last measured systolic and diastolic blood pressure readings (but not the pulse) when a memory recall switch is pressed, while the CH-611C is able to display the last seven measured systolic and diastolic blood pressure readings (but not the pulse) as well as an average of the seven.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Citizen CH-611B and CH-611C Digital Wrist Sphygmomanometers:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Predicate Device Performance)Reported Device Performance (Citizen CH-611B/C)
Blood Pressure Measurement Accuracy: ± 3mm Hg or ± 2% of the reading (whichever is greater)Blood Pressure Measurement Accuracy: ± 3mm Hg or ± 2% of the reading (whichever is greater)
Pulse Measurement Range: 40 to 200 pulses/minutePulse Measurement Range: 40 to 200 pulses/minute
Pulse Measurement Accuracy: ± 5% of the measured pulse frequencyPulse Measurement Accuracy: ± 5% of the measured pulse frequency
Wrist Cuff Circumference: 135 mm to 195 mmWrist Cuff Circumference: 135 mm to 195 mm
Oscillometric measurement methodOscillometric measurement method
Microcomputer controlled, digital, auto-inflateMicrocomputer controlled, digital, auto-inflate

Note: The document explicitly states that the Citizen devices and the predicate (Omron HEM-601) have the same accuracy and range specifications for blood pressure and pulse measurements. The acceptance criteria are therefore directly derived from the predicate device's established performance. The study aimed to demonstrate that the Citizen device met these same specifications.

2. Sample Size Used for the Test Set and Data Provenance

The provided text only states: "Citizen submitted laboratory testing data to establish the accuracy of its Sphygmomanometers for comparison to the Omron predicate device."

  • Sample Size: The specific sample size used for this laboratory testing is not mentioned in the provided document.
  • Data Provenance: The data was generated through "laboratory testing," implying a controlled environment. The country of origin is implicitly Japan, as the submitter, Citizen Watch Company, Ltd., is located in Tokyo, Japan. The data is prospective in the sense that it was specifically generated to support this 510(k) submission for the new devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not provided in the document. For blood pressure monitors, the "ground truth" for accuracy is typically established by trained technicians or clinicians using a calibrated reference sphygmomanometer (e.g., a mercury sphygmomanometer) simultaneously with the device under test. The document does not detail the methodology for establishing this ground truth.

4. Adjudication Method for the Test Set

This information is not provided in the document. Adjudication methods are typically relevant for studies where multiple human readers interpret data, and their findings need to be reconciled. Since this is a device performance study for a blood pressure monitor, an adjudication method in the traditional sense (e.g., 2+1 for image interpretation) would not be applicable.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of results (e.g., imaging devices). The Citizen CH-611B/C are automated blood pressure monitors, not devices requiring human interpretation in this manner.

6. Standalone (i.e. algorithm only without human-in-the-loop performance) Performance

Yes, a standalone performance study was done. The entire premise of the submission is to demonstrate the accuracy of the algorithm and hardware of the CH-611B and CH-611C devices themselves, without operator intervention once the measurement process is initiated, against established accuracy standards. The "laboratory testing data" directly assesses the device's ability to measure blood pressure and pulse accurately.

7. Type of Ground Truth Used

The type of ground truth used would have been a reference measurement from a highly accurate and calibrated blood pressure measurement device, typically a mercury sphygmomanometer or an equivalent validated device operated by a trained professional. The text does not explicitly state "mercury sphygmomanometer" but "laboratory testing data to establish the accuracy" implies comparison to a gold standard.

8. Sample Size for the Training Set

The concept of a "training set" in the context of machine learning is not directly applicable here. While the oscillometric method involves algorithms, the document describes a traditional device validation approach rather than a machine learning model that would require a separate training phase with a distinct dataset. Therefore, no "training set sample size" is mentioned or implied in the document.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a training set in the machine learning sense is not applicable. The device's underlying oscillometric algorithm would have been developed and refined based on engineering principles and potentially internal testing, but this document focuses on the validation of the final device's performance against established standards.

{0}------------------------------------------------

K9736013

APR 1 4 1998

510(k) SUMMARY

Citizen Watch Company, Ltd. CH-611B and CH-611C Digital Wrist Sphygmomanometers (Blood Pressure Meters)

Submitter's Name, Address, Telephone Number, Contact Person and Date PREPARED

Citizen Watch Company, Ltd. Medical Device Section, NP Development Department 1-12. 6-Chome, Hon-cho Tanashi-shi, Tokyo, 188 Japan

Contact Person:Joseph D. Edmondson, Jr., Esq.Yoichuro Yamaguchi, Esq.
Phone:(202) 672-5300
Fax:(202) 672-5399

