K Number
K973506

Validate with FDA (Live)

Device Name
ACTIVATOR II
Date Cleared
1998-01-27

(133 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Activator II® is indicated for chiropractic adjustment of the spine and extremities.

Device Description

The Activator II ® is a chiropractic adjusting instrument for use in spinal manipulative therapy. The instrument is approximately seven (7) inches in length and approximately 2.5 inches wide at the handles. The major components of the device include handles for holding the device and applying thrust, a thrust element for delivering an input force to the patient, a body contact member, a spring for propelling the thrust element, and an adjustment knob. The body contact member is manufactured from silicone rubber. The adjustment knob is used for manually adjusting the amount of potential energy imposed on the spring and for controlling the magnitude of the resulting input force. When the handle is depressed, the body contact member applies the force in a specific line of drive at a rapid speed. This action is similar to a practitioner manually applying thumb thrust at specific contact points to achieve chiropractic adjustment.

AI/ML Overview

This document describes the Activator II®, a chiropractic adjusting instrument, and the basis for its substantial equivalence to predicate devices, thus demonstrating it meets acceptance criteria for market clearance.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly derived from the comparison to the predicate device, the Activator I, and the demonstration of similar performance characteristics. The reported device performance is based on the Activator II®'s ability to cause vertebral movement at very low energy, as detailed in the study mentioned.

Acceptance Criteria (characteristics compared to predicate Activator I)Reported Device Performance (Activator II®)
Indicated for chiropractic adjustment of the spine?Yes
Hand held adjusting device?Yes
Impact force derived by spring energy?Yes
Adjustable impact force?Yes
Silicone rubber body contact member?Yes
Performance aspect: Ability to cause vertebral movement at very low energyDemonstrated (results show effectiveness)

2. Sample size used for the test set and the data provenance

The study referenced, "Accuracy of Piezoelectric Accelerometers Measuring Displacement of a Spinal Adjusting Instrument," involved testing the Activator II® on:

  • One human subject
  • One animal subject

The data provenance is from a published study in the Journal of Manipulative and Physiological Therapeutics, Volume 9, Number 1, March 1986. The country of origin of the data is not explicitly stated, but given the journal and the company's location (Phoenix, Arizona), it is likely to be prospective data collected in the USA.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not explicitly state the number of experts or their qualifications used to establish ground truth for this particular study. The study focuses on the accuracy of piezoelectric accelerometers measuring displacement, suggesting a technical assessment rather than a subjective expert assessment of chiropractic effect. The "ground truth" in this context would likely be the objective displacement measurements themselves.

4. Adjudication method for the test set

Not applicable. The study appears to be a technical measurement of instrument displacement, not a clinical assessment requiring adjudication of observer interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not a study involving human readers or AI assistance. It describes a medical device for chiropractic adjustments.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device. The study describes the performance of a physical chiropractic adjusting instrument.

7. The type of ground truth used

The ground truth used in the referenced study appears to be objective displacement measurements obtained through an "electronic system to measure the displacement of the Activator II® instrument." This is a form of measurement/instrumentation data.

8. The sample size for the training set

Not applicable. This device does not involve a "training set" in the context of machine learning or algorithms. It is a mechanical device.

9. How the ground truth for the training set was established

Not applicable. As above, there is no training set for this mechanical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image is a black and white logo. The logo is circular and contains a stylized mountain range at the top. Below the mountain range are several horizontal lines that could represent water or land. The overall design is simple and abstract.

DR. ARLAN W. FUHR, PRESIDENT AND CO-FOUNDER

JAN 27 1998

510(k) Summary- Activator II(8)

September 11, 1997

Manufacturer:

Activator Methods, Inc. 3714 E. Indian School Road Phoenix, Arizona 85060-0317

Contact:

Arlan W. Fuhr, D.C. phone: (602) 224-0220 fax: (602) 224-0230

chiropractic adjusting device

Plunger-like joint manipulator

Trade name:

Common name:

Classification name:

Device Description:

.

Activator® Substantially equivalent to: preamendments device Activator Methods Phoenix, Arizona Integrator K950646 HRI, Inc. Arthrostim K930431

Activator II®

The Activator II ® is a chiropractic adjusting instrument for use in spinal manipulative therapy. The instrument is approximately seven (7) inches in length and approximately 2.5 inches wide at the handles. The major components of the device include handles for holding the device and applying thrust, a thrust element for delivering an input force to the patient, a body contact member, a spring for propelling the thrust element, and an adjustment knob. The body contact member is manufactured from silicone rubber. The adjustment knob is used for manually adjusting the amount of potential energy imposed on the spring and for controlling the magnitude of the resulting input force. When the handle is depressed, the body contact member applies the force in a specific line of drive at a rapid speed. This action is similar to a practitioner manually applying thumb thrust at specific contact points to achieve chiropractic adjustment.

Freeman Procedure Seminars

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Activator Methods, Inc. The logo includes a circular graphic on the left and the company name in bold, black letters. Below the company name is the text "DR. ARLAN W. FUHR, PRESIDENT AND CO-FOUNDER".

Intended Use:

chiropractic adjustment of the spine and extremities.

Comparison to predicate device:

Indicated for chiropracticadjustment of the spine?yesyes
Hand held adjusting device?yesyes
Impact force derived byspring energy?yesyes
Adjustable impact force?yesyes
Silicone rubber body contactmember?yesyes
Weighted anvil tip?noyes

Table 1. Substantial Equivalence Comparison

Summary of data upon which substantial equivalence was based:

As a result of the research on the Activator 10 which included a 46 gram weighted anvil tip, the formation of the Activator II(8) evolved. This Activator II's was tested on a human subject and an animal subject. Performance data for the Activator II 6 has been published in the Journal of Manipulative and Physiological Therapeutics, Volume 9, Number 1, March 1986. In this study, entitled "Accuracy of Piezoelectric Accelerometers Measuring Displacement of a Spinal Adjusting Instrument," an electronic system to measure the displacement of the Activator II'® instrument was examined/ Results demonstrate the effectiveness of the instruments to cause vertebral movement for chiropractic adjustment at very low energy.

very low energy.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles a bird or a series of human profiles. The symbol is composed of three curved lines that create a sense of movement or flow.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 27 1998

Arlan W. Fuhr, D.C. · President Activator Methods, Inc. 3714 East Indian School Road Phoenix, Arizona 85018

Re: K973506 Activator II® Requlatory Class: Unclassified Product Code: LXM Dated: December 4, 1997 December 16, 1997 Received:

Dear Dr. Fuhr:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ನ್ನ substantially equivalent determination assumes compliance with -----the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements Please note: concerning your device in the Federal Reqister. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Arlan W. Fuhr, D.C.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number:________________________________________________________________________________________________________________________________________________________________

Device Name:__________________________________________________________________________________________________________________________________________________________________

and the comments of the

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

,

Indications for Use: The Activator II® is indicated for chiropractic adjustment of the spine

and extremities. and extremit

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CRDH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberL473506