K Number
K971145

Validate with FDA (Live)

Date Cleared
1997-09-17

(173 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GDS® GlucoSite® Test System is a device for in vitro diagnostic use only. The GDS® GlucoSite® Test Card is ultagi looks are quantitative determination of glucose in whole blood. It is intended for use with the Stat-Site® Meter.

Device Description

This whole blood glucose test system consists of a GlucoSite® Test Card, test card specific Test Module, and a portable, hand held Stat-Site® electronic reflectance photometer (Meter). The test method employs a dry reagent technology based on the glucose oxidase method and is specific for D-Glucose. When a drop of whole blood is applied to the top opening on the test card, the red blood cells are separated allowing only the plasma to pass through. Glucose oxidase catalyzes the oxidation of the glucose in the plasma to produce gluconic acid and hydrogen peroxide in the presence of atmospheric oxygen. The second enzyme peroxidase then catalyzes the reduction of hydrogen peroxide in presence of 4-amino-antipyrine with concomitant oxidation of the chromogen TOOS [N-ethyl-N-(2-hydroxy-3sulfopropyl)-m-toluidine] to produce a purple color that is measured at 660 nm. The intensity of color developed is correlated with the glucose concentration in the whole blood sample. The Stat-Site® Meter measures the reflectance of the color produced at 660 nm and converts the reflectance reading to corresponding plasma glucose values.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the GlucoSite® Test System:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device PerformanceComments
Correlation with YSI 2300 STAT PLUS Analyzer (Predicate Method):Slope: 1.008Intercept: -2.996Correlation Coefficient (r): 0.99These values indicate a strong linear relationship and close agreement with the predicate device. An 'r' value of 0.99 signifies excellent correlation.
Correlation with Hexokinase Method (Reference Method):Slope: 1.000Intercept: -11.425Correlation Coefficient (r): 0.99Another strong linear relationship and excellent agreement with the Hexokinase reference method, providing further validation of accuracy.
Correlation of Finger Stick Blood (GlucoSite) with Venous Plasma (YSI):Slope: 0.999Intercept: 5.473Correlation Coefficient (r): 0.965This demonstrates a good correlation between the device's finger stick results and established plasma glucose values, indicating clinical utility for the intended sample type.
Total Coefficient of Variation (T-CV) for Precision:Below 5% for glucose concentrations of 50, 80, 150, and 350 mg/dLThis demonstrates good precision across the measurable range, meaning the device provides consistent results.
Intended Use Conditions:The device was evaluated under intended use conditions (POL/physician's office settings).The study confirms the device performs as expected in its target environment.
Targeted Population:Patients except neonates and children under 6 months.The study included samples from diabetic and non-diabetic subjects, aligning with the targeted population.

Study Details

  • Sample Size Used for the Test Set:

    • 152 venous whole blood clinical samples
    • 152 capillary finger stick blood samples
    • 10 additional venous whole blood samples (spiked to 3 different glucose concentrations)
    • Total: 314 samples
  • Data Provenance: Retrospective and prospective. Clinical samples were obtained from 3 different physician's office sites (indicates prospective collection). The additional 10 spiked samples were generated "in house." Country of origin is not explicitly stated but implied to be the US, given the FDA submission.

  • Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications: Not applicable. For glucose meters, the ground truth is established by a reference laboratory method (Hexokinase) or a highly accurate predicate device (YSI 2300 STAT PLUS Analyzer), not by human experts interpreting results.

  • Adjudication Method for the Test Set: Not applicable. Glucose measurements are quantitative and are directly compared to the reference methods, not adjudicated by human consensus.

  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No, an MRMC study was not done. This type of study is typically performed for imaging or interpretive diagnostic devices where human reader performance is a key factor. This is a standalone diagnostic device.

  • Standalone (i.e., algorithm only without human-in-the-loop performance) Study: Yes, this was a standalone study. The performance of the Stat-Site®/GlucoSite® Test System was directly compared to the reference methods. Human interpretation or assistance is not part of the device's function or the study design.

  • Type of Ground Truth Used:

    • Reference Method: Hexokinase plasma glucose method.
    • Predicate Device: YSI 2300 STAT PLUS Analyzer (also a highly accurate glucose measurement device).
  • Sample Size for the Training Set: Not explicitly stated as a separate "training set." For this type of device, calibration and internal algorithm development would occur during the device's R&D phase, likely using a different set of samples. The described study focuses on validation using clinical and spiked samples.

  • How the Ground Truth for the Training Set Was Established: Not detailed in this summary. However, typically, the ground truth for calibrating such a device would also be established using highly accurate laboratory reference methods like Hexokinase on a large set of samples covering the intended measurement range.

