K Number
K971522

Validate with FDA (Live)

Date Cleared
1998-03-03

(312 days)

Product Code
Regulation Number
866.3740
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Abbott TestPack Plus Strep A with On Board Controls (OBC) II is a rapid immunoassay for the qualitative detection of Group A Streptococcal (Group A Strep) antigen in throat swab specimens from patients with suspected Group A Strep associated pharyngitis and for confirmation of presumptive Group A Strep colonies isolated on culture plates.

Device Description

TestPack Plus Strep A with OBC II is a rapid immunoassay for the qualitative detection of Group A Streptococcal antigen from throat swab specimens or confirmation of presumptive Group A Strep colonies recovered from culture. TestPack Plus Strep A with OBC II utilizes an internal on board control system consisting of six control features.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Abbott TestPack® Plus™ Strep A with OBC® II, based on the provided text:

Acceptance Criteria and Device Performance

The provided text doesn't explicitly state pre-defined acceptance criteria (e.g., "The device must achieve sensitivity of at least X%"). Instead, it compares the device's performance to that of the predicate methods (SBA culture and SSA culture) to demonstrate substantial equivalence. The reported performance of the device serves as the basis for this comparison.

Table of Acceptance Criteria (Implied by Predicate Performance) and Reported Device Performance:

Performance MetricImplied Acceptance Criteria (via Predicate: SBA Culture)Reported Device Performance (TestPack Plus Strep A with OBC® II vs. SBA Culture)Implied Acceptance Criteria (via Predicate: SSA Culture)Reported Device Performance (TestPack Plus Strep A with OBC® II vs. SSA Culture)
Sensitivity(Performance similar to SBA culture)92.9% at End of Assay (EOA), 94.6% at 10 minutes(Performance similar to SSA culture)90.4% at EOA, 92.2% at 10 minutes
Specificity(Performance similar to SBA culture)97.9% at EOA, 97.0% at 10 minutes(Performance similar to SSA culture)97.9% at EOA, 97.0% at 10 minutes
Agreement(Performance similar to SBA culture)96.6% at EOA, 96.4% at 10 minutes(Performance similar to SSA culture)96.0% at EOA, 95.7% at 10 minutes

Note: The acceptance criteria are "implied" because the document focuses on demonstrating substantial equivalence rather than meeting pre-defined quantitative thresholds. The goal was to show that the new device performs comparably to established methods.

Study Details:

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 445 throat swabs.
    • Data Provenance:
      • Country of Origin: U.S. (collected at five U.S. clinical sites).
      • Retrospective or Prospective: Prospective. The swabs were collected from "patients seeking medical attention for pharyngitis," and then used to inoculate culture plates within 30 minutes. This indicates a forward-looking collection as part of the study.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that "Suspect beta-hemolytic colonies were confirmed as Group A Strep by a latex agglutination test." This implies a standard laboratory procedure performed by trained personnel, but specific numbers or credentials are not provided.
  3. Adjudication Method for the Test Set:

    • The document does not describe a specific adjudication method like 2+1 or 3+1. The ground truth was established by laboratory confirmation (latex agglutination test) of suspect colonies from the culture plates, which is a standard diagnostic procedure rather than an expert consensus/adjudication process in the typical sense for imaging or clinical decision scenarios.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This study is for an in-vitro diagnostic (IVD) device that detects a specific antigen, not an imaging or an AI-assisted diagnostic tool that would involve multiple human readers interpreting results with and without AI.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, this was a standalone performance study. The TestPack Plus Strep A with OBC II assay itself provides the result, which is then compared to the culture results. There is no "human-in-the-loop" aspect to the device's performance that is being evaluated in this context. The device directly detects the antigen.
  6. The Type of Ground Truth Used:

    • Culture confirmation/Pathology: The primary ground truth was established by isolating Group A Strep colonies on 5% Sheep Blood Agar (SBA) and Strep Selective Sheep Blood Agar (SSA) culture plates, followed by confirmation using a latex agglutination test. This is a microbiological gold standard for identifying Group A Streptococcus.
  7. The Sample Size for the Training Set:

    • The document does not mention a training set. This is because the TestPack Plus Strep A with OBC II is a rapid immunoassay (a rapid diagnostic test), not an AI/machine learning model that requires a training set. The device operates based on a pre-defined chemical/biological reaction, not learnable parameters.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there was no training set for this type of device.

{0}------------------------------------------------

K711822

MAR - 3 1998 510(k) Summary Abbott TestPack® Plus™ Strep A with OBC® II

Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination

The following information as presented in the Premarket Notification [510(k)] for Abbott TestPack® Plus™ Strep A with On Board Controls (OBC®) II constitutes data supporting a substantially equivalent determination.

