K Number
K972873

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-03-17

(225 days)

Product Code
Regulation Number
866.5240
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ImmuSTRIP® HAMA IgG ELISA Test System is a direct enzyme-linked immunosorbent assay for the detection and semi-quantitation of human antibodies to mouse IgG1. Significant titers of human anti-mouse antibody (HAMA) have been associated with patients receiving injections of murine monoclonal antibody for diagnostic and/or therapeutic purposes2 .

Device Description

The ImmuSTRIP" HAMA IgG ELISA Test System is a direct enzyme linked immunosorbent assay for the semi-quantitative detection of human antibodies to mouse IgG (HAMA).

The HAMA assay is a two-stage test carried out in plastic microwell strips which have been coated with mouse IgG, whole molecule.

In the first stage, mouse IgG conjugated to horseradish peroxidase (conjugate) is added to the microwell. Diluted test sample is then added and incubated for a specified length of time. If antibody to the mouse IgG is present in the test sample, bridging will occur with the solid phase mouse IgG, the test sample antibody, and the conjugate. If antibody is not present in the test sample, the unbound conjugate will be removed in the subsequent washing step.

In the second stage, enzyme substrate is added to the microwell. If bound conjugate is present, the substrate will be reduced; the reduced end product of the catalytic reaction oxidizes the colorless chromogen resulting in a colored end product. Acid is added to stop the reaction and fix the color.

The color intensity is proportional to the amount of bound conjugate and, therefore, to the amount of precipitable antibody present in the sample. The color intensity is measured with a microwell strip reader.

The ImmuSTRIP" HAMA ELISA Test System may be performed manually or with existing microtiter equipment. Results, which are determined by optical density at 488-492 nm, are available in less than one hour. The test has been standardized against primate anti-mouse IgG serum and has a sensitivity of approximately 40 ng/ml. Final values are reported as nanograms of precipitable antibody equivalents per ml.

The assay system is a rapid, efficient, and semi-quantitative method. Specific reagent formulation has eliminated the problem of background interference currently existing in this type of assay.

AI/ML Overview

Acceptance Criteria and Device Performance Study for ImmuSTRIP® HAMA IgG ELISA Test System

1. Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (from text, inferred)Reported Device Performance
Reproducibility
Within-run CV≤ 10%Low Control: 4.3%
Medium Control: 3.4%
High Control: 1.9%
Between-run CV< 5%Low Control: 4.3%
Medium Control: 3.4%
High Control: 1.9%
Lot-to-lot ReproducibilityMaximum difference in values across lots for controls and normal samples to be within an acceptable range for a semi-quantitative assay (no explicit numeric acceptance criteria provided in the text, but the data is presented to demonstrate consistency)For 15 normal serum samples: max difference of 6.6 ng/ml across 3 lots. For 6 serum controls (50-400 ng/ml HAMA): max differences ranged from 1.4 ng/ml to 44 ng/ml across 3 lots.
SensitivityLabeled sensitivity of 37 ng precipitable antibody equivalents/ml37 ng precipitable antibody equivalents/ml
Specimen Dilution LinearityAverage %CV for expected vs. actual results to be within an acceptable range for a semi-quantitative assay (no explicit numeric acceptance criteria provided in the text, but the data is presented to demonstrate linearity)Average CV of 3.08% (range 0.4% to 6.8%) for 4 positive HAMA specimens diluted.
RecoveryRecovery percentages to be within an acceptable range for a semi-quantitative assay (no explicit numeric acceptance criteria provided in the text, but the data is presented to demonstrate accurate recovery)Range of 93.3% to 106.2% across concentrations from 50 ng/ml to 400 ng/ml.
Specificity (Normal Population)High percentage of healthy individuals to test negative for HAMA (< 74 ng/ml)98% of males, 100% of non-pregnant females, and 96.1% of pregnant females were negative (< 74 ng/ml).
Specificity (Rheumatoid Factor Samples)The device should demonstrate adequate specificity, acknowledging potential for interference from RF but providing data on its extent (no explicit numeric acceptance criteria provided for RF interference, but the data informs the understanding of specificity).21 (36.8%) of 57 RF samples resulted in falsely elevated HAMA test results (> 74 ng/ml). No apparent correlation between RF level and interference severity.
Shelf Life StabilityValid assay performance after specified storage period based on criteria: R² ≥ 0.950, Slope: 0.005 to 0.015, Y-intercept: ≤ 0.200Supported 12-month shelf life at 2° to 8°C for 12 lots, with most lots remaining "Valid" throughout the tested period.
Clinical Utility (HAMA detection in patients receiving MAb)Detection rate of HAMA in patients receiving murine-derived monoclonal antibodies consistent with existing literature (30-50% documented in literature) for any assay method.80.8% of 26 patients (CMMI cohort) positive post-injection. 41.2% of 51 patients (VA Medical Center cohort) positive post-injection.
Clinical Utility (Comparison to Reference Method)Agreement with values obtained by a double-antigen, radiometric assay method.Post-injection samples (n=42): - ImmuSTRIP HAMA IgG: 71.4% positive (30/42) - Reference Method: 64.3% positive (27/42) - With Reference Method as GT: Sensitivity = 85.2%, Specificity = 66.7%, Accuracy = 78.6%, PPV = 82.1%
Clinical Utility (Impact on Pharmacokinetics/Dosimetry)Demonstrate relationship between HAMA levels and impact on pharmacokinetics and dosimetry.Study demonstrated that at HAMA titers below 300 ng/ml, no effect on antibody clearance rates was apparent. Above 300 ng/ml, a rapidly increasing plasma and whole-body clearance rate was observed, reflected by decreasing red marrow and whole-body doses.

2. Sample Size Used for the Test Set and Data Provenance

  • Reproducibility (Within-run and Between-run): Not explicitly stated how many individual replicates were performed, but CVs were calculated for replicates of three for low, medium, and high control samples.
  • Lot-to-lot Reproducibility: 6 prepared serum controls and 15 normal serum samples.
  • Specimen Dilution: 4 positive HAMA specimens.
  • Recovery: A normal human serum pool spiked at 6 different HAMA concentrations (50 ng/ml to 400 ng/ml), assayed in quadruplicate for each concentration.
  • Specificity (Normal Population): 464 healthy individuals (250 males, 110 non-pregnant females, 104 pregnant females). Data Provenance: Dianon Systems, Inc. (Stratford, CT), retrospective.
  • Specificity (Rheumatoid Factor Samples): 57 serum samples with varying concentrations of rheumatoid factors. Data Provenance: Samples sequestered from Universal Reagents, Inc., and Scantibodies Laboratories, Inc., assayed by Dianon Systems, Inc.
  • Clinical Utility (HAMA detection in patients receiving MAb):
    • Retrospective Study: 26 patients from Center for Molecular Medicine and Immunology (CMMI), Newark, NJ, and 51 patients from VA Medical Center, Buffalo, NY. A total of 305 pre- and post-injection samples from 77 patients.
    • Prospective Study: 240 patients (out of 382 entered) from CEA-Scan Phase III study who had baseline and at least one follow-up measurement.
    • Data Provenance: Clinical studies were conducted at two medical institutions (CMMI and State University of New York at Buffalo, Dept. of Nuclear Medicine) and one reference laboratory (Dianon Systems, Inc.). The original submission states retrospectively for determination of HAMA incidence in normal population and in patients, and prospectively for patients receiving CEA-Scan.
  • Clinical Utility (Comparison to Reference Method): 61 samples (19 pre-injection, 42 post-injection) from 28 patients. Data Provenance: Samples provided by Dr. Robert Sharkey of CMMI, assayed by Dr. M. Khazaeli of HAMA-KINE, Inc.
  • Clinical Utility (Impact on Pharmacokinetics/Dosimetry): 57 patients with CEA-expressing tumors. Data Provenance: From a published study in The Journal of Nuclear Medicine, June 1997.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not explicitly state that experts were used to establish the ground truth for the test set. Instead, the primary "ground truth" or reference for performance characteristics relies on:

