K Number
K970902

Validate with FDA (Live)

Device Name
MEDILOG FD4
Date Cleared
1997-05-05

(55 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medilog FD4 Hoher monitor is intended to be used in the diagnosis of various cardiac disorders by the recording of ECG in an ambulatory mode. These disorders principally include cardiac arrhythmia's and myocardial ischeamia.

The ECG data is used by the physician as an aid in the diagnosis of the patients condition. In some circumstances the patient may be hospitalised as part of the medical evaluation and Holter monitoring may be performed in conjunction with other non-invasive or invasive procedures. In such cases there is no interaction between the FD4 and other equipment.

Device Description

Decer poton it looks like an MR63. The difference being the analogue cassette mechanism has been replaced with a digital acquisition module containing a slot for a PCMCIA Flash ATA card.

The patient is connected to the FD4 via disposable electrodes attached to the chest. These detect the electrical signals generated by the heart, which are then transferred to the recorder via the electrode leads and patient cable. In the recorder, these signals are amplified, filtered, digitised and compressed without loss before being stored on a removable PCMCIA Flash ATA card. At the completion of recording the flash memory card is removed and the recorded ECG data is replayed and analysed on the Oxford Instruments ECG Analysis System.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance (FD4)
Performance Bench TestsInput dynamic range meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
Gain accuracy meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
System noise meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
Multichannel crosstalk meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
Frequency response meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
Minimum feature size meets American National Standard EC38 requirements.FD4 meets all requirements of American National Standard EC38.
Meets draft British Standard Electrical Equipment - part 2: Particular requirements for safety - section 2.47: specification for ambulatory electrocardiographic monitors.FD4 meets all requirements of the draft British Standard.
Trial PerformanceRecording lasted for its intended duration.All recordings (29 total) met this criterion.
Patient events detected.All recordings (29 total) met this criterion.
Signal quality and gain comparable with the predicate device (Medilog MR63), where applicable.All recordings (29 total) met this criterion.
No problems with either the recording or the replay.All recordings (29 total) met this criterion.
Successful review by Oxford Instruments clinical specialist.All recordings (29 total) met this criterion.
Technical SpecificationsImproved, updated standards including EC38 performance (compared to MR63).FD4 complies with IEC 601-1, UL 2601-1, AAMI EC38, 93/42/EEC (MDD). The MR63 complied with older standards (IEC 601-1, UL 544, 93/42/EEC).
Improved operating range (temperature).FD4: 0 °C to 45 °C (compared to MR63: 5 °C to 40 °C).
Improved frequency response (due to digital recording).FD4: 0.05Hz - 40 Hz (compared to MR63: 0.5Hz - 25Hz).
Lower typical peak-to-peak noise over input bandwidth.FD4: <12 μV (compared to MR63: <50 μV).
Higher resolution patient event timing.FD4: 1/128th sec (compared to MR63: 1 second).
Improved dynamic range.FD4: 10 μV to 10 mV (compared to MR63: 50 μV to 10 mV).
Provides patient isolation by virtue of plastic optical fiber.FD4: Automatically provided by plastic fibre optic cable (compared to MR63: XE-45 write-coupler).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 29 recordings. 17 of these were performed in tandem with the predicate device (MR63).
  • Data Provenance: The data came from a variety of sources:
    • A programmable simulator (giving "researchiseble" EKGs - likely representative samples)
    • An EKG generator (providing real patient data from a database of recordings - likely retrospective, but no specific country of origin is mentioned)
    • Human volunteers (prospective data, but no specific country of origin is mentioned)

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: The document states "Successful review by Oxford Instruments clinical specialist." This implies at least one clinical specialist was involved.
  • Qualifications of Experts: The qualification provided is "clinical specialist" at Oxford Instruments. Further specifics (e.g., years of experience, specific medical specialty like cardiology or electrophysiology) are not provided.

4. Adjudication Method for the Test Set

  • The document mentions "Successful review by Oxford Instruments clinical specialist." This suggests a single expert review. There is no indication of an adjudication method involving multiple experts (e.g., 2+1, 3+1). It appears to be a single-reader review based on the stated acceptance criterion.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done in the context of human readers improving with AI vs. without AI assistance.
  • The study compared the FD4 device's performance to its predicate device (MR63), but this was a technical/signal comparison, not a human reader performance study. The FD4 is a data acquisition device, not an AI interpretation system.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • This device (Medilog FD4) is a Holter monitor designed for recording ECGs, not for automated interpretation or diagnosis. It records and stores data for later analysis by a physician using a separate "Oxford Instruments ECG Analysis System" (Medilog Optima & Excel 2).
  • Therefore, the concept of "standalone algorithm performance" as typically applied to AI-based diagnostic devices does not directly apply to the FD4. The performance testing focuses on the quality of the recorded signal and its comparability to the predicate device.

