IS-ANTI-SSA TEST SYSTEM
K970218 · Diamedix Corp. · LLL · Apr 8, 1997 · Immunology
Device Facts
| Record ID | K970218 |
| Device Name | IS-ANTI-SSA TEST SYSTEM |
| Applicant | Diamedix Corp. |
| Product Code | LLL · Immunology |
| Decision Date | Apr 8, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5100 |
| Device Class | Class 2 |
Intended Use
The assay is intended for use in detecting antibodies to SSA (Ro) antigen in a single human serum sample. The results of the assay are to be used as an aid in the diagnosis of autoimmune disorders.
Device Story
Is-anti-SSA Test System is an ELISA kit for detecting IgG antibodies to SSA (Ro) antigen in human serum. Purified SSA (Ro) antigen is immobilized on microtiter wells; patient serum is added; if anti-SSA antibodies are present, they bind to the antigen. After washing, enzyme-labeled anti-human immunoglobulin conjugate is added, followed by a substrate solution. The substrate converts to a colored product proportional to the amount of bound antibody. Color intensity is measured photometrically. The device is used in clinical laboratories, either manually or via the MAGO automated processing system. Results provide an indirect measure of specific antibody levels, aiding clinicians in diagnosing autoimmune disorders.
Clinical Evidence
Clinical evaluation compared the Is-anti-SSA Test Kit to a commercial predicate using 165 samples (100 normal, 65 autoimmune). Manual testing showed 93% relative sensitivity, 99% relative specificity, and 96% agreement. MAGO automated testing showed 93% relative sensitivity, 98% relative specificity, and 96% agreement. Precision testing (intra/inter-assay CVs) and cross-reactivity studies with other autoimmune specificities (SSB, Sm, RNP, Jo-1, Scl-70) were performed. Correlation between manual and MAGO methods was 0.85 (Pearson).
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) using purified SSA (Ro) antigen immobilized on microtiter wells. Employs enzyme-labeled anti-human immunoglobulin conjugate and substrate for colorimetric detection. Photometric measurement of color intensity. Compatible with manual processing or the MAGO automated system. Connectivity: Standalone/Instrument-integrated.
Indications for Use
Indicated for the detection and semi-quantitation of IgG antibodies to SSA (Ro) antigen in human serum to aid in the diagnosis of autoimmune disorders.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
- Helix Diagnostics Enzyme Immunoassay Anti-SS-A/Ro Antibody Test Kit
Related Devices
- K970237 — IS-ANTI-SSB TEST SYSTEM (720-260) · Diamedix Corp. · Apr 8, 1997
- K993636 — MDI SS-A TEST · Micro Detect, Inc. · Jan 10, 2000
- K970220 — IS-ANTI-SM TEST SYSTEM · Diamedix Corp. · Apr 8, 1997
- K142781 — ImmuLisa Enhanced SS-A (Ro) Antibody ELISA, ImmuLisa Enhanced SS-B (La) Antibody ELISA, ImmuLisa Enhanced Sm Antibody ELISA, ImmuLisa Enhanced RNP Antibody ELISA · Immco Diagnostics, Inc. · Mar 31, 2015
- K990627 — AUTOSTAT II ANTI-SS-A/RO ELISA · Cogent Diagnotics , Ltd. · Apr 28, 1999
Submission Summary (Full Text)
{0}
K970218
# APR - 8 1997
# 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: ________________
## Applicant Information:
Date Prepared: January 15, 1997
Name: Diamedix Corporation
Address: 2140 N. Miami Avenue
Miami, FL 33127
Contact Person: Dr. Lynne Stirling
Phone Number: 305-324-2354
Fax Number: 305-324-2585
## Device Information:
Trade Name: Is-(immunosimplicity)-anti-SSA Test System
Common Name: Anti-SSA EIA Test
Classification Name: Extractable Antinuclear Antibody
## Equivalent Device:
Helix Diagnostics Enzyme Immunoassay Anti-SS-A/Ro Antibody Test Kit
## Device Description:
The Is-anti-SSA Test Kit System is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of IgG to SSA (Ro) antigen in human serum.
## Intended use:
The assay is intended for use in detecting antibodies to SSA (Ro) antigen in a single human serum sample. The results of the assay are to be used as an aid in the diagnosis of autoimmune disorders.
## Comparison to Predicate Device:
The Is-anti-SSA Test System is an enzyme-linked immunosorbent assay to detect IgG to SSA (Ro) in human serum. Purified SSA (Ro) antigen is attached to a solid phase microtiter well. Diluted test sera are added to each well. If antibodies which recognize the SSA antigen are present in the patient sample, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme-labeled anti-human immunoglobulin (conjugate) is added to each test well. If antibody is present the enzyme-linked antibody will bind to it. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is then added to each well. If enzyme is present from the prior step, the substrate will be converted to produce a colored product. The reaction is stopped and the color intensity is measured photometrically providing an indirect measure of the specific antibody present in the patient sample.
000192
{1}
Summary of Safety and Effectiveness
## Performance Characteristics
### A. Comparison Testing
The Diamedix Is-anti-SSA Test Kit was evaluated relative to another commercially available anti-SSA ELISA test kit using 100 sera from normal blood donors and 65 sera from autoimmune patients. The results are summarized in Table 1 below.
| Table 1 | Manual | | | MAGO | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Number of Sera | % | 95% Confidence | Number of Sera | % | 95% Confidence |
| Relative Sensitivity | 64/69 | 93 | 84-98 | 64/69 | 93 | 84-98 |
| Relative Specificity | 92/93 | 99 | 94-100 | 91/93 | 98 | 92-100 |
| Agreement | 156/162* | 96 | 92-99 | 155-162* | 96 | 91-98 |
* Three borderline samples were excluded from the calculation.
Five sera negative by Is-anti-SSA (manual and MAGO) and positive by the comparative method were negative when tested by a third method. For manual testing, one serum positive by Is-anti-SSA and negative by the comparative method was negative when tested by a referee method. For MAGO testing, two sera were positive by Is-anti-SSA and negative by the comparative method. When tested by a third method, one was negative and one was positive.
### B. Linearity
Figures 1 and 2 show typical examples of Is-anti-SSA Test Kit linearity. The figures depict the results of the Calibrator tested by Is-anti-SSA after serial two-fold manual dilutions in Sample Diluent. Separate dilutions were tested both manually and with MAGO. The results demonstrate a high degree of linearity for the Is-anti-SSA Test Kit throughout the testing range.

