K Number
K970383

Validate with FDA (Live)

Device Name
SHARPS CONTAINER
Date Cleared
1997-07-03

(150 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Our Horizontal Entry Sharps Containers provide a receptacle for used, contaminated medical Sharps and for enclosure during transport to ultimate disposal.

Device Description

Our Horizontal Entry Sharps Container consists of an injectionmolded polypropylene container fitted with a snap-on, leakresistant, poly-propylene lid. The lid has a "mail-box" style raised opening into which the sharps are dropped horizontally. When the container is full, a unique butterfly closure system closes to prevent over-filling and to prevent access to the sharps inside.

AI/ML Overview

This document describes a 510(k) submission for a Horizontal Entry Sharps Container, focusing on its substantial equivalence to existing devices rather than a detailed study proving performance against acceptance criteria. There is no information in the provided text that describes the actual acceptance criteria, a study proving the device meets those criteria, or any of the detailed study parameters requested in your prompt.

The 510(k) summary (Attachment 5, Rev. 1) and the FDA letter explicitly state that the device is being submitted for "substantially equivalent" status. This means the manufacturer is asserting that their device is as safe and effective as a legally marketed predicate device, rather than providing new clinical or non-clinical study data to prove novel performance claims against specific acceptance criteria.

Therefore, I cannot populate the table or answer most of the questions you've posed based on the provided text.

Here's what can be inferred from the document regarding the device's intended claims and comparison:

1. A table of acceptance criteria and the reported device performance

No explicit acceptance criteria or reported device performance are provided in the document. The submission is based on substantial equivalence.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. No test set or data provenance from a performance study is mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No ground truth determination or expert involvement for a study is mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No adjudication is mentioned.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/imaging device, and no MRMC study is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. No ground truth for a study is mentioned.

8. The sample size for the training set

Not applicable. No training set for an algorithm is mentioned.

9. How the ground truth for the training set was established

Not applicable. No ground truth for a training set is mentioned.


Summary of Device and 510(k) Submission:

  • Device Name: Horizontal Entry Sharps Container
  • Manufacturer: Custom Medical Plastics, Inc.
  • Intended Use: Provides a receptacle for used, contaminated medical sharps and for enclosure during transport to ultimate disposal.
  • Classification: Class II accessory to a hypodermic needle (21 CFR Section 880.5570), Product Code: FMI.
  • Basis for Marketing: Substantial equivalence to Horizontal Entry Sharps Containers manufactured by Sage Products Inc. and Sherwood Medical.
  • Key Features (described in 510(k) summary):
    • Injection-molded polypropylene container with a snap-on, leak-resistant polypropylene lid.
    • "Mail-box" style raised opening for horizontal sharps insertion.
    • Unique butterfly closure system to prevent over-filling and access once full.
    • Wall mount bracket and locking mechanism (similar to predicates).
  • Key Claim: "Our Horizontal Entry Sharps Container will be manufactured per specifications and good manufacturing practices that ensure the device is safe and effective for its intended use." (This is a general quality statement, not a specific performance metric).
  • Cautionary Statement: "container is puncture resistant, but not entirely puncture-proof. To avoid the possibility of accidental sticks, do not overfill, inspect unit before handling and handle with care."

The provided text focuses on the description of the device and the justification for substantial equivalence to existing devices, which is the core of a 510(k) submission. It does not contain the specifics of a detailed performance study as you've outlined. For physical devices like this, manufacturing specifications, material testing, and design validation (often demonstrated through comparison with predicate devices and adherence to relevant standards, though not explicitly detailed here) would form the basis of safety and effectiveness, rather than clinical study-style acceptance criteria.

{0}------------------------------------------------

K970383 ATTACHMENT 5, Rev. 1

70383 3 JUL * 3 1997

510 (k) SUMMARY

THIS SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION IS BEING SUBMITTED IN ACCORDANCE WITH THE REQUIREMENTS OF SMDA 1990.

Summary of Safety & Efficacy: Horizontal Entry Sharps 510 (k) Container

Custom Medical Plastics, Inc. 11811 Calhoun Road Submitted By: Omaha, NE 68122 (402) 691-2855 Daniel R. Brown, President Contact: Date Prepared: December 24, 1996 (Ammended April 4, 1997)

Our Horizontal Entry Sharps Container consists of an injectionmolded polypropylene container fitted with a snap-on, leakresistant, poly-propylene lid. The lid has a "mail-box" style raised opening into which the sharps are dropped horizontally. When the container is full, a unique butterfly closure system closes to prevent over-filling and to prevent access to the sharps inside.

Sharps Containers are classified as a Class II accessory to a hypodermic needle (21 CFR Section 880.5570); Panel 80 FMI.

Our Horizontal Entry Sharps Container is substantially equivalent to Horizontal Entry Sharps Containers manufactured by Sage Products Inc. and Sherwood Medical. These containers are all made of polypropylene by the injection molding process. They feature a mechanism for inserting the sharps in a horizontal plane and a lid to secure the opening when the container has been filled. Our Horizontal Entry Sharps Containers, along with the Sage and Sherwood Containers also feature a wall mount bracket and a locking mechanism to secure the container, thus preventing unauthorized access or removal.

Our Horizontal Entry Sharps Container provides a receptacle for used, contaminated sharps and for enclosure during transport to ultimate disposal. The product labeling includes a cautionary statement that the "container is puncture resistant, but not entirely puncture-proof. To avoid the possibility of accidental sticks, do not overfill, inspect unit before handling and handle with care." Final disposal is typically accomplished by incineration. When incinerated in compliant medical waste incinerators, these sharps containers do not present any environmental hazards resulting from plastics combustion.

Our Horizontal Entry Sharps Container will be manufactured per specifications and good manufacturing practices that ensure the device is safe and effective for its intended use.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUM SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines forming its head and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Daniel R. Brown ·President Custom Medical Plastics, Incorporated and 14315 "C" Circle Omaha, Nebraska 68144

JUL - 3 1997

Re : K970383 Trade Name: Sharps Container Requlatory Class: II Product Code: FMI Dated: May 14, 1997 Received: May 20, 1997

Dear Mr. Brown:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{2}------------------------------------------------

Page 2 - Mr. Brown

through 542 of the Act for devices under the Electronic chrough Siz on on ontrol provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The ▽DA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on complianse as (advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note one regulation entitled, "Misbranding by reference to
the regulation entitled, "Misbranding by reference to information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaa.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K970383 ATTACHMENT 4, Rev. 1

Page 1 of 1

K970383 510(k) Number:

Device Name: Sharps Container

Indications For Use:

Our Horizontal Entry Sharps Containers provide a receptacle for used, contaminated medical Sharps and for enclosure during transport to ultimate disposal.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CORH, office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Dental, infection Cotro!,
and General Hospita! Devices
510(k) Number K970383

Prescription Use __ OR Over-the-Counter Use X
(Per 21 CFR 801.109) (Optional Format 1-2-96)

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).