K Number
K970139

Validate with FDA (Live)

Date Cleared
1998-04-30

(470 days)

Product Code
Regulation Number
870.2100
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BP-2BIO Digital Blood Pressure Monitor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

Device Description

The BP-2BIO Digital Blood Pressure Monitor is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a noninvasive technique which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressures is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than stethoscope and mercury manometer. The sensor converts tiny alteration in cuff pressure to electrical signals; by analyzing those signals to define the systolic, diastolic and calculating pulse rate is a well known technique in the market so called "oscillometric method".

AI/ML Overview

The MicroLife Manual Blood Pressure Monitor, Model BP-2BIO, has shown to meet acceptance criteria through various tests, primarily guided by the ANSI/AAMI SP-10-1992 "National Standard for Electronic or Automated Sphygmomanometers".

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

CriteriaAcceptance StandardDevice Performance
Electrical, Mechanical, and Environmental PerformanceRequirements outlined in the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND.Met all relevant requirements. Specific tests conducted include: - General Functions Test - Reliability Test-Operation Conditions - Reliability Test-Drop Testing - Reliability Test-Storage - Reliability Test-Vibration Testing - EMC Test Report - EN1060 Regulations and Test Report - CE Certification
Blood Pressure Monitoring AccuracyANSI/AAMI SP-10-1992 "National Standard for Electronic or Automated Sphygmomanometers"Met all relevant requirements. The device uses an oscillometric method similar to the predicate device (Omron HEM-405C) but with an improved software interface to the sensor. The software utilizes a split slope resolution (0-100mmHg, 100-200mmHg, 200-300mmHg) to reduce nonlinearity error from 1% to 0.33%, enhancing accuracy.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the sample size for the clinical test set used for the ANSI/AAMI SP-10-1992 standard. It also does not mention the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not provide information regarding the number of experts used to establish the ground truth for the test set or their qualifications. The ANSI/AAMI SP-10-1992 standard typically mandates specific protocols for clinical validation, often involving trained human observers using mercury sphygmomanometers as a reference, but these details are not present in this submission.

4. Adjudication Method for the Test Set

The document does not describe the adjudication method used for the test set.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

An MRMC comparative effectiveness study was not performed or reported in this submission. The evaluation focuses on the standalone performance of the device against a standard and its substantial equivalence to a predicate, not on human-AI collaboration.

6. Standalone Performance Study

Yes, a standalone study was performed. The "Discussion of Clinical Tests Performed" section explicitly states: "ANSI/AAMI SP-10-1992 'National Standard for Electronic or Automated Sphygmomanometers' testing was performed. All relevant sections were addressed and testing conducted. The BP-2BIO met all relevant requirements of this standard." This indicates an evaluation of the algorithm's performance in determining blood pressure and pulse rate without human-in-the-loop assistance during the measurement process itself.

7. The Type of Ground Truth Used

Based on the reference to ANSI/AAMI SP-10-1992, the ground truth for blood pressure measurements would typically be established by expert auscultation using a mercury sphygmomanometer, considered the gold standard for non-invasive blood pressure measurement at the time. This standard defines precise protocols for comparing automated device readings against human-observed readings by trained professionals.

8. The Sample Size for the Training Set

The document does not provide any information about a training set or its sample size. This device appears to be a traditional medical device (hardware with embedded software for signal processing) rather than an AI/ML-driven device that requires a separate "training set" in the modern sense. The software's accuracy is improved by a "split slope resolution" that maps to the sensor's output, a calibration approach rather than a machine learning training process.

9. How the Ground Truth for the Training Set was Established

As no "training set" is explicitly mentioned in the context of machine learning, there is no information on how ground truth for such a set would have been established. The device relies on a "well known oscillometric method" and a software interface designed to optimize accuracy against the sensor's output through calibration, rather than learning from a diverse dataset.

{0}------------------------------------------------

APR 3 0 1998

ATTACHMENT #12b Page 1 of 3

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K970139

1. Submitter's Identification:

Micro Weiss Electronics 89 Bell Street West Babvlon, NY 11704

Date Summary Prepared:

August 22, 1997

2. Name of the Device:

MicroLife Manual Blood Pressure Monitor, Model BP-2BIO

3. Predicate Device Information:

The MicroLife Manual Blood Pressure Monitor, Model BP-2BIO is substantially equivalent to the Omron Oscillometric Digital Blood Pressure Monitor, Model HEM-405C. K#903133.