Figure 1 Manual Linearity

Figure 2 MAGO Linearity
000193
{2}
C. Precision Testing
The precision of the Is-anti-SSA Test Kit was determined at Diamedix by testing six different sera and kit Calibrator and controls in two runs on three different days. The intra- and interassay precision is shown in Table 2 below.
| Table 2 | anti-SSA PRECISION | | | | |
| --- | --- | --- | --- | --- | --- |
| SERUM | Overall MEAN EU/ml | MANUAL | | MAGO | |
| | | INTRA-CV% | INTER-CV% | INTRA-CV% | INTER-CV% |
| 1 (NEG) | 2.0 | 14.5 | 22.2 | 16.3 | 13.0 |
| 2 (NEG) | 2.4 | 6.5 | 20.0 | 11.1 | 12.5 |
| 3 (POS) | 46.7 | 4.1 | 6.3 | 3.9 | 5.2 |
| 4 (POS) | 23.6 | 6.5 | 8.5 | 6.6 | 10.4 |
| 5 (POS) | 65.6 | 5.0 | 8.2 | 2.3 | 5.2 |
| 6 (POS) | 104.5 | 4.4 | 7.5 | 2.4 | 4.4 |
| CAL | 100.1 | 4.8 | 5.8 | 6.2 | 8.6 |
| POS CTRL | 42.6 | 3.1 | 5.0 | 5.0 | 7.4 |
| NEG CTRL | 1.5 | 9.1 | 21.4 | 24.9 | 25.0 |
D. Crossreactivity
Twenty-four sera positive for the six autoimmune specificities were tested in $\mathcal{I}$ anti-SSA Test Kit. The results are shown in Table 5. Since anti-SSB antibodies are nearly always accompanied by anti-SSA antibodies, samples 5-8 are positive in the $\mathcal{I}$ anti-SSA test due to the additional presence of anti-SSA antibodies.
Table 3 Crossreactivity
| Sample | Is-anti-SSA EU/ml | Interp | Specificity |
| --- | --- | --- | --- |
| 1 | 154.8 | POS | SSA |
| 2 | 204.0 | POS | SSA |
| 3 | 54.8 | POS | SSA |
| 4 | 203.6 | POS | SSA |
| 5 | 203.7 | POS | SSB |
| 6 | 204.7 | POS | SSB |
| 7 | 203.9 | POS | SSB |
| 8 | 203.7 | POS | SSB |
| 9 | 31.8* | POS | Sm |
| 10 | 4.4 | NEG | Sm |
| 11 | 1.8 | NEG | Sm |
| 12 | 3.6 | NEG | Sm |
| 13 | 6.0 | NEG | RNP |
| 14 | 18.2** | EQ | RNP |
| 15 | 8.0 | NEG | RNP |
| 16 | 6.8 | NEG | RNP |
| 17 | 2.5 | NEG | Jo-1 |
| 18 | 5.1 | NEG | Jo-1 |
| 19 | 4.3 | NEG | Jo-1 |
| 20 | 2.6 | NEG | Jo-1 |
| 21 | 4.1 | NEG | Scl-70 |
| 22 | 6.4 | NEG | Scl-70 |
| 23 | 2.1 | NEG | Scl-70 |
| 24 | 6.5 | NEG | Scl-70 |
*Confirmed positive for anti-SSA by an alternative ELISA method.
** Repeated equivocal in Is-anti-SSA.
000194
{3}
000195
# E. Expected Values
The expected values in the normal population were determined by assaying 100 normal donor sera collected in South Florida. Figures 3 and 5 show the distribution of SSA results in the normal population performed manually and on MAGO respectively.
The distribution of EU/ml values for 65 clinically characterized sera along with the 100 normal donor sera are shown in Figures 4 and 6 performed manually and on MAGO respectively.

Figure 3. Manual Normals

Figure 4. Manual Expected Values

Figure 5. MAGO Normals

Figure 6. MAGO Expected Values
# F. Correlation of Manual and MAGO Results
Numerical comparison of EU/ml values, between manual and MAGO results for 165 samples in the Is-anti-SSA Test Kit showed a correlation of 0.85 (Pearson).