K Number
K971993

Validate with FDA (Live)

Date Cleared
1997-11-12

(167 days)

Product Code
Regulation Number
888.3030
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ODi Compression Hip Screw System will be used on indications that are common with presently marketed compression hip screw systems. The primary indications are for fixation/stabilization of intertrochanteric fractures, some high subtrochanteric fractures and intracapsular fractures, provided that weight bearing is not allowed until the fracture shows evidence of union. The device is intended to stabilize fragments of the fracture until bony union can occur.

Device Description

The Orthopedic Designs, Inc. Compression Hip Screw System is a cannulated lag and compression screw system to be used with a side plate for fracture fixation and stabilization of the proximal femur until bony union can occur. The system includes a lag screw that may be utilized alone with a side plate, or for compression of a fracture, may be used with a compression screw. Each system shall consist of the same material when used and shall not be mixed. Titanium Alloy and Stainless Steel within the same use. The distal end of the lag screws are keyed, but not threaded, thereby allowing the screw to slide within the barrel of the side plate while preventing rotation of the lag screw within the head of the femur. Immediate compression may be obtained with the use of a compression screw threaded into the inner canal of the lag screw. A compression screw locking mechanism called a jam screw is also available, providing a safe mechanism to lock the compression screw in place. The jam screw will prevent a common natural migration of the compression screw from the lag attachment. The side plates, lag screws, compression screws, cortical bones screws and instruments are similar in design, materials and intended use to other 510(k) cleared devices/instruments which are in commercial distribution. A range of system component sizes and lengths will be made available, offering a range of fracture compression. The components are designed to be manually or mechanically inserted using common (e.g. American Orthopedic) manual and power instrumentation.

AI/ML Overview

The provided K971993 510(k) Premarket Notification for the ODi Compression Hip Screw System is a regulatory submission for medical device clearance, not a clinical study report. As such, it does not contain the detailed information requested about acceptance criteria, device performance, test set characteristics, expert qualifications, adjudication methods, or training set details that would typically be found in a detailed study report.

The document primarily focuses on establishing substantial equivalence to predicate devices. Here's what can be extracted and what is explicitly not available:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Functional Equivalence to Predicate Devices: - Indications for use - Device design - Materials - Method of manufacture - Method of sterilizationThe ODi Compression Hip Screw System is described as "similar in design, materials and intended use to other 510(k) cleared devices/instruments which are in commercial distribution." Specifically, the document states: "Orthopedic Designs, Inc. believes the ODi Compression Hip Screw System is substantially equivalent to the products described herein with respect to indications for use, device design, materials, method of manufacture and method of sterilization."
Material Standards: - Chemical and mechanical requirements for titanium alloy 6AL-4V ELI - Chemical and mechanical requirements for 316-L stainless steelMeets ASTM F136-92 for titanium alloy 6AL-4V ELI. Meets ASTM F138-92 for 316-L stainless steel.
No FDA Performance Standards: - No specific FDA performance standards for fixation bone screws.No FDA performance standards were applicable to this device at the time of submission.

Missing Information:

  • Specific numerical performance metrics (e.g., fracture stability strength, fatigue life numbers, screw pull-out force) that might be expected from a detailed engineering or biomechanical study are not reported or referenced in this summary. The assessment of "equivalence" is qualitative based on design, materials, and intended use as compared to predicate devices.

2. Sample Size Used for the Test Set and the Data Provenance

Information Not Provided: The 510(k) summary does not describe a clinical "test set" in the context of an AI/algorithm evaluation. The comparison appears to be based on design specifications, material properties, and intended use against existing predicate devices, rather than a prospective or retrospective study with a specific patient dataset.

  • Sample Size (Test Set): Not applicable in this document.
  • Data Provenance (Country, Retrospective/Prospective): Not applicable; no patient data described.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Information Not Provided: This document does not describe the use of experts to establish ground truth for a test set. This type of information is pertinent to an AI or diagnostic device study, which this is not.


4. Adjudication Method (E.g. 2+1, 3+1, None) for the Test Set

Information Not Provided: Not applicable; no test set or adjudication process described.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Information Not Provided: Not applicable. This device is a mechanical orthopedic implant, not an AI or diagnostic imaging device that would involve human readers or AI assistance.


6. If a Standalone (I.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

Information Not Provided: Not applicable, as this is not an algorithm-based device.


