Intended Use

THE TILTMASTER KIT DESIGN IS INTENDED FOR USE ON A POWER AND MANUAL WHEELCHAIR TO PROVIDE PATIENTS WITH POSTERIOR TILT FOR POSITIONING AND SEATING PRESSURE RELIEF.

Device Description

THE TILT MASTER CG SYSTEMS (CENTER OF GRAVITY), IS AN AFTERMARKET KIT CONSISTING OF A LINEAR ACTUATOR WITH 900 POUNDS CAPACITY, ALUMINUM AND STEEL BRACKETS, WIRING HARNESS TO SUPPLY VOLTAGE TO THE ACTUATOR, AND A MERCURY SWITCH DRIVER LOCKOUT WHICH RENDERS THE WHEELCHAIR INOPERABLE WHEN TILTED BEYOND 15 DEGREES. THE CG SYSTEM KIT IS DESIGNED TO CONVERT A POWERED OR MANUAL STANDARD WHEELCHAIR TO A POWER RECLINE. THE TILT MASTER CG SYSTEM USES A CENTER OF GRAVITY DESIGN TO MOVE THE PATIENTS CENTER OF GRAVITY FORWARD 6" AS THE SYSTEM TILTS. THIS PROVIDES ADDED SECURITY WHEN IN THE TILTED POSITION. THE TILT MASTER CG SYSTEM ALLOWS THE USER TO HAVE 5 TO 55 DEGREE RANGE OF TILT. THE TILT MASTER CG SYSTEM IS DESIGNED SO IF ANY CATASTROPHIC FAILURE OCCURS WHILE THE SEAT IS IN THE TILT POSITION, THE SAFETY NUT ON THE ACTUATOR WILL ALLOW THE DRIVER TO PUT THE SEAT BACK ON THE DRIVE POSITION AND WILL DISABLE THE TILT MECHANISM. THE DRIVER WILL BE UNABLE TO PUT THE SYSTEM IN THE TILT POSITION ONCE THIS OCCURS.

AI/ML Overview

The provided document is a 510(k) premarket notification for the "TILT MASTER CG SYSTEMS (CENTER OF GRAVITY)" and primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study demonstrating performance against those criteria.

Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies is not available in the provided text.

The document states "NON CLINICAL TESTING: PLEASE REFER TO THE FOLLOWING ATTACHMENT," indicating that the actual performance study details were provided in a separate attachment, which is not included in the provided text. The text only mentions the device's technical characteristics and intended use.

Based on the provided text, the following information can be extracted/inferred:

  • Device Name: TILT MASTER CG SYSTEMS (CENTER OF GRAVITY)
  • Intended Use: To provide power recline for posterior tilt for positioning and seating pressure relief on power or manual standard wheelchairs.
  • Technological Characteristics Comparison (to E&J System - a predicate device):
    • TILT MASTER CG SYSTEM has a 6" linear motion (vs. 3" for E&J).
    • TILT MASTER CG SYSTEM uses an oil light bearing roller through a trough means (vs. rolls on cam means for E&J).
    • TILT MASTER CG SYSTEM uses a "higher quality made actuator by LINAK which also has an extra 2 year warranty" (vs. actuator by Motion for E&J).
    • TILT MASTER CG SYSTEM includes a padded contoured back and headrest.
  • Safety Feature: A mercury switch driver lockout renders the wheelchair inoperable when tilted beyond 15 degrees. If a catastrophic failure occurs while tilted, a safety nut on the actuator allows the seat to be put back on the drive position and disables the tilt mechanism.
  • Tilt Range: 5 to 55 degrees.

Therefore, a table of acceptance criteria and reported device performance, and details about the study proving it, cannot be generated from the given text. The document focuses on the description of the device and its intended use for regulatory clearance.

