K Number
K971420

Validate with FDA (Live)

Date Cleared
1997-10-22

(188 days)

Product Code
Regulation Number
884.1690
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microendoscope system is used to apply and maintain pneumoperitoneum for laparoscopic procedures and to provide access and visualization for diagnostic and therapeutic purposes for laparoscopic and hysteroscopic procedures.

Device Description

Cannula and tubes are made of medical grade stainless steel and chrome plated brass.

Telescopes are made of medical grade stainless steel (sleeve), glass-lenses, glass-fibers, brass chrome plated (housing), and plastic (eyepiece).

As the intended use describes, the set is miniaturized in diameter and length for minimally invasive laparoscopy. The veress cannula and the tubes are larger in diameter and longer in working length. The telescopes are shorter in length and smaller in diameter.

AI/ML Overview

The provided text is a 510(k) summary for the Richard Wolf Mini Laparoscopes and Microendoscope System. It details the device's description, intended use, and claims of substantial equivalence to existing devices. However, the document explicitly states that "No clinical tests performed" and does not provide any acceptance criteria or performance data from a study to prove these criteria were met.

Therefore, I cannot populate the requested table or sections with information that is not present in the provided text.

Here's a summary of what is available regarding testing and performance, based on the input:

SectionInformation from Document
Acceptance Criteria & Reported Device PerformanceNot specified. The document states a claim of "substantial equivalence" to predicate devices.
Sample size for the test set and data provenanceNot applicable, as no clinical tests were performed.
Number of experts used to establish ground truth & qualificationsNot applicable, as no clinical tests were performed.
Adjudication method for the test setNot applicable, as no clinical tests were performed.
Multi-Reader Multi-Case (MRMC) comparative effectiveness studyNo. The document explicitly states, "No clinical tests performed."
Standalone (algorithm only) performance studyNot applicable, as this is a medical instrument (laparoscope/microendoscope system), not an algorithm or AI device. The document explicitly states, "No clinical tests performed."
Type of ground truth usedNot applicable, as no clinical tests were performed.
Sample size for the training setNot applicable, as this is a medical instrument (laparoscope/microendoscope system) and not an AI or algorithm-based device that would require a training set. The document explicitly states, "No clinical tests performed."
How the ground truth for the training set was establishedNot applicable, as this is a medical instrument (laparoscope/microendoscope system) and not an AI or algorithm-based device that would require a training set. The document explicitly states, "No clinical tests performed."

Further details from the document:

  • Study (or lack thereof): The document explicitly states in section 6.0, "Clinical Tests: No clinical tests performed."
  • Performance Data: Section 5.0, "Performance Data," states: "Instruments have been tested to assure the function of the spring loaded cannula and the possibility of steam autoclaving." This indicates engineering/functional testing was done, but no clinical performance data is provided or referenced, nor are specific acceptance criteria for these functional tests detailed.
  • Conclusion: The conclusion drawn (Section 7.0) is that the devices are "substantially equivalent to existing devices on the market" and "They have been tested to allow safe usage with no significant changes during the standard lifetime when used according to instructions." This reinforces the focus on functional testing and substantial equivalence rather than a clinical performance study against specific acceptance criteria.

{0}------------------------------------------------

Richard Wolf Medical Instruments Corporation

OCT 2 199

Telenhone: 847 Fax: 847,913,1488

510(k) Summary of Safety and Effectiveness
Submitter:Date of Preparation:April 16, 1997
Company / Institution name:FDA establishment registration number:
RICHARD WOLF MEDICAL INSTRUMENTS CORP.14 184 79
Division name (if applicable):Phone number (include area code):
N.A.(847) 913-1113
Street address:FAX number (include area code):
353 Corporate Woods Parkway(847) 913-0924
City:Vernon HillsState/Province:IllinoisCountry:USAZIP / Postal Code:60061
Contact name:Mr. Robert L. Casarsa
Contact title:Quality Assurance Manager
Product Information:
Trade name:Mini LaparoscopesModel number:8755.xxx
Common name:Classification name:
Veress Cannula Spring Loaded, TelescopeSet Laparoscopy
Information on devices to which substantial equivalence is claimed:
510(k)Trade or proprietary or model nameManufacturer
Number
1 K9627991 Mini Laparoscopes1 Richard Wolf M.I.C.
22 Surgy Needle VN-15002 IMAGYN
33 Veress Needle AN3MM3 Ethicon
44 OMS-PNS Pixie Microendoscope4 Origin

1.0 Description

Cannula and tubes are made of medical grade stainless steel and chrome plated brass.

Telescopes are made of medical grade stainless steel (sleeve), glass-lenses, glass-fibers, brass chrome plated (housing), and plastic (eyepiece).

2.0 Intended Use

The Microendoscope system is used to apply and maintain pneumoperitoneum for laparoscopic procedures and to provide access and visualization for diagnostic and therapeutic purposes for laparoscopic and hysteroscopic procedures.

Image /page/0/Figure/9 description: The image shows the logo for Richard Wolf. The text "RICHARD" is stacked on top of the text "WOLF". To the right of the text is a square with rounded corners that contains a stylized image. The image appears to be a drawing of a person.

{1}------------------------------------------------

Richard Wolf Medical Instruments Corporation

353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Telephone: 847.913.1113 Fax: 847.913.1488

K971420
P2972

Technological Characteristics 3.0

As the intended use describes, the set is miniaturized in diameter and length for minimally invasive laparoscopy. The veress cannula and the tubes are larger in diameter and longer in working length. The telescopes are shorter in length and smaller in diameter.

Substantial Equivalence 4.0

The submitted devices are substantially equivalent to existing pre-enactment devices sold by Richard Wolf and 510(k) devices sold by Imagyn, Ethicon, and Origin.

Performance Data 5.0

Instruments have been tested to assure the function of the spring loaded cannula and the possibility of steam autoclaving.

6.0 Clinical Tests

No clinical tests performed.

Conclusions Drawn 7.0

The mentioned R. Wolf Mini Laparoscopy devices are substantially equivalent to existing devices on the market. They have been tested to allow safe usage with no significant changes during the standard lifetime when used according to instructions.

By: Robert J. Casassa

Date: APR 15, 97

Robert L. Casarsa Quality Assurance Manager

Image /page/1/Picture/16 description: The image shows the words "RICHARD WOLF" in a bold, sans-serif font. The words are stacked on top of each other, with "RICHARD" on top and "WOLF" on the bottom. To the right of the words is a square with rounded corners, inside of which is a stylized drawing of the letters "A" and "I".

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a seal for the Department of Health and Human Services. The seal is circular and contains an emblem in the center. The emblem consists of a stylized representation of an eagle or other bird with its wings spread. The words "DEPARTMENT OF HEALTH - HUMAN" are arranged around the top of the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 22 1997

Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, Illinois 60061

Re: K971420

Microendoscope System (Mini Lap) for OB/GYN Use Dated: July 31, 1997 Received: August 1, 1997 Regulatory Class: II 21 CFR §884.1690 & 884.1720 Product Code: 85 HIH & HET

Dear Mr. Casarsa

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

h.T.Liau Yri

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K971420 510(k) Number (if known): _

7

Microendoscope System (Mini Lap) for OB/GYN Use Device Name:_

Intended Use:

The Microendoscope system is used to apply and maintain pneumoperotoneum for laparoscopic procedures and to provide access and visualization for diagnostic and therapeutic purposes for laparoscopic and hysteroscopic procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE)

Robert R. Sather/
(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 1971420

510(k) Number K94 1142

Prescription Use Per 21 CFR 801.109 L

OR

Over-The Counter

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.