4. . Device Description:

The BP-2BIO Digital Blood Pressure Monitor is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a noninvasive technique which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressures is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than stethoscope and mercury manometer. The sensor converts tiny alteration in cuff pressure to electrical signals; by analyzing those signals to define the systolic, diastolic and calculating pulse rate is a well known technique in the market so called "oscillometric method".

{1}------------------------------------------------

ATTACHMENT #12b Page 2 of 3

5. Intended Use:

The BP-2BIO Digital Blood Pressure Monitor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

ଚ. Comparison to Predicate Devices:

Both devices use the well known oscillometric method within the software algorithm to determine the systolic, diastolic and pulse rate. A similar arm cuff is inflated automatically, deflate rate is controlled but a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Each device uses a similar capacitance-type pressure sensor to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Once the reading is determined each unit operates a solenoid valve to release the pressure to zero. Our Digital Blood Pressure Monitor, Model BP-B21O, differs from the predicate device in the software interface to the sensor which determines the system accuracy.

The interface between the sensor and the microprocessor determines the system's accuracy. For our Digital Blood Pressure Monitor, Model BP-B2IO, the software is capable of a split slope resolution to improve accuracy over the entire range. Since the range is "split" into three sections (0 to 100mmHg) (100 to 200mmHq) (200 to 300mmHg) error due to nonlinearity is reduced by the ability to adjunct the slope to best fit the output curve. A nonlinearity of 1 % is reduced to .33% by splitting the span into three separate linear relations. This way the sensor is matched to the software by using a series of jumpers that profile the slopes to the output of the sensor.

The Digital Blood Pressure Monitor Model BP-B210 has field calibration access; this mode is initiated by depressing both the start and the on button for over 5 seconds. A sleeve on the input tube fitting can be removed so that the fitting can be inserted into the unit in a manner to bypass the pressure bleed valve to close the pressure system and enable it to respond to the input pressure for calibration check.

{2}------------------------------------------------

ATTACHMENT #12b Page 3 of 3

Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:

Testing information demonstrating safety and effectiveness of the MicroLife Manual Blood Pressure Monitor (Manual Inflate) Model BP-2BIO) in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.

The following testing was conducted:

  • General Functions Test a.
  • Reliability Test-Operation Conditions b.
  • Reliability Test-Drop Testing C.
  • Reliability Test-Storage d.
  • Reliability Test-Vibration Testing e.
  • EMC Test Report f.
  • EN1060 Regulations and Test Report g.
  • CE Certification h.

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the MicroLife Manual Blood Pressure Monitor (manual inflate), Model BP-2BIO tested met all relevant requirements of the aforementioned tests.

8. Discussion of Clinical Tests Performed:

ANSI/AAMI SP-10-1992 "National Standard for Electronic or Automated Sphygmomanometers" testing was performed. All relevant sections were addressed and testing conducted. The BP-2BIO met all relevant requirements of this standard.

9. Conclusions:

We have demonstrated that the MicroLife Manual Blood Pressure Monitor, Model BP-2BIO is as safe and effective as the predicate, the Omron Oscillometric Digital Blood Pressure Monitor, Model HEM-405C based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions," and, the ANSI/AAMI Voluntary Standard, SP-10-1992 testing results.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is centered on the image and is the only element present.

Public Health Service

Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized symbol consisting of three overlapping profiles of human faces.

APR 3 0 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Susan D. Goldstein-Falk Official Correspondent for Micro Weiss Electronics 89 Bell Street W. Babylon, NY 11704

K970139 Re:

MicroLife Manual Blood Pressure Monitor (Manual Inflate) Regulatory Cľass: II (Two) Product Code: DPW Dated: April 13, 1998 Received: April 15, 1998

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Susan Goldstein-Falk

This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html."

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

ATTACHMENT #12a Page_1________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): K970139

MicroLife Manual Blood Pressure Monitor, Device Name: Manual Inflate, Model BP-2BIO

Indications For Use:

The BP-2BIO Digital Blood Pressure Monitor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number. K970139

ription Use OR Over-The-Counter Use

Prescription Use (Per 21 CFR 801.109)

. .

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).