  • Known concentrations: For reproducibility, specimen dilution, and recovery, purified HAMA was added to serum to create known concentrations.
  • Reference Method: For comparative studies, a "double-antigen, radiometric assay method," referred to as "The Reference Method," was used as a comparator (implicitly considered a reference for comparison, not necessarily a 'ground truth' established by experts in the typical sense of diagnostic imaging reader studies). The assay was performed by Dr. M. Khazaeli, of HAMA-KINE, Inc. (no specific qualifications other than being responsible for that lab's testing).
  • Clinical Outcomes/Diagnosis: For clinical utility sections, the patient's status (e.g., received murine monoclonal antibody, diagnosed with rheumatoid arthritis) serves as context for HAMA detection, rather than an expert ground truth for individual HAMA presence.
  • Pathology/Outcomes Data: For the study on pharmacokinetic/dosimetry impact, clinical outcomes data (e.g., antibody clearance rates, red marrow and whole-body doses) linked to HAMA levels were used.

4. Adjudication Method for the Test Set

No explicit adjudication method is mentioned for establishing ground truth or resolving discrepancies, as the studies primarily rely on quantitative measurements against known standards or a reference assay. For the comparative study with the reference method, discrepancies between the ImmuSTRIP® ELISA and the RIA method are reported as false positives/negatives, suggesting a direct comparison rather than an adjudicated outcome being derived.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The device is an ELISA assay, not an imaging or diagnostic algorithm typically evaluated with MRMC studies involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies presented are all standalone validations of the ImmuSTRIP® HAMA IgG ELISA test system. The device is an in-vitro diagnostic test, and its performance is evaluated directly through laboratory measurements, not as an AI algorithm that assists human interpretation.

7. The Type of Ground Truth Used

  • Known concentrations: For reproducibility, specimen dilution, and recovery, the ground truth was established by spiking normal human serum with purified HAMA to create samples with known concentrations of HAMA.
  • Reference Assay: For the comparative study, the "double-antigen, radiometric assay method" (referred to as "The Reference Method") served as the comparative baseline for evaluating agreement, implicitly acting as a form of ground truth for comparing detection rates.
  • Clinical Status/Literature: For specificity in a normal population, the absence of known exposure to murine antibodies was considered "negative." For clinical utility in exposed patients, the patient's history of receiving murine monoclonal antibodies was the context, with elevated HAMA levels being the documented finding. Discussion also includes agreement with "distribution of HAMA values from similar populations cited in numerous publications."
  • Pathology/Outcomes Data: For pharmacokinetics and dosimetry, the clinical parameters and physiological responses (clearance rates, doses) were the ground truth for assessing the impact of HAMA.

8. The Sample Size for the Training Set

The document does not describe a "training set" in the context of machine learning or AI. This is a traditional IVD device. However, the reference standard and controls used for calibrating and evaluating the assay were developed from baboons immunized with murine anti-CEA monoclonal antibody (IMMU-4 IgG). The process involved:

  • Immunization of baboons to produce anti-mouse antibodies.
  • Collection of serum samples during immunization.
  • Testing by RID until significant levels of anti-mouse antibody were achieved.
  • Pooling collected sera and freezing at -80°C.
  • Determining HAMA activity by RID.
  • Preparing a dilution of the pooled sera to 220 ng of precipitable antibody equivalents/ml to serve as the reference standard.

9. How the Ground Truth for the Training Set Was Established

As noted above, there isn't a "training set" in the AI sense. However, the reference standard for the ImmuSTRIP® HAMA assay (which guides the assay's performance and quantification) was established by:

  • Immunizing baboons with a known murine monoclonal antibody (IMMU-4 IgG) to induce the production of anti-mouse antibodies (HAMA).
  • Monitoring the development of anti-mouse antibodies in the baboon sera using Radial Immunodiffusion (RID) until significant levels were reached.
  • Quantifying the HAMA activity in the pooled baboon sera via RID to precisely determine the concentration of precipitable antibody equivalents. This allowed for the standardization of the reference material at 220 ng precipitable antibody equivalents/ml. Thus, the ground truth for the reference standard's concentration was established through a validated immunological method (RID) on a well-characterized biological source.

{0}------------------------------------------------

K972873

IV SUMMARY OF SAFETY AND EFFECTIVENESS

General Information

MAR 17 1998

Enzyme Linked Immunosorbent Assay for the Detection and Generic Name: Semi-Quantitation of Human Antibodies to Mouse IgG (HAMA).

ImmuSTRIP HAMA IgG Trade Name:

Applicant's Name and Address:

Immunomedics, Inc. 300 American Road Morris Plains, NJ 07950

ImmuSTRIP" is a registered trademark of Immunomedics, Inc. All references herein to ImmuSTRIP" are to the registered trademark.

INDICATIONS FOR USE A.

The ImmuSTRIP HAMA IgG ELISA Test System is a direct enzymelinked immunosorbent assay for the detection and semi-quantitation of human antibodies to mouse IgG1. Significant titers of human anti-mouse antibody (HAMA) have been associated with patients receiving injections of murine monoclonal antibody for diagnostic and/or therapeutic purposes2 .

Background

The expanding use of murine-derived monoclonal antibodies for in-vivo diagnostic and therapeutic procedures has resulted in an increased incidence of HAMA titers in patients receiving such antibodies. Circulating levels of HAMA may bind with the injected antibody, forming complexes that may adversely affect the biodistribution and pharmacokinetics of the injected antibody 1. In order to avoid injecting murine monoclonal antibodies into such patients who would receive no benefit, there is a need for an assay to predict potential complexations 10. Additionally, HAMA has been shown to significantly interfere with many commercial assays utilizing murinederived monoclonal antibodies, resulting in both false-positive and falsenegative results 11-18.

