(89 days)
The HI STAR system is indicated for use a general purpose MR imager that produces transverse, sagittal, coronal, and oblique cross-sectional images of internal structures of the human body, head, and extremities. Contrast in these images reflect the intrinsic oroperties of the hydrogen proton, proton density, T1, and T2, and the extrinsic parameters of the applied imaging sequences. The indications for the HI STAR do not differ significantly from the predicate, or legally marketed devices.
The Health Images HI STAR is a proton (H-1) Magnetic Resonance Imaging device. The HI STAR has been designed to produce transverse, sagittal, coronal, and oblique section images of the head, body and extremities of the human body. Contrast in these images is produced primarily by T1, T2, and proton density tissue characteristics. This imager employs conventional MRI imaging techniques in which the application of a combination of pulsed radio frequency energy and magnetic gradient fields is applied to the body to generate MR proton echo signals. These multidimensionally coded signals are converted by Fourier transform into 2D and 3D image sets.
Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets those criteria:
It's important to note that this document is a 510(k) Summary from 1997, specifically for a Magnetic Resonance Imager (MRI). The nature of acceptance criteria and supporting studies for such a device at that time, and compared to a more modern AI/software-as-a-medical-device (SaMD) submission, will be different. This document focuses on hardware performance and safety rather than diagnostic accuracy of an AI algorithm.
1. A table of acceptance criteria and the reported device performance
The document provides "Safety Parameters" and "Imaging Performance Parameters" which function as acceptance criteria for the MRI device.
| Parameter Criterion | Acceptance Criteria (FDA Limit/Standard) | Reported Device Performance (HI STAR) |
|---|---|---|
| Safety Parameters | ||
| Maximum Static Magnetic Field | ≤ 2 tesla | 0.6T |
| Rate of Magnetic Field Strength change | ≤ 20 T/second | < 20 T/s |
| SAR (whole body, First Controlled) | < 1.5 W/kg, averaged over whole body | < 1.5 W/kg |
| SAR (whole body, Second Controlled) | 1.5 ≤ SAR < 4.0 W/kg, averaged over whole body | 1.5 ≤ SAR < 4.0 W/kg |
| Acoustic Noise Levels | < 105 dB (A weighted average over 1 hour) | < 92 dBA |
| Imaging Performance Parameters | Standard | Data (HI STAR) |
| Specification Volume (Head) | Manufacture | 25.6 cm DSV |
| Specification Volume (Body) | Manufacture | 48 cm DSV |
| Signal-to-noise Ratio (Body Coil) | NEMA MS-1 | 103 |
| Signal-to-noise Ratio (Head Coil) | NEMA MS-1 | 62 |
| Signal-to-noise Ratio (MRA Coil) | NEMA MS-6 | 50 |
| Signal-to-noise Ratio (Wrist Coil) | NEMA MS-6 | 117 |
| Signal-to-noise Ratio (Neck Coil) | NEMA MS-6 | 108 |
| Signal-to-noise Ratio (Spine Coil) | NEMA MS-6 | 62 |
| Signal-to-noise Ratio (Breast Coil) | NEMA MS-6 | 100 |
| Signal-to-noise Ratio (Bilateral TMJ) | NEMA MS-6 | 133 |
| Signal-to-noise Ratio (Extremity Coil) | NEMA MS-6 | 49 |
| Signal-to-noise Ratio (Shoulder Coil) | NEMA MS-6 | 108 |
| Signal-to-noise Ratio (GPS Coils) | NEMA MS-6 | 95 (29.2x15.2cm), 98 (19x19cm), 80 (12.7x12.7cm), 133 (6.4x6.4cm) |
| Image Uniformity (Body Coil) | NEMA MS-2 | 12%, Specification Volume 48 cm |
| Image Uniformity (Head Coil) | NEMA MS-2 | 6%, Specification Volume 25.6 cm |
| Slice Thickness, Separation | AAPM-28 | 10% of specified |
| Resolution (Body) | AAPM-28 | 1.5 mm |
| Resolution (Head) | AAPM-28 | 1 mm |
| Geometric Distortion (Body) | NEMA #2 | < 5% @ 31 cm |
| Geometric Distortion (Head) | NEMA #2 | < 5% @ 15.5 cm |
2. Sample size used for the test set and the data provenance
The document states: "Sample images included with this submission were obtained on R &D systems at Health Sample Includio war this oubmisers. Each volunteer signed a consent form. There were no adverse reactions noted on any of the participants."
