K Number
K964815

Validate with FDA (Live)

Date Cleared
1997-09-04

(279 days)

Product Code
Regulation Number
882.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AP-4000, Air Pulse Sensory Stimulator, is intended to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve. The structures being stimulated in the area of the Upper Airway innervated by the Superior Laryngeal Nerve are: the Left and Right Anterior Wall of the Pyriform Sinus and the Left and Right Aryepiglottic Folds. The device is intended to be used with the Pentax FNL-10AP or FNL-10RAP, Fiber Naso -Pharyngo- Laryngo Scope, introduced per nasally in Adult and Pediatric patient populations with suspected Dysphagia.

Device Description

The AP-4000, Air Pulse Sensory Stimulator, is an endoscopic accessory that must be used with the Pentax FNL-10AP or FNL-10RAP, Fiber Naso-Pharyngo-(Laryngo) Scopes. The AP-4000 is a software controlled device. The device contains an Air Pump and the mechanics to control the pressure of a 5 0 millisecond (ms) air pulse, delivered through the FNL-10AP working channel. The device is connected to the endoscope working channel with an Air Supply Tube (two pieces are included with the device as standard set components). The device front panel has controls and displays to allow a desired pressure for the 50 ms air pulse to be selected (in the range 2 mm Hg to 10 mm Hg), the air pulse to be initiated, and the actual pressure of the delivered air pulse to be displayed. There are connections for the Air Supply Tube (to the endoscope working channel), a pneumatic footswitch (alternate control to initiate air pulse, one piece is included as a standard set component), and an RS-232 connection so that unit displays may be available to other system peripheral devices (Addon Camera).

AI/ML Overview

The provided document is a 510(k) summary for the AP-4000, Air Pulse Sensory Stimulator. It describes the device, its intended use, and comparison to predicate devices, but does not include any information regarding acceptance criteria or a study proving the device meets said criteria.

The document explicitly states: "The submission for substantial equivalence was not based on an assessment of clinical performance data." This indicates that no clinical study was performed or submitted for this specific 510(k) clearance process to demonstrate performance against acceptance criteria.

Therefore, I cannot provide the requested information for the following sections based on the input document:

  • A table of acceptance criteria and the reported device performance
  • Sample size used for the test set and the data provenance
  • Number of experts used to establish the ground truth for the test set and their qualifications
  • Adjudication method for the test set
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
  • If a standalone (algorithm only) performance study was done
  • The type of ground truth used
  • The sample size for the training set
  • How the ground truth for the training set was established

The document focuses on demonstrating substantial equivalence to predicate devices based on design, specifications, and intended use, rather than presenting a performance study with defined acceptance criteria.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body or wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

MAY 1 1 2017

Mr. Paul Silva Regulatory Affairs Coordinator Pentax Precision Instrument Corporation 3117 Commerce Parkway Miramar, Florida 33025

Re: K964815

Trade/Device Name: AP-4000, Air Pulse Sensory Stimulator Regulation Number: 21 CFR 882.1200 Regulation Name: Two-point discriminator Regulatory Class: Class I Product Code: LLN Dated: August 12, 1997 Received: August 13, 1997

Dear Mr. Silva:

This letter corrects our substantially equivalent letter of September 4, 1997.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

Page 2 - Mr. Paul Silva

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Michae

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Page _ 1 of _ 1 . .

510(k) Number (if Known):

AP-4000, Air Pulse Sensory Stimulator Device Name:

Indications for IJse: ....... . .... ..... ...... . . . . . . . . . . .

The AP-4000, Air Pulse Sensory Stimulator, is intended to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve. The structures being stimulated in the area of the Upper Airway innervated by the Superior Laryngeal Nerve are: the Left and Right Anterior Wall of the Pyriform Sinus and the Left and Right Aryepiglottic Folds. The device is intended to be used with the Pentax FNL-10AP or FNL-10RAP, Fiber Naso -Pharyngo- Laryngo Scope, introduced per nasally in Adult and Pediatric patient populations with suspected Dysphagia.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Thomas J. Callahan

(Division Sign-Off) Division of C:nthinvascular, Respiratory, and Neurological Devices 510(k) Number_1964815

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

{3}------------------------------------------------

K464815

KS053 DOCUMENT NUMBER: 03-19-97 DATE PREPARED:

PAGE 1 OF 1 PAUL SILVA PREPARED BY:

510(K) SUMMARY: AP-4000, AIR PULSE SENSORY STIMULATOR

SUBMITTER INFORMATION: Paul Silva

Pentax Precision Instrument Corporation 30 Ramland Road SEP - 4 1997 Orangeburg, NY, 10962 TEL: (914)-365-0700

NAME OF DEVICE:

Trade Name: AP-4000. Air Pulse Sensory Stimulator Classification Name: Endoscope and Accessories (78KOG) {876.1500}

PREDICATED DEVICE(S) INFORMATION:

FNL-15P2, Fiber NasoPharyngo(Laryngo)scopePentaxK921707
Reflex Hammer, Manual*Class 1 device, no 510(k) clearance
Keeler Non Contact TonometerKeeler InstrumentsK870750
Companion 335, Nasal CPAP SystemNellcor Puritan BennettK942210
Motility CatheterMAIK823701
Infusion PumpMAIK823700
Synetics liberty System, PC PolygraphSyneticsK904625

DEVICE DESCRIPTION:

The AP-4000, Air Pulse Sensory Stimulator, is an endoscopic accessory that must be used with the Pentax FNL-10AP or FNL-10RAP, Fiber Naso-Pharyngo-(Laryngo) Scopes. The AP-4000 is a software controlled device. The device contains an Air Pump and the mechanics to control the pressure of a 50 millisecond (ms) air pulse, delivered through the FNL-10AP working channel. The device is connected to the endoscope working channel with an Air Supply Tube (two pieces are included with the device as standard set components). The device front panel has controls and displays to allow a desired pressure for the 50 ms air pulse to be selected (in the range 2 mm Hg to 10 mm Hg), the air pulse to be initiated, and the actual pressure of the delivered air pulse to be displayed. There are connections for the Air Supply Tube (to the endoscope working channel), a pneumatic footswitch (alternate control to initiate air pulse, one piece is included as a standard set component), and an RS-232 connection so that unit displays may be available to other system peripheral devices (Addon Camera).

INTENDED USE:

The AP-4000, Air Pulse Sensory Stimulator, is intended to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve. The structures being stimulated in the area of the Upper Airway innervated by the Superior Laryngeal Nerve are: the Left and Right Anterior Wall of the Pyriform Sinus and the Left and Right Aryepiglottic Folds. The device is intended to be used with the Pentax FNL-10AP or FNL-10RAP, Fiber Naso -Pharyngo- Laryngo Scope, introduced per nasally in Adult and Pediatric patient populations with suspected Dysphagia.

COMPARISON TO PREDICATED DEVICE(S):

The submission for substantial equivalence included AP-4000, Air Pulse Sensory Stimulator, specifications, literature, the identification of standard set components and optional accessories. Comparison tables were provided to illustrate the comparisons to the predicated devices. The submission for substantial equivalence was not based on an assessment of clinical performance data.

§ 882.1200 Two-point discriminator.

(a)
Identification. A two-point discriminator is a device with points used for testing a patient's touch discrimination.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.