{1}------------------------------------------------

B. DEVICE DESCRIPTION

The ImmuSTRIP" HAMA IgG ELISA Test System is a direct enzyme linked immunosorbent assay for the semi-quantitative detection of human antibodies to mouse IgG (HAMA).

The HAMA assay is a two-stage test carried out in plastic microwell strips which have been coated with mouse IgG, whole molecule.

In the first stage, mouse IgG conjugated to horseradish peroxidase (conjugate) is added to the microwell. Diluted test sample is then added and incubated for a specified length of time. If antibody to the mouse IgG is present in the test sample, bridging will occur with the solid phase mouse IgG, the test sample antibody, and the conjugate. If antibody is not present in the test sample, the unbound conjugate will be removed in the subsequent washing step.

In the second stage, enzyme substrate is added to the microwell. If bound conjugate is present, the substrate will be reduced; the reduced end product of the catalytic reaction oxidizes the colorless chromogen resulting in a colored end product. Acid is added to stop the reaction and fix the color.

The color intensity is proportional to the amount of bound conjugate and, therefore, to the amount of precipitable antibody present in the sample. The color intensity is measured with a microwell strip reader.

The ImmuSTRIP" HAMA ELISA Test System may be performed manually or with existing microtiter equipment. Results, which are determined by optical density at 488-492 nm, are available in less than one hour. The test has been standardized against primate anti-mouse IgG serum and has a sensitivity of approximately 40 ng/ml. Final values are reported as nanograms of precipitable antibody equivalents per ml.

The assay system is a rapid, efficient, and semi-quantitative method. Specific reagent formulation has eliminated the problem of background interference currently existing in this type of assay.

{2}------------------------------------------------

C.

ALTERNATIVE PRACTICES AND PROCEDURES

Current Methods to Measure HAMA

Currently, the research methods that are available to measure HAMA are (1) Radioimmunoassay (RIA), (2) radial immunodiffusion (RID), (3) Ouchterlony, (4) rocket electrophoresis, (5) precipitation analysis and (6) passive hemagglutination. The aforementioned methodologies are currently being performed on samples from patients who have been injected with murine monoclonal antibodies 17.3. These methods may be insensitive, labor intensive, and/or require as many as 2-3 days for the results to be available.

D. MARKETING HISTORY

From 1988 to 1994. Immunomedics, Inc. has manufactured the ImmuSTRIP" HAMA IgG ELISA Test System and it has been available for research use only. Since 1994 Scimedx Corp., 400 Ford Road, Denville, New Jersey 07839, has manufactured the ImmuSTRIP" HAMA IgG ELISA for Immunomedics, Inc.

POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH E.

When this device is used according to the instructions provided, accurate assay results should be obtained. An error in the assay, however, which would produce a falsely low result could adversely affect the biodistribution and pharmacokinetics of the subsequent antibody injection. In this case, the quality of the imaging or therapy intended could be adversely affected. A falsely elevated result could lead to a medical decision resulting in the delay of a diagnostic or therapeutic procedure utilizing a murine-based monoclonal antibody.

{3}------------------------------------------------

SUMMARY OF STUDIES F.

1. Pertinent Publications

  • Detection of Human Anti-Murine Kaladas PM, et al. a. Antibody (HAMA) Following Infusion of OncoScint" CR103. Comparison of ImmuSTRIP" ELISA with a Double Antigen Radiometric Assay. Antib Immunoconj Radiopharm 1993; 4:309-317.
  • Massuger L. et al. Specific and Nonspecific Immunoassays b. to Detect HAMA After Administration of Indium-111-Labeled OV-TL 3 F(ab'), Monoclonal Antibody to Patients with Ovarian Cancer. J Nucl Med 1992; 33:1958-1963.
  • Seybold K, Trinkler M, Frey L, et al. Antigenicity of C. Different Antigranulocytes Antibodies Assessed by HAMA Follow-up in Patients Undergoing Immunoscintigraphic Detection of Infections. Presented at the German Nuclear Medicine Congress, Cologne, Germany, April 15-17, 1993.
  • Behr TM, et al. Phase I/II Clinical Radioimmunotherapy ત . with an Iodine-131-Labeled Anti-Carcinoembryonic Antigen Murine Monoclonal Antibody IgG. J Nucl Med 1997; 38: 858-870.

{4}------------------------------------------------

  • Summary of non-clinical studies 2.
    Preclinical laboratory studies were conducted to determine the purity and specificity of the reagents.

  • Production of the baboon anti-mouse antibody for the a. ImmuSTRIP" HAMA IgG Reference Standard.

Immunogen Preparation

Description of Antibody

The Immunomedics anti-CEA monoclonal antibody used as the immunogen was NP-4, later referred to as IMMU-4. IMMU-4 is a Class-III anti-CEA antibody of the immunoglobulin IgG, subclass. It is specific for CEA, not reacting with antigens that share CEA-related epitopes, such as meconium antigen and normal cross-reactive antigens. To prepare the IgG, ascites was produced in viral antibodyfree mice with the IMMU-4 hybridoma cell line. The ascites was aseptically removed, centrifuged to remove cells, and the supernatant solution was frozen and stored at -80° C. After thawing, the ascites was further clarified by passing through an ion-exchange column, using pH and ionic conditions that prevented binding of the IMMU-4 to the ion-exchange matrix. IgG was isolated from the clarified supernatant solution by Protein A affinity chromatography, and further purified by ion-exchange chromatography. Purity and identity of the purified IMMU-4 IgG were proven by immunoelectrophoresis and SDS gel electrophoresis. (Refer to Section VI, Volume III).

Lampire Biologicals, located at 217 Farmschool Road, Ottsville, PA was contracted by Immunomedics to perform the immunization of the baboons, and to maintain the animals.

{5}------------------------------------------------

Immunization Schedule

Immunogen: 0.5 ml immunogen (IMMU-4 IgG, 1 mg/ml) was emulsified in 0.5 incomplete Freund's adjuvant and divided into aliquots for four subcutaneous injection sites per animal.

ScheduleBoostBleed
Week of
10/30/861.0 ml50 ml
11/06/861.0 ml0 ml
11/13/861.0 ml0 ml
11/20/861.0 ml0 ml
11/27/861.0 ml50 ml
12/25/861.0 ml50 ml
12/11/861.0 ml50 ml
01/08/871.0 ml50 ml
01/22/871.0 ml50 ml
02/05/871.0 ml50 ml
11/09/871.0 ml20 ml
11/28/871.0 ml100 ml
12/07/871.0 ml100 ml
12/21/871.0 ml100 ml

Preparation of the ImmuSTRIP* HAMA reference standard b.

Serum samples from the immunized baboons were collected during the immunization procedure. These samples were tested by RID until significant levels of anti-mouse antibody were achieved. Large bleeds of the animals were then performed and serum samples were pooled and frozen at -80°C. Aliquots of the pooled sera were assayed by RID and the HAMA activity was determined. A dilution of the pooled sera, to a concentration of 220 ng of precipitable antibody equivalents/ml, was prepared in phosphate buffered saline, sterile filtered and maintained as a reference standard.