- Sample Size: The exact number of "volunteers" is not specified. It refers to "sample images" and "each volunteer," implying a limited, unnamed number.
- Data Provenance: The images were obtained on R&D systems at Health Images itself. It was collected from "volunteers" who signed consent forms, suggesting these were healthy volunteers rather than patients with specific pathologies. The study appears to be prospective in nature, as images were "obtained" for the submission, rather than retrospective analysis of existing data. The country of origin is implicitly the United States, given the submitter's address and FDA context.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. For an MRI device submission of this nature (hardware performance), the "ground truth" for imaging parameters (SNR, uniformity, resolution, distortion) is typically established through quantitative measurements using phantoms and standardized protocols (e.g., NEMA standards), not expert interpretation of clinical images. The safety parameters are also measured quantitatively against established limits.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable and therefore not provided. Adjudication methods are relevant for studies requiring multiple expert interpretations to establish a consensus ground truth for diagnostic AI, which is not the type of study described here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable and therefore not provided. This document pertains to the performance of an MRI machine itself, not an AI-assisted diagnostic tool. There is no mention of human readers or AI assistance in the context of diagnostic interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable and therefore not provided. This document is for an MRI hardware device, not a standalone AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance metrics appears to be based on:
- Physical measurements against industry standards/specifications: For imaging parameters like SNR, image uniformity, slice thickness, resolution, and geometric distortion, the device's performance is compared against NEMA and AAPM standards. These involve quantitative measurements using phantoms, not clinical ground truth like pathology or expert consensus.
- Quantitative measurements against regulatory limits: For safety parameters such as magnetic field strength, SAR, and acoustic noise, physical measurements are taken and compared directly to FDA limits and IEC standards.
8. The sample size for the training set
This is not applicable and therefore not provided. This document describes an MRI hardware device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable and therefore not provided. As there is no training set for an AI algorithm mentioned, there is no ground truth established for one.
{0}------------------------------------------------
FEB 2 4 1997
Attachment H
510 (k) Summary / Safety and Effectiveness
- 510 (k) Summary, Safety and Effectiveness
{1}------------------------------------------------
The following information is in accordance with 21 CFR 807.92
Submitter's Name, Address, Telephone Number , Contact Person, Preparation Date
Health Images, Inc. 8601 Dunwoody Place Building 200 Atlanta, GA 30350 (770) 642-9869 ext. 2057 Richard Halbach, Dir. of Clinical Research & Regulatory Affairs
November 25, 1996
Name of the Device and Classification Name/Number
HI STAR, Magnetic Resonance Imager. The FDA has reclassified these devices, 90LNH, to Class II.
Predicate Devices
| Manufacturer | Commercial Name | Description |
|---|---|---|
| Health Images | HI-Standard | 0.6T whole body MRI imager,superconducting magnet |
| Health Images | HI-Standard/CalibreIRIS sequence | 0.6T whole body MRI imager,superconducting magnet, IRISSilicone Specific Sequence |
| Health Images | HI STAR | 0.6T whole body MRI imager,superconducting magnet |
| Elscint | Gyrex 2T Prestige | Spectral Fat Suppression, FSE |
| Philips | Gyroscan NT, 1T | Spectral Fat Suppression |
Description of the Device
The Health Images HI STAR is a proton (H-1) Magnetic Resonance Imaging device. The HI STAR has been designed to produce transverse, sagittal, coronal, and oblique section images of the head, body and extremities of the human body. Contrast in these images is produced primarily by T1, T2, and proton density tissue characteristics. This imager employs conventional MRI imaging techniques in which the application of a combination of pulsed radio frequency energy and magnetic gradient fields is applied to the body to generate MR proton echo signals. These multidimensionally coded signals are converted by Fourier transform into 2D and 3D image sets.
{2}------------------------------------------------
194466
.