Date Prepared: September 19, 1997

NAME OF DEVICE

Trade Name:CH-611B and CH-611C Digital Wrist Sphygmomanometers (Blood Pressure Meters)
Common Name:Wrist Sphygmomanometer (blood pressure meter)
Classification Name:Noninvasive blood pressure measuring system, per 21 C.F.R. § 870.1130

PREDICATE DEVICES

(1) Omron HEM-601 Wrist Digital Blood Pressure Monitor

INTENDED USE

The Citizen CH-611B and CH-611C digital wrist sphygmomanometers (blood pressure meters) are intended to be used for the oscillometric measurement of systolic and diastolic blood They are intended to be sold over-the-counter and to health care pressure and pulse. professionals, and not primarily under the order or direction of a physician.

DEVICE DESCRIPTION

The Citizen CH-611B and CH-611C digital wrist sphygmomanometers are small handheld noninvasive blood pressure measurement systems that measure systolic and diastolic blood pressure and pulse from the user's left wrist. The units are contained in a hard plastic housing that contains a user interface panel to which an adjustable wrist cuff is attached by rubber tubing. The user interface panel has a power switch, and a liquid crystal display

{1}------------------------------------------------

("LCD") for displaying the systolic and diastolic blood pressure and pulse, and a memory switch.

The device measures blood pressure through the use of an automatically-inflating wrist cuff. The cuff automatically deflates during blood pressure measurement. The only difference between the CH-611B and the CH-611C is that the CH-611B has a memory function which is able to display the last measured systolic and diastolic blood pressure readings (but not the pulse) when a memory recall switch is pressed, while the CH-611C is able to display the last seven measured systolic and diastolic blood pressure readings (but not the pulse) as well as an average of the seven.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

Both the Citizen Sphygmomanometers and the Omron predicate device measure the diastolic and systolic blood pressure from the left wrist using oscillometric methods. Both systems are microcomputer controlled, digital, auto-inflate, wrist sphygmomanometers. The Omron predicate device does not have a memory function.

Like the Omron predicate, the Citizen sphygmomanometers display systolic and diastolic pressure ranges from 0 and 280 mm Hg. Both devices have a blood pressure measurement accuracy of the greater of ± 3mm Hg or ± 2% of the reading. The pulse measurement range of the Citizen devices and the Omron predicate are also the same, from 40 to 200 pulses/minute. The accuracy of the pulse measurements for both the Citizen sphygmomanometers and the Omron predicate device are + 5% of the measured pulse frequency. Citizen submitted , laboratory testing data to establish the accuracy of its Sphygmomanometers for comparison to the Omron predicate device.

Both the Citizen and the Omron devices utilize an auto-inflate wrist cuff system for blood pressure measurement. The Citizen and Omron devices are designed to measure blood pressure and pulse from the left wrist. The wrist cuff for both systems is designed to accommodate wrists with a circumference of between 135 mm and 195 mm. The wrist cuff on the Citizen device and the predicate device are both adjusted by velcro. Inflation of the wrist cuff for the Citizen devices and the Omron predicate are accomplished with an electric pump and pressure is released during deflation by an automatic air-release valve.

The Citizen devices, like the Omron predicate, have a "POWER" switch, a "START" switch, and an LCD display. The operating environment for the Citizen sphygmomanometers and the Omron predicate are also the same: 50°F and 104°F and 30% to 85% relative humidity. Although the Citizen devices use four "AAA" size batteries, the Omron predicate device uses four "AA" batteries. Any minor differences in the appearance, technology, or manufacture of the Citizen devices and the predicate device do not raise any new questions of safety or effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains a single letter, "S", in a simple, sans-serif font. The letter is black and stands out against a white background. The letter is in uppercase.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Public Health Service

APR 1 4 1998

Citizen Watch Company, Ltd c/o Mr. Joseph D. Edmondson, Jr. Foley & Lardner 3000 K Street, N.W. Suite 500 Washington, DC 20007

K973613 Re : CH-611B and CH-611C Digital Wrist Sphygmomanometer Regulatory Class: II (Two) Product Code: DXN Dated: January 16, 1998 Received: January 16, 1998

Dear Mr. Edmondson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Joseph D. Edmondson, Jr.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation - --------Center for Devices and -------Radiological Health

Enclosure

{4}------------------------------------------------

of of of Page

510(k) Number (if known):__________

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

STATEMENT OF INDICATIONS FOR USE

The Citizen CH-611B and CH-611C digital wrist sphygmomanometers (blood pressure meters) are intended to be used for oscillometric measurement of systolic and diastolic blood pressure and pulse and are intended to be sold over-the-counter and to health care professionals, and not primarily under the order or direction of a physician.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Thomas L. Callahon

Prescription Use_ (Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).