{6}------------------------------------------------

  • Performance Characteristics c.

Reproducibility 1 .

Within run and between run assay reproducibilities were evaluated by performing the ImmuSTRIP® HAMA IgG assay utilizing low, medium and high control samples. The control samples were prepared by enriching normal human serum with purified The assigned control values were as HAMA. follows:

low control = 50 ng/ml medium control = 75 ng/ml high control = 200 ng/ml

Within run coefficients of variations (percent CVs) were calculated for replicates of three.

Between run CVs were calculated for replicates of three, from assays performed on different days. The CVs for within run values were ≤10%, and between run CVs were < 5%.

Results for within run and between run evaluations are summarized below:

Run #Low Control(50 ng/ml)Medium Control(75 ng/ml)High Control(200 ng/ml)
ng/mlSDCV (%)ng/mlSDCV (%)ng/mlSDCV (%)
153.4$\pm$ 5.29.776.3$\pm$ 2.12.8186.6$\pm$ 5.93.2
256.0$\pm$ 2.54.572.5$\pm$ 0.50.7192.8$\pm$ 17.79.2
351.4$\pm$ 1.83.571.5$\pm$ 3.24.5193.3$\pm$ 19.410.0
Mean53.6$\pm$ 2.34.373.4$\pm$ 2.53.4190.9$\pm$ 3.71.9

{7}------------------------------------------------

Lot to lot reproducibility was evaluated by assaying six prepared serum controls and 15 normal serum samples using three lots of ImmuSTRIP® HAMA IgG reagents. The HAMA results were obtained from three standard curves prepared from the reference standard provided with each lot; all results are expressed as a mean of triplicate values. For the 15 normal serum samples, negative for HAMA (≤ 74 ng/ml), a maximum difference of 6.6 ng/ml was obtained across the three lots.

Six serum controls were prepared by enriching negative serum with varying concentrations of purified HAMA. The range in concentration was 50 ng/ml to 400 ng/ml, and maximum differences across the three lots were as follows:

HAMA ConcentrationMaximum Difference
50 ng/ml9.2 ng/ml
75 ng/ml17 ng/ml
100 ng/ml1.4 ng/ml
200 ng/ml33.8 ng/ml
300 ng/ml19.4 ng/ml
400 ng/ml44 ng/ml

{8}------------------------------------------------

2. Sensitivity

The sensitivity of the ImmuSTRIP® HAMA IgG is 37 of precipitable antibody equivalents/ml. ng Recommended dilutions of the 220 ng precipitable antibody equivalent/ml reference standard results in a four point curve, prepared as follows:

StandardConcentration ngAntibody/mLVolume ofHAMARef STD (µL)Volume ofSpecimenDiluent (µL)TestDilution
(S1)220--Undiluted
(S2)1102002001:2
(S3)551003001:4
(S4)371005001:6

The lowest concentration prepared from the reference standard is equal to 37 ng precipitable antibody equivalents/ml (S4), and is the labeled sensitivity of the assay. The recommended dilution of test sample is 1:2, therefore, a "negative" HAMA result has been defined as ≤ 74 ng, precipitable antibody per mL, (37 X 2 [dilution factor]).

{9}------------------------------------------------

Specimen Dilution 3.

Multiple dilution of four positive HAMA specimens were prepared in specimen diluent and assayed with the ImmuSTRIP* HAMA IgG test system.

The following table illustrates the results of this study:

Sample #DilutionExpected(ng/ml)Actual(ng/ml)%CV
1--1:21:4*N/A109.054.5217.9102.950.92.93.4
2--1:21:4*N/A188.294.1376.3186.498.40.42.2
3--1:21:4*N/A167.383.7334.6153.073.04.56.8
4--1:21:4*N/A89.744.9179.386.147.02.02.4

"Not Applicable.

The ImmuSTRIP" HAMA IgG assays requires an initial sample dilution of 1:2. The first HAMA value recorded in the "actual" column is the HAMA result of this initial sample dilution. Further dilutions of the initial sample dilution occurred; (1:2 and 1:4); and expected vs. actual results were The percent CV of the mean was compared. calculated for expected vs. actual results. The average CV for the four specimens was 3.08 with a range of 0.4% to 6.8%.

$\uparrow$
$\downarrow$

{10}------------------------------------------------

  • Recovery 4.
    A normal human serum pool was spiked with known levels of purified HAMA ranging from 50 ng/ml to 400 ng/ml. Each spiked sample was assayed in quadruplicate with the ImmuSTRIP" HAMA IgG ELISA assay. For each known concentration of HAMA, the following values resulted, calculated from the mean of the quadruplicate values at each concentration.

ImmuSTRIP® HAMA IgG I of 1628

Recovery (%)

Lot 1628
HAMA Concentrationng/mlng/mlRecovery %CV %
50 ng/ml53.1106.26.7
75 ng/ml76.0101.32.1
100 ng/ml98.698.64.2
200 ng/ml186.693.32.4
300 ng/ml290.896.92.6
400 ng/ml373.993.50.6

{11}------------------------------------------------

5. Specificity

Discussion and Tabulation of Results from a Normal. Apparently Healthy Population

Dianon Systems, Inc. Stratford, CT, determined the presence of HAMA in serum samples from an apparently healthy population consisting of 250 males and 214 females, 104 of which were pregnant, with ImmuSTRIP" HAMA IgG.

In this patient population, 98% of the males, 100% of the non-pregnant females and 96.1% of the pregnant females were negative (< 74 ng/ml) with the ImmuSTRIP HAMA IgG assay. Naturally occurring HAMA, in varying concentrations and frequency of occurrence, has been reported in normal, apparently healthy populations 8. 20-23. The degree of frequency has been suggested to be dependent on the sensitivity of the assay method used to determine the presence of HAMA35.

Rheumatoid Factor Samples

An evaluation of the specificity of ImmuSTRIP* HAMA IgG was performed by assaying 57 serum samples having various concentrations of rheumatoid factors (RF). RF are autoantibodies directed to the Fc fragment of human and other mammalian IgG molecules 9. In this study, serum RF concentrations, expressed in International Units (IU), varied in the range from negative (normal) to 3000 IU (high). All positive samples were from patients diagnosed with rheumatoid arthritis. Dianon Systems, Inc. sequestered these samples from Universal Reagents, Inc., and Scantibodies Laboratories, Inc. and assayed these RF samples following the ImmuSTRIP" HAMA IgG manufacturer's directions. All samples were run in duplicate, the mean of the duplicate values was reported. The patient population from which these samples had been collected either had no previous exposure to mouse protein or such exposure was unknown. Twenty-one (36.8%) of the RF samples resulted in falsely elevated HAMA test results (> 74 ng/ml). There was no apparent correlation between the level of RF in the sample and the severity of interference with the ImmuSTRIP" HAMA IgG assay.