Device Characteristics
| Parameter | Example |
|---|---|
| Type of magnet | Superconducting,passive/active shim |
| Static field strength | 0.6T |
| Type of RF coils | Transmit/Receive - quadratureBody |
| Receive - quadratureHead, MRA, Neck,Spine, WristReceive - linearBilateral TMJ, VolumeExtremity, Bilateral Breast,ShoulderGeneral Purpose Surface(body & extremity)29.2 cm x 15.2 cm(11.5 x 6.0 inch)19.0 cm x 19.0 cm(7.5 x 7.5 inch)12.7 cm x 12.7 cm(5.0 x 5.0 inch)6.4 cm x 6.4 cm(2.5 x 2.5 inch) | |
| Data acquisition modes | Single slice, Multi-slice, Multi-angle oblique, 3D Volume,Physiologic Gating |
| Pulse Sequences | Spin Echo, Field Echo, TurboField Echo, Multi-spin Echo,Inversion Recovery, MRA, FSE |
| Additional Contrast Techniques | IRIS (Silicone Specific), FatSat,Magnetization Transfer |
| Reconstruction Technique | Fourier transform |
| Type of installation: | Mobile, Fixed |
| Slice selection method: | Selective Gradient,3D phase encode |
| Display Matrix | Variable up to 1024 x 1024 |
{3}------------------------------------------------
In plane 64 - 512, increment of Acquisition Matrix 8 2D, 1 - 256 slices, increment of Number of slices, 1 3D, 12 - 256 increment of 2 Range and Increment Window and level Post processing Rotate image Mirror image Zoom image Pan image Add and Subtract images HI-Q image processing noise reduction function Transverse, Sagittal, Coronal, Slice orientation Single and multiple oblique From Contiguous up to 320 mm, Interslice spacing, minimum, maximum, increment of 0.5 mm
....
{4}------------------------------------------------
Intended Uses
The HI STAR system is indicated for use a general purpose MR imager that produces transverse, sagittal, coronal, and oblique cross-sectional images of internal structures of the human body, head, and extremities. Contrast in these images reflect the intrinsic oroperties of the hydrogen proton, proton density, T1, and T2, and the extrinsic parameters of the applied imaging sequences. The indications for the HI STAR do not differ significantly from the predicate, or legally marketed devices.
Safety Parameters
| Parameter | FDA Limit | HI STAR | Test Standard |
|---|---|---|---|
| Maximum Static MagneticField | ≤ 2 tesla | 0.6T | Manufacturer |
| Rate of Magnetic FieldStrength change | ≤ 20 T/second | < 20 T/s | IEC 601-2-33 |
| SAR | < 1.5 W/kg, averaged overwhole body (First ControlledOperating Level) | < 1.5 W/kg | IEC 601-2-33 |
| 1.5 ≤ SAR < 4.0 W/kg,averaged over wholebody (Second ControlledOperating Level) | 1.5 ≤ SAR < 4.0W/kg | IEC 601-2-33 | |
| Acoustic Noise Levels | < 105 dB (A weighted averageover 1 hour) | < 92 dBA | NEMA MS-4 |
Imaging Performance Parameters
| Parameter | Data | Standard |
|---|---|---|
| Specification Volume | Head 25.6 cm DSVBody 48 cm DSV | Manufacture |
| Signal-to-noise Ratio | ||
| Body Coil | 103 | NEMA MS-1 |
| Head Coil | 62 | NEMA MS-1 |
| MRA Coil | 50 | NEMA MS-6 |
| Wrist Coil | 117 | NEMA MS-6 |
| Neck Coil | 108 | NEMA MS-6 |
| Spine Coil | 62 | NEMA MS-6 |
| Breast Coil | 100 | NEMA MS-6 |
| Bilateral TMJ Coil | 133 | NEMA MS-6 |
| Extremity Coil | 49 | NEMA MS-6 |
| Shoulder Coil | 108 | NEMA MS-6 |
{5}------------------------------------------------
| Parameter | Data | Standard |
|---|---|---|
| General PurposeSurface Coils29.2 cm x 15.2 cm(11.5 x 6.0 inch)19.0 cm x 19.0 cm(7.5 x 7.5 inch)12.7 cm x 12.7 cm(5.0 x 5.0 inch)6.4 cm x 6.4 cm(2.5 x 2.5 inch) | 959880133 | NEMA MS-6 |
| Image Uniformity | ||
| Body Coil | 12%, Specification Volume 48 cm | NEMA MS-2 |
| Head Coil | 6%, Specification Volume 25.6 cm | NEMA MS-2 |
| Slice Thickness, Separation | 10% of specified | AAPM-28 |
| Resolution | Body 1.5 mmHead 1 mm | AAPM-28 |
| Geometric Distortion | Body < 5% @ 31 cmHead < 5% @ 15.5 cm | NEMA #2 |
Sample images included with this submission were obtained on R &D systems at Health Sample Includio war this oubmisers. Each volunteer signed a consent form. There were no adverse reactions noted on any of the participants.
re
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.