Substantial equivalence has been demonstrated between TestPack Plus Strep A with OBC II and the results obtained through isolation of Group A Strep colonies on sheep blood agar medium followed by a serological grouping of presumptive Group A Strep.

TestPack Plus Strep A with OBC II is a rapid immunoassay for the qualitative detection of Group A Streptococcal antigen from throat swab specimens or confirmation of presumptive Group A Strep colonies recovered from culture.

The intended use of both is for the qualitative detection of Streptococcus pyogenes. Both can be used with throat swab specimens. TestPack Plus Strep A with OBC II detects Strep A specific antigen, whereas, culture and confirmation by typing identify growth of the Streptococcus pyogenes organisms. TestPack Plus Strep A with OBC II can be used for both the qualitative identification of Group A Strep antigen and the confirmation of suspect colonies from culture, whereas, culture and confirmation by typing can only be used for the qualitative detection of Streptococcus pyogenes organisms. TestPack Plus Strep A with OBC II utilizes an internal on board control system consisting of six control features, however, culture does not contain any internal controls. TestPack Plus Strep A with OBC II requires processing of the throat swab specimen, however, culture does not require processing of the throat swab specimen.

TestPack Plus Strep A OBC II 510(k) April 1997 ebc310ka.lwp

Attachment A Page 1 of 2

{1}------------------------------------------------

Clinical testing was performed on 445 throat swabs collected at five U.S. clinical sites from patients seeking medical attention for pharyngitis. Each swab was used to inoculate 5% Sheep Blood Agar (SBA) and Strep Selective Sheep Blood Agar (SSA) culture plates within 30 minutes of collection. The culture plates were aerobically incubated at 35 to 37℃. Suspect beta-hemolytic colonies were confirmed as Group A Strep by a latex agglutination test. When compared to SBA culture, TestPack Plus Strep A with OBC II assay demonstrated a sensitivity of 92.9% at End of Assay (EOA) and 94.6% at 10 minutes after specimen addition, a specificity of 97.9% at EOA and 97.0% at 10 minutes after specimen addition, and an agreement of 96.6% at EOA and 96.4% at 10 minutes after specimen addition for all sites combined. When compared to SSA culture, TestPack Plus Strep A with OBC II assay demonstrated a sensitivity of 90.4% at EOA and 92.2% at 10 minutes after specimen addition, a specificity of 97.9% at EOA and 97.0% at 10 minutes after specimen addition, and an agreement of 96.0% at EOA and 95.7% at 10 minutes after specimen addition for all sites combined.

In conclusion, these data demonstrate that the TestPack Plus Strep A with OBC II assay is as safe and effective as, and is substantially equivalent to the results obtained through isolation of Group A Strep colonies on SBA culture plates or SSA culture plates.

Prepared and Submitted April 24, 1997 by:

Grace LeMieux Regulatory Specialist Tel. #: (847) 937-0165 ADD Regulatory Affairs Abbott Laboratories 200 Abbott Park Road Abbott Park, IL 60064-3537

TestPack Plus Strep A OBC II 510(k) April 1997 obc510ka.lwp

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of an eagle or bird-like figure.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR - 3 1998

Grace LeMieux Sr. Regulatory Specialist ADD Regulatory Affairs, Dept. 9V6 Bldg. AP31 Abbott Laboratories 200 Abbott Park Road Abbott Park, IL 60064-3537

Re: K971522

Trade Name: Abbott TestPack® PLUS™ Strep A with On Board Controls (OBC®) II Regulatory Class: I Product Code: GTZ Dated: December 18, 1997 Received: December 19, 1997

Dear Ms. LeMieux:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely vours.

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K971522

Device Name: Abbott TestPack® Plus™ Strep A with On Board Controls (OBC*) II

Indication For Use:

21 CFR 866.3740 Streptococcus spp. serological reagents.

"Streptoccus spp. serological reagents are devices that consist of antigen and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on skin), urinary tract infections, rheumatic fever, and kidney diseases."

Abbott TestPack Plus Strep A with On Board Controls (OBC) II is a rapid immunoassay for the qualitative detection of Group A Streptococcal (Group A Strep) antigen in throat swab specimens from patients with suspected Group A Strep associated pharyngitis and for confirmation of presumptive Group A Strep colonies isolated on culture plates.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

(Per 21 CFR 801.109)

the Arr

Division of Clinical Lab

510(k) Number.

(Optional Format 1-2-96)

8

§ 866.3740

Streptococcus spp. serological reagents.(a)
Identification. Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identifyStreptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.