7. The Type of Ground Truth Used

  • The ground truth for the trial results was based on several metrics:
    • Technical/Operational Criteria: Intended recording duration, successful download, signal quality/gain comparability with predicate, absence of recording/replay problems.
    • Clinical Relevance: Detection of patient events, heart rate variability monitored and logged for comparison.
    • Expert Review: Successful review by an Oxford Instruments clinical specialist.
  • Given the nature of the device (a recorder), the "ground truth" is less about a definitive medical diagnosis and more about the fidelity and completeness of the ECG data captured and reviewed by an expert. It's a combination of objective technical performance and subjective expert assessment of data utility.

8. The Sample Size for the Training Set

  • The document does not mention a training set for the FD4 device. This is because the FD4 is a hardware device for acquiring data, not a machine learning algorithm that requires training. The "Medilog Optima & Excel 2" analysis system might potentially incorporate algorithms, but the FD4 itself is a data acquisition unit.

9. How the Ground Truth for the Training Set Was Established

  • As no training set is mentioned for the FD4 device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "OXFORD" in a stylized, sans-serif font. The letter "O" is a perfect circle, and the "X" is formed by two intersecting lines. The "F" is a straight line with a curved shape attached to the right side. The second "O" has a crescent shape inside of it. The "R" and "D" are more traditional sans-serif letters.

ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ

MAY - 5 1997 - 5 1997


K970902

:

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "OXFORD" in a stylized font. Each letter is distinct and bold. The letters are evenly spaced and the word is horizontally aligned.

Oxford Instruments Modical Systems Division 1 1 526 53rd Street N Clearwater, Florida 34620 Telephone (813) 573-4500 Fax (813) 572-6836

510(k) Summary

Date of summary :

Contact : Charles Holz Regulatory Affairs Manager

Name of device : Medilog FD4 Classification name : Monitor - Electrocardiograph Name of predicate : Medilog MR63 K943862

Description

Decer poton it looks like an MR63. The difference being the analogue cassette mechanism has been replaced with a digital acquisition module containing a slot for a PCMCIA Flash ATA card.

The patient is connected to the FD4 via disposable electrodes attached to the chest. These detect the electrical signals generated by the heart, which are then transferred to the recorder via the electrode leads and patient cable. In the recorder, these signals are amplified, filtered, digitised and compressed without loss before being stored on a removable PCMCIA Flash ATA card. At the completion of recording the flash memory card is removed and the recorded ECG data is replayed and analysed on the Oxford Instruments ECG Analysis System.

Intended use

The Medilog FD4 Hoher monitor is intended to be used in the diagnosis of various cardiac disorders by the recording of ECG in an ambulatory mode. These disorders principally include cardiac arrhythmia's and myocardial ischeamia.

The ECG data is used by the physician as an aid in the diagnosis of the patients condition. In some circumstances the patient may be hospitalised as part of the medical evaluation and Holter monitoring may be performed in conjunction with other non-invasive or invasive procedures. In such cases there is no interaction between the FD4 and other equipment.

FD4 performance testing.

Performance bench tests have been carried out on FD4 in accordance with American National Standard EC38 "Ambulatory electrocardiographs" and draft British Standard Electrical Equipment- part 2: Particular requirements for safety- section 2.47: specification for ambulatory electrocardiographic monitors. The performance parameters tested were; input dynamic range, gain accuracy, system noise, multichannel crosstalk, frequency response, and minimum feature size. The FD4 was found to meet all requirements.

Summary of FD4 trial results

A total of 29 recordings, 17 of which were performed in undern with an MR63, were carried out from a variety of sources, with a view to establishing that the FD4 recorder was equivalent to its predicate device, the Medilog MR63. The recording sources were: a programmable simulator (giving researchiseble EKGs); an EKG generator (providing real patient data from a database of recordings); and human volunteers. The recordings were checked for signal quality, reliability of recording and download, detection of patient events and recording duration. In addition the heart rate variability was monitored and logged for comparison with the predicate device. All recordings were archived to optical disk.

All recordings met the criteria for a successful recording as defined below.