7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, Etc.)

Information Not Provided: For this mechanical device, "ground truth" would typically relate to engineering specifications (e.g., material strength, dimensional accuracy, fatigue performance) or successful long-term clinical outcomes from predicate devices. However, this level of detail is not provided in the summary. The "ground truth" for the substantial equivalence claim relies on documented characteristics and performance of the predicate devices.


8. The Sample Size for the Training Set

Information Not Provided: Not applicable. This document pertains to regulatory clearance for a mechanical medical device, not an AI or algorithm-based device that would require a training set.


9. How the Ground Truth for the Training Set Was Established

Information Not Provided: Not applicable. No training set is involved.


Summary of the Study Proving Device Meets Acceptance Criteria:

The "study" presented in this 510(k) Premarket Notification is a substantial equivalence comparison to legally marketed predicate devices, rather than a de novo clinical trial or performance study generating new data for this specific device.

The premise for regulatory clearance (as detailed in the 510(k) process) is that the device is as safe and effective as a predicate device already on the market. The evidence used to support this typically includes:

  • Comparison of Design and Materials: The ODi Compression Hip Screw System utilizes materials (Titanium Alloy 6AL-4V ELI or 316-L Stainless Steel) that meet established ASTM standards (F136-92 and F138-92), which are widely accepted for orthopedic implants. The design is explicitly stated as "similar in design, materials and intended use to other 510(k) cleared devices/instruments which are in commercial distribution."
  • Comparison of Intended Use and Indications: The indications for use of the ODi system are "common with presently marketed compression hip screw systems" and align with those of the listed predicate devices (fixation/stabilization of intertrochanteric, high subtrochanteric, and intracapsular fractures).
  • Reference to Predicate Devices: A comprehensive list of approximately 20 predicate devices from companies like Howmedica, Synthes, Ace, Richards, Zimmer, OFCO, Wright, and Depuy is provided (e.g., Howmedica Compression Hip Screw System K781762, Synthes DHS Hip Screw K791619). The existence and successful use of these devices form the basis of the claim of substantial equivalence.

In essence, the "study" proving the device meets acceptance criteria is the demonstration to the FDA that the ODi Compression Hip Screw System is fundamentally similar to devices already recognized as safe and effective, and that its materials meet recognized industry standards. No new performance data from a clinical trial or a specific algorithmic validation study is detailed in this summary.

{0}------------------------------------------------

K971993

Premarket Notification

510(k) Summary of Safety and Effectiveness Information

NOV | 2 |997

For Release Upon Request Only

Date of Preparation:September 17, 1997
Regulatory Authority:Safe Medical Devices Act of 1990, 21 CFR 807.92
Company Name / Contact:
Company:ODi (Orthopedic Designs, Inc.)5501-D Airport BoulevardTampa, FL 33634
Contact:Randy Lawson, Vice President, COO(813) 889-9194
Establishment Registration Number:The registration number forOrthopedic Designs, Inc. ispending
Classification Name:Appliance, Fixation, Nail/Blade/PlateCombination, Multiple Component
Classification Reference:21 CFR § 888.3030
Common Used Name:Orthopaedic Compression Bone Screw
Device Product Code:KTT
Classification Panel:87 - Orthopaedic and Rehabilitation Devices
Trade Proprietary Name:ODi Compression Hip Screw System

Proposed Regulatory Class:

The FDA has classified similar products as a Class II device by the Orthopedic Device Section of the Surgical and Rehabilitation Devices Panel at Section 888-3030. The product code generally referred to is KTT ( Product Code: KTT ), and ODi submits this application under this designation.

{1}------------------------------------------------

Performance Standards:

No performance standards applicable to the fixation bone screw have been established by the FDA. However, the titanium alloy 6AL-4V ELI alloy or the 316-L stainless steel used to manufacture the ODi Compression Hip Screw System meets the chemical and mechanical requirements in voluntary standards established by the American Society for Testing and Materials (ASTM F136-92 and ASTM F138-92), respectively.

Package and Labeling:

Package labeling has been developed to industry standards. Packaging is also standard commercially available type quality and is stored in a fashion which prevents damage to the container or package.

Device Description:

The Orthopedic Designs, Inc. Compression Hip Screw System is a cannulated lag and compression screw system to be used with a side plate for fracture fixation and stabilization of the proximal femur until bony union can occur. The system includes a lag screw that may be utilized alone with a side plate, or for compression of a fracture, may be used with a compression screw.