{0}------------------------------------------------

"510K SUMMARY"

AUG 25 8

TILT-N-SPACE

DALVA R. ALEXANDER MECHANICAL APPLICATION DESIGNS, INC. 6819 HWY 90 BLVD.. SUITE 680 KATY, TEXAS 77491 (281) 391-8898 (281) 391-5585 FAX

K972564

CONTACT: DALVA R. ALEXANDER DATE PREPARED: MARCH 15, 1997

NAME OF DEVICE:TILT MASTER CG SYSTEMS (CENTER OFGRAVITY)
TRADE NAME:TILT-N-SPACE CENTER OF GRAVITYSYSTEM
PROPRIETARY NAME:CG SYSTEM
CLASSIFICATION NAME:PHYSICAL MEDICINE/POWEREDWHEELCHAIR
PRODUCT CODE:890-3860
SUBSTANTIALEQUIVALENCE:E&J-VORTEX TILT-N-SPACECENTER OF GRAVITY
FOLIO-TS90, TS95
VECTORMOBILITY-BASE LINK

DESCRIPTION: GRAVITY), IS AN AFTERMARKET KIT CONSISTING OF A LINEAR ACTUATOR WITH 900 POUNDS CAPACITY, ALUMINUM AND STEEL BRACKETS, WIRING HARNESS TO SUPPLY VOLTAGE TO THE ACTUATOR, AND A MERCURY SWITCH DRIVER LOCKOUT WHICH RENDERS THE WHEELCHAIR INOPERABLE WHEN TILTED BEYOND 15 DEGREES. THE CG SYSTEM KIT IS DESIGNED TO CONVERT A POWERED OR MANUAL STANDARD WHEELCHAIR TO A POWER RECLINE. THE TILT MASTER CG SYSTEM USES A CENTER OF GRAVITY DESIGN TO MOVE THE PATIENTS CENTER OF GRAVITY FORWARD 6" AS THE SYSTEM TILTS. THIS PROVIDES ADDED SECURITY WHEN IN THE TILTED POSITION. THE TILT MASTER CG SYSTEM ALLOWS THE USER TO HAVE 5 TO 55 DEGREE RANGE OF TILT. THE TILT MASTER CG SYSTEM IS DESIGNED SO IF ANY CATASTROPHIC FAILURE OCCURS WHILE THE SEAT IS IN THE TILT POSITION, THE SAFETY NUT ON THE ACTUATOR WILL ALLOW THE DRIVER TO PUT THE SEAT BACK ON THE DRIVE POSITION AND WILL DISABLE THE TILT MECHANISM. THE

{1}------------------------------------------------

DRIVER WILL BE UNABLE TO PUT THE SYSTEM IN THE TILT POSITION ONCE THIS OCCURS.

INTENDED USE: THE TILT MASTER CG SYSTEM DESIGN IS INTENDED FOR USE ON A POWER OR MANUAL STANDARD WHEELCHAIR TO PROVIDE POWER RECLINE FOR POSTERIOR TILT FOR POSITIONING AND SEATING PRESSURE RELIEF.

TECHNOLOGICAL CHARACTERISTICS: THE E&J SYSTEM HAS A LIMIT OF 3" OF LINEAR TRAVEL. THEIR SYSTEM ROLLS ON CAM MEANS (ROLLS ON A MACHINED OUT GROOVE). E&J USES AN ACTUATOR BY MOTION. THE TILTMASTER CG SYSTEM HAS A 6" LINEAR MOTION, HAS AN OIL LIGHT BEARING ROLLER THROUGH A TROUGH MEANS, USES A HIGHER QUALITY MADE ACTUATOR BY LINAK WHICH ALSO HAS AN EXTRA 2 YEAR WARRANTY. TILTMASTER ALSO SENDS OUT A PADDED COUNTOURED BACK AND HEADREST WITH ALL OUR SYSTEMS.

NON CLINICAL TESTING: PLEASE REFER TO THE FOLLOWING ATTACHMENT.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is simple and monochromatic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Dalva R. Alexander Chief Executive Officer Mechanical Application Designs, Inc. 6819 Highway 90 Boulevard, Suite 680 Katy, Texas 77494

AUG 2 5 1997

Re: K972564 Tilt Master CG System (Center of Gravity) Tilt-N-Space-Seat Regulatory Class: II Product Code: ITI Dated: March 15, 1997 Received: July 9, 1997

Dear Ms. Alexander:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Dalva R. Alexander

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

ia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

"STATEMENT FOR INDICATION FOR USE"

THE TILTMASTER KIT DESIGN IS INTENDED FOR USE ON A POWER AND MANUAL WHEELCHAIR TO PROVIDE PATIENTS WITH POSTERIOR TILT FOR POSITIONING AND SEATING PRESSURE RELIEF.

Over-the-Counter Use

De Colle

Divisi eneral Restorative De 510(k) Numb

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).