{12}------------------------------------------------

Real time stability data were determined on 12 lots of ImmuSTRIP® HAMA IgG kits stored at 2° to 8°C. These data are reported as valid or invalid based on the criteria for a valid assay." These stability data support a shelf life of 12 months.

ImmuSTRIP®HAMA IgG Lot #Day 03 Months6 Months9 Months12 Months
1604ValidValidValidValidValid
1606ValidValidValidValidValid
1607ValidValidValidValidValid
1608ValidValidValidValidValid
1611ValidValidNDValidValid
1613ValidValidValidInvalidValid
1620ValidValidNDNDValid
1623ValidValidValidValidValid
1631ValidValidValidValidND
1633ValidValidValidValidValid
1634ValidValidValidValidValid
1636ValidValidValidTBDTBD
NOTE: The studies are on-going for lot 1636.
----------------------------------------------------

ND = Not Done TBD = To be determined

*Criteria for a Valid Assay: Correlation coefficient (R2): ≥ 0.950 Slope: 0.005 to 0.015 Y-intercept: ≤ 0.200

{13}------------------------------------------------

3. Clinical Studies

Clinical studies were conducted at two medical institutions and at one reference laboratory. The investigators and their institutions were: Robert M. Sharkey, Ph.D., Center for Molecular Medicine and Immunology (CMMI), H. Abdel-Nabi, M.D., State University of New York at Buffalo, Dept. of Nuclear Medicine, and Dianon Systems, Inc., Stratford, CT (a commercial reference laboratory).

The objectives of the clinical studies were to:

  • retrospectively determine the incidence of HAMA in serum samples a. from a normal, apparently healthy population.
  • retrospectively evaluate the ImmuSTRIP® HAMA IgG assay to b. detect and quantitate HAMA in patients receiving murine-derived monoclonal antibodies for diagnostic and/or therapeutic purposes.
  • prospectively evaluate the ImmuSTRIP" HAMA IgG assay to detect C. and quantitate HAMA in patients receiving a murine-derived monoclonal antibody fragment for imaging colorectal cancer, (CEA-Scan", Arcitumomab).
  • compare the ImmuSTRIP" HAMA IgG assay values to those d. obtained with a reference method.
  • determine the clinical impact of elevated HAMA levels on e. pharmacokinetics and dosimetry in patients receiving 13'1-labeled murine anti-carcinoembryonic antigen (CEA) monoclonal antibody.

The results of each of the above studies are summarized on the following pages.

{14}------------------------------------------------

Summary of Clinical Studies:

Distribution of HAMA values in a normal population a.

Blood samples obtained from 464 healthy individuals (250 males and 214 females, 104 of which were pregnant) were assayed with ImmuSTRIP HAMA IgG by Dianon Systems, Inc., Stratford, CT.

Results were obtained following the manufacturer's directions for use; all samples were assayed in duplicate and final values are expressed as the mean of the duplicate results.

The following table summarizes the distribution of HAMA values in an apparently healthy population.

Healthy Subjects< 10ng/ml10-74ng/ml75-99ng/ml100-199ng/ml200-299ng/ml
250 Males41.256.80.80.40.8
110 Females49.150.9000
104 Pregnant Females35.660.5112

PERCENT (%)

The ImmuSTRIP" HAMA IgG assays defines a "negative" as ≤ 74 ng/ml. In this healthy population, 98% of the males, 100% of the non-pregnant females and 96.1% of the pregnant females were negative for HAMA. The presence of antibodies to mouse immunoglobulin in the serum of healthy individuals has been documented in several studies . 21623. Thompson et al. found interfering heterophilic antibodies in 92 samples (9.1%)20. Boscato and Stuart, using an assay specifically designed to detect the heterophilic antibodies, demonstrated their presence in 40% of 66 normal serum samples21. Courtenay-Luck et al. reported antimurine antibody immune reactivity in sera from all the healthy controls (n =24) included in their study22. The high variability of the reported incidences of heterophilic antibodies may be a result of the different sensitivities of the assays used23 .

{15}------------------------------------------------

A second clinical study was performed to determine the clinical b. utility of the ImmuSTRIP HAMA IgG ELISA assay to detect and semi-quantitate HAMA in patients receiving injections of murinederived monoclonal antibodies.

Twenty-six patients were provided by Dr. Robert Sharkey of the Center for Molecular Medicine and Immunology (CMMI), Newark, NJ and 51 patients were provided by Dr. Hani Abdel-Nabi of the VA Medical Center, Buffalo, NY.

All patients evaluated in the study had received at least one injection of a murine-derived, intact IgG antibody for imaging and/or therapeutic purposes. The total number of patients required for clinical utility study was determined based on the assumption that 30-50% of elevated HAMA levels can be detected by any assay method as documented in the literature. 4, 7, 8, 18, 21 To ensure that there is a 95% chance of estimating the detection rate with ImmuSTRIP* HAMA IgG to be no less than 10% of the minimum rate, for any method, a sample size of at least 65 HAMA determinations would be required.

A total of 305 pre- and post-injection specimens from 77 patients were assayed with ImmuSTRIP HAMA IgG ELISA test system. Seventy-seven of the 305 samples were pre-injection samples collected from patients with no known exposure to murine-derived protein prior to injection of murine-derived monoclonal antibody. One pre-injection sample, per patient, was collected and assayed with ImmuSTRIP" HAMA IgG ELISA test system.

A total of 228 post-injection samples were collected from these 77 patients at various intervals after the injection of a murine-derived monoclonal antibody. Although an effort was made to collect postinjection samples within the specified intervals, it was not always possible due to death, loss to follow-up, or unavailability of the patients. Alternatively, a substantial number of patients had samples collected in excess of the required intervals for postinjection testing. In either event, all assay results have been reported.

{16}------------------------------------------------

# of Patients(Pre-injectionSamples)# of Post-1stinjectionIntervals# of Post-1stinjectionSamples# of Pre-2ndinjectionSamples# of Post-2ndinjectionSamples# of Pre-3rdinjectionSamples# of Post-3rdinjectionSamples
111113512
252503611
3239626
5420
3515
166
77 Total PatientsTotalSamples198TotalSamples8TotalSamples17TotalSamples2TotalSamples3
GRAND TOTAL PRE-AND POST-INJECTION SAMPLES = 305

The following table represents the number of pre- and post-injection samples collected at various intervals:

Of the 11 patients with one post-1"-injection sample, three patients went on to receive a second injection of murine-derived antibody. There were three pre-2nd injection samples; two of the three patients had two post-200injection samples, and one patient had one post-20d- injection sample, for a total of five post-2nd-injection samples. One of the three patients went on to receive a third injection of murine-derived antibody. There was one pre-3rd-injection sample; that patient had two post-36-injection samples, for a total of two post-3rd-injection samples.

Of the 25 patients with two post-18-injection samples, three patients went on to receive a second injection of murine-derived antibody. There were three pre-20d-injection samples and all patients had two post-20d-injection samples, for a total of six post-200 injection samples. One of the three patients received a third injection of murine-derived antibody; one pre-30injection sample and one post-3" injection sample were assayed.