{0}------------------------------------------------

. 1 . 1 . 1 . 1 . 1 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . 2 . Division of GDS Technology, Inc 25235 Leer Drive PO. Box 473 Elkhart, IN 46515

(219) 262-0109

SEP 1 7 1997

510(k) Summary

Submitter Identification:

GDS® Diagnostics A Division of GDS® Technology, Inc. 25235 Leer Drive Elkhart, IN 46514 (219) 264-7384 Fax: (219) 262-0109

John W. Swatosh March 27, 1997

GlucoSite® Test System Blood Glucose Test System Glucose Test System YSI 2300 STAT PLUS Analyzer

Contact: Date:

Trade Name: Common Name: Classification Name: Comparison Device:

Device Description:

This whole blood glucose test system consists of a GlucoSite® Test Card, test card specific Test Module, and a portable, hand held Stat-Site® electronic reflectance photometer (Meter). The test method employs a dry reagent technology based on the glucose oxidase method and is specific for D-Glucose. When a drop of whole blood is applied to the top opening on the test card, the red blood cells are separated allowing only the plasma to pass through. Glucose oxidase catalyzes the oxidation of the glucose in the plasma to produce gluconic acid and hydrogen peroxide in the presence of atmospheric oxygen. The second enzyme peroxidase then catalyzes the reduction of hydrogen peroxide in presence of 4-amino-antipyrine with concomitant oxidation of the chromogen TOOS [N-ethyl-N-(2-hydroxy-3sulfopropyl)-m-toluidine] to produce a purple color that is measured at 660 nm. The intensity of color developed is correlated with the glucose concentration in the whole blood sample. The Stat-Site® Meter measures the reflectance of the color produced at 660 nm and converts the reflectance reading to corresponding plasma glucose values.

Intended Use:

The Stat-Site®/GlucoSite® Blood Glucose System is intended for quantitative determination of glucose in whole blood in a point-of-care setting such as a physician's office or hospital point of care sites. The test system provides glucose values equivalent to plasma glucose values. The test is not to be used at home or for neonates or children under the age of six months.

{1}------------------------------------------------

Comparison:

iparioon.
To verify safety and effectiveness of the Stat-Site®/GlucoSite® Glucose Test when used under intended use conditions such as a POL setting, the device was compared with the Hexokinase plasma glucose method and the YSI 2300 STAT PLUS Analyzer. A total of 152 venous whole blood clinical samples and 152 capillary finger stick blood samples were obtained at 3 different physician's office sites. Half of the samples were from diabetic subjects. Ten (10) more venous whole blood samples obtained in house were spiked to 3 different concentrations of glucose and were included in the study to demonstrate maximum reportable range. Whole blood samples were tested on GlucoSite® Test Cards on the Stat-Site® Meter and the results obtained were compared with the glucose values obtained when plasma from the whole blood samples was analyzed on the YSI 2300 STAT PLUS Analyzer and with Hexokinase method on the Olympus Reply Instrument. The following regression parameters were obtained comparing (A) Stat-Site®/GlucoSite Test System (y) with YSI(x) and (B) Stat-Site®/GlucoSite® Test System(y) with Hexokinase results (x).

  • Stat-Site®/GlucoSite® vs. YSI A. Slope = 1.008 Intercept = - 2.996 Correlation Coefficient (r) = 0.99
  • Stat-Site®/GlucoSite® vs. Hexokinase B. Slope = 1.000 Intercept = -11.425 Correlation Coefficient (r) = 0.99

When the glucose data from the finger stick blood on Stat-Site®/GlucoSite® device was compared with plasma glucose values of the same subject obtained from venous blood using YSI, the predicate method, the following regression equation was obtained with 'r' value of 0.965:

y = 0.999 (x) + 5.473

Where y = Stat-Site®/GlucoSite® glucose result and

x = Plasma glucose result from YSI.

A similar regression was obtained when compared with Hexokinase method with 'r' value of 0.965.

Precision evaluation was performed at glucose concentrations of 50, 80, 150 and 350 mg/dL on 3 Stat-Site®/GlucoSite® Test Systems. Total CV's were below 5% demonstrating the performance of the Stat-Site®/GlucoSite® Glucose Test to be comparable to the claimed performance of similar meters currently in commercial distribution.

Conclusion:

The data demonstrated that blood glucose results from Stat-Site®/GlucoSite® Test System, when used in a point of care setting such as a POL or physician's office, compare well to those of laboratory instruments such as YSI and Hexokinase method when determining plasma glucose from normal or diabetic patients.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged on a single line, with "Public" and "Health" being slightly larger than "Service". The text is black against a white background.

Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

SEP 17 1997

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Regulatory Affairs
GDS Diagnostics
25235 Leer Drive
Elkhart, IN 46515
Re:K971145/S1
GlucoSite® Test System/GlucoSite® Test Cards and Test Module
Regulatory Class:II
Product Code:CGA
Dated:June 4, 1997
Received:June 10, 1997

Dear Mr. Swatosh:

Mr. John W. Swatosh

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

GlucoSite® 510(k) Submission
GlucoSite® 510(k) Submission
1000007 - 1007 Glucoon
March 27, 1997 Page 7 of 91

Indications for Use Statement 8.1

Pursuant to the Notice of 2/6/96 regarding listing of Indications for Use on a separate sheet, the following is provided.

(To be assigned) 510(k) Number __

GlucoSite® Test System Device Name:

Indications for Use:

The GDS® GlucoSite® Test System is a device for in vitro diagnostic use only. The GDS® GlucoSite® Test Card is ultagi looks are quantitative determination of glucose in whole blood. It is intended for use with the Stat-Site® Meter.

Patients except neonates and children under the Targeted population: age of 6 months.

Physician's Office or other Professional Point of Environment of Use: Care Setting

(Division Sign-Off)
Division of Clinical Laboratory Zevices
510(k) Number 477/14

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.