  • Recording lasted for its intended duration ●
  • Patient events detected 0
  • Signal quality and gain comparable with the predicate device, where applicable �
  • No problems with either the recording or the replay ●
  • Successful review by Oxford Instruments clinical specialist .

{2}------------------------------------------------

Comparison table of FD4/MR63 features & specification

..

MIR63FD4Impact oneffectivenessImpact on safety
Type of deviceHolter ECG recorderHolter ECG recorderN/AN/A
Device classnon-criticalRegulatory Class IInon-criticalRegulatory Class IIno changeno change
designed tocomply with.....IEC 601-1, UL 544,93/42/EEC (MDD)IEC 601-1, UL 2601-1,AAMI EC38,93/42/EEC (MDD)Improved, updatedstandards includingEC38 performance.none
EMCIEC 601-1-2IEC 601-1-2no changeno change
Environmental5 °C to 40 °C10 to 95% RH0 °C to 45 °C10 to 95% RHImproved operatingrange.none
Recording duration24 hours24 hoursno changeno change
Mounting methodBelt-mounted pouchBelt-mounted pouchno changeno change
CasePolycarbonate / ABSPolycarbonate / ABSno changeno change
Power source2xAA 1.5v primaryalkaline or 1.2v NiMhrechargeable2xAA 1.5v primaryalkaline or 1.2v NiMhrechargeableno changeno change
Electrode system3/5/7 leadconfigurations usingyoke cable andindividual electrodeleads3/5/7 leadconfigurations usingyoke cable andindividual electrodeleadsno changeno change
Recording mediumC-60 cassette20Mb flash ATA cardimproved frequencyresponse (no tapevariations)none
Recording methodDirectDigitised andcompressed withoutloss.Improved (no replayvariations)none
Display4 digit + colon 7segment LCD displaystime4 digit + colon 7segment LCD displaystime and additionalerror codesMore effective use ofdisplay to showuseful informationnone
Audible sounder4kHz bleeper used toprovide feedbackduring time/datesetting and patienteventsFrequency ofbleeping variableProvide moreeffective feedbackand to distinguishbetween varioussoundsnone
Patient eventmarkingencoded onto datatrack of cassette.Resolution 1 secondStored in file header.Resolution 1/128 thsecHigher resolutionpatient event timingan advantagenone
External controlsPatient event button(also used fortime/date setting).Head arm switchPatient event button(also used fortime/date setting)plus ON/OFF slideswitch underrecorder lidNeed ON/OFF switchto replace head armswitch of tapemechanism.none
Signals recorded3 x analogue ECG3 x digitised ECGimprovednone
Setting upAnalogue monitorsocket (not isolated)for all 3 ECG channelsFibre optic outputproducing ECGsignals in digitalformat at real-timeAttaches directly toexisting OXFORDreplay systems. Noneed for separatewrite-out unit.Provides patientisolation by virtue ofplastic optical fibre.
Method ofisolation duringmonitoringXE-45 write-couplerAutomaticallyprovided by plasticfibre optic cablenonePlastic fibre opticinherently safer thanrelying on use of safewriter coupler.
Data recordedTime/Date & patienteventsTime/date & patienteventsno changeno change
Data formatPulse widthmodulatedDigital format in fileheader.Better reliability-notsensitive to tapevariationsnone
Channel gainadjustmentHardware gainadjustment usingdigitalSoftware gainadjustment to giveoptimum signalEnsures high fidelityreproduction of small& large ECG signalsnone
Dynamic rangepotentiometers.50 μV to 10 mVoptimum signal amplitude10 μV to 10 mVImproved. Allows accurate recording of larger range of ECG signals & large ECG signals amplitudes.none

{3}------------------------------------------------

MR63FD4Impact oneffectivenessImpact on safety
Overallrecorder/replaybandwidth0.5Hz - 25Hz0.05Hz - 40 HzNo frequencyresponse limitationdue to taperecording/replay.Frequency responseonly due to hardwarefilters.none
Typical pk-pk noiseover inputbandwidth<50 μV<12 μVLack of tapeeliminates tape noisefrom overall systemnoisenone
Data compressionanalogue- noneLossless compression.No data is lost - justpacked moreefficiently onto diskEnsures 24 hours ofdata fit onto 20Mbflash card. Nodegradation ofsignal.none
Replay & analysissystemMedilog Optima &Excel 2Medilog Optima &Excel 2nonenone
Analysis featuresin recordernonenoneN/AN/A
Alarms, warningsnonenoneN/AN/A
control over otherequipmentnonenoneN/AN/A

:


:

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).