Of the 32 patients with three post-18-injection samples, two patients went on to receive a second injection of murine-derived antibody. There were two pre-2nd injection samples; one patient had two post-2004-injection samples, and one patient had four post-20d-injection samples, for a total of six post-2nd-injection samples.

In summary, there were eight patients of the 77 who received a second injection of murine-derived antibody; two of the eight patients went on to receive a third injection of murine-derived antibody.

{17}------------------------------------------------

The results of both retrospective clinical studies to determine the utility of ImmuSTRIP" HAMA IgG in patients receiving murine-derived monoclonal antibodies are summarized as follows:

Of the 26 patients provided by Dr. Robert Sharkey of the Center for Molecular Medicine and Immunology (CMMI), Newark, NJ, 21 patients (80.8%) were positive for HAMA post-injection with murinederived antibodies. The ImmuSTRIP HAMA IgG assay result was positive (> 74 ng/ml) at least once during the post-injection test interval(s).

Of the 51 patients provided by Dr. Hani Abdel-Nabi of the VA Medical Center, Buffalo, NY, 21 patients (41.2%) were positive for HAMA post-injection with murine-derived antibodies. Dr. Nabi's laboratory prepared a 1:10 dilution of patient sample with specimen diluent prior to assay with ImmuSTRIP* HAMA IgG. Immunomedics, Inc. initially recommended a 1:10 dilution, but subsequently changed the recommended sample dilution to 1:2. Specimen dilution studies resulted in acceptable expected vs. actual values and percent CV's, therefore no difference was expected in 1:2 vs. 1:10 sample dilution HAMA values. The ImmuSTRIP" HAMA IgG assay was positive (> 400 ng/ml: sensitivity 40 ng/ml X 10 [dilution factor]), at least once during the post-injection test interval(s).

c. ImmuSTRIP® HAMA and ImmuSTRIP HAMA Fragment assay results for patients receiving CEA-Scan", (Arcitumomab).

Two HAMA enzyme immunoassays, ImmuSTRIP" HAMA IgG and ImmuSTRIP HAMA Fragment, were used to quantify HAMA in patient sera obtained in the CEA-Scan Phase III study. CEA-Scan is an FDA-approved (BLA #1205) imaging agent for colorectal cancer. (Please note that the ImmuSTRIP" HAMA Fragment ELISA assay kit is the subject of a separate 510(k) submission, therefore, clinical results are provided here for informational purposes only.) Sera subsets were also assayed for drug-reactive HAMA with an HPLC assay that is used, and has been validated, to determine purity immunoreactivity, and stability of CEA-Scan". HAMA determined by this assay will be referred to as "drug-reactive HAMA." To perform the drug-reactive HAMA assay, a vial of CEA-Scan" is labeled with 990"Tc-pertechnetate according to the package insert. Twenty ng of the labeled drug contained in one ul of saline-HSA (1% human serum albumin in saline) is added to 100 ul of patient serum. The specimen is incubated for one hour at 37°C, and then analyzed for drug-reactive HAMA by application of 50 ul of the serum onto a HPLC size exclusion column, and continuously monitoring the radioactivity eluting from the column. If drug-reactive HAMA is not present in the serum, the primary peak of radioactivity elutes at a time consistent with its molecular weight of

{18}------------------------------------------------

50,000 daltons. If drug-reactive HAMA is present in the sera, the radiolabeled drug is bound by the antibody, and the resultant complex that has a molecular weight of 200,000 daltons or greater, elutes from the column prior to the elution of the 50,000 Dalton Fab'-SH. To determine if drug-reactive HAMA is directed against the constant region of the Fab', or the variable region of the Fab', 10 ug of an irrelevant murine Mab is added to the 100 ul of serum prior to addition of the labeled drug, and the assay repeated. Thus, HAMA reactive with the constant region of Fab' is neutralized and an antibody-drug complex is not formed. HAMA reactive with the variable region is not neutralized and the elution time of the radioactivity is not changed by addition of the irrelevant MAb.

Results:

Of 382 patients entered into the Phase III trials, 240 of the patients had a baseline determination and at least one follow-up measurement with ImmuSTRIP" HAMA-IgG and ImmuSTRIP" HAMA Fragment Assays (data on file at Immunomedics).

Of the 240 patients, 213 (89%) had negative baseline HAMA-IgG and HAMA Fragment test result, and remained negative after infusion of CEA-Scan ; these patients are listed in Table VII. Sera of 24 of these patients were assayed for drug-reactive HAMA (Vol. 2, Section VI, pages 140-143, Table VII, underlined patient numbers) 2-3 months post-infusion, and all of the sera were negative in this assay. Seventeen of these 213 patients received a second infusion of CEA-Scan and all of these patients remained negative with the HAMA-IgG and HAMA Fragment Assays (Vol. 2, Section VI, Table VIII, pages 144-145). Twenty sera from 12 of these patients that received a second infusion of CEA-Scan" were assayed for drug-reactive HAMA after the second administration of the drug, and all remained negative for drug-reactive HAMA (Vol.2, Section VI, Table VIII, pages 144-145).

Of the 240 patients, 24 (10%) patients had at least one serum sample that tested positive with the ImmuSTRIP" HAMA IgG Assay and negative with the ImmuSTRIP" HAMA Fragment Assay (Vol. 2, Section VI, Table IX. page 146). Twenty-one sera from ten of these patients were tested for drug-reactive HAMA. With the exception of patient 040950, all tested negative. The baseline serum and the two post-infusion sera of patient 040950 bound approximately 50% of the labeled drug, and this HAMA was neutralized by irrelevant IgG. Therefore, this preexistent drug-reactive HAMA is directed against the constant region of the Fab'. Administration of CEA-Scan" did not significantly boost the HAMA titer in this patient. One of the ImmuSTRIP" HAMA IgG positive, ImmuSTRIP" HAMA Fragment negative patients (181801) received a second injection of CEA-Scan

{19}------------------------------------------------

(Vol. 2. Section VI, Table VIII, page 145). The baseline HAMA-IgG was 209 ng/mL the day the second dose of CEA-Scan" was given, and the HAMA-IgG in the two post-infusion sera were essentially unchanged (183 ng/mL and 170 ng/mL). Both of these post-infusion sera were negative with the drug-reactive HAMA method.

Only three (1.3%) of the 240 patients tested positive with the ImmuSTRIP" HAMA Fragment Assay at any time (patients 040405, 050602, and 051803); one other patient tested positive for HAMA-Fragment (050991), but was determined to have been injected with OncoScint CR 103 shortly before being given CEA-Scan" (Vol 2, Section VI, Table X, page 147). OncoScint CR 103 contains an intact IgG-MAb, and has been established to induce HAMA-IgG levels of >400 ng/mL in over 30% of patients after a single drug-infusion PM. et al., Antibody Immunoconjugates (Kaladas and Radiopharmaceuticals 4:309-317:1991).