Each system shall consist of the same material when used and shall not be mixed. Titanium Alloy and Stainless Steel within the same use.

The distal end of the lag screws are keyed, but not threaded, thereby allowing the screw to slide within the barrel of the side plate while preventing rotation of the lag screw within the head of the femur. Immediate compression may be obtained with the use of a compression screw threaded into the inner canal of the lag screw.

A compression screw locking mechanism called a jam screw is also available, providing a safe mechanism to lock the compression screw in place. The jam screw will prevent a common natural migration of the compression screw from the lag attachment.

The side plates, lag screws, compression screws, cortical bones screws and instruments are similar in design, materials and intended use to other 510(k) cleared devices/instruments which are in commercial distribution.

A range of system component sizes and lengths will be made available, offering a range of fracture compression. The components are designed to be manually or mechanically inserted using common (e.g. American Orthopedic) manual and power instrumentation.

{2}------------------------------------------------

Indications for Use:

The ODi Compression Hip Screw will be used on indications that are common with presently marketed compression hip screw systems. The primary indications are for fixation/stabilization of intertrochanteric fractures, some high subtrochanteric fractures, and intracapsular fractures, provided that weight bearing is not allowed until the fracture shows evidence of union. The device is intended to stabilize fragments of the fracture until bony union can occur.

Contra-indications for Use:

The ODi Compression Hip Screw System is not intended for use in patients with the following conditions:

    1. Active local Infection.
    1. Metal sensitivity or allergic reaction to foreign bodies.
    1. Other conditions that may place the patient at risk.
    1. When quality of bone stock prevents secure seating of the screws.

Substantial Equivalent Devices:

Orthopedic Designs, Inc. believes the ODi Compression Hip Screw System is substantially equivalent to the products described herein with respect to indications for use, device design, materials, method of manufacture and method of sterilization. Within the proposed class, the following devices are used as predicate devices for comparison:

Howmedica Compression Hip Screw System(K781762)
Howmedica Compression Hip Screw Sideplate(K823548)
Howmedica Alta Lag Screw & Compression Screw(K900584)
Howmedica Omega Compression Screw System(K850886),(K872223)
Howmedica Omega+ Compression Hip Screw System(K955306)
Howmedica Omega Plus Compression Hip System(K922295)
Synthes DHS Hip Screw(K791619)
Ace Cannulated Hip Screw and Captured Hip Screw System
Richards KeyLock System
Richards Kwik-Key
Richards Ambi Compression Hip Screw
Richards Classic
Zimmer Compression Hip Screw with ECT
Zimmer Free-Lock
Zimmer Versa-Fx
Orthopedic Equipment Company's (OFCO C.) Compression Hip Screw Plate

{3}------------------------------------------------

Wright Hip Compression Screw/Plate Wright Concise Compression Hip Screw System Wright Cannulated Plus Screw System Depuy Combined NoLok /Keyed Compression Hip Screw System (K861178), (K946156)

Each of these products are commercially available and marketed Class II devices indicated for use for intertrochanteric fractures, some high subtrochanteric fractures, and intracapsular fractures. The ODi Compression Hip Screw System is also labeled for similar use.

{4}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 2 1997

Mr. Randy Lawson ·Vice President, COO Orthopedic Designs, Inc. 5501-D Airport Boulevard Tampa, Florida 33634-5303

Re : K971993 Trade Name: ODi Compression Hip Screw System Requlatory Class: II Product Code: KTT Dated: September 18, 1997 Received: September 22, 1997

Dear Mr. Lawson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The qeneral controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਚੇ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. Randy Lawson

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

C. Colin M. Wittner, Ph.D.

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K971993 510(k) Number:

Device Name: ODi Compression Hip Screw System

Indications For Use:

The ODi Compression Hip Screw System will be used on indications that are common with presently marketed compression hip screw systems. The primary indications are for fixation/stabilization of intertrochanteric fractures, some high subtrochanteric fractures and intracapsular fractures, provided that weight bearing is not allowed until the fracture shows evidence of union. The device is intended to stabilize fragments of the fracture until bony union can occur.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK971893
Prescription Use (per 21 CFR 801.109)X
------------------------------------------

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.