One of the three Phase III patients (051803) that was positive with both the ImmuSTRIP* HAMA IgG and the ImmuSTRIP* HAMA Fragment also received a second injection of CEA-Scan", a year after the first administration (Vol. 2, Section VI, Table X, page 147). The baseline sera at the time of the second injection was negative for both HAMA-IgG and HAMA-Fragment. Both post-infusion sera remained negative for HAMA-Fragment, and only the one-month post-infusion sera demonstrated a low amount of HAMA-IgG (91.7 ng/mL). However, both post-infusion sera were strongly positive with the drug-reactive HAMA test method, and HAMA in neither of the sera was neutralized with irrelevant MAb-IgG, putative evidence that this HAMA is directed against the variable region of the Fab'.

In summary, with the exception of the patient that had received OncoScint CR 103, none of the 240 patients that were monitored in this study for induction of HAMA appear to have developed HAMA.

All clinical samples were run in duplicate, the mean of the duplicate values was reported.

{20}------------------------------------------------

  • d. Comparison of the ImmuSTRIP" HAMA IgG ELISA test system values with values obtained with a double-antigen, radiometric assay method referred to as "The Reference Method."7
    Samples, provided by Dr. Robert Sharkey of CMMI, Newark, NJ, from the 28 patients evaluated in the pre- and post- injection HAMA study were assayed by Dr. M. Khazaeli, of HAMA-KINE, Inc, 1075 Thirteenth Street South, Birmingham, Alabama. There were a total of 61 samples (19 pre-injection samples and 42 post-injection samples) from the CMMI study with sufficient quantity to perform comparative testing with the Reference Method.

The results of the ImmuSTRIP" HAMA IgG ELISA and Reference Method comparative study are summarized as follows:

{21}------------------------------------------------

Of the 19 pre-injection of murine-derived antibody samples, one sample was positive (> 74 ng/ml) with the ImmuSTRIP* HAMA IgG ELISA assay and three were positive (> 10 ng/ml) with the Reference Method. Each test gave a positive result on sample #36; the Reference Method resulted in two additional positive results, samples #42 and #47.

In the post-injection of murine-derived antibody samples (n=42), there were 30 positive (> 74 ng/ml) results (71.4%) with the ImmuSTRIP* HAMA IgG assay method and 27 positive (> 10 ng/ml) results (64.3%) with the Reference Method.

  • Comparison of ImmuSTRIP* HAMA IgG (ELISA) assay Table 1. results vs. double-antigen, Radiometric, (RIA) assay results in pre- and post-injection samples.
    Pre-iniection (n=19)
ళ్ళారPositive
1---****------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------* ***BIRDINGS-------81--------------------------------------
Administration---------------------------****

"ELISA positive > 74 ng/ml, RIA positive > 10 ng/ml.

Post-injection (n=42)

% Positive*
ELISA71.4 (30/42)
RIA64.3 (27/42)

"ELISA positive > 74 ng/ml, RIA positive > 10 ng/ml.

{22}------------------------------------------------

Summary of and post- injection with murine-derived antibody (n=42)
Direction of the more of the block to the redicametric organ (D) - Summary of and post "injection will merican" and metric assay (RIA)

ELISA

NEGATIVEPOSITIVE
NEGATIVE23.8%(10/42)11.9%(5/42)
POSITIVE9.5%(4/42)54.8%(23/42)
True Positive (TP)= 23
False Positive (FP)= 5
True Negative (TN)= 10
False Negative (FN)= 4
Sensitivity=$\frac{TP}{TP + FN}$=85.2%
Accuracy=$\frac{TP + TN}{TP+TN+FP+FN}$=78.6%
Specificity=$\frac{TN}{TN + FP}$=66.7%
Positive Predictive Value(PPV)=$\frac{TP}{TP + FP}$=82.1%

RIA

:共

{23}------------------------------------------------

  • e. Clinical impact of elevated HAMA levels on pharmacokinetics and dosimetry in patients receiving 131I-Jabeled murine anticarcinoembryonic antigen (CEA) monoclonal antibody.
    A total of 57 patients with CEA-expressing tumors received therapeutic doses (4-23 mg of protein and 44-268 mCi of 1311) of anti-CEA (NP-4) IgG antibody. HAMA levels were determined preinjection and post-injection on a weekly basis until 6 weeks posttherapy and then monthly thereafter. The ImmuSTRIP" HAMA IgG assay was used to determine the HAMA values.

The above study was published in The Journal of Nuclear Medicine, June 1997. The publication is provided in this Section for the convenience of the reviewer (page 27). The impact of elevated HAMA levels on pharmacokinetics and dosimetry is discussed in detail under the section "Pharmacokinetics." The study demonstrated that at HAMA titers below 300 ng/ml, no effect on the clearance rates of the (injected) antibody from the blood and whole-body was apparent, whereas with titers above this threshold a rapidly increasing plasma and whole-body clearance rate was observed. which was reflected by decreasing red marrow and whole-body doses (r = - 0.6; significantly different from zero at p < 0.001). Further discussion on elevated HAMA levels may be found in the "Dosimetry" section of the publication.

{24}------------------------------------------------

CONCLUSIONS DRAWN FROM THE STUDIES G.

The safety and effectiveness evaluations of ImmuSTRIP* HAMA IgG demonstrated the ability of the device to detect and semi-quantitate Human Anti-Mouse Antibody (HAMA) in human serum. The preclinical studies resulted in performance specifications for sensitivity, specificity, reproducibility, dilution linearity and stability that are within acceptable limits for devices of this type.

Clinical studies with regard to the distribution of HAMA values in the normal, apparently healthy population and in the pre- and post- injection with murine-derived antibody populations demonstrate agreement with the distribution of HAMA values from similar populations cited in numerous publications' 23. The clinical impact of elevated HAMA levels on pharmacokinetics and dosimetry has been demonstrated by Behr TM, et al.24

Lastly, the comparative study with ImmuSTRIP" HAMA IgG vs. a double antigen radiometric assay method (Reference Method) demonstrated agreement with values obtained on the same samples in the two methods, and with values previously reported.7

{25}------------------------------------------------

H. REFERENCES FOR SAFETY AND EFFECTIVENESS SECTION

  • し、 La Fontaine GS, Hansen HJ, Weiss BF, et al. Enzyme Immunoassay for the Detection of Circulating Immunoglobulins in Human to Mouse Monoclonal Antibody (HAMA). Presented at the Third International Conference of Monoclonal Antibody Immunoconjugates for Cancer, San Diego, February 4-6, 1988.
    1. Courtenay-Luck NS. Epenetos AA, Moore R, et al. Development of Primary and Secondary Immune Responses to Mouse Monoclonal Antibodies used in the Diagnosis and Therapy of Malignant Neoplasms. Cancer Res. 1986; 46:6489-6493.
    1. Davies AG, Bourne SP, Richardson RB, et al. Pre-existing Anti-Mouse Immunoglobulin in a Patient Receiving 131-Murine Monoclonal Antibody for Radioimmunolocalization. Br J Cancer 1986; 53:289-292.
    1. Klein JL, Sandoz JW, Kopher MS, et al. Detection of Specific Anti-Antibodies in Patients Treated with Radiolabeled Antibody. Int J Radiat Oncol Biol Phys 1986: 12:939-943.
    1. Pimm MV, Perkins AC, Armitage NC, et al. The Characteristics of Blood-Borne Radiolabels and the Effect of Anti-Mouse IgG Antibodies on Localization of Radiolabeled Monoclonal Antibody in Cancer Patients. J Nucl Med 1985; 26:1011-1023.
    1. Schroff RW, Foon KA, Beatty SM, et al. Human Anti-Murine Immunoglobulin Responses in Patients Receiving Monoclonal Antibody Therapy. Cancer Res 1985; 45:879-885.
    1. Kaladas PM, et al. Detection of Human Anti-Murine Antibody (HAMA) Following Infusion of OncoScint CR103. Comparison of ImmuSTRIP" ELISA with a Double Antigen Radiometric Assay. Antib Immunoconj Radiopharm 1993; 4:309-317.
    1. Massuger L, et al. Specific and Nonspecific Immunoassays to Detect HAMA After Administration of Indium-111-Labeled OV-TL 3 F(ab'), Monoclonal Antibody to Patients with Ovarian Cancer. J Nucl Med 1992; 33:1958-1963.

{26}------------------------------------------------

  • Sevbold K. Trinkler M. Frey L. et al. Antigenicity of Different の Antigranulocvtes Antibodies Assessed by HAMA Follow-up in Patients Undergoing Immunoscintigraphic Detection of Infections. Presented at the German Nuclear Medicine Congress, Cologne, Germany, April 15-17, 1993.
  • Sharkey RM, Hansen HJ, Vagg R, et al. Determination of 10. Human Anti-Mouse Antibody Levels that Alter the Clearance of Radiolabeled Antibodies. Presented as a poster at the American Association for Cancer Research, Orlando, Florida, May 1993.
  • Hansen HJ, La Fontaine G, Newman ES, Schwartz MK, Malkin 11. A. Moiziski K. Martin, EW, Goldenberg DM. Solving the Problem of Antibody Interference in Commercial "Sandwich"-Type Immunoassays of Carcinoembryonic Antigen. Clin Chem 1989; 35/1:146-151.
  • Price T, Beatty BG, Beatty JD, McNally AJ. Human Anti-12. Murine Antibody Interference in Measurement of Carcinoembryonic Antigen Assessed with a Double-Antibody Enzyme Immunoassay. Clin Chem 1991; 37/1:51-57.
  • Morton, BA, O'Connor-Tressel M, Beatty, BG, Shively JE, 13. Beatty D. Artifactual CEA Elevation due to Human Anti-Mouse Antibodies. Arch Surg 1988; 123:1242-1246.
  • McCarthy RC, Ryan FJ, McKenzie CA. Interference in 14. Immunoenzymometric Assays Caused by IgM Anti-Mouse IgG Antibodies. Arch Pathol Lab Med 1988; 112:901-907.
  • ાર . Kricka LJ, Schmerfeld-Pruss D, Senior M, Goodman DBP, Kaladas P. Interference by Human Anti-Mouse Antibody in Two-Site Immunoassays. Clin Chem 1990; 36/6:892-894.
    1. Vaidya HC, Beatty BG. Eliminating Interference from Heterophilic Antibodies in a Two-Site Immunoassay for Creatine Kinase MB by using F(ab'), Conjugate and Polyclonal Mouse IgG. Clin Chem 1992; 38/9:1737-1742.

0062

{27}------------------------------------------------

    1. Ferroni P. Milenic DE. Roselli M. Carrasquillo JA, Raubitschek A. Schlom J. Colcher D. Potential for Artifacts in Monitoring for the Detection of Tumor Associated Antigens (TAG-72 and CEA) in Serum from Patients undergoing MAb-based Diagnostic and Therapeutic Protocols. Int J Biol Mark 1990; 5/4:166-176.
    1. Morrissey, NE. et al. Modified Method for Determining Carcinoembryonic Antigen in the Presence of Human Anti-Murine Antibodies. Clin Chem 1993; 39:522-531.
    1. Tuomi T. Which Antigen to Use in the Detection of Rheumatoid Factors? Comparison of Patients with Rheumatori Arthritis and Subject with 'False Positive' Rheumatoid Factor Reactions. Clin Exp. Immunol. 1989; 77:349-355.
    1. Thompson RJ, Jackson AP, Langlis N. Circulating Antibodies to Mouse Monoclonal Immunoglobulins in Normal Subjects -Incidence, Species Specificity, and Effects on a Two-Site Assay for Creatine Kinase-MB Isoenzyme. Clin Chem 1986; 32:476-481.
  • Boscato LM, Stuart MC. Heterophilic Antibodies: A Problem 21. for all Immunoassays. Clin Chem 1988; 34:27-33.
    1. Courtenay-Luck NS, Epenetos AA, Winearls CG, et al. Preexisting Human Anti-Murine Immunoglobulin Reactivity due to Polyclonal Rheumatoid Factors. Cancer Res 1987; 47:4520-04525.
    1. Boerman OC, Segers MFG, Poels LG, et al. Heterophilic Antibodies in Human Sera Causing Falsely Increased Results in the CA 125 Immunofluorometric Assay. Clin Chem 1990; 36:888-891 .
  • Behr TM, Sharkey RM, Juweid ME, et al. Phase I/II Clinical 24. Radioimmunotherapy with an Iodine-131-Labeled Anti-Carcinoembryonic Antigen Murine Monoclonal Antibody IgG. J Nucl Med, June 1997; 38/6: 858-870.

{28}------------------------------------------------

Image /page/28/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 17 1998

Joseph E. Presslitz, Ph.D. Vice President, Requlatory Affairs Immunomedics®, Inc. 300 American Road Morris Plains, New Jersey 07950

Re : K972873/S2 ImmuSTRIP® HAMA IgG - ELISA Test System for Human Trade Name: Antibodies to Mouse IgG Regulatory Class: II Product Code: DAK Dated: February 9, 1998 February 13, 1998 Received:

Dear Dr. Presslitz:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{29}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{30}------------------------------------------------

INDICATION FOR USE STATEMENT

510(k) Number (if known): 19728 73

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The ImmuSTRIP® HAMA IgG ELISA Test System is a direct enzyme-linked immunosorbent assay for the detection and quantitation of human antibodies to mouse IgG (HAMA).

Peter Naim

(Division Sign-Off)
Division of Clinical Laboratory DeviceK970873
510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

III.

§ 866.5240 Complement components immunological test system.

(a)
Identification. A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.(b)
Classification